Arterial blood pressure monitoring in stroke cohorts: the impact of reduced sampling rates to optimise remote patient monitoring. Blood Press. Monit. (IF 1.2) Pub Date : 2024-08-12 James D Ball, Ronney B Panerai, Tim Henstock, Jatinder S Minhas
Remote patient monitoring (RPM) beat-to-beat blood pressure (BP) provides an opportunity to measure poststroke BP variability (BPV), which is associated with clinical stroke outcomes. BP sampling interval (SI) influences ambulatory BPV, but RPM BP SI optimisation research is limited. SI and RPM device capabilities require compromises, meaning SI impact requires investigation. Therefore, this study
-
Validation of the AViTA upper-arm blood pressure monitor BP636 with multiple cuff sizes in the general population according to the ANSI/AAMI/ISO 81060-2:2018/Amd1:2020. Blood Press. Monit. (IF 1.2) Pub Date : 2024-08-09 Chung-Yu Lin, I-Chih Huang
To determine the accuracy of the AViTA BP636 upper-arm home blood pressure (BP) monitor in general populations with small to extra-large arms according to the ANSI/AAMI/ISO Universal Standard (ISO 81060-2 : 2018/Amd1 : 2020).
-
Validation of the JAMR upper-arm pressure monitor B23 according to the AAMI/ESH/ISO Universal Standard (ISO 81060- 2:2018/AMD 1:2020). Blood Press. Monit. (IF 1.2) Pub Date : 2024-07-30 Jing Li, Li-Hong Han, Ru Feng, Li-Hao Wu, Jie Cai, Tian-Ying Zeng, Hui Liu, Lei Wang, Ju-Yun Li, Zhong-Qun Zhan
The aim of this study was to evaluate the accuracy of the JAMR upper-arm blood pressure monitor B23 in the general population according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2 : 2018/AMD 1 : 2020). The study recruited participants who met the criteria of the AAMI/ESH/ISO Universal Standard in terms of their number, sex, age, limb size, and blood pressure (BP) distribution. The study involved
-
Psychological distress, forced awakening, and morning blood pressure surge. Blood Press. Monit. (IF 1.2) Pub Date : 2024-07-22 Yeonsu Kim, Jill Howie Esquivel, Meghan Kathleen Mattos, Eric M Davis, Jeongok Logan
Morning blood pressure surge (MBPS) has been recognized as an independent predictor of cardiovascular disease events. Psychological distress, including anxiety, depression, and perceived stress, and behavioral risk factors, such as poor sleep quality, have been associated with increased MBPS. Elevations in sympathetic activity induced by forced awakening may also contribute to further increases in
-
Validation of the validity of the new wrist-type fully automatic blood pressure monitor: DBP-2242 according to ISO 81060-2:2018+AMD 1:2020. Blood Press. Monit. (IF 1.2) Pub Date : 2024-07-16 Yuehong Jin, Xiaoming Bei, Ying Jin
Evaluation of the accuracy of the wrist-type fully automatic blood pressure (BP) monitor: DBP-2242 in adolescent and adult populations according to ISO 81060-2 : 2018+Amd.1 : 2020.
-
Validation of the DBP-8278B Wrist-type Fully Automatic Blood Pressure Monitor in the general population according to the ISO 81060-2:2018/AMD 1:2020 protocol. Blood Press. Monit. (IF 1.2) Pub Date : 2024-07-16 Qi Zhou
The aim of this study was to evaluate the accuracy of the DBP-8278B Wrist-type Fully Automatic Blood Pressure Monitor in the general population according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2 : 2018/AMD 1 : 2020). The test monitor was used to measure SBP and DBP and pulse rate in the wrists using the oscillometric method. According to this protocol using the same-arm sequential blood
-
Validation of two automatic sphygmomanometers according to the modified International Standardization Organization 81060-2:2018 protocol in adults with a mid-upper arm circumference of 22 centimeters or less. Blood Press. Monit. (IF 1.2) Pub Date : 2024-07-02 Igor N Posokhov, Evgeny A Praskurnichiy, Olga S Orlova, Ekaterina A Sukhodolova, Aleksandra V Lystseva
A recent International Organization for Standardization (ISO) Task Group report calls for research to investingate potential special populations in validation studies of automated blood pressure (BP) devices. Accordingly, we aimed to determine the accuracy of two previously validated BP monitors passed in a general population when measured in adults with a mid-upper arm circumference (MUAC) ≤ 22 cm
-
Home blood pressure measurement days and changes in urinary sodium-to-potassium ratio, estimated salt and potassium intakes and blood pressure: 1-year prospective study. Blood Press. Monit. (IF 1.2) Pub Date : 2024-06-26 Minako Kinuta, Takashi Hisamatsu, Mari Fukuda, Kaori Taniguchi, Noriko Nakahata, Hideyuki Kanda
Current international guidelines recommend home blood pressure (BP) measurement and low sodium and high potassium intakes for the management of hypertension. We hypothesized that increased home BP measurement may result in more effective management of sodium and potassium intakes and BP.
-
Could soluble ST2 levels be used to detect nondipper hypertensive subgroup in newly diagnosed hypertension patients. Blood Press. Monit. (IF 1.2) Pub Date : 2024-06-24 Emrah Ozdemir, Berna Stavileci, Bahar Ozdemir, Faik Alper Aksoy, Ceyla Zeynep Colakoglu Gevher, Ali Dogan, Murat Ziyrek
ST2 receptor is a member of toll-like/interleukin-1 receptor family. After the activation of IL-33/ST2 signaling pathway clinically detectable amount of soluble form of ST2 (sST2) is released into the circulation. Previous studies showed that sST2 levels were significantly higher in hypertension patients than in controls. In this prospective study, we aimed to analyze this relation and test the predictive
-
Long-term blood pressure variability and risk of cardiovascular diseases in populations with different blood pressure status: an ambispective cohort study. Blood Press. Monit. (IF 1.2) Pub Date : 2024-06-21 Jinnan Li, Ting Qu, Ying Li, Pengcheng Li, Benmai Luo, Yue Yi, Aibin Shi, Zhixin Pang, Yuting Chu, Yuxin Zhao, Li Yang, Shaopeng Xu, Juan Xie, Hong Zhu
We aimed to investigate the correlation between long-term blood pressure variability (BPV) and the risk of cardiovascular diseases (CVDs) among population with different blood pressure statuses (normotension, well-controlled hypertension, and uncontrolled hypertension).
-
High-sensitivity C-reactive protein predicts microalbuminuria progression in essential hypertensive patients: a 3-year follow-up study. Blood Press. Monit. (IF 1.2) Pub Date : 2024-06-21 Yan Yang, Xiao-Feng Tang, Yan Wang, Jian-Zhong Xu, Ping-Jin Gao, Yan Li
To determine the independent effect of high-sensitivity C-reactive protein (hs-CRP) and the combined effects of hs-CRP and other traditional risk factors on microalbuminuria in hypertensive patients during the 3-year follow-up period.
-
Smart solutions in hypertension diagnosis and management: a deep dive into artificial intelligence and modern wearables for blood pressure monitoring. Blood Press. Monit. (IF 1.2) Pub Date : 2024-06-17 Anubhuti Juyal, Shradha Bisht, Mamta F Singh
Hypertension, a widespread cardiovascular issue, presents a major global health challenge. Traditional diagnosis and treatment methods involve periodic blood pressure monitoring and prescribing antihypertensive drugs. Smart technology integration in healthcare offers promising results in optimizing the diagnosis and treatment of various conditions. We investigate its role in improving hypertension
-
Blood pressure variability at rest and during pressor challenges in patients with acute ischemic stroke. Blood Press. Monit. (IF 1.2) Pub Date : 2024-06-04 Sankanika Roy, Man Y Lam, Ronney B Panerai, Thompson G Robinson, Jatinder S Minhas
Patients with acute ischemic stroke (AIS) have elevated blood pressure (BP) variability (BPV) and reduced baroreflex sensitivity (BRS) at rest for several days after initial stroke symptoms. We aimed to assess BPV and BRS in AIS patients during pressor challenge maneuvers in the acute and subacute phases of stroke. Pressor challenge maneuvers simulate day-to-day activities and can predict the quality
-
Inflammation-based markers, especially the uric acid/albumin ratio, are associated with non-dipper pattern in newly diagnosed treatment-naive hypertensive patients. Blood Press. Monit. (IF 1.2) Pub Date : 2024-05-22 Kaya Özen, Kenan Toprak, Mesut Karataş, Ayten Dursun
Physiologically, at night, blood pressure (BP) is expected to decrease by at least 10% in hypertensive individuals. The absence of this decrease, called non-dipper hypertension, is associated with increased end-organ damage and cardiovascular mortality and morbidity in hypertensive individuals. It is known that increased inflammatory process plays an important role in the etiopathogenesis of non-dipper
-
Validating the accuracy of Omron HEM-7372T1-AZAZ (BP5460) in monitoring blood pressure according to the ISO 81060-2:2018+Amd 1:2020 protocol in the general population. Blood Press. Monit. (IF 1.2) Pub Date : 2024-05-21 Hakuo Takahashi, Shingo Yamashita, Nobuki Yakura
The performance of the Omron HEM-7372T1-AZAZ (BP5460) in monitoring blood pressure (BP) in the upper arm was validated in accordance with the International Organization for Standardization (ISO) 81060-2:2018 (E) and amendment (Amd)1:2020 protocol.
-
Accuracy of the OMRON HEM-7361T blood pressure monitor in the differentiation between atrial fibrillation and sinus rhythm. Blood Press. Monit. (IF 1.2) Pub Date : 2024-05-20 Yi Zhou, Jia-Hui Xia, Wei Zhang, Yan Li, Ji-Guang Wang
We investigated the accuracy of the OMRON HEM-7361T automated oscillometric blood pressure (BP) monitor in the differentiation between atrial fibrillation and sinus rhythm.
-
Validation of a novel professional automated auscultatory upper-arm cuff blood pressure monitor in a general population according to the AAMI/ESH/ISO Universal Standard: KOROT V2 Doctor (InBody BPBIO280KV). Blood Press. Monit. (IF 1.2) Pub Date : 2024-05-17 Aikaterini Theodosiadi, Vasileios Ntousopoulos, Angeliki Ntineri, Panagiota Stathopoulou, Ariadni Menti, Konstantinos G Kyriakoulis, Anastasios Kollias, George S Stergiou
A novel automated auscultatory upper arm cuff blood pressure (BP) monitor KOROT V2 Doctor (InBody BPBIO280KV) was developed for professional use. An electronic stethoscope embedded in the device cuff records the Korotkoff sounds, which are graphically displayed during deflation allowing visual evaluation by the healthcare professional. The device provides automated measurements of BP and this study
-
Validation of the Mindray VS9 Vital Signs Monitor in a combined adult and pediatric population according to ISO Standard 81060-2:2018. Blood Press. Monit. (IF 1.2) Pub Date : 2024-03-25 Na Guo, Yihan Zhang, Weiqiang Chen, Hexian Zhong, Liping Li, Hanbin Xie, Wenxiu Zhu, Jun Liu, Shangrong Li
We aimed to validate the accuracy of the Mindray VS9 Vital Signs Monitor, which features the Mindray TrueBP inflation algorithm for oscillometric blood pressure (BP) measurement, to check if it complies with the International Organization for Standardization Standard (ISO 81060-2:2018) in a combined adult and pediatric population. A total of 86 participants, including both adult and pediatric subjects
-
Risk of heart disease and stroke among individuals with prehypertension or blood pressure progression: a national population-based cohort study. Blood Press. Monit. (IF 1.2) Pub Date : 2024-03-11 Qingdong Jin, Yanqing Chen
The purpose of our study was to analyze the association of blood pressure and blood pressure progression with heart disease and stroke among Chinese population.
-
Solution is not simple; sodium-glucose cotransporter-2 inhibitor use in Conn syndrome. Blood Press. Monit. (IF 1.2) Pub Date : 2024-03-11 Utku Soyaltin
In patients with bilateral primary hyperaldosteronism (PA) and those with unilateral PA who are unwilling or unable to undergo adrenalectomy an increase in plasma renin activity (PRA) provided by mineralocorticoid receptor antagonists (MRAs) therapy reflects sufficient antagonism for elevated aldosterone. Sodium-glucose cotransporter-2 inhibitors (SGLT2-i) have cardiovascular, renal protective properties
-
Validation of the Meditech ABPM-06 24-hour blood pressure monitoring system in a pediatric population according to International Organization for Standardization Protocol 81060-2:2018. Blood Press. Monit. (IF 1.2) Pub Date : 2024-03-11 Andrea E Jakab, Edit Horváth, Dorottya Molnár, Mátyás Bukva, Csaba Bereczki
Ambulatory blood pressure monitoring (ABPM) devices play a crucial role in diagnosing hypertension, not only in adults but also in pediatric patients. ABPM-06, the latest oscillometric device from Meditech Ltd. (Budapest, Hungary), is the focus of this study. The objective was to validate the ABPM-06 device using the International Organization for Standardization (ISO) 81060-2 : 2018 standard.
-
A comparison among oscillometric waveforms in healthy nonpregnant women, pregnancy and hypertensive disorders of pregnancy. Blood Press. Monit. (IF 1.2) Pub Date : 2024-03-05 Jennifer S Ringrose, Sangita Sridar, Patricia Araneta, Lillian Chan, Jalisa Kassam, Mira Wirzba, Kate Greeff, Gillian Ramsay, Winnie Sia, Rshmi Khurana, Erin Bader, Raj Padwal
Understanding of how oscillometric waveforms (OMW) vary between pregnant and nonpregnant individuals remains low. An exploratory analysis was completed to assess for quantitative and qualitative changes in OMW and oscillometric envelope features in pregnancy.
-
Validation of the Raycome model M2 ambulatory blood pressure monitor in the general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard (ISO 81060-2:2018). Blood Press. Monit. (IF 1.2) Pub Date : 2024-02-23 Shijie Yang, Zhanyang Zhou, Huanhuan Miao, Hongye Zhang, Qiong Zhou, Mei Zhai, Yuqing Zhang
The aim of this study was to evaluate the accuracy of the Raycome model M2 oscillometric upper-arm blood pressure (BP) monitor developed for ambulatory BP measurement in the general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018)
-
Accuracy of the KOROT P3 Accurate automated auscultatory blood pressure measuring device for professional use in people with extra-large arms. Blood Press. Monit. (IF 1.2) Pub Date : 2024-02-23 Claudio Fania, Antonella Giletto, Alice Niero, Paolo Palatini
To determine the accuracy of the KOROT P3 Accurate (previously InBody BPBIO480KV) monitor, an automated auscultatory blood pressure (BP) measuring device developed for professional use, in people with extra-large arms according to the ISO81060-2 2018 protocol.
-
Validation of noninvasive blood pressure monitoring function of EDAN elite V5 patient monitor with reference invasive measurement according to the ISO 81060-2:2018 standard. Blood Press. Monit. (IF 1.2) Pub Date : 2024-02-19 Yi Wan, Duan Li, Jin-Feng Chen, Yang Zhang
To validate the noninvasive blood pressure monitoring function of the EDAN elite V5 patient monitor with reference invasive blood pressure monitoring equipment for clinical use in adults, adolescents or children according to the International Organization for Standardization (ISO) 81060-2:2018 standard.
-
Noninvasive pediatric blood pressure assessment: exploring the clinicians' perspective. Blood Press. Monit. (IF 1.2) Pub Date : 2024-02-19 Natalie P Stewart, Catherine Quinlan, Stephanie Best, Jonathan P Mynard
Obtaining accurate and reliable blood pressure (BP) readings in pediatric patients is challenging, given difficulties in adhering to measurement guidelines, limited device validation and variable patient cooperation. This study aimed to investigate clinicians' perspectives surrounding noninvasive pediatric BP assessment to identify opportunities for improvement in BP technology and clinical practice
-
Relationship between the behavior change model and salt intake in hypertensive patients: a single non-specialized hypertension clinic prospective observational study. Blood Press. Monit. (IF 1.2) Pub Date : 2024-02-02 Hiromitsu Sekizuka, Toshiya Ishii, Hitoshi Miyake
We investigated whether changes in salt reduction readiness are associated with changes in estimated daily salt intake and blood pressure (BP). We divided 86 hypertensive patients into groups with high and low readiness for salt-reducing behavior [an up (UP) and a down (DN) groups, respectively] based on the transtheoretical model (TTM) over a 12-month observation period. We then investigated the relationships
-
Feasibility of Arteriograph 24 for evaluation of 24-hour pulse wave velocity and central blood pressure in patients with type 2 diabetes and non-diabetic controls. Blood Press. Monit. (IF 1.2) Pub Date : 2024-01-15 Esben Laugesen, Amalie Nørkjær Svendsen, Liv Vernstrøm, Lene Halkjær, Anna Dons-Jensen, Kristian L Funck, Klavs Würgler Hansen, Per Løgstrup Poulsen
The objective of this study was to assess the feasibility of the Arteriograph 24 device to measure 24-hour PWV and central systolic blood pressure (cSBP) in patients with type 2 diabetes (T2DM) and non-diabetic controls and compare daytime and nighttime characteristics in the two groups. Twenty-four-hour PWV and cSBP was measured in 58 patients with T2DM (mean age: 66 ± 9 years, 50% women, mean duration
-
Lowering blood pressure by exercise: investigating the effect of sweating. Blood Press. Monit. (IF 1.2) Pub Date : 2024-01-15 Rastegar Hoseini, Rasha Raed Hamid
High blood pressure (hypertension), is a common medical condition, affecting millions of people and is associated with significant health risks. Exercise has been suggested to manage hypertension by inducing sweating and the corresponding loss of sodium and water from the body.Thus, a variety of epidemiological and clinical studies have been conducted to investigate the relationship between sweating
-
Blood pressure responsiveness to resistance training in the hypertensive older adult: a randomized controlled study. Blood Press. Monit. (IF 1.2) Pub Date : 2024-01-15 Odilon Abrahin, Rejane Pequeno Abrahin, Mayko Guimarães, Vitor Bruno Teixeira de Holanda, Fernanda Andreza De Pinho Lott Figueiredo, Bruno Viana Rosa, Ivo Vieira de Sousa Neto, Nicholas Rolnick, Gislane Ferreira de Melo, Emanuelle Fernandes Prestes, Dahan da Cunha Nascimento
Different lifestyle changes have been employed to improve clinical hypertension. However, there is scarce evidence on the blood pressure responsiveness to resistance training (RT) in hypertensive older adults. Consequently, little is known about some participants clinically reducing blood pressure and others not. Thus, we investigate the effects and responsiveness of RT on blood pressure in hypertensive
-
Validation of the oscillometric home blood pressure monitor Braun BUA4000 with wide-range cuff in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard. Blood Press. Monit. (IF 1.2) Pub Date : 2023-11-15 Vasileios Ntousopoulos, Angeliki Ntineri, Aikaterini Theodosiadi, Panagiota Stathopoulou, Ariadni Menti, Konstantinos G Kyriakoulis, Anastasios Kollias, George S Stergiou
To evaluate the blood pressure (BP) measurement accuracy of the Braun BUA4000 automated oscillometric upper-arm cuff device for self-home use in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) and its Amendment 1.2020-01
-
Validation of the Wellvii VitalDetect blood pressure monitor in general population according to the International Standardization Organization 81060-2:2018. Blood Press. Monit. (IF 1.2) Pub Date : 2023-11-15 Lin-Yi Li, Jin-Feng Chen, Sen Bing, Kui Liu, Rui Cheng, Yi Wan
To evaluate the accuracy of the Wellvii VitalDetect automated oscillometric finger blood pressure monitor (single cuff size) for self/home blood pressure measurement according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018).
-
Automated office blood pressure measurement with the assistance of an instructional video in patients with hypertension. Blood Press. Monit. (IF 1.2) Pub Date : 2023-11-15 Ayumi Toba, Joji Ishikawa, Kazumasa Harada
Automated office blood pressure (AOBP) measurement often requires assistance. Accompanied by an instructional video, AOBP measurement could be performed independently by patients.
-
Validation of the automated oscillometric upper-arm cuff home blood pressure monitor AVITA BPM82 in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard. Blood Press. Monit. (IF 1.2) Pub Date : 2023-11-06 Ariadni Menti, Vasileios Ntousopoulos, Aikaterini Theodosiadi, Panagiota Stathopoulou, Anastasios Kollias, George S Stergiou
This study evaluated the accuracy of the automated oscillometric upper-arm cuff device AVITA BPM82 intended for home use in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) and its Amendment 1.2020-01.
-
Osteoprotegerin is associated with subclinical left ventricular systolic dysfunction in non-dipper hypertensive patients: a 2D speckle tracking echocardiographic study. Blood Press. Monit. (IF 1.2) Pub Date : 2023-11-06 Elnur Alizade, Muzaffer Kahyaoglu, Ismail Balaban, Servet Izci, Ahmet Guler
Past studies have shown that non-dipper hypertensive patients have more frequent subclinical left ventricular (LV) systolic dysfunction compared to dippers. Many different parameters have been examined to predict subclinical LV dysfunction. The role of osteoprotegerin (OPG) in the pathogenesis of heart failure and LV systolic dysfunction through different mechanisms had well described. In the present
-
Comparison of direct intra-arterial pressure and ClearSight finger cuff arterial pressure measurements in elderly patients undergoing transcatheter aortic valve replacement. Blood Press. Monit. (IF 1.2) Pub Date : 2023-09-21 Musashi Yahagi, Momoko Sasaki
This study aimed to assess the agreement between direct intra-arterial pressure and ClearSight finger cuff arterial pressure measurements in elderly patients undergoing transcatheter aortic valve replacement (TAVR).
-
Race differences in ambulatory blood pressure monitoring parameters. Blood Press. Monit. (IF 1.2) Pub Date : 2023-09-21 Swati Sakhuja, Byron C Jaeger, Yuichiro Yano, Daichi Shimbo, Cora E Lewis, Donald Clark Iii, Gabriel S Tajeu, Shakia T Hardy, Norrina B Allen, James M Shikany, Joseph E Schwartz, Anthony J Viera, Paul Muntner
Mean systolic and diastolic blood pressure (SBP and DBP) on ambulatory blood pressure (BP) monitoring (ABPM) are higher among Black compared with White adults. With 48 to 72 BP measurements obtained over 24 h, ABPM can generate parameters other than mean BP that are associated with increased risk for cardiovascular events. There are few data on race differences in ABPM parameters other than mean BP
-
Positive blood pressure response may predict the recovery of renal function after transcatheter aortic valve implantation. Blood Press. Monit. (IF 1.2) Pub Date : 2023-09-14 Orhan Ince, Kamil Gulsen, Sevgi Ozcan, Sevil Tugrul, Fahrettin Katkat, Serkan Karahan, Esra Donmez, Alper Kepez, Irfan Sahin, Ertugrul Okuyan
Transcatheter aortic valve implantation (TAVI) may ameliorate renal function and increase blood pressure (BP). We aimed to investigate the association between increased BP and improved renal function (IRF) after TAVI.
-
Comparison of the Omron HeartGuide to the Welch Allyn ProBP 3400 blood pressure monitor. Blood Press. Monit. (IF 1.2) Pub Date : 2023-09-14 Brianna D Harfmann, Sylvia E Neph, Mason M Gardner, Audrey A Plouffe, Jennifer R Vranish, Alexander H K Montoye
Hypertension affects approximately 100 million U.S. adults and is the leading single contributing risk factor to all-cause mortality. Accurate blood pressure (BP) measurement is essential in the treatment of BP, and a number of devices exist for monitoring. Recently, a new watch-type design was released, the Omron HeartGuide (BP8000), with claims to provide clinically accurate BP measurement while
-
Mean arterial pressure and mortality in patients with heart failure: a retrospective analysis of Zigong heart failure database. Blood Press. Monit. (IF 1.2) Pub Date : 2023-09-14 Hangkun Ma, Haibo Li, Song Sheng, Longfang Quan, Zhixu Yang, Fengqin Xu, Wenying Zeng
It is commonly observed that a higher target of mean arterial pressure (MAP) is in previous studies. This study assessed the association of MAP with short-term mortality in heart failure (HF) patients.
-
Validation of the iHealth Track upper-arm blood pressure monitor KN-550BT in general population according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018+AMD1:2020). Blood Press. Monit. (IF 1.2) Pub Date : 2023-09-14 Lei Shi, Xin-Yi Zhang, Jia-Yi Sun, Jin-Feng Chen, Yi-Fei Ma, Kai-Yuan Zhou
To validate the iHealth Track KN-550BT oscillometric upper-arm blood pressure monitor in general population according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018+AMD1:2020).
-
Ambulatory blood pressure monitoring in pediatric patients with sickle cell anemia. Blood Press. Monit. (IF 1.2) Pub Date : 2023-09-14 Gustavo Adolfo Guerrero Tinoco, Zilac Espitaleta Vergara, Maria José Daniels García, Alex Domínguez-Vargas
Sickle cell anemia (SCA) is a hemoglobinopathy presenting severe endothelial damage associated with increased prevalence of hypertension (HTN). Few studies have used ambulatory blood pressure monitoring (ABPM) in pediatric patients with SCA. The aim of this study was to characterize the ABPM profile in children with SCA.
-
Screening for hypertension in adults - the use of tonometric blood pressure monitoring. Blood Press. Monit. (IF 1.2) Pub Date : 2023-09-05 Bodil G Hornstrup, Jeppe B Rosenbæk, Nikolai Hoffmann-Petersen, Jesper N Bech
Arterial hypertension increases the risk of developing cardiovascular disease. Reliable screening tools for diagnosing hypertension are important to ensure correct risk stratification of subjects. In this study, we aimed to analyse if a wrist-worn device using a tonometric technique for measuring of 24-hour blood pressure could be used to diagnose hypertension and non-dipping. A conventional device
-
Evaluation of stroke volume estimation during orthostatic stress: the utility of Modelflow. Blood Press. Monit. (IF 1.2) Pub Date : 2023-08-24 Vera-Ellen M Lucci, Iain T Parsons, Brooke C D Hockin, Victoria E Claydon
Advanced blood pressure monitoring devices contain algorithms that permit estimation of stroke volume (SV). Modelflow (Finapres Medical Systems) is one common method to non-invasively estimate beat-to-beat SV. However, Modelflow accuracy during profound reductions in SV is unclear. We aimed to compare SV estimation by Modelflow and echocardiography, at rest and during orthostatic challenge. We tested
-
Screening for orthostatic hypotension in the geriatric population in a real-world primary care setting reduces prescribed antihypertensive medications. Blood Press. Monit. (IF 1.2) Pub Date : 2023-08-24 Matthew G Kaye, James Rutowski, Hamza Aftab, Rohan Pandey, Raheel Khan, Mohamad A Kalot, Renata Anand, Susan P Graham
To determine if outpatient screening for orthostatic hypotension (OH) in the geriatric population results in fewer prescribed antihypertensive medications and if a relationship exists between OH and specific pharmacologic classes of antihypertensive medications.
-
Central blood pressure obtained by cuff-based oscillometry as a determinant of left ventricular hypertrophy in hypertensive patients. Blood Press. Monit. (IF 1.2) Pub Date : 2023-08-14 Mohamed Aboel-Kassem F Abdelmegid, Doaa A Fouad, Nardeen W M Beshay, Dalia G Mahran, Hamdy Shams-Eddin
Limited and conflicting data have been reported on the prognostic relevance of central blood pressure (CBP) compared with brachial blood pressure (BP) in the anticipation of hypertension-mediated organ damage and the majority of data derived using applanation tonometry with its known complexities. The objective of the present study was to investigate the diagnostic factors of left ventricular hypertrophy
-
Impact of home blood pressure monitors on self-monitoring and control of blood pressure in vulnerable adults. Blood Press. Monit. (IF 1.2) Pub Date : 2023-08-14 Yufei Shi, Carrie McAdam-Marx, Jessica M Downes
To evaluate associations between home blood pressure monitoring (HBPM) and blood pressure (BP) in vulnerable adults during the COVID-19 pandemic, when access to in-person care was restricted. A retrospective cohort study was conducted in adults with hypertension or elevated BP given a home BP monitor vs. usual care. Change in BP from baseline to follow-up was compared between groups, controlling for
-
Factors related to medication adherence in patients with hypertension in Iran: a systematic review study. Blood Press. Monit. (IF 1.2) Pub Date : 2023-08-11 Golara Maleki, Rohollah Norian, Babak Moeini, Majid Barati, Shahnaz Maleki, Maryam Afshari
This study aimed to review the available studies of the factors in Iranian hypertensive patients' adherence to drug treatment.
-
Validation of the JOYTECH DBP-6279B blood pressure monitor in adults and adolescents according to the AAMI/ESH/ISO universal standard(ISO 81060-2:2018 + Amd.1:2020). Blood Press. Monit. (IF 1.2) Pub Date : 2023-07-12 Baojing Duan, Le He, Jianling Zhang, Lixiu Fang, Gang Li
This study aimed to validate the accuracy of DBP-6279B, an automated inflationary oscillometric upper-arm blood pressure (BP) monitor, in the sitting position according to the AAMI/ESH/ISO (81060-2 : 2018 + Amd.1 : 2020) universal standard protocol. SBPs and DBPs were measured simultaneously on the same arm in 88 adults (female : male = 47 : 41) with a mean age of 56.85 years using a mercury sphygmomanometer
-
Validation of the U60EH Wrist Electronic Blood Pressure Monitor in general population according to the ISO 81060-2:2018/AMD 1:2020 Protocol. Blood Press. Monit. (IF 1.2) Pub Date : 2023-07-12 Sicong Mu, Min Zhu, Liping Ma, Lixiu Fang
To evaluate the accuracy of the U60EH Wrist Electronic Blood Pressure Monitor in general population according to the Universal Standard (ISO 81060-2:2018/AMD 1:2020).
-
Evaluation of hypertension control among hemodialysis patients in Pakistan: a prospective follow-up study. Blood Press. Monit. (IF 1.2) Pub Date : 2023-07-06 Kiran Hameed, Sampath Chinnam, Khanzada Sheraz Khan, Saima Mushtaq, Nafees Ahmad, Ammar Ali Saleh Jaber, Zeeshan Anwar, Inayat Ur-Rehman, Amjad Khan
Hypertension is highly prevalent and uncontrolled among hemodialysis patients. In Pakistan published data does not provide enough information about the management and factors associated with uncontrolled hypertension in hemodialysis patients.
-
Cost-utility analysis of diagnostic methods for arterial hypertension in primary care for Brazil: ABPM vs. OBPM vs. HBPM. Blood Press. Monit. (IF 1.2) Pub Date : 2023-06-22 Rosana Lima Garcia, Marcus Tolentino Silva, Amaury Zatorre Amaral, Giovanio Vieira Silva
Arterial hypertension (AH) is diagnosed using three methods: office blood pressure measurement (OBPM), home blood pressure monitoring (HBPM), and ambulatory blood pressure monitoring (ABPM). No economic studies have evaluated the impact of incorporating these strategies for AH diagnosis into the Brazilian public health system.
-
Different time points, different blood pressures: complexity of blood pressure measurement in hemodialysis patients. Blood Press. Monit. (IF 1.2) Pub Date : 2023-06-22 Shan Lan, Yazhen Zhang, Jing Wang, Zhibin Wu, Shanying Chen
We used our established database to investigate predialysis blood pressure (BP) measurements at different time points.
-
Validation of the YuWell YE660D oscillometric upper-arm blood pressure monitor for clinic and home in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard (International Organization for Standardization 81060-2:2018) and the International Organization for Standardization Blood Press. Monit. (IF 1.2) Pub Date : 2023-06-22 Hui-Jie Zhang, Yue Peng, Juan Zhang, Jin Zhang, Li-Ni Teng, Shu-Juan Zhang, De-Jun Zhou, Ming-Zhi Long
To evaluate the accuracy of the YuWell YE660D oscillometric upper-arm blood pressure (BP) monitor in general population (for clinic and home BP measurements in adults) according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2 : 2018) and its Amendment 1. 2020
-
Glyceryl trinitrate-induced blood pressure variability decrease during head-up tilt test predicts vasovagal response. Blood Press. Monit. (IF 1.2) Pub Date : 2023-06-19 Bernard Meglič, Aljoša Danieli
Glyceryl trinitrate (GTN) provoked cardioinhibitory syncope during the head-up tilt test is preceded by a period of disrupted blood pressure variability (BPV). Endogenous nitric oxide (NO) attenuates BPV independently of blood pressure (BP). We hypothesized that exogenous NO donor GTN might decrease BPV during the presyncope period. A decrease in BPV may predict the tilt outcome.
-
Comparison of blood pressure measurements on the bare and sleeved arms - what does it uncover? Blood Press. Monit. (IF 1.2) Pub Date : 2023-06-19 Rotem Tal-Ben Ishay, Adi Leiba, Vladimir Rappoprt, Avital Angel-Korman, Zeev Katzir
Accurate office blood pressure (BP) measurement remains crucial in diagnosing and managing hypertension. In this study, we aimed to compare BP measurements done over a bare arm versus a sleeved arm, while controlling all other possible sources of variance. We collected BP measurements of 100 hypertensive patients visiting a nephrology and hypertension clinic between January 2019 and December 2023.
-
The evidence of effectiveness of isometric resistance training on the management of hypertension in adults: an umbrella review. Blood Press. Monit. (IF 1.2) Pub Date : 2023-06-13 Abbey Wehrmann, Esther J Tian, Elizabeth Liz Tyack, Saravana Kumar
Hypertension is the leading risk factor for cardiovascular disease yet also one of the most readily preventable causes of death. Isometric resistance training (IRT) has gained increasing popularity in recent times as a viable nonpharmacological management option for hypertension. Whilst there have been several reviews on this topic, with varying findings, this umbrella review aimed to summarize the
-
Ambulatory daytime blood pressure versus tonometric blood pressure measurements in the laboratory: effect of posture. Blood Press. Monit. (IF 1.2) Pub Date : 2023-06-08 Emmi Värri, Lauri Suojanen, Jenni K Koskela, Manoj K Choudhary, Antti Tikkakoski, Mika Kähönen, Pasi I Nevalainen, Jukka Mustonen, Ilkka Pörsti
To compare blood pressure (BP) in tonometric radial artery recordings during passive head-up tilt with ambulatory recordings and evaluate possible laboratory cutoff values for hypertension.
-
Waist circumference and blood pressure in Brazilian children. Blood Press. Monit. (IF 1.2) Pub Date : 2023-06-07 Vanessa Heidemann Grawe, Silmara Salete de Barros Silva Mastroeni, Zaíne Glaci Duarte Corrêa, Marco Fabio Mastroeni
We aimed to examine the association between waist circumference and blood pressure (BP) in children at 9 years of age.
-
The relationship between vitamin D and short-term blood pressure variability. Blood Press. Monit. (IF 1.2) Pub Date : 2023-06-07 Amir Baniasad, Abnoos Mokhtari Ardekan, Mohammad Javad Najafzadeh, Fatemeh Mousavi Mehdiabadi
This study aimed to evaluate the relationships between serum levels of 25-hydroxyvitamin D (25(OH)D) and short-term blood pressure variability (BPV) in newly diagnosed hypertensive patients.