Applied Magnetic Resonance ( IF 1.1 ) Pub Date : 2021-06-29 , DOI: 10.1007/s00723-021-01373-8 V. V. Davydov , V. I. Dudkin , N. S. Myazin , R. V. Davydov
The necessity to develop a magnetometer for variations research in the mid-field magnetic strength with a relative error of 10–6 is justified. A new design of nuclear magnetometer which uses maser with flowing liquid is proposed. The block diagram of the magnetometer is presented, and the principle of its operation is described in detail. Furthermore, the conditions of occurrence of maser generation are established. The measurement errors of magnetic field variations in the developed magnetometer design are calculated. The capabilities of the developed magnetometer are determined. Finally, the results of experimental investigations of various variations of magnetic fields are presented.
中文翻译:
基于核磁共振的磁力计新设计研究中场磁场强度的变化
开发磁力计用于中场磁场强度变化研究的必要性是合理的,相对误差为 10 –6。提出了一种利用脉瑟流动液体的核磁力计的新设计。给出了磁力计的框图,并详细描述了其工作原理。此外,还建立了脉泽发生的条件。计算开发的磁力计设计中磁场变化的测量误差。确定开发的磁力计的能力。最后,介绍了各种磁场变化的实验研究结果。