Contamination of dental unit waterlines: assessment of three continuous water disinfection systems
BDJ Open ( IF 2.5 ) Pub Date : 2016-12-16 , DOI: 10.1038/bdjopen.2016.7
Damien Offner 1, 2, 3 , Florence Fioretti 1, 2, 3 , Anne-Marie Musset 2, 3
Affiliation  

Objectives:

To assess the efficacy of three continuous water disinfection systems for dental units under real conditions of dental care.

Design and settings:

A prospective study carried out from 45 days to 20 months on the water microbial quality of the dental units is benefited from three different systems: two chemical treatment systems (IGN EVO/Calbenium/IGN Cartridge and Sterispray) and one physical treatment system (BacTerminator). Studied items were six dental units of the Dental Medicine and Oral Surgery Center within the University Hospital of Strasbourg (HUS), France.

Results and disucussion:

The IGN EVO/Calbenium/IGN Cartridge and Sterispray systems showed an immediate and long-term efficacy on contaminated dental unit waterlines. However, the first system offers ergonomic advantages (automatic system, action on the water from the water supply network). The BacTerminator system took longer to be effective and was less effective than the other two.



中文翻译:


牙科综合治疗台水管污染:三个连续水消毒系统的评估


 目标:


评估牙科治疗台的三种连续水消毒系统在真实牙科护理条件下的功效。

 设计与设置:


一项为期 45 天至 20 个月的关于牙科治疗台水微生物质量的前瞻性研究受益于三种不同的系统:两种化学处理系统(IGN EVO/Calbenium/IGN Cartridge 和 Sterispray)和一种物理处理系统(BacTerminator) 。研究对象是法国斯特拉斯堡大学医院 (HUS) 牙科医学和口腔外科中心的六个牙科单位。

 结果与讨论:


IGN EVO/Calbenium/IGN Cartridge 和 Sterispray 系统对受污染的牙科综合治疗台水管显示出即时和长期的功效。然而,第一个系统具有人体工程学优势(自动系统,作用于来自供水网络的水)。 BacTerminator 系统需要更长的时间才能发挥作用,而且效果不如其他两个系统。

更新日期:2019-05-16
down
wechat
bug