Prospective study to assess the tissue response to HPC-coated p48 flow diverter stents compared to uncoated devices in the rabbit carotid artery model.
European Radiology Experimental ( IF 3.7 ) Pub Date : 2019-12-05 , DOI: 10.1186/s41747-019-0128-z Tim Lenz-Habijan 1 , Pervinder Bhogal 2 , Catrin Bannewitz 1 , Ralf Hannes 1 , Hermann Monstadt 1 , Andreas Simgen 3 , Ruben Mühl-Benninghaus 3 , Wolfgang Reith 3 , Hans Henkes 4, 5
中文翻译:
与兔颈动脉模型中的未涂层器械相比,评估对HPC涂层p48分流器支架的组织反应的前瞻性研究。
更新日期:2019-12-05
European Radiology Experimental ( IF 3.7 ) Pub Date : 2019-12-05 , DOI: 10.1186/s41747-019-0128-z Tim Lenz-Habijan 1 , Pervinder Bhogal 2 , Catrin Bannewitz 1 , Ralf Hannes 1 , Hermann Monstadt 1 , Andreas Simgen 3 , Ruben Mühl-Benninghaus 3 , Wolfgang Reith 3 , Hans Henkes 4, 5
Affiliation
Background
Flow diverters (FDs) are widely used in the treatment of intracranial aneurysms, but the required medication increases the risk of haemorrhagic complications and limits their use in the acute setting. Surface modified FDs may limit the need for dual antiplatelet therapy (DAPT). Hydrophilic polymer coating (HPC) may reduce the need of medication.Methods
This explorative study, approved by the local authorities and the local welfare committee, compared stent behaviour and overall tissue response between HPC-coated FDs and uncoated FDs, both implanted into the common carotid arteries of eight New Zealand white rabbits. Endothelialisation, inflammatory response, and performance during implantation were assessed. Angiographic follow-up was performed to observe the patency of the devices after implantation and after 30 days. Histological examinations were performed at 30 days to assess foreign body reaction and endothelialisation. Kruskal-Wallis and Wilcoxon tests were used to compare non-parametric variables.Results
Angiography showed that both coated and uncoated FDs performed well during implantation. All devices remained patent during immediate follow-up and after 30 days. Histopathology showed no significant difference in inflammation within the vessel wall between the two cohorts (2.12 ± 0.75 vs. 1.96 ± 0.79, p = 0.7072). Complete endothelialisation of the stent struts was seen with very similar (0.04 ± 0.02 mm vs. 0.04 ± 0.03 mm, p = 0.892) neoendothelial thickness between the two cohorts after 30 days.Conclusion
Taking into account the limitation in sample size, non-significant differences between the HPC-coated and uncoated FDs regarding implantation, foreign body response, and endothelialisation were found.中文翻译: