Validation of a commercial human ELISA to measure hyaluronic acid concentration in feline plasma
The Journal of Veterinary Diagnostic Investigation ( IF 1.2 ) Pub Date : 2021-09-13 , DOI: 10.1177/10406387211044718
Kaela E Shaw 1 , Alexa M Bersenas 1 , Shane W Bateman 1 , Shauna L Blois 1 , R Darren Wood 2
Affiliation  

Our goal was to validate a human hyaluronic acid (HA) ELISA (Hyaluronic acid plus ELISA; TECOmedical Group) for use in feline plasma. Plasma from 5 healthy cats and 5 critically ill cats was used for validation of the assay. Validation methods performed included intra- and inter-assay variability, spike-and-recovery, and dilutional linearity. All measurements were performed in duplicate. The precision study revealed good intra-assay CV of 7.4–8.9%; inter-assay CV was 3.4–4.2%. Extraction efficiency via spiking tests yielded mean recovery of 89.6%. The assay met criteria for acceptable linearity using 3 serial dilutions. Our results demonstrate that this commercial HA ELISA had acceptable analytical performance using feline plasma and could be a useful tool in the veterinary clinical research setting.