An upgraded lab-based method to determine natural γ-ray emitters in NORM samples by using Ge detectors
Measurement ( IF 5.6 ) Pub Date : 2021-09-11 , DOI: 10.1016/j.measurement.2021.110153
A. Barba-Lobo 1 , F. Mosqueda 1 , J.P. Bolívar 1
Affiliation  

The regulation existing on NORM activities requires the accurate and quick measurement of natural radionuclides in a wide range of matrices by gamma spectrometry. The aim of this work has been to develop a comprehensive and accurate methodology to determine natural long-lived radionuclides (210Pb, 228,226Ra, 234,228Th and 40K) by gamma spectrometry with Ge detectors in NORM samples. An exhaustive calibration method to obtain the full-energy peak efficiency (FEPE), for each gamma emission energy, as a function of the thickness, apparent density and matrix composition of the cylindrical sample, has been developed. The selected certified calibration standards contain only natural radionuclides belonging to the 238U- and 232Th-series as well as 40K (codes RGU-1, RGTh-1 and RGK-1 from IAEA). Then, the obtained FEPE curves were validated using Genie 2000 simulations and Certified Reference Materials, for which the self-attenuation corrections were considered. Finally, a study on 222Rn losses was done for several samples.



中文翻译:

使用 Ge 探测器确定 NORM 样品中天然 γ 射线发射体的升级版实验室方法

现有的关于 NORM 活动的法规要求通过伽马能谱法准确、快速地测量各种基质中的天然放射性核素。这项工作的目的是开发一种全面而准确的方法,通过伽马光谱法和GE探测器在 NORM 样品中测定天然长寿命放射性核素(210 Pb、228,226 Ra、234,228 Th 和40 K)。已经开发了一种详尽的校准方法,用于获得每个伽马发射能量的全能峰值效率 ( FEPE ),作为圆柱形样品的厚度、表观密度和基体组成的函数。选定的认证校准标准只包含属于238 U系列和232 Th 系列以及40 K(来自 IAEA 的代码 RGU-1、RGTh-1 和 RGK-1)。然后,使用 Genie 2000 模拟和认证参考材料验证获得的FEPE曲线,其中考虑了自衰减校正。最后,对几个样品进行了222 Rn 损失的研究。

更新日期:2021-09-22
down
wechat
bug