Accuracy analysis on concentration determination of molecular iodine in gamma irradiated solutions
Journal of Radioanalytical and Nuclear Chemistry ( IF 1.6 ) Pub Date : 2021-07-27 , DOI: 10.1007/s10967-021-07912-8
Minsik Kim 1 , Jei-Won Yeon 1
Affiliation  

The gamma irradiated aqueous samples containing I2 and I were measured by two different analysis methods using an UV–VIS spectrophotometer. One is the I2 concentration was determined by using the equilibrium relationship between I3 and I. The other is I2 concentration was measured after extracted by toluene. In the I2 and total I measurements, both methods were confirmed to have an accuracy within 4%. The I2 concentrations measured by the toluene extraction method were always lower for the identical samples than those measured by the equilibrium relationship method. This was understood as a loss occurring in the toluene extraction step.



中文翻译:

γ辐照溶液中分子碘浓度测定的准确度分析

使用紫外-可见分光光度计通过两种不同的分析方法测量含有 I 2和 I -的伽马辐射水样。一种是通过使用I 3 -和I -之间的平衡关系来确定I 2浓度。另一种是甲苯萃取后测定的I 2浓度。在I 2和总I 测量中,两种方法的准确度均在4% 以内。我2对于相同的样品,甲苯萃取法测得的浓度总是低于平衡关系法测得的浓度。这被理解为甲苯萃取步骤中发生的损失。