Medical Physics ( IF 3.8 ) Pub Date : 2021-07-05 , DOI: 10.1002/mp.15080
Jasper N Smit 1 , Koert F D Kuhlmann 1 , Bart R Thomson 1 , Niels F M Kok 1 , Matteo Fusaglia 1 , Theo J M Ruers 1, 2
Affiliation  

Registration of pre- and intraoperative images is a crucial step of surgical liver navigation, where rigid registration of vessel centerlines is currently commonly used. When using 3D ultrasound (US), accuracy during navigation might be influenced by the size of the intraoperative US volume, yet the relationship between registration accuracy and US volume size is understudied. In this study, we specify an optimal 3D US volume size for registration using varying volumes of liver vasculature. While previous studies measured accuracy at registered fiducials, in this work, accuracy is determined at the target lesion which is clinically the most relevant structure.