Intrinsic detector sensitivity analysis as a tool to characterize ArcCHECK and EPID sensitivity to variations in delivery for lung SBRT VMAT plans
Journal of Applied Clinical Medical Physics ( IF 2.0 ) Pub Date : 2021-05-05 , DOI: 10.1002/acm2.13221
Thahabah Alharthi 1, 2, 3, 4 , Phil Vial 3, 4, 5 , Lois Holloway 1, 3, 4, 5, 6 , David Thwaites 1
Affiliation  

To investigate intrinsic sensitivity of an electronic portal imaging device (EPID) and the ArcCHECK detector and to use this in assessing their performance in detecting delivery variations for lung SBRT VMAT. The effect of detector spatial resolution and dose matrix interpolation on the gamma pass rate was also considered.

中文翻译:

内在检测器灵敏度分析作为一种工具来表征 ArcCHECK 和 EPID 对肺部 SBRT VMAT 计划交付变化的敏感性

研究电子入口成像设备 (EPID) 和 ArcCHECK 检测器的内在灵敏度,并使用它来评估它们在检测肺 SBRT VMAT 输送变化方面的性能。还考虑了探测器空间分辨率和剂量矩阵插值对伽马通过率的影响。