Study of after-pulses in the 20-inch HQE-MCP-PMT for the JUNO experiment
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment ( IF 1.5 ) Pub Date : 2021-04-16 , DOI: 10.1016/j.nima.2021.165351
Qi Wu , Sen Qian , Lishuang Ma , Jingkai Xia , Bayarto K. Lubsandorzhiev , Zhigang Wang , Yao Zhu , Haitao Li , Nikita Ushakov , Andrei Sidorenkov , Qianyu Hu , Jianning Sun , Shuguang Si

This manuscript reports a study of the after-pulse properties of different types of 20-inch photomultiplier tubes (PMTs), especially for the 20-inch microchannel-plate-PMTs (MCP-PMTs) for JUNO experiment. In the study, the after-pulse rate (APR) is calculated from PMT waveforms obtained by a CAEN DT5751 waveform digitizer. We tested the after-pulse characteristics of both Dynode-PMTs and MCP-PMTs and made a thorough study on the after-pulses of high quantum efficiency (HQE) MCP-PMT for JUNO experiment, including their time distribution, after-pulse rate and the causes of different after-pulse groups through flight time analysis. Based on a fast test method, individual tests on 28 HQE-MCP-PMTs on the after-pulses were conducted at the North Night Vision Tech. Ltd. (NNVT) in Nanjing in December of 2019. This paper reports the average APR under two different definitions for HQE-MCP-PMTs and the proportions of the five different after-pulse groups.



中文翻译:

用于JUNO实验的20英寸HQE-MCP-PMT中的后脉冲研究

该手稿报告了对不同类型的20英寸光电倍增管(PMT),特别是用于JUNO实验的20英寸微通道板-PMT(MCP-PMT)的脉冲后特性的研究。在这项研究中,从CAEN DT5751波形数字化仪获得的PMT波形计算出后脉冲率(APR)。我们测试了Dynode-PMT和MCP-PMT的脉冲后特性,并对JUNO实验的高量子效率(HQE)MCP-PMT脉冲后进行了深入研究,包括它们的时间分布,脉冲后速率和通过飞行时间分析了解不同脉冲群的原因。基于快速测试方法,在North Night Vision Tech上对28个HQE-MCP-PMT的后脉冲进行了单独测试。有限公司(NNVT)于2019年12月在南京。

更新日期:2021-04-23
down
wechat
bug