Stabilization Features of Mercury Capillary Lamp Radiation in the Mercury Vapor Analyzer Based on Zeeman Atomic Absorption Spectroscopy
Russian Physics Journal ( IF 0.4 ) Pub Date : 2021-04-12 , DOI: 10.1007/s11182-021-02273-x
A. I. Abramochkin , V. V. Tatur , A. A. Tikhomirov

The possibility of increasing the sensitivity of the mercury vapor analyzer in ambient air using a capillary lamp with a natural mercury isotopic mixture in the presence of the transverse Zeeman effect as a radiation source has been experimentally studied. Lamp radiation is stabilized by separate control of temperatures of the capillary and lamp bulb. The obtained sensitivity of the analyzer reached 14 ng/m3 during more than 40 min.