Optimal Dataset Sizes for Constant Optimization in Published Theoretical Optical Formulae
Current Eye Research ( IF 1.7 ) Pub Date : 2021-08-24 , DOI: 10.1080/02713683.2021.1900272
Achim Langenbucher 1 , Michael Schwemm 1 , Timo Eppig 1 , Simon Schröder 1 , Alan Cayless 2 , Nóra Szentmáry 3, 4
Affiliation  

ABSTRACT

Purpose: To determine the optimal number of data points required for optimization of formulae for classical lens power calculation.

Methods: A large dataset of preoperative biometric values was used to assess the convergence of formula constants in a number of established intraocular lens power calculation formulae.

Results: In formulae with a single constant, 80–100 clinical data points are sufficient to obtain convergence. The Haigis formula (three constants) requires 200–300 data points although refractive error converges more rapidly.

Conclusions: In all formulae, 80–100 clinical data points are sufficient to achieve a stable mean refractive error.