Whole anterior segment and retinal swept source OCT for comprehensive ocular screening
Biomedical Optics Express ( IF 2.9 ) Pub Date : 2021-02-05 , DOI: 10.1364/boe.414592
Ana Rodríguez-Aramendía , Fernando Diaz-Douton , José Fernández-Trullàs , Pol Falgueras , Laura Gonzalez , Jaume Pujol , Ireneusz Grulkowski , Jose Güell

Whole eye visualization and morphometry are of high relevance in clinical practice. However, most standard ophthalmic OCT instruments are dedicated either to retinal or to anterior segment imaging. We demonstrate a swept source optical coherence tomography system (SS-OCT) that images both the whole anterior segment and the retina alternately using a single source and detector. A pilot population was imaged with the proof of concept prototype. We demonstrate the clinical potential of whole eye OCT screening for the description and early detection of relevant clinical features in the anterior segment and retina of several patients.

中文翻译:

整个前节段和视网膜扫描源OCT进行全面的眼部筛查

全眼可视化和形态测量在临床实践中具有高度相关性。但是,大多数标准眼科OCT仪器专用于视网膜或前段成像。我们展示了一种扫频光源光学相干断层扫描系统(SS-OCT),该系统使用单个光源和检测器交替成像整个前节和视网膜。用概念证明原型对试点人群进行了成像。我们证明了全眼OCT筛查的临床潜力,可用于描述和早期检测几名患者的前节段和视网膜中的相关临床特征。