Impact of prompt gamma emission of 44Sc on quantification in preclinical and clinical PET systems
Applied Radiation and Isotopes ( IF 1.6 ) Pub Date : 2021-01-21 , DOI: 10.1016/j.apradiso.2021.109599
Florian Rosar 1 , Hendrik Bohnenberger 2 , Euy Sung Moon 3 , Frank Rösch 3 , Achim Denig 4 , Daniel Vincenz-Zörner 5 , Manuela A Hoffmann 6 , Fadi Khreish 2 , Samer Ezziddin 2 , Mathias Schreckenberger 6 , Hans-Georg Buchholz 6 , Andrea Schaefer-Schuler 2
Affiliation  

44Sc is an increasingly investigated positron emitter for use in positron emission tomography (PET) imaging. However, 44Sc is a non-pure positron emitter, since prompt photons are co-emitted during the decay process. This study investigates coincidence energy spectra of 44Sc and its impact on PET quantification on a preclinical and clinical PET system in comparison with 18F. The raw data of the coincidence events revealed characteristic differences comparing the photon energy distribution of 44Sc and 18F. Due to prompt gamma emission of 44Sc, activity recovery is underestimated on PET systems. However, clinical PET imaging of 44Sc with acceptable quantitative accuracy appears feasible by using a single, constant correction factor.