Task group 284 report: magnetic resonance imaging simulation in radiotherapy: considerations for clinical implementation, optimization, and quality assurance
Medical Physics ( IF 3.2 ) Pub Date : 2021-01-01 , DOI: 10.1002/mp.14695
Carri K Glide-Hurst 1 , Eric S Paulson 2 , Kiaran McGee 3 , Neelam Tyagi 4 , Yanle Hu 5 , James Balter 6 , John Bayouth 1
Affiliation  

The use of dedicated magnetic resonance simulation (MR-SIM) platforms in Radiation Oncology has expanded rapidly, introducing new equipment and functionality with the overall goal of improving the accuracy of radiation treatment planning. However, this emerging technology presents a new set of challenges that need to be addressed for safe and effective MR-SIM implementation. The major objectives of this report are to provide recommendations for commercially available MR simulators, including initial equipment selection, siting, acceptance testing, quality assurance, optimization of dedicated radiation therapy specific MR-SIM workflows, patient-specific considerations, safety, and staffing. Major contributions include guidance on motion and distortion management as well as MRI coil configurations to accommodate patients immobilized in the treatment position. Examples of optimized protocols and checklists for QA programs are provided. While the recommendations provided here are minimum requirements, emerging areas and unmet needs are also highlighted for future development.

中文翻译:

任务组 284 报告:放射治疗中的磁共振成像模拟:临床实施、优化和质量保证的考虑因素

放射肿瘤学中专用磁共振模拟 (MR-SIM) 平台的使用迅速扩大,引入了新的设备和功能,总体目标是提高放射治疗计划的准确性。然而,这项新兴技术提出了一系列新的挑战,需要解决这些挑战才能安全有效地实施 MR-SIM。本报告的主要目标是为商用 MR 模拟器提供建议,包括初始设备选择、选址、验收测试、质量保证、专用放射治疗特定 MR-SIM 工作流程的优化、患者特定注意事项、安全性和人员配备。主要贡献包括关于运动和变形管理的指导以及 MRI 线圈配置,以适应固定在治疗位置的患者。提供了 QA 程序的优化协议和清单示例。虽然此处提供的建议是最低要求,但也强调了未来发展的新兴领域和未满足的需求。
更新日期:2021-01-01
down
wechat
bug