Reduction of contaminants originating from primary beam by improving the beam stoppers in GARIS-II
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment ( IF 1.5 ) Pub Date : 2020-12-31 , DOI: 10.1016/j.nima.2020.164996 S. Kimura , D. Kaji , Y. Ito , H. Miyatake , K. Morimoto , P. Schury , M. Wada
中文翻译:
通过改进GARIS-II中的束塞,减少源自初级束的污染物
更新日期:2021-01-18
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment ( IF 1.5 ) Pub Date : 2020-12-31 , DOI: 10.1016/j.nima.2020.164996 S. Kimura , D. Kaji , Y. Ito , H. Miyatake , K. Morimoto , P. Schury , M. Wada
Two independent beam stoppers have been developed for improving the beam separation of the gas-filled recoil ion separator GARIS-II. Performance of these supplemental beam stoppers was investigated by using the 208Pb () 223Th reaction. A160-fold enhancement of the signal-to-noise ratio at the GARIS-II focal plane was observed.