A Chemiluminescent Immunoassay for Osteocalcin in Human Serum and a Solution to the “Hook Effect”
Journal of Analytical Methods in Chemistry ( IF 2.3 ) Pub Date : 2020-12-09 , DOI: 10.1155/2020/8891437
Shuang Han 1, 2, 3 , Yifeng Xue 4 , Junlan Zhang 1, 2, 3 , Jianrong Huang 4 , Xiuxia Liu 1, 2, 3 , Yankun Yang 1, 2, 3 , Zhonghu Bai 1, 2, 3 , Chunxin Wang 4
Affiliation  

A chemiluminescent immunoassay for human serum osteocalcin, or bone Gla protein, was established using a double-antibody sandwich model. Examination of the hook effect revealed that it was significantly reduced, and no hook effect was observed at an osteocalcin concentration of 4000 ng/mL. The established method showed good analytical performance and thermal stability. The limit of detection was 0.03 ng/mL. The intra-assay coefficient of variation (CV) was 3.22%–5.64%, the interassay CV was 4.42%–7.25%, and the recovery rate was 93.22%–107.99%. Cross-reactivity (CR) was not observed with bovine, rat, mouse, rabbit, or porcine samples. The developed method showed a good correlation with the N-MID product from Roche. In total, 1069 clinical patient samples were measured using the reagent kit developed in this study.

中文翻译:

人血清中骨钙素的化学发光免疫分析及“钩效应”的解决方案

使用双抗体夹心模型建立了针对人血清骨钙素或骨Gla蛋白的化学发光免疫分析法。钩效应的检查显示其显着降低,并且在骨钙素浓度为4000 ng / mL时未观察到钩效应。建立的方法显示出良好的分析性能和热稳定性。检测下限为0.03 ng / mL。批内变异系数(CV)为3.22%-5.64%,批间变异系数为4.42%-7.25%,回收率为93.22%-107.99%。在牛,大鼠,小鼠,兔或猪的样品中未观察到交叉反应性(CR)。所开发的方法与罗氏公司的N-MID产品具有良好的相关性。使用本研究开发的试剂盒,总共测量了1069个临床患者样品。