当前位置: X-MOL 学术Instrum. Exp. Tech. › 论文详情
Our official English website, www.x-mol.net, welcomes your feedback! (Note: you will need to create a separate account there.)
Calibration of Lateral Scanning in Optical Coherence Tomography Devices
Instruments and Experimental Techniques ( IF 0.4 ) Pub Date : 2020-09-23 , DOI: 10.1134/S0020441220040296
S. Yu. Ksenofontov , A. A. Moiseev , V. A. Matkivsky , P. A. Shilyagin , T. V. Vasilenkova , V. M. Gelikonov , G. V. Gelikonov

Methods for determining the dependence of the lateral-scan coordinate on the A-scan number from tomographic images of test samples are investigated and techniques for compensating for horizontal distortions of tomographic images caused by uneven speed of the probing-beam displacement are proposed. An opal-glass Ronchi ruling and an inclined flat surface of the volume-scattering plate were used as test samples .

中文翻译:

光学相干断层扫描设备中的横向扫描校准

研究了从测试样品的断层图像确定横向扫描坐标对A扫描次数的依赖性的方法,并提出了补偿由于探测光束位移速度不均匀而导致的断层图像水平失真的技术。使用蛋白石玻璃Ronchi规则和体积散射板的倾斜平面作为测试样品
更新日期:2020-09-23
down
wechat
bug