Calibration of Lateral Scanning in Optical Coherence Tomography Devices
Instruments and Experimental Techniques ( IF 0.4 ) Pub Date : 2020-09-23 , DOI: 10.1134/S0020441220040296 S. Yu. Ksenofontov , A. A. Moiseev , V. A. Matkivsky , P. A. Shilyagin , T. V. Vasilenkova , V. M. Gelikonov , G. V. Gelikonov
Instruments and Experimental Techniques ( IF 0.4 ) Pub Date : 2020-09-23 , DOI: 10.1134/S0020441220040296 S. Yu. Ksenofontov , A. A. Moiseev , V. A. Matkivsky , P. A. Shilyagin , T. V. Vasilenkova , V. M. Gelikonov , G. V. Gelikonov
Methods for determining the dependence of the lateral-scan coordinate on the A-scan number from tomographic images of test samples are investigated and techniques for compensating for horizontal distortions of tomographic images caused by uneven speed of the probing-beam displacement are proposed. An opal-glass Ronchi ruling and an inclined flat surface of the volume-scattering plate were used as test samples
.