当前位置: X-MOL 学术Int. J. Appl. Electromagn. Mech. › 论文详情
Our official English website, www.x-mol.net, welcomes your feedback! (Note: you will need to create a separate account there.)
Design and development of wireless power transfer system for implantable cardioverter defibrillator
International Journal of Applied Electromagnetics and Mechanics ( IF 1.1 ) Pub Date : 2020-10-05 , DOI: 10.3233/jae-201524
K. Aditya 1 , Phaneendra Babu Bobba 1 , Y. Satyavani 1
Affiliation  

Wireless power transfer system is playing an important role in biomedical applications by transferring power wirelessly to the implants present in the body. In this paper, magnetic coupling resonance wireless power transfer system is used to recharge battery of implantable cardioverter defibrillator. The proposed system operates at 300 KHz frequency. The LCC-C compensation topology is used to tune the coils for operating the system at stable resonant frequency and to obtain higher efficiency. In this paper, we have investigated the performance of different coil structures in multiple layer configurations and we offer recommendations on the best suitable configuration for effective transfer of power.

中文翻译:

植入式心脏复律除颤器无线功率传输系统的设计与开发

无线功率传输系统通过将功率无线传输到体内的植入物,在生物医学应用中发挥着重要作用。本文采用磁耦合共振无线电力传输系统对植入式心脏复律除颤器的电池进行充电。提议的系统以300 KHz的频率运行。LCC-C补偿拓扑用于调整线圈,以使系统以稳定的谐振频率工作并获得更高的效率。在本文中,我们研究了多层结构中不同线圈结构的性能,并针对有效传输功率的最佳配置提供了建议。
更新日期:2020-10-07