Water calorimetry in MR-linac: Direct measurement of absorbed dose and determination of chamber KQ mag.
Medical Physics ( IF 3.2 ) Pub Date : 2020-09-24 , DOI: 10.1002/mp.14468
Mark D'Souza 1 , Humza Nusrat 2 , Viktor Iakovenko 2 , Brian Keller 2 , Arjun Sahgal 2 , James Renaud 3, 4 , Arman Sarfehnia 1, 2, 5
Affiliation  

To use a portable 4°C cooled MR‐compatible water calorimeter to measure absorbed dose in a magnetic resonance‐guided radiation therapy (MRgRT) system. Furthermore, to use the calorimetric dose results and direct cross‐calibration to experimentally measure the combined beam quality and magnetic field correction factor (urn:x-wiley:00942405:media:mp14468:mp14468-math-0002) of a clinically used reference‐class ionization chamber placed under the same radiation field.