Thermoluminescence properties of potassium fluoride
Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms ( IF 1.4 ) Pub Date : 2020-09-24 , DOI: 10.1016/j.nimb.2020.09.009
F.O. Ogundare , D.E. Folley , M.L. Chithambo , T.O. Arise

This study is designed to examine the thermoluminescence characteristics of potassium fluoride for possible use as a thermoluminescence dosemeter. Thermoluminescence measurements were carried out at doses up to 20 Gy and heating rates between 0.2 and 4 °C/s. The glow curve of the fluoride, readout at 1°Cs−1, exhibited two peaks at 130 and 250 °C. In addition, two shoulder peaks appeared at 70 and 200 °C. The kinetic parameters of the peaks were determined using glow curve deconvolution, peak shape and variable heating rate methods. The intensity of the 130 °C peak exhibited linearity at low doses and sublinearity at high doses. The dose response of the 250 °C peak is supralinear at low doses but becomes linear with further increase in dose. The intensity of 130 °C peak is affected by thermal quenching. Glow curve deconvolution indicated the presence of four peaks in the glow curve.