A low-noise current-sensitive preamplifier for X-ray computed tomography with applying a charge-sensitive preamplifier
Journal of Nuclear Science and Technology ( IF 1.5 ) Pub Date : 2020-09-07 , DOI: 10.1080/00223131.2020.1806136 Ikuo Kanno 1 , Hideaki Onabe 2
Journal of Nuclear Science and Technology ( IF 1.5 ) Pub Date : 2020-09-07 , DOI: 10.1080/00223131.2020.1806136 Ikuo Kanno 1 , Hideaki Onabe 2
Affiliation
ABSTRACT A low-noise current-sensitive preamplifier was developed for aiming low-dose X-ray computed tomography with applying a charge-sensitive preamplifier. We compared the number of X-ray photons needed for current measurement with the new preamplifier with that needed with a conventional preamplifier. Also, the dose rates for above conditions were measured by a commercial dose rate meter. Results show that the new preamplifier reduced the number of X-ray photons needed by two orders of magnitude compared with the conventional one.
中文翻译:
一种应用电荷敏感前置放大器的用于 X 射线计算机断层扫描的低噪声电流敏感前置放大器
摘要开发了一种低噪声电流敏感前置放大器,用于通过应用电荷敏感前置放大器来瞄准低剂量 X 射线计算机断层扫描。我们将使用新型前置放大器进行电流测量所需的 X 射线光子数量与使用传统前置放大器所需的 X 射线光子数量进行了比较。此外,上述条件下的剂量率是通过商业剂量率计测量的。结果表明,与传统前置放大器相比,新型前置放大器减少了两个数量级所需的 X 射线光子数量。