Comparison of central, peripheral, and weighted size-specific dose in CT.
Journal of X-Ray Science and Technology ( IF 3 ) Pub Date : 2020-05-20 , DOI: 10.3233/xst-200667
Choirul Anam 1 , Dwi Adhianto 1 , Heri Sutanto 1 , Kusworo Adi 1 , Mohd Hanafi Ali 2 , William Ian Duncombe Rae 2 , Toshioh Fujibuchi 3 , Geoff Dougherty 4
Affiliation  

The objective of this study is to determine X-ray dose distribution and the correlation between central, peripheral and weighted-centre peripheral doses for various phantom sizes and tube voltages in computed tomography (CT). We used phantoms developed in-house, with various water-equivalent diameters (Dw) from 8.5 up to 42.1 cm. The phantoms have one hole in the centre and four holes at the periphery. By using these five holes, it is possible to measure the size-specific central dose (Ds,c), peripheral dose (Ds,p), and weighted dose (Ds,w).The phantoms are scanned using a CT scanner (Siemens Somatom Definition AS), with the tube voltage varied from 80 up to 140 kVps. The doses are measured using a pencil ionization chamber (Ray safe X2 CT Sensor) in every hole for all phantoms. The relationships between Ds,c, Ds,p, and Ds,w, and the water-equivalent diameter are established. The size-conversion factors are calculated. Comparisons between Ds,c, Ds,p, and Ds,ware also established. We observe that the dose is relatively homogeneous over the phantom for water-equivalent diameters of 12–14 cm. For water-equivalent diameters less than 12 cm, the dose in the centre is higher than at the periphery, whereas for water-equivalent diameters greater than 14 cm, the dose at the centre is lower than that at the periphery. We also find that the distribution of the doses is influenced by the tube voltage. These dose distributions may be useful for calculating organ doses for specific patients using their CT images in future clinical practice.

中文翻译:

CT 中中心、外周和加权大小特异性剂量的比较。

本研究的目的是确定 X 射线剂量分布以及计算机断层扫描 (CT) 中各种体模尺寸和管电压的中心、外周和加权中心外周剂量之间的相关性。我们使用内部开发的体模,具有从 8.5 到 42.1 厘米的各种水当量直径 (Dw)。幻影在中心有一个孔,在外围有四个孔。通过使用这五个孔,可以测量特定尺寸的中心剂量 (Ds,c)、周边剂量 (Ds,p) 和加权剂量 (Ds,w)。使用 CT 扫描仪 (Siemens Somatom 定义 AS),管电压从 80 到 140 kVps 不等。使用铅笔电离室(射线安全 X2 CT 传感器)在所有体模的每个孔中测量剂量。Ds,c, Ds,p 和 Ds,w 之间的关系,并确定水当量直径。计算尺寸转换因子。还建立了 Ds,c、Ds,p 和 Ds,ware 之间的比较。我们观察到,对于 12-14 cm 的水当量直径,剂量在体模上相对均匀。当水当量直径小于 12 cm 时,中心剂量高于周边,而当水当量直径大于 14 cm 时,中心剂量低于周边剂量。我们还发现剂量的分布受管电压的影响。这些剂量分布可能有助于在未来的临床实践中使用他们的 CT 图像计算特定患者的器官剂量。我们观察到,对于 12-14 cm 的水当量直径,剂量在体模上相对均匀。当水当量直径小于 12 cm 时,中心剂量高于周边,而当水当量直径大于 14 cm 时,中心剂量低于周边剂量。我们还发现剂量的分布受管电压的影响。这些剂量分布可能有助于在未来的临床实践中使用他们的 CT 图像计算特定患者的器官剂量。我们观察到,对于 12-14 cm 的水当量直径,剂量在体模上相对均匀。当水当量直径小于 12 cm 时,中心剂量高于周边,而当水当量直径大于 14 cm 时,中心剂量低于周边剂量。我们还发现剂量的分布受管电压的影响。这些剂量分布可能有助于在未来的临床实践中使用他们的 CT 图像计算特定患者的器官剂量。中心剂量低于周边剂量。我们还发现剂量的分布受管电压的影响。这些剂量分布可能有助于在未来的临床实践中使用他们的 CT 图像计算特定患者的器官剂量。中心剂量低于周边剂量。我们还发现剂量的分布受管电压的影响。这些剂量分布可能有助于在未来的临床实践中使用他们的 CT 图像计算特定患者的器官剂量。
更新日期:2020-06-30
down
wechat
bug