Introduction of a Measurement Setup to Monitor the Pressure Applied During Handheld Ultrasound Elastography.
Ultrasound in Medicine & Biology ( IF 2.4 ) Pub Date : 2020-06-15 , DOI: 10.1016/j.ultrasmedbio.2020.04.017
Katrin Skerl 1 , Bernd Eichhorn 2 , Romans Poltorjanoks 3 , Sandy Cochran 3 , Andrew Evans 4
Affiliation  

Shear-wave elastography may produce misleadingly high values if too much pressure is applied during the imaging process. However, in clinical routine there is presently no way to monitor the pressure applied during the measurements. In this work we introduce a novel measurement setup which can directly be attached to an ultrasonic imaging transducer and allows observation of the applied pressure in real time. The setup supports free-hand imaging according to the clinical standard. We tested the setup by carrying out shear-wave elastography under varying pressures on ex vivo animal tissue. The values increased with pressure, as was expected. Thus, the setup is a possible solution for measuring applied pressure in real time.



中文翻译:

引入测量设置以监测手持式超声弹性成像期间施加的压力。

如果在成像过程中施加过大的压力,剪切波弹性成像可能会产生误导性的高值。然而,在临床常规中,目前没有办法监测在测量期间施加的压力。在这项工作中,我们引入了一种新颖的测量装置,它可以直接连接到超声成像换能器上,并允许实时观察施加的压力。根据临床标准,该设置支持徒手成像。我们通过在离体动物组织的不同压力下进行剪切波弹性成像来测试设置。正如预期的那样,这些值随着压力而增加。因此,该设置是实时测量施加压力的可能解决方案。