Sensitivity and specificity of a commercial ELISA test for anti-MAG antibodies in patients with neuropathy
Journal of Neuroimmunology ( IF 3.3 ) Pub Date : 2020-08-01 , DOI: 10.1016/j.jneuroim.2020.577288
Giuseppe Liberatore 1 , Claudia Giannotta 1 , Blesson Punnen Sajeev 1 , Emanuela Morenghi 2 , Fabrizia Terenghi 1 , Francesca Gallia 1 , Pietro Emiliano Doneddu 1 , Fiore Manganelli 3 , Dario Cocito 4 , Massimiliano Filosto 5 , Giovanni Antonini 6 , Giuseppe Cosentino 7 , Girolama Alessandra Marfia 8 , Angelo Maurizio Clerici 9 , Giuseppe Lauria 10 , Tiziana Rosso 11 , Guido Cavaletti 12 , Eduardo Nobile-Orazio 13
Affiliation  

For the diagnosis of anti-MAG polyneuropathy the commercial ELISA manufacturer currently recommends a cut-off of 1000 Bühlmann Titer Units (BTU). We analyzed sera from 80 anti-MAG neuropathy patients and 383 controls (with other neuropathies or healthy controls) to assess the ELISA sensitivity and specificity at different thresholds. A better combination of sensitivity/specificity was found at a threshold >1500 BTU than at >1000 BTU. The best value of specificity was obtained at threshold >7000 BTU. There was a diagnostic grey area between 1500 and 7000 BTU in which the clinical phenotypes as well as electrophysiological studies need to be carefully assessed particularly to differentiate CIDP and anti-MAG neuropathy.

中文翻译:

用于神经病患者抗 MAG 抗体的商业 ELISA 检测的敏感性和特异性

对于抗 MAG 多发性神经病的诊断,商业 ELISA 制造商目前建议临界值为 1000 Bühlmann Titer Units (BTU)。我们分析了 80 名抗 MAG 神经病患者和 383 名对照(与其他神经病或健康对照)的血清,以评估不同阈值下的 ELISA 敏感性和特异性。在阈值 >1500 BTU 时比在 >1000 BTU 时发现了更好的敏感性/特异性组合。在阈值 >7000 BTU 时获得最佳特异性值。在 1500 到 7000 BTU 之间存在诊断灰色区域,需要仔细评估临床表型和电生理学研究,特别是要区分 CIDP 和抗 MAG 神经病变。