当前位置: X-MOL 学术J. Instrum. › 论文详情
Our official English website, www.x-mol.net, welcomes your feedback! (Note: you will need to create a separate account there.)
Design and implementation of a time measurement system for Micromegas detector based on the TPC-like method
Journal of Instrumentation ( IF 1.3 ) Pub Date : 2020-05-25 , DOI: 10.1088/1748-0221/15/05/p05020
X.X. Wang 1, 2 , F. Li 1, 2 , Z.L. Zhang 1, 2 , Z.Y. Zhang 1, 2 , L. Han 1, 2 , G. Jin 1, 2
Affiliation  

The Micromegas detector is widely used in particle physics experiments due to its high count rate, good spatial resolution and other characteristics. In this article, we developed a drift time measurement system to reconstruct the hit position through the μTPC method. Our system is expected to run on a 768-channel prototype Micromegas detector being developed at the University of Science and Technology of China. It consists of two main parts: the data collection board receiving the trigger signal from the trigger detector to record the trigger time and the front end electronics receiving hit signals from the Micromegas detector to record the stop drift time of the ionized electrons through the VMM3 chip module. The electronics test results show that the system can measure time with a resolution of less than 1 ns. With this system, it is possible to reconstruct hit positions with large incident angles (for angles above 15 o ) in a Micromegas de...
更新日期:2020-05-25