当前位置: X-MOL 学术Sci. Rep. › 论文详情
Our official English website, www.x-mol.net, welcomes your feedback! (Note: you will need to create a separate account there.)
Instrumentation for quantitative analysis of volatile compounds emission at elevated temperatures. Part 1: Design and implementation.
Scientific Reports ( IF 3.8 ) Pub Date : 2020-05-26 , DOI: 10.1038/s41598-020-65472-5
Célia Lourenço 1 , Sarah Bergin 1, 2 , Jane Hodgkinson 1 , Daniel Francis 1 , Stephen E Staines 1 , John R Saffell 3 , Christopher Walton 4 , Ralph P Tatam 1
Affiliation  

A novel suite of instrumentation for the characterisation of materials held inside an air-tight tube furnace operated up to 250 °C has been developed. Real-time detection of released gases (volatile organic compounds (VOCs), CO2, NO, NO2, SO2, CO and O2) was achieved combining commercial off-the-shelf (COTS) gas sensors and sorbent tubes for further qualitative and semi-quantitative analysis by gas chromatography-mass spectrometry coupled to thermal desorption (TD-GC-MS). The test system was designed to provide a controlled flow (1000 cm3 min−1) of hydrocarbon free air through the furnace. The furnace temperature ramp was set at a rate of 5 °C min−1 with 10 min dwell points at 70 °C, 150 °C, 200 °C and 250 °C to allow time for stabilisation and further headspace sampling onto sorbent tubes. Experimental design of the instrumentation is described here and an example data set upon exposure to a gas sample is presented.



中文翻译:

用于定量分析高温下挥发性化合物排放的仪器。第1部分:设计和实现。

已开发出一套新颖的仪器,用于表征在最高250°C的气密管式炉内所盛物料的特性。结合商用现货(COTS)气体传感器和吸附剂管,可以实时检测释放的气体(挥发性有机化合物(VOC),CO 2,NO,NO 2,SO 2,CO和O 2)气相色谱-质谱联用热脱附(TD-GC-MS)进行定性和半定量分析。该测试系统被设计为提供通过炉子的受控的不含烃的空气流(1000 cm 3 min -1)。炉温梯度设定为5°C min -1在70°C,150°C,200°C和250°C的温度下具有10分钟的驻留点,以留出稳定时间并在吸附剂管上进一步进行顶空进样。这里介绍了仪器的实验设计,并介绍了暴露于气体样品后的示例数据集。

更新日期:2020-05-26
down
wechat
bug