Dose calculation errors as a component of failing IROC lung and spine phantom irradiations.
Medical Physics ( IF 3.2 ) Pub Date : 2020-05-25 , DOI: 10.1002/mp.14258
Sharbacha S Edward 1, 2, 3 , Mallory C Glenn 1, 2, 3 , Christine B Peterson 1, 4 , Peter A Balter 1, 3 , Julianne M Pollard-Larkin 1, 3 , Rebecca M Howell 1, 3 , David S Followill 1, 2, 3 , Stephen F Kry 1, 2, 3
Affiliation  

Between July 2013 and August 2019, 22% of the imaging and radiation oncology core (IROC) spine, and 15% of the moving lung phantom irradiations have failed to meet established acceptability criteria. The spine phantom simulates a highly modulated stereotactic body radiation therapy (SBRT) case, whereas the lung phantom represents a low‐to‐none modulation moving target case. In this study, we assessed the contribution of dose calculation errors to these phantom results and evaluated their effects on failure rates.