当前位置: X-MOL 学术Braz. J. Infect. Dis. › 论文详情
Our official English website, www.x-mol.net, welcomes your feedback! (Note: you will need to create a separate account there.)
Cost minimization analysis of an in-house molecular test for cytomegalovirus in relation to a commercial molecular system.
The Brazilian Journal of Infectious Diseases ( IF 3.0 ) Pub Date : 2020-05-23 , DOI: 10.1016/j.bjid.2020.04.015
Cássia F B Caurio 1 , Odelta S Allende 2 , Roger Kist 1 , Izadora C S Vasconcellos 1 , Francieli P Rozales 2 , Maikel Reck-Kortmann 2 , Daiane F Dalla Lana 3 , Ana Paula Alegretti 4 , Giácomo B Neto 5 , Alessandro C Pasqualotto 1
Affiliation  

Introduction

Cytomegalovirus may cause severe disease in immunocompromised patients. Nowadays, quantitative polymerase chain reaction is the gold-standard for both diagnosis and monitoring of cytomegalovirus infection. Most of these assays use cytomegalovirus automated molecular kits which are expensive and therefore not an option for small laboratories, particularly in the developing world.

Objective

This study aimed to optimize and validate an in-house cytomegalovirus quantitative polymerase chain reaction test calibrated using the World Health Organization Standards, and to perform a cost-minimization analysis, in comparison to a commercial cytomegalovirus quantitative polymerase chain reaction test.

Study design

The methodology consisted of determining: optimization, analytical sensitivity, analytical specificity, precision, curve variability analysis, and inter-laboratorial reproducibility. Patients (n = 30) with known results for cytomegalovirus tested with m2000 RealTime System (Abbott Laboratories, BR) were tested with the in-house assay, as well as patients infected with other human herpes virus, in addition to BK virus. A cost-minimization analysis was performed, from a perspective of the laboratory, assuming diagnostic equivalence of the methodologies applied in the study.

Results

The in-house assay had a limit of detection and quantification of 60.3 IU/mL, with no cross-reactivity with the other viral agents tested. Moreover, the test was precise and had a R2 of 0.954 when compared with the m2000 equipment. The cost analysis showed that the assay was economically advantageous costing a median value of 37.8% and 82.2% in comparison to the molecular test in use at the hospital and the m2000 equipment, respectively.

Conclusions

These results demonstrated that in-house quantitative polymerase chain reaction testing is an attractive alternative in comparison to automated molecular platforms, being considerably less expensive and as efficacious as the commercial methods.