Comment on: “Slope ratio calibration for analysis of plant leaves by laser-induced breakdown spectroscopy” by Lidiane C. Nunes, Fabio R. P. Rocha and Francisco J. Krug, JAAS, 2019, 34, 2314
Journal of Analytical Atomic Spectrometry ( IF 3.1 ) Pub Date : 2020-05-22 , DOI: 10.1039/c9ja00381a
Vincenzo Palleschi 1, 2, 3, 4
Affiliation  

In the paper “Slope ratio calibration for analysis of plant leaves by laser-induced breakdown spectroscopy”, recently published on JAAS, the authors present an allegedly new calibration method for LIBS analysis, called Slope Ratio Calibration (SRC). In this short comment we will demonstrate that this method is completely equivalent to Single-Point Calibration (SPC), when a fair comparison between the two methods is done.