Bioimpedance Spectroscopy of the Breast.
Lymphatic Research and Biology ( IF 1.6 ) Pub Date : 2020-10-19 , DOI: 10.1089/lrb.2019.0087
Leigh C Ward 1 , Amy C Degnim 2 , Elizabeth S Dylke 3 , Sharon L Kilbreath 3
Affiliation  

Background: Bioimpedance spectroscopy (BIS) measurements of breast lymphedema poses practical and technical challenges, in particular the determination of the resistance at zero frequency (R0), the index of change in breast lymph content. Conventionally, R0 is calculated from data analysis by using a procedure eponymously known as Cole modeling, a method that is error-prone in the breast. The aim of this study was to evaluate polynomial curve fitting as an alternative analytic procedure.