Acceptance experimentation and quality monitor of x-ray radiography units
Radiation Physics and Chemistry ( IF 2.9 ) Pub Date : 2020-07-01 , DOI: 10.1016/j.radphyschem.2020.108810
Ammar A. Oglat

Abstract This study presents the 2019 findings of the acceptance testing and routine quality control (QC) for the general radiographic X-ray equipment such as film/screen fixed systems in Malaysia. Acceptance testing and routine QC of the diagnostic X-ray imaging equipment are the requirements for the standard radiation apparatus and should be compared to the requirements for European and RADIATION SAFETY STANDARD HR001:2010 legislation, which were considered the standard values of all general radiographic X-ray equipment since it identical to the manufacture's data. The general radiographic X-ray system test was conducted in the Biophysics lab, University Science Malaysia, and it has failed to meet the required QC guidelines, whereby one or more faults were identified. The major failures identified were minor, requiring attention by the suppliers at the next routine service. The suppliers were requested to investigate the issues and take the necessary corrective action. A review of the QC results highlights the need to perform comprehensive acceptance and routine testing of the systems.

中文翻译:

X射线摄影机验收实验及质量监控

摘要 本研究介绍了 2019 年马来西亚胶片/屏幕固定系统等通用 X 射线设备的验收测试和常规质量控制 (QC) 的结果。诊断 X 射线成像设备的验收测试和常规 QC 是标准辐射设备的要求,应与欧洲和辐射安全标准 HR001:2010 立法的要求进行比较,这些要求被认为是所有通用射线照相 X 的标准值- 射线设备,因为它与制造商的数据相同。一般的射线照相 X 射线系统测试是在马来西亚科学大学的生物物理实验室进行的,它未能满足要求的 QC 指南,因此发现了一个或多个故障。确定的主要故障是轻微的,要求供应商在下一次例行服务中注意。供应商被要求调查问题并采取必要的纠正措施。对 QC 结果的审查强调了对系统进行全面验收和常规测试的必要性。
更新日期:2020-07-01
down
wechat
bug