当前位置: X-MOL 学术Optik › 论文详情
Our official English website, www.x-mol.net, welcomes your feedback! (Note: you will need to create a separate account there.)
Intra-cavity multi-point sensor for C2H2 detection in photoacoustic spectroscopy based on acousto-optic Q-switched fiber laser
Optik Pub Date : 2020-01-21 , DOI: 10.1016/j.ijleo.2020.164277
Qinduan Zhang , Jun Chang , Zhenhua Cong , Zongliang Wang , Fupeng Wang , Jiachen Sun , Cunguang Zhu

In this paper, we reported a new multi-point sensor for C2H2 detection based on photoacoustic spectroscopy and acousto-optic Q-switched fiber laser. Three transmission-type PAS cells were placed inside the fiber laser cavity and used for C2H2 detection simultaneously to demonstrate the potential of spatially resolved measurements. A maximum pulse peak power of ∼544 mW was generated by the acousto-optic Q-switched fiber laser. The signal-to-noise ratios (SNR) calculated from the experimental results was 57.7 dB, 45.1 dB and 57.2 dB. This resulted in a minimum detection limit (MDL) for C2H2 of 1.73 ppmv, 7.82 ppmv and 1.93 ppmv for PAS cell 1, PAS cell 2 and PAS cell 3, respectively.



中文翻译:

基于声光调Q光纤激光器的光声光谱中腔内多点C 2 H 2检测传感器

在本文中,我们报道了一种基于光声光谱和声光调Q光纤激光器的用于C 2 H 2检测的新型多点传感器。将三个透射型PAS单元放置在光纤激光腔内,并同时用于C 2 H 2检测,以证明空间分辨测量的潜力。声光调Q光纤激光器产生的最大脉冲峰值功率约为544 mW。根据实验结果计算出的信噪比(SNR)为57.7 dB,45.1 dB和57.2 dB。这导致PAS单元1,PAS单元2和PAS单元3的C 2 H 2的最低检测限(MDL)分别为1.73 ppmv,7.82 ppmv和1.93 ppmv。

更新日期:2020-01-21