British Journal of Anaesthesia ( IF 9.1 ) Pub Date : 2020-01-14 , DOI: 10.1016/j.bja.2019.11.030 Douglas C Crockett 1 , Minh C Tran 2 , Federico Formenti 3 , John N Cronin 4 , Göran Hedenstierna 5 , Anders Larsson 6 , Phi A Phan 1 , Andrew D Farmery 1
Background
Bedside lung volume measurement could personalise ventilation and reduce driving pressure in patients with acute respiratory distress syndrome (ARDS). We investigated a modified gas-dilution method, the inspired sinewave technique (IST), to measure the effective lung volume (ELV) in pigs with uninjured lungs and in an ARDS model.
Methods
Anaesthetised mechanically ventilated pigs were studied before and after surfactant depletion by saline lavage. Changes in PEEP were used to change ELV. Paired measurements of absolute ELV were taken with IST (ELVIST) and compared with gold-standard measures (sulphur hexafluoride wash in/washout [ELVSF6] and computed tomography (CT) [ELVCT]). Measured volumes were used to calculate changes in ELV (ΔELV) between PEEP levels for each method (ΔELVIST, ΔELVSF6, and ΔELVCT).
Results
The coefficient of variation was <5% for repeated ELVIST measurements (n=13 pigs). There was a strong linear relationship between ELVIST and ELVSF6 in uninjured lungs (r2=0.97), and with both ELVSF6 and ELVCT in the ARDS model (r2=0.87 and 0.92, respectively). ELVIST had a mean bias of –12 to 13% (95% limits=±17 – 25%) compared with ELVSF6 and ELVCT. ΔELVIST was concordant with ΔELVSF6 and ΔELVCT in 98–100% of measurements, and had a mean bias of –73 to –77 ml (95% limits=±128 – 186 ml) compared with ΔELVSF6 and –1 ml (95% limits ±333 ml) compared with ΔELVCT.
Conclusions
IST provides a repeatable measure of absolute ELV and shows minimal bias when tracking PEEP-induced changes in lung volume compared with CT in a saline-lavage model of ARDS.
中文翻译:
验证启发性正弦波技术以测量猪肺损伤模型中“婴儿肺”的体积。
背景
床旁肺体积测量可以个性化通气并降低急性呼吸窘迫综合征(ARDS)患者的驾驶压力。我们研究了一种改良的气体稀释方法,即正弦波技术(IST),以测量未受伤肺和ARDS模型中猪的有效肺体积(ELV)。
方法
在用盐水冲洗去除表面活性剂之前和之后,对麻醉的机械通气猪进行了研究。PEEP的变化用于改变ELV。使用IST(ELV IST)对绝对ELV进行配对测量,并与金标准测量(六氟化硫冲洗/冲洗[ELV SF6 ]和计算机断层扫描(CT)[ELV CT ])进行比较。测量体积用于在ELV(ΔELV)为每个方法PEEP水平之间的变化计算(ΔELV IST,ΔELV SF6和ΔELV CT)。
结果
重复ELV IST测量的变异系数<5%(n = 13头猪)。在未受伤的肺中,ELV IST和ELV SF6之间存在很强的线性关系(r 2 = 0.97),在ARDS模型中,ELV SF6和ELV CT两者之间有很强的线性关系(r 2分别为0.87和0.92)。与ELV SF6和ELV CT相比,ELV IST的平均偏差为–12至13%(95%限值=±17 – 25%)。ΔELV IST与ΔELV一致SF6和ΔELV CT在测量98-100%,并且具有-73到-77毫升(95%限度=±128 - 186毫升)的平均偏压ΔELV相比SF6与ΔELV比较和-1毫升(95%的范围±333毫升)CT。
结论
与ARDS盐水灌洗模型中的CT相比,IST提供了可重复测量的绝对ELV,并且在追踪PEEP诱导的肺体积变化时显示出最小的偏差。