Analytical performance of a high-repetition rate laser head (500 Hz) for HR LA-ICP-QMS imaging
Journal of Analytical Atomic Spectrometry ( IF 3.1 ) Pub Date : 2020-01-07 , DOI: 10.1039/c9ja00421a
Martin Šala 1, 2, 3, 4 , Vid Simon Šelih 1, 2, 3, 4 , Ciprian Cosmin Stremtan 5, 6, 7 , Johannes Teun van Elteren 1, 2, 3, 4
Affiliation  

With the advent of low dispersion, fast aerosol transport technologies used in laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS), the need for high-repetition rate lasers has become crucial. This work reports on the analytical performance of a LA unit with an experimental 500 Hz laser head, interfaced with a quadrupole ICP-MS instrument, and comprising a commercial two-volume ablation chamber (HelEx II) and low-dispersion aerosol transport device (ARIS). The system was tested for potential problems related to particle saturation, elemental fractionation, ionisation efficiency, etc., prior to validation as a high-resolution LA-ICP-QMS imaging tool with a high mapping rate.