Meticulous plasma isolation is essential to avoid false low-level viraemia in Roche Cobas HIV-1 viral load assays.
Antiviral Therapy ( IF 1.3 ) Pub Date : 2017-10-25 , DOI: 10.3851/imp3203
Virginie Mortier 1 , Leen Vancoillie 1 , Kenny Dauwe 1 , Delfien Staelens 1 , Els Demecheleer 1 , Marlies Schauvliege 1 , Sylvie Dinakis 1 , Tom Van Maerken 2 , Géraldine Dessilly 3 , Jean Ruelle 3 , Chris Verhofstede 1
Affiliation  

Pre-analytical sample processing is often overlooked as a potential cause of inaccurate assay results. Here we demonstrate how plasma, extracted from standard EDTA-containing blood collection tubes, may contain traces of blood cells consequently resulting in a false low-level HIV-1 viral load when using Roche Cobas HIV-1 assays.

中文翻译:

在Roche Cobas HIV-1病毒载量测定中,认真进行血浆分离对于避免虚假的低水平病毒血症至关重要。

分析前样品处理经常被忽视,因为其可能导致测定结果不准确。在这里,我们演示了从含标准EDTA的采血管中提取的血浆中可能含有微量血细胞,从而在使用Roche Cobas HIV-1检测法时导致虚假的低水平HIV-1病毒载量。