Validation of two accelerated 4D flow MRI sequences at 3 T: a phantom study.
European Radiology Experimental ( IF 3.7 ) Pub Date : 2019-02-26 , DOI: 10.1186/s41747-019-0089-2 Sebastian Ebel 1 , Lisa Hübner 1 , Benjamin Köhler 2 , Siegfried Kropf 3 , Bernhard Preim 2 , Bernd Jung 4 , Matthias Grothoff 1 , Matthias Gutberlet 1
中文翻译:
在3 T验证两个加速的4D流MRI序列:一项幻像研究。
更新日期:2019-02-26
European Radiology Experimental ( IF 3.7 ) Pub Date : 2019-02-26 , DOI: 10.1186/s41747-019-0089-2 Sebastian Ebel 1 , Lisa Hübner 1 , Benjamin Köhler 2 , Siegfried Kropf 3 , Bernhard Preim 2 , Bernd Jung 4 , Matthias Grothoff 1 , Matthias Gutberlet 1
Affiliation
Background
Four-dimensional (4D) flow magnetic resonance imaging (MRI) sequences with advanced parallel imaging have the potential to reduce scan time with equivalent image quality and accuracy compared with standard two-dimensional (2D) flow MRI. We compared 4D flow to standard 2D flow sequences using a constant and pulsatile flow phantom at 3 T.Methods
Two accelerated 4D flow sequences (GRAPPA2 and k-t-GRAPPA5) were evaluated regarding the concordance of flow volumes, flow velocities, and reproducibility as well as dependency on measuring plane and velocity encoding (Venc). The calculated flow volumes and peak velocities of the phantom were used as reference standard. Flow analysis was performed using the custom-made software “Bloodline”.Results
No significant differences in flow volume were found between the 2D, both 4D flow MRI sequences, and the pump reference (p = 0.994) or flow velocities (p = 0.998) in continuous and pulsatile flow. An excellent correlation (R = 0.99–1.0) with a reference standard and excellent reproducibility of measurements (R = 0.99) was achieved for all sequences. A Venc overestimated by up to two times had no impact on flow measurements. However, misaligned measuring planes led to an increasing underestimation of flow volume and mean velocity in 2D flow accuracy, while both 4D flow measurements were not affected. Scan time was significantly shorter for k-t-GRAPPA5 (1:54 ± 0:01 min, mean ± standard deviation) compared to GRAPPA2 (3:56 ± 0:02 min) (p = 0.002).Conclusions
Both 4D flow sequences demonstrated equal agreement with 2D flow measurements, without impact of Venc overestimation and plane misalignment. The highly accelerated k-t-GRAPPA5 sequence yielded results similar to those of GRAPPA2.中文翻译: