当前位置: X-MOL 学术Opt. Lett. › 论文详情
Our official English website, www.x-mol.net, welcomes your feedback! (Note: you will need to create a separate account there.)
Auto-phase-locked measurement of time-gated luminescence spectra with a microsecond delay
Optics Letters ( IF 3.1 ) Pub Date : 2018-05-23 , DOI: 10.1364/ol.43.002575
Zece Zhu , Xuewen Shu

Time-resolved techniques are widely used in measuring the spectra and lifetimes of the excited states of molecules. However, the relative apparatus always requires gated detector and phase-matching circuitry, which is expensive to implement and maintain. Herein, a novel auto-phase-locked method for time-gated luminescence (TGL) spectra measurement was developed by adjusting the exciting and detecting optical paths to pass through the same chopper wheel, which simultaneously acted as a pulse generator and detecting shutter. This low-cost system needs no phase-matching circuitry or control system. It can detect TGL spectra with a delay time of only microseconds, demonstrating a high temporal resolution for thermally activated delayed fluorescence detection.

中文翻译:

具有微秒延迟的时控发光光谱的自动锁相测量

时间分辨技术广泛用于测量分子激发态的光谱和寿命。然而,相关设备总是需要门控检测器和相位匹配电路,这对于实施和维护而言是昂贵的。在本文中,通过调节激发光和检测光路通过同一斩波轮,开发了一种新颖的自动锁相方法,用于测量时间门控发光(TGL)光谱,该光路同时用作脉冲发生器和光闸。这种低成本系统不需要相位匹配电路或控制系统。它可以以仅微秒的延迟时间检测TGL光谱,从而证明了热激活延迟荧光检测的高时间分辨率。
更新日期:2018-06-01
down
wechat
bug