The use of artificial intelligence (AI) in medical imaging is steadily growing1,2,3. Currently, task-specific AI applications are able to match and occasionally exceed human intelligence, and it has been predicted that AI will surpass the skills of a radiologist in the next 50 years4. The development of such technology requires advances in computing power and algorithms, but also advances in curation and imaging capabilities. X-ray detectors with increased resolution can, in particular, provide improved medical images that further enhance the benefits of AI technologies5. X-ray detectors can be divided into two main classes:6,7,8 indirect-conversion detectors and direct-conversion detectors. Indirect converters exhibit high sensitivity but suffer from low spatial resolution9. Direct converters can capture high-resolution images (up to 10–15 line pairs per millimetre (lp mm−1)) but at a relatively high applied electric field (1–10 V μm−1) and relatively low sensitivity10,11,12,13,14. The requirement is an X-ray detector that combines high resolution (to enhance AI performance) with high sensitivity (to reduce patient X-ray dose).

Hybrid inorganic–organic perovskites, such as methylammonium lead iodide (CH3NH3PbI3, abbreviated as MAPbI3), offer high electron and hole diffusion lengths due to their high charge carrier mobilities (µ) and long carrier lifetimes (τ), and have demonstrated promising characteristics for use as direct X-ray converters15,16,17,18,19,20,21. For example, polycrystalline and single-crystalline hybrid inorganic–organic perovskites have exhibited a μτ product of 2.0 × 10–4 and 1.2 × 10–2 cm2 V−1, respectively, which are in the same range as polycrystalline cadmium zinc telluride (CZT)13,15,16,17,18,19,20,21. The high X-ray absorption coefficient of MAPbI3 over large parts of the energy spectrum used in healthcare also makes it an ideal candidate for use in next-generation imaging systems. However, the integration of direct X-ray converter layers onto a pixelated electrode substrate (often referred to as a backplane), which converts the generated X-ray signal of each pixel into a two-dimensional digital image, can be challenging from a manufacturing perspective. Thus, while X-ray detection with direct-conversion perovskites has been demonstrated, the integration of perovskite detecting layers with pixelated backplanes has only received limited attention17,22.

In this Article, we report a two-step manufacturing process to create X-ray flat-panel detectors that combine high spatial resolution and high sensitivity. Our approach separates the fabrication of an X-ray absorber layer (which is made of microcrystalline MAPbI3) with a thickness of several hundreds of micrometres from integration onto the backplane. The integration is subsequently performed at room temperature, and thus, backplane limitations regarding the temperature budget are not a factor. Our approach also allows for the independent optimization of the MAPbI3 absorber formation.

X-ray imager architecture

The layer stack of the imaging X-ray detector is shown in Fig. 1a. The glass-based backplane is a self-aligned dual-gate indium gallium zinc oxide thin-film transistor array, which is described and characterized elsewhere23,24. The range of applicable voltages on the backplane is limited due the missing high-voltage protection of the pixels. Bottom electrodes are made of a molybdenum–chromium alloy. On top of the backplane, a grid structure made of an approximately 10-µm-thick photoresist is used as a mechanical anchoring structure for the thick absorber layer. Without this grid, the mechanical adhesion between MAPbI3 and backplane was found to be poor, indicated by the release of the wafer after a few days. However, with the grid, no spontaneous detachment can be seen in several months. Pull tests revealed a tensile strength of 100 mN mm−2. The grid is filled with liquid MAPbI3 that acts—after its recrystallization—as an adhesion promoter for the attachment of the X-ray absorber, which consists of a 230-µm-thick MAPbI3 layer; this thickness was chosen with respect to the limited applicable bias voltage at the imager. As the cathode, a chromium (Cr) layer is deposited on top of MAPbI3. The active area of the imager is encapsulated with a barrier foil to avoid environmental influences, as shown in Fig. 1b.

Fig. 1: Perovskite X-ray imaging detector.
figure 1

a, Exploded view of the different elements of the X-ray imager. b, Photograph of the final detector with chip-on-glass gate driver integrated circuits (bottom part) and chip-on-flex ROIC (right side); the MAPbI3 (MAPI) layer attached on the backplane is encapsulated by a laminated metallic foil (centre); and flex bonds connect the array with the reading and driving printed circuit boards. The glass substrate size is 102 × 124 mm2. c, X-ray image of a resolution phantom up to 6 lp mm–1 structures. d, X-ray response as a function of dose. Inset: SNR for different pulses with a detection limit of 0.22 nGyair per frame. Each frame is 35 ms.

This direct-conversion X-ray detector can capture objects with a very high resolution of 6 lp mm−1 (Fig. 1c) and shows an unprecedented low detection limit of 0.22 nGyair per frame at an applied electrical field of 0.03 V µm−1. The resolution capability was tested by X-ray imaging a phantom having structures made of three lines at different spacings of 5.0, 5.5 and 6.0 lp mm−1. The detection limit is calculated taking—for each frame—the average signal of pixels from a region of interest (ROI) for a 1-s-long exposure at different doses and considering a signal-to-noise ratio (SNR) of >3, as shown in Fig. 1d25. The beam quality was generated using an X-ray source with an anode bias of 70 peak kilovoltage (kVp), and filtering the resulting radiation with 21 mm aluminium (Al) and additional attenuation with 1 mm Pb (Supplementary Information provides details about the X-ray setup). The signal is presented in the least significant bit (LSB), the smallest possible unit of the analogue-to-digital converter in the ROIC, which—in the used settings—is equivalent to 48 electrons.

The dose per frame was varied between 0.33 and 20.61 nGyair, and the SNR was determined from the resulting height (H) of the pulses and variance of the background noise (h) according to the following equation: SNR = 2H/h ≥ 3 (ref. 25). The SNR is inversely proportional to the dose (Fig. 1d, inset) and can be linearly fitted, which leads to a dose of 0.22 nGyair per frame at SNR = 3. This is due to the impressive stability of the mean value of the ROI in the range of 0–2 LSB in the observed period. Electrical stability is reached after 90 min of biasing of the imager at an electrical field of 0.03 V µm−1, which is the time required to achieve an almost ionic equilibrium state (Supplementary Fig. 1a). With a frame rate of 28.6 frames per second, the detection limit is 6.3 nGyair s−1, which is 20% lower compared with the best reported detection limit of perovskites: Wei and co-workers made a 1-mm-thick single crystal of methylammonium lead tribromide (MAPbBr3) alloyed with Cl, achieving the lowest detectable dose rate of 7.6 nGyair s−1 for a photon energy of 8 keV (ref. 26). Previously reported limits for MAPbI3 are even higher: 19.1 µGyair s−1 for single-crystal MAPbI3 (ref. 19). With A-site cation variation, Huang et al. could reduce the detection limit to 16.9 nGyair s−1 (ref. 20).

Two-step manufacturing process

The manufacturing process of our perovskite X-ray detector consists of two phases: the first focuses on the production of the X-ray absorber layer. We chose a mechanical soft-sintering process, which is described in detail in ref. 18. The used microcrystalline MAPbI3 powder is commercially available, where the grain size varies between 0.1 and 100.0 µm, whereas 85% of the measured sizes are below 5.0 µm (Supplementary Fig. 2a,b). For the investigation of the structural properties of the powder, an X-ray diffraction measurement has been performed (Supplementary Fig. 2c). The mean diffraction peaks at 14.1°, 23.4°, 24.5°, 28.1° and 28.4° are related to the lattice planes (110), (211), (202), (004) and (220), respectively, and fit well with the values of the tetragonal phase with the I4cm space group of MAPbI3 found in the literature27,28. To form a freestanding stable wafer, the powder was compacted for 30 min at room temperature using a hydraulic press with a pressure of 75.5 MPa. The compactness of these wafers with a resulting thickness of 230 µm is 88% of the theoretical limit calculated from the simulated lattice parameters27,28 and is in good agreement with the density versus pressure curve shown in Supplementary Fig. 2 (generated as the calibration plot).

The second manufacturing phase is shown in Supplementary Fig. 4. First, a photoresist grid with 10 µm height is build up on the backplane by photolithography. A top view of the regular grid structure with a pixel pitch of 50 µm and fill factor of 58% is shown in the scanning electron microscopy (SEM) image in Fig. 2a. The total active area of the X-ray detector is 3.2 × 2.4 cm2. Figure 2b shows a micrograph of the grid structure, in which the bottom electrode array can be seen as bright areas. In the second step, the grid on the backplane is filled with MAPbI3 powder and then liquefied under a methylamine atmosphere. More details on this reaction can be found elsewhere29,30,31. In the timescale of seconds, the previously prepared MAPbI3 wafer is placed on the liquid phase. Fixation occurs after the liquefied MAPbI3 recrystallizes due to the evaporation of excessive methylamine during the annealing step at 50 °C, acting as an adhesion promoter between the wafer and backplane, as shown in Fig. 2c.

Fig. 2: Characterization of the manufacturing process.
figure 2

a, SEM image of the grid photoresist structure. The bird’s eye view shows the regular shape of the grid. The width of the structure is 10 µm. The pixel pitch is 50 µm. b, Optical microscopy image of the grid structure. The bright areas are the bottom electrodes defining the pixel size. c, Photograph of the MAPbI3 absorber layer attached to the backplane. The size of the MAPbI3 absorber layer is 3 × 4 cm2 and fully covers the area of the electrode array. df, Cross-sectional SEM images of different MAPbI3 layers: bottom part of an 880-µm-thick absorber layer above an indium tin oxide glass with the photoresist grid visible and indicated with arrows (d); a closer look into the wafer shows grains with sizes between 0.5 and 5.0 µm and pores between them (e); grains of the recrystallized MAPbI3 layer with smaller grains up to 2 µm (f).

To investigate the recrystallized grains, SEM cross-section images were obtained, as shown in Fig. 2d–f. During the bonding process, the liquification of the MAPbI3 wafer due to methylamine vapours is not excluded, but it will be very limited since the thickness of the recrystallized layer slightly exceeds the grid height. The bottom part of the MAPbI3 wafer bonded to an indium tin oxide glass with grid structures at the glass interface is shown in Fig. 2d. The surface of the cross section is rough because of the breaking mechanism of the wafer and glass. Morphology differences between the recrystallized and soft-sintered perovskite can be identified. A closer look inside the soft-sintered absorber is shown in Fig. 2e. The grain size varies between 0.5 and 5.0 µm and some pores are visible between the powder grains, which are attributed to mechanical soft sintering. In contrast, recrystallized MAPbI3 has fewer pores between the grains (Fig. 2f), which has a smaller size of up to 2 µm. The sensitivity and thus the charge transport properties of methylamine-treated MAPbI3 wafer show no notable change compared with a non-treated one. Nevertheless, additional defect states may have been introduced by changing the grain size32,33,34.

A great advantage of the presented manufacturing process of the X-ray imager is the possibility to execute quality control of the MAPbI3 wafer before attachment to the backplane. In fact, freestanding wafers can be X-rayed themselves, and the homogeneity of X-ray absorption can be controlled using a commercially available flat-panel detector (Supplementary Fig. 5a–c).

Characterization of freestanding wafers

For a better understanding of the imaging X-ray detector properties, a closer look at the compact wafer is required. Therefore, we have performed impedance measurements on a single device made of a soft-sintered MAPbI3 wafer with platinum (Pt) and Cr electrodes (Supplementary Fig. 6a,b). A schematic of the wafer stack is shown in Supplementary Fig. 11a. According to the resulting geometrical capacitance and sample dimensions, the dielectric constant of the wafer is ε = 75. This is in good agreement with previously reported values35,36,37. For dark current measurements, a 977-µm-thick, 86% dense wafer has been used.

The Cr electrode was grounded during the measurements, and the applied voltage was in the range of −200 to +200 V, corresponding to an electrical field of ±0.2 V µm−1. The electrical field was at first decreased from 0 to −0.200 V µm−1 in steps of 0.001 V µm−1; thereafter, it was increased from 0 to +0.2 V µm−1. In the current density versus electrical field (JE) plot (Fig. 3a), the dark current density reaches the maximum value of 8.40 × 10–4 mA cm−2 for negative fields, the so-called reverse bias direction, and 1.98 × 10–3 mA cm−2 for positive (forward) bias. This leads to a dark resistivity of ~2.4 × 109 Ω cm, which is in the same order of magnitude to the values reported in ref. 38. This small rectifying behaviour could be caused by the used electrodes and their different work functions of ϕm,Cr = 4.5 eV and ϕm,Pt = 5.7 eV (ref. 39) or ions and their associated vacancies32.

Fig. 3: Optoelectronic properties of MAPbI3 wafers.
figure 3

a, Current density measurement under dark condition for an MAPbI3 wafer. The applied voltage ranges from −200 to +200 V, which corresponds to an electrical field of ±0.2 V µm−1. b, Calculated sensitivity (S) of the X-ray responses in the RQA5 spectrum and a dose of 213 µGyair s−1. c, EHP creation energy for the measurement series of 52 and 4 µGyair s−1. The limit is 4.5 eV according to the Klein rule40. In the double logarithmic representation (inset), the hyperbolic behaviour of W± is visible. d, EHP creation energy versus exposed X-ray spectrum. The MAPbI3 wafer was exposed with a dose of 40 µGyair in the RQA5 spectrum and an applied electrical field of 0.043 V µm−1. The X-ray tube voltage was varied from 50 to 120 kVp (black points) and back to 50 kVp thereafter (red dots).

The response of the wafer to 2-s-long irradiation with the RQA5 (according to the IEC 61267:2005 standard) X-ray spectrum and a dose rate of 213 µGyair s−1 at different bias voltages was captured with a Keithley 2400 source meter. This results in the maximum value of 9,300 μC Gyair−1 cm−2 for the sensitivity of our wafer at an electrical field of 0.17 V µm−1 (Fig. 3b). Fitting the data by the Hecht equation10 (red curve in Supplementary Fig. 11b) results in a µτ product of approximately 4 × 104 cm2 V−1. A comparison with previously reported perovskite materials shows that the measured sensitivity is in good agreement with printed MAPbI3 samples17, 3.5 times higher than the soft-sintered MAPbI3 wafers18 and over 100 times better than single-crystal MAPbBr3 (ref. 16).

Another important figure of merit in comparing X-ray detectors with different thicknesses and materials is the electron–hole pair (EHP) generation energy W±. W± is the amount of absorbed radiation energy needed to create a single free EHP (Supplementary Information provides more information)40,41. In Fig. 3c, the values of W± are plotted as a function of the applied bias for two dose rates. W± decreases with an approximately hyperbolic (f(x) = 1/x) behaviour with increasing electrical field and approaches the empirical limit of W± = 3EG ≈ 4.5 eV (EG is the bandgap), which is known as the Klein rule40. W± values (on a linear scale) for a total of six different dose rates are shown in Supplementary Fig. 12a. Within the measurement accuracy, W± is independent of the dose rate for electrical fields higher than 0.05 V µm−1, indicating an almost full extraction of the generated charges.

To investigate the dependencies of W± with the X-ray photon energy, we performed a series of X-ray response measurements with different photon spectra obtained by varying the anode voltages on the X-ray tube. The simulated X-ray photon spectra are shown in Supplementary Fig. 12b. With the help of these photon energy densities, the absorbed energy for the MAPbI3 wafer can be determined: for an anode voltage of 50 kVp, the wafer absorbed 98.63% of the emitted X-ray spectrum. The absorbed energy decreases to 71.33% for increasing anode voltages up to 120 kVp (Supplementary Table 1). The wafer was irradiated for 1 s each time with a dose of 40 µGyair under an applied electrical field of 0.043 V µm−1. For a better statistical evaluation and excluding possible degradation factors, the anode voltage was initially increased from 50 to 120 kVp and then decreased from 120 to 50 kVp. The corresponding X-ray pulses show an increasing pulse height with higher anode voltages with the maximum value at around 100 kVp (Supplementary Fig. 12c). The increase is proportional to the photon energy fluence of the X-ray tube. The resulting average W± values are 5.99 ± 0.20 eV (Fig. 3d), that is, W± is independent of the X-ray photon energy. This is of great importance since X-ray photons with a lower energy carry higher diagnostic information.

Furthermore, no degradation caused by irradiation has been observed after a cumulative dose of 11 Gyair (Supplementary Fig. 13). Similar to MAPbI3, the theoretical limit of W± in amorphous selenium (a-Se) ranges between 5 and 6 eV. In practice, however, values of 40–50 eV at an electrical field of 10 V µm−1 are achieved42. In addition, a-Se shows decreasing W± for increasing X-ray photon energy42. All the results presented here with MAPbI3 wafer devices show the great potential of this material class for X-ray detector application.

Characterization of X-ray detectors

With the knowledge of freestanding wafer measurements, we were able to manufacture a pixelated MAPbI3 X-ray imaging detector with outstanding performance. X-ray characterization was performed by varying the dose range over five orders of magnitude initially using the RQA5 spectrum and adding different filters later. For the corresponding photon energy densities, see Supplementary Fig. 14a. The level of absorption of these three spectra for different thicknesses of MAPbI3 is shown in Supplementary Fig. 14b. With the effective thickness of our wafer (red dashed line), we obtained theoretical absorption values of 50%, 46% and 36% for the RQA5, RQA5 with 15 mm Al filter and RQA5 + 1 mm Pb spectra, respectively.

For the investigation of the imager response to 1-s-long X-ray pulses, video sequences were taken with an integration time of 35 ms per frame. Frame sequences were acquired in the rolling shutter mode without synchronization between the X-ray source and X-ray detector. The offset-corrected signals under the maximum applicable electrical field of 0.03 V µm−1 using the RQA5 spectrum with doses ranging from 44 to 8,390 nGyair per frame are presented in Fig. 4a. The pulses show good time response, as indicated by the steep increase and decrease. To evaluate the linearity of the imager responses for a wider dose range, we used the RQA5 spectrum in combination with additional 15 mm Al filtration. This detector is impressive, with good linearity over a dose range of three orders of magnitude (Fig. 4b and Supplementary Fig. 15a). Supplementary Fig. 15b shows a close-to-ideal dose dependence (slope, 0.99).

Fig. 4: X-ray imaging properties of MAPbI3 detector at RQA5.
figure 4

a, X-ray response of the pixelated MAPbI3 imager under an electrical field of 0.03 V μm−1 at different doses. b, Linearity plot of the extracted charges versus dose in the RQA5 (blue) and 15 mm Al-filtered RQA5 (red) spectrum. c, MTF curves of the pixelated MAPbI3 imager for two different doses (blue and green) and MTF of commercial indirect-conversion detector based on the complementary metal–oxide–semiconductor technology (red). d, X-ray image with the resolution of phantom structures taken by the indirect-conversion (left) and MAPbI3 detector (MAPI; right). e, Photograph of part of a hearing aid (left) and a metallic coronary stent (right). f, X-ray image of the hearing aid (left) and coronary stent (right) taken by the pixelated MAPbI3 detector at a dose of 4.17 µGyair per frame. Even the mesh of the stent with a cross-section size of 100 μm × 200 μm can be resolved.

Image lag describes the amount of charge that was carried over from the previous to the next image frame21. In Supplementary Fig. 15c, image lags for different doses after 1 frame (35 ms), 5 frames (175 ms) and 10 frames (350 ms) are shown. In comparison to a-Se and amorphous silicon, MAPbI3 shows higher lag: lags of around 1.2% for amorphous silicon and 0.7% for a-Se are detected after 330 ms (refs. 43,44). The image lag of MAPbI3 after 1 frame is similar to polycrystalline CZT and better for higher frames14. Since the applied electrical field (0.03 V µm−1) on our detector is rather low, we expect a much lower image lag for higher electric fields. The sensitivity varies between 1,010 and 1,060 μC Gyair−1 cm−2 (Supplementary Fig. 16a). A comparison with previously reported detector materials shows that the measured sensitivity is four times higher than CZT (refs. 12,13) and up to 60 times better than a-Se (ref. 10).

Another key parameter for the imager performance is the modulation transfer function (MTF), that is, the spatial resolution or relative response as a function of the spatial frequency44. The MTF plot (Fig. 4c) shows that for the measurement range used here, the MTF is independent of the X-ray dose. For comparing with another X-ray imaging detector mostly used for radiography application, an indirect-conversion complementary metal–oxide–semiconductor detector (Xineos-2222HS, Teledyne DALSA) was chosen. The associated MTF (Fig. 4c, red curve) is obviously worse than that of the MAPbI3 detector. A further comparison of the two detectors is shown in Fig. 4d. The image made by the indirect-conversion detector is blurred, and none of the resolution phantoms can be resolved. In contrast, 5.0 lp mm−1 can be resolved using our MAPbI3 detector.

To prove the good resolution, part of a hearing aid and a coronary stent (diameter, 2 mm; length, 15 mm; rectangular mesh cross section, 100 µm x 200 µm (Supplementary Fig. 17)) were X-rayed with an exposure of 4.17 µGyair per frame in the RQA5 spectrum. The photograph of the objects is shown in Fig. 4e. The resulting X-ray images offer a detailed view of the structures examined; even the mesh of the stent (Fig. 4f, right) is clearly visible. These results show the outstanding potential of the MAPbI3 imaging detector fabricated by our two-step manufacturing process for use in the medical field.

Conclusions

We have reported a two-step manufacturing process for MAPbI3 X-ray flat-panel detectors based on the mechanical sintering of a freestanding absorber layer and integration of this layer on a pixelated backplane. A photoresist grid functions as a mechanical anchor and recrystallized MAPbI3 acts as the adhesion promoter for the soft-sintered thick MAPbI3 absorber layer. We used our approach to create a pixelated X-ray detector with a resolution of 6 lp mm−1, sensitivity of 1,060 μC Gyair−1 cm−2 and detection limit of 0.22 nGyair per frame at an applied electrical field of 0.03 V µm−1. The EHP creation energy W± of 12.4 eV is still higher than the empirical limit of 4.5 eV given by the Klein rule40, but we have shown—via measurements on freestanding wafers—that a W± value of 5.99 eV should be achievable when applying an electrical field higher than 0.05 V µm−1. These freestanding wafers were produced with the same soft-sintering approach as the X-ray detector and confirmed the electrical transport properties of MAPbI3. A value of 9,300 μC Gyair−1 cm−2 could be achieved for sensitivity and 4 × 10–4 cm2 V−1 for the µτ product. We also showed that the W± value of our freestanding MAPbI3 devices is independent of the X-ray photon energy. Hence, the freestanding compact MAPbI3 wafer has a very high potential for stable and excellent detection over the whole energy range of X-ray applications.

We have illustrated that this technology can be scaled to large detection areas via integration on a 508 pixels per inch backplane with 640 × 480 pixels. To further improve the performance (sensitivity and dynamic behaviour) of the pixelated MAPbI3 detector to the level shown for freestanding wafers, the backplane must be tailored to accommodate a higher electric field. Additional interlayer engineering of the detector stack is required to further reduce the dark current. The development of such X-ray detectors with high resolution and sensitivity can—we believe—speed up the translation of AI to routine clinical practice in X-ray imaging applications and help improve healthcare outcomes. In the short term, applications in the field of general radiography, fluoroscopy, angiography and neurology could benefit from improved resolution, which today is an option only in mammography. Moreover, mammography will benefit from improved sensitivities, which are typical for indirect-conversion detectors. Improved sensitivities can also lead to lower X-ray doses but with similar or improved image quality. Currently, these different applications rely on different technologies—our MAPbI3 X-ray imaging detector could potentially provide a single technology for all of them.

Methods

Device manufacturing

For this paper, two different devices were used.

For the first device, freestanding CH3NH3PbI3 (MAPbI3) wafers with a diameter of 15 mm and thickness between 880 and 1,000 μm have been produced using commercially available MAPbI3 powder (Xi’an Polymer Light Technology). The powder was sieved with a 50 μm mesh. The powder was filled into a height-adjustable powder container and a cylinder of stainless steel with a polished surface was placed above this. The hydraulic press (PerkinElmer) can apply up to 9 t, which corresponds to a pressure of 495 MPa for the wafers. The soft sintering of the wafers was done at a pressure of 110 MPa (2 t), applied for 30 min at a temperature of 70 °C. On the two-wafer surfaces, electrodes with an area of 1 cm2 were deposited via physical vapour deposition. For that purpose, 100 nm Cr on one side and 100 nm Pt on the other side were sputtered.

The second device is the X-ray image sensor having the following parts. The backplane consists of a indium gallium zinc oxide thin-film transistor (TFT) array comprising 640 × 480 pixels with a pixel pitch of 50 µm and a resolution of 508 pixels per inch. The negative photoresist grid, with a height of 10 µm, comprised a mixture of SU-8 50 (Kayaku Advanced Materials and distributed by micro resist technology) and cyclopentanone (Sigma-Aldrich) (5:1); it was deposited on the backplane by spin coating at 3,000 rounds per minute and structured by optical lithography. MAPbI3 powder was filled into the grid-like structure and liquefied under a methylamine gas (Sigma-Aldrich) atmosphere. Fabrication of the absorber layer was performed using a slightly modified procedure presented in ref. 1. The wafer was pressed at 9 t (75.5 MPa) and at room temperature for 30 min. On top of this wafer, an 80-nm-thick Cr electrode and 100 nm gold (Au) were sputtered. The manufactured wafer was placed on the liquid MAPbI3 and fixed after its recrystallization. To ensure good evaporation of excess methylamine gas, the sample was placed on a heating plate at 50 °C for 30 min. The stacked layers were encapsulated by laminating a high barrier film (TESA 61572) to avoid degradation due to external stimuli.

Image readout and processing for pixelated X-ray detector

Images from the 480 × 640 pixels TFT panel were read by a commercially available ROIC (AD71124, Analog Devices). The signal at the input was simultaneously integrated, amplified, low-pass filtered and converted from analogue to digital with a 16-bit converter. The integrator feedback capacitance Cf was 0.125 pF and the integration time was 35 ms. The readout occurred by the rolling shutter principle, and the X-ray tube and detector were not synchronized. The acquiring frame rate was ~28.6 frames per second. To eliminate fixed pattern noise (offset compensation), an offset map was generated by averaging 100 dark images, which was subtracted from the images. The mean value and standard deviation were deduced from the ROI (Supplementary Fig. 18). The standard deviation in the ROI represents the electronic noise in LSB (~11 LSB). Noise in the image sensor in terms of electrons (considering 48 electrons per LSB) is 535 electrons. X-ray recordings from objects were offset and compensated by subtracting the offset map and flat-field corrected by calculating the gain factor of each pixel for a flat-field image. The MTF is determined by the slanted-edge method45. First, an object with a sharp tungsten edge is placed on the X-ray detector and the edge profile is derived from the resulting X-ray image. The line-spread function is derived by differentiating the edge profile. The Fourier transform of the line-spread function defines the MTF.

Electrical characterization of freestanding wafers

The current density measurement was performed with a Keithley 2400 source meter sampling at 10 Hz that was connected to the sample holder filled with argon gas. The MAPbI3 wafers are measured in an inert atmosphere and therefore protected from moisture and light. For measurement during X-ray exposure, the sample holder is placed in an Al box underneath a MEGALIX Cat Plus 125/40/90 (Siemens Healthineers AG) X-ray source with a tungsten anode. The distance between the sample and X-ray source is 120 cm. The X-ray dose was varied by changing the tube current by over two orders of magnitude, measured with a PTW Diados T11003-001896 dosimeter and adjusted with correction factors provided in its datasheet for anode voltages other than 70 kVp. For more details, see Supplementary Information.

SEM

To obtain high-resolution SEM images of the MAPbI3 perovskite layers, a Schottky field-emission SEM (JEOL JSM-7610F) was used at an acceleration voltage of 15 kV. For the cross-sectional image, the glass of the imager was scratched with a diamond pen and then broken by hand. SEM images of the grid photoresist were obtained by an FEI Quanta 3D FEG microscope, using a 5 kV electron beam and a secondary electron detector.

X-ray diffraction measurement

The structural analysis of the MAPbI3 powder consists of XRD measurements performed by classical ex situ Bragg–Brentano geometry using a Panalytical X’pert powder diffractometer with filtered Cu Kα radiation and an X’Celerator solid-state strip detector.