Abstract
Aims/hypothesis
We aimed to compare the effects of intermittently scanned continuous glucose monitoring (isCGM) and carbohydrate counting with automated bolus calculation (ABC) with usual care.
Methods
In a randomised, controlled, open-label trial carried out at five diabetes clinics in the Capital Region of Denmark, 170 adults with type 1 diabetes for ≥1 year, multiple daily insulin injections and HbA1c > 53 mmol/mol (7.0%) were randomly assigned 1:1:1:1 with centrally prepared envelopes to usual care (n = 42), ABC (n = 41), isCGM (n = 48) or ABC+isCGM (n = 39). Blinded continuous glucose monitoring data, HbA1c and patient-reported outcomes were recorded at baseline and after 26 weeks. The primary outcome was change in time in range using isCGM vs usual care.
Results
Baseline characteristics were comparable across arms: mean age 47 (SD 13.7) years, median (IQR) diabetes duration 18 (10–28) years and HbA1c 65 (61–72) mmol/mol (8.1% [7.7–8.7%]). Change in time in range using isCGM was comparable to usual care (% difference of 3.9 [−12–23], p = 0.660). The same was true for the ABC and ABC+isCGM arms and for hypo- and hyperglycaemia. Also compared with usual care, using ABC+isCGM reduced HbA1c (4 [95% CI 1, 8] mmol/mol) (0.4 [0.1, 0.7] %-point) and glucose CV (11% [4%, 17%]) and improved treatment satisfaction, psychosocial self-efficacy and present life quality. Treatment satisfaction also improved by using isCGM alone vs usual care. Statistical significance was maintained after multiple testing adjustment concerning glucose CV and treatment satisfaction with ABC+isCGM, and treatment satisfaction with isCGM. Discontinuation was most common among ABC only users, and among completers the ABC was used 4 (2–5) times/day and the number of daily isCGM scans was 5 (1–7) at study end.
Conclusions/interpretation
isCGM alone did not improve time in range, but treatment satisfaction increased in technology-naive people with type 1 diabetes and suboptimal HbA1c. The combination of ABC+isCGM appears advantageous regarding glycaemic variables and patient-reported outcomes, but many showed resistance towards ABC.
Trial registration
ClinicalTrials.gov NCT03682237.
Funding
The study is investigator initiated and financed by the Capital Region of Denmark.
Graphical abstract
Similar content being viewed by others
Introduction
Adjustment of bolus insulin dosage based on carbohydrate intake and frequent blood glucose monitoring (BGM) or continuous glucose monitoring (CGM) are suggested approaches to obtain optimal glycaemic control in type 1 diabetes treated with multiple daily insulin injections (MDI) [1,2,3,4,5].
Advanced carbohydrate counting, defined as bolus calculation based on carbohydrate ratios and insulin sensitivity [6], is known to improve HbA1c and treatment satisfaction compared with experience-based insulin bolus dosing [7,8,9]. Training in automated bolus calculation (ABC), rather than manual bolus calculation, may be even more efficient [8]. Conflicting results are available regarding influence on hypoglycaemia rate [10]. Sustained HbA1c reduction after introduction of ABC has been demonstrated even in a routine care setting [11].
Intermittently scanned CGM (isCGM) provides CGM data on request but has, until recently, had no alarms. Several studies have evaluated isCGM in terms of glycaemic effectiveness and person-related outcomes [12,13,14,15,16]. IsCGM has been found to decrease hypoglycaemia and glycaemic variability and increase treatment satisfaction in people with near-optimal HbA1c [17]. Predictors of HbA1c reduction are high baseline HbA1c [14, 16, 18,19,20] and frequent scans [18, 21].
Despite promising effects of several interventions, only a minority of people with type 1 diabetes achieve glycaemic goals [22]. While ABC presupposes user engagement and skills, isCGM is easily usable and widely desired. In our local setting, ABC is broadly introduced, while isCGM is only reimbursed in certain subpopulations. On this background, we intended to investigate the abovementioned interventions in the vast group of people with type 1 diabetes who have not yet obtained target glycaemic control, practised ABC or fulfilled the strict Danish criteria for isCGM reimbursement.
The primary aim of this trial therefore was to examine the effect of isCGM on glycaemic control compared with usual care, i.e. experience-based insulin bolus decision and BGM, in people with suboptimal HbA1c and naive to diabetes technology and advanced carbohydrate counting. In addition, we aimed to assess patient-reported outcomes among isCGM users compared with those allocated usual care, whether there is an additive effect of introducing ABC on top of isCGM and whether isCGM is superior to ABC.
Methods
Study design and participants
The study protocol was previously published, including details on usual care, interventional devices, screening and randomisation procedures, etc. [23].
The study is a randomised, controlled, multicentre, open-label trial carried out in five diabetes clinics in the Capital Region of Denmark. Inclusion criteria were age ≥ 18 years, type 1 diabetes ≥1 year, HbA1c > 53 mmol/mol (7.0%) and MDI with basal insulin ≥30% of total insulin dose. Exclusion criteria were ongoing practice of daily advanced carbohydrate counting, CGM/isCGM, NPH insulin, pregnancy, breastfeeding, gastroparesis, severe diabetes complications (i.e. proliferative retinopathy, myocardial infarction within the last 6 months), other medical or psychological conditions judged unsuitable for participation, participation in other diabetes-related clinical research projects, other drugs than insulin affecting glucose metabolism or the inability to understand the information and to give informed consent. The inclusion period was 1 October 2018 to 30 April 2020.
The trial was carried out in accordance with the Helsinki Declaration after approval by the Capital Regional Scientific Ethics Committee (H-17040573). The data collection was performed in accordance with the General Data Protection Regulation (P-2019-107). The trial is registered at ClinicalTrials.gov (registration no. NCT03682237).
Screening and randomisation
Possible candidates with interest in participation attended a screening visit, and those eligible were randomised 1:1:1:1 to usual care, ABC, isCGM or ABC+isCGM, with centrally prepared envelopes left unopened until after baseline blinded CGM data had been obtained. In the case of dropout, the participant’s allocation was not replaced [23].
Intervention
All participants attended a group course (4.5 h duration, 4–6 participants, week 0), five individual study visits (weeks 2, 4, 12, 24 and 26) and two telephone consultations (weeks 8 and 17) according to treatment allocation (electronic supplementary material [ESM] Fig. 1) [23]. All group courses were held by a study nurse and a study dietitian, each contributing to their areas of expertise. The screening and the midway visits were held with a study physician. Participants allocated ABC had a joint nurse–dietitian appointment for the first follow-up visit. The remaining study visits were held by a nurse solely for all participants. Participants not allocated ABC received common dietary advice for people with diabetes but were not taught bolus calculation based on carbohydrate ratios and insulin sensitivity. The ABC device selected was the Food and Drug Administration (FDA)-approved and Conformité Européenne (CE)-marked mobile phone application mySugr (mySugr, Austria) consisting of a bolus calculator adjustable on carbohydrate ratio, insulin sensitivity, target glucose value and insulin duration time. Before meals, the current glucose value and estimated carbohydrate intake are entered, and the user receives the suggested insulin bolus dose. Similarly, to correct hyperglycaemia outside meals, the user enters the current glucose value and receives the suggested corrective insulin dose. The isCGM device was the FDA-approved and CE-marked FreeStyle Libre Flash CGM system (Abbott Diabetes Care, Alameda, CA, USA), consisting of an upper arm sensor with a 14 day lifespan and a reading device (either an independent reader or a mobile phone application: FreeStyle Libre Link, Abbott Diabetes Care). The user scans the sensor with the reading device for on-demand values. Automatically obtained glucose values are continuously gathered for later review. The isCGM has a built-in bolus calculator, which can be used only with BGM measurements and not with scanned isCGM sensor values, which is why another bolus calculator was chosen for the study.
Hence, the group course elements were: general diabetes education (usual care and isCGM arms); training in carbohydrate counting, bolus calculation and mySugr application (ABC and ABC+isCGM arms); training in FreeStyle Libre Flash (isCGM and ABC+isCGM arms); and training in how to incorporate glucose trend arrows to adjust experience-based bolus dosing (isCGM arm) or the mySugr application settings (ABC+isCGM arm). Participants were also encouraged to continuously refer to isCGM data, i.e. current and average sensor glucose, time in range (TIR), as well as hypoglycaemic events on a weekly basis, and to adjust insulin or bolus calculator settings if needed. These different elements were also in focus at all study visits, where the study personnel titrated insulin doses based on the different types of glucose values, information on hypoglycaemic events, planned physical activity, etc. The overall aim was fasting and preprandial glucose values 4–6 mmol/l, postprandial glucose values <10 mmol/l, to prevent glucose values <3 mmol/l and to minimise glucose values <4 mmol/l [24]. Participants were encouraged to perform CGM scans ≥5 times daily (isCGM and ABC+isCGM arms, to decrease the risk of missing values, unless scanned every 8 h) or BGM ≥4 times daily (usual care and ABC arms, to secure a glucose measurement prior to all main meal boluses and before bedtime) and to use the ABC before carbohydrate intake and to correct hyperglycaemia (ABC and ABC+isCGM arms). Basal and bolus insulin doses, and ABC settings for those allocated ABC, were repeatedly evaluated during participation and adjusted if needed [6, 8]. Study healthcare professionals had received adequate education in the different group course contents and treatment modalities before first screening [23].
Sample size
The intervention was designed to have 80% power to detect a difference in mean TIR between isCGM and usual care arms of 75 min/day [25] with an SD of 120 min [2], and a two-sided α level of 0.05. A sample size of 160 (40 per arm) was needed and was further increased to 180 to account for discontinuation.
Data collection
At baseline and study end the participants wore a 14 day blinded CGM device (the FreeStyle Libre Professional CGM system, Abbott Diabetes Care, Oxon, UK) (participants in the isCGM arms wore both types of sensors for the last 2 weeks of participation), had blood and urine samples taken, had HbA1c locally analysed with high-performance liquid chromatography on a Tosoh G7 (Tosoh Bioscience, Japan), had body weight measured and filled in work sheets for daily reporting of insulin doses and the following questionnaires: the Problem Areas in Diabetes Questionnaire (PAID), the Diabetes Treatment Satisfaction Questionnaire (‘status’ and ‘change’ versions; DTSQs and DTSQc) [26], the Diabetes Empowerment Scale (DES short form) and Diabetes quality of life (ADDQoL-19) [27, 28]. At baseline, hypoglycaemia awareness was also assessed [29, 30]. IsCGM (scans/day) and ABC (application use/day) data and the number of weekly mild symptomatic hypoglycaemic episodes were collected at 2, 12 and 26 weeks. Adverse events were consecutively recorded.
Outcome measures
The primary outcome was the difference in change from baseline to study end between isCGM and usual care arms in TIR (minutes/day, obtained by blinded CGM). Due to consensus guidelines on glucose values obtained with CGM published after trial initiation, the range for TIR was redefined from 4–10 to 3.9–10 mmol/l [25].
Predefined and analysed secondary outcomes are: difference between groups in change in HbA1c (mmol/mol, %), TIR, time in hypo- (<3, <3.9 mmol/l) and hyperglycaemia (>10 mmol/l) (minutes/day), glycaemic variability (SD), total insulin dose (U kg-1 day-1), basal insulin dose (U kg-1 day-1), body weight (kg), urinary albumin/excretion rate (normo-, micro- and macroalbuminuria), diabetes distress, diabetes treatment satisfaction, diabetes psychosocial self-efficacy, diabetes quality of life and last mild symptomatic hypoglycaemia (<3 mmol/l, events/week) throughout participation.
In addition to the predefined outcomes, blinded CGM mean glucose (mmol/l), time in hyperglycaemia (>13.9 mmol/l), glucose CV, and TIR separately for daytime (06:00 hours to midnight) and night-time (midnight to 06:00 hours) [31] have also been analysed.
Statistical analysis
Changes in outcomes within arms over the intervention period and in comparison with the control arm (effects of the treatments) were modelled by linear mixed-effects models with random intercepts for participants and centre. Treatment arms and visits were included as fixed effects. The baseline level of the outcome was retained as an outcome in the model. The exact times of measurements were used. Outcomes for which the model residuals were not normally distributed were log-transformed prior to analyses. In the analysis of body weight, we adjusted for sex. Analyses were performed as intention to treat including all available data. Per-protocol (defined as ≥5 daily isCGM scans and ≥3 daily ABC insulin administrations) analyses were further performed for TIR, time in hypoglycaemia (<3.9 mmol/l), time in hyperglycaemia (>10 mmol/l) and HbA1c. Statistical significance was inferred at a two-tailed p < 0.05. The original p values are shown. Statistical significance for the secondary outcomes after adjustment for multiple testing (Benjamini–Hochberg method) [32] is also reported. Statistical analyses were conducted in R version 3.6.1 (The R Foundation for Statistical Computing, www.R-project.org) and SAS version 9.4 (SAS Institute, Cary, NC, USA).
Results
Population characteristics
In total, 184 MDI-treated adults with type 1 diabetes attended the screening visit, whereof 14 (8%) did not meet inclusion criteria due to HbA1c ≤ 53 mmol/mol (7.0%) (n = 8), use of glucose-lowering drugs other than insulin (n = 4), severe diabetes complications (n = 1) and conditions judged unsuitable for participation (n = 1).
The 170 eligible people were randomly assigned to usual care (n = 42), ABC (n = 41), isCGM (n = 48) or ABC+isCGM (n = 39). The slightly unbalanced numbers were due to unopened randomisation envelopes. The participants were followed at the diabetes centres: Steno Diabetes Center Copenhagen (55%); Nordsjællands Hospital, Hillerød (19%); Bispebjerg and Frederiksberg Hospital (14%); Amager and Hvidovre Hospital (9%); and Rigshospitalet (4%).
Baseline characteristics were comparable across arms: median (IQR) diabetes duration 18 (10–28) years, HbA1c 65 (61–72) mmol/mol (8.1% [7.7–8.7%]), mean BMI 26.7 (SD 3.9) kg/m2, age 47 (13.7) years and sex (63% male participants), apart from higher prevalence of antihypertensive medication (p = 0.010) and lower prevalence of unaware participants according to the Hillerød method (p = 0.013) in the ABC+isCGM arm (Table 1). The proportion of participants with TIR >70% at baseline was 5%, 3%, 13% and 0% for usual care, ABC, isCGM and isCGM+ABC (p = 0.056 for group difference).
Discontinuation and compliance
The discontinuation rate was 17% in the whole cohort. The rate was higher in the ABC arm (39%) compared with usual care (17%), isCGM (4%) and ABC+isCGM (10%) (p ≤ 0.026). The most frequent reasons for withdrawal from the ABC arm were time demands, workload and technical challenges (ESM Fig. 1). The ABC was used for bolus calculation 4 (3–6) times daily in the ABC arm and 3 (1–5) times daily in the ABC+isCGM arm at study end without group difference (4 [2–5] times/day at study end among all participants allocated ABC). The numbers of daily isCGM scans were 9 (7–12), 8 (6–12) and 5 (1–7) at 2 weeks after initiation, midway and at study end (isCGM and ABC+isCGM arms), without differences between groups, but with reduction over time in both arms (p < 0.001). The 141 completed participants were followed for 26 (24–28) weeks. The proportions of per-protocol users among the completers in the different intervention arms were 84% (ABC), 93% (isCGM) and 66% (isCGM+ABC). In the isCGM+ABC arm, 66% used the ABC and 97% used the isCGM per-protocol. Recording durations for blinded CGM were 13 (12–14) and 14 (12–14) days at baseline and study end.
Time in ranges
There was no significant improvement in the primary endpoint change in TIR from baseline to study end in the isCGM arm vs the usual care arm (% difference of 3.9 [−12–23], p = 0.660) (Fig. 1). The same was true for ABC and ABC+isCGM compared with usual care (p ≥ 0.061). Changes in time in hypo- and hyperglycaemia and in mean glucose were also comparable to usual care in all three intervention arms (p ≥ 0.105). Daytime TIR increased among ABC only participants (% difference of 22 [95% CI 1, 47]) and night-time hyperglycaemia increased among isCGM only participants (% difference of 258 [40, 814]) compared with usual care. In per-protocol analyses, TIR increased (% difference of 25 [2, 54]), but time in hypoglycaemia (<3.9 mmol/l) did not change significantly (p = 0.057) among ABC only users compared with usual care.
Findings on changes in TIR and time in hypo- (<3.9 mmol/l) and hyperglycaemia (>10 mmol/l) in those allocated ABC+isCGM vs those allocated isCGM only were comparable (p ≥ 0.241). Similarly, the results were comparable in those allocated isCGM only vs ABC only (p ≥ 0.118). The results on TIR and time in hypo- and hyperglycaemia among all participants using isCGM (isCGM or ABC+isCGM) were also comparable to those in the usual care arm (p ≥ 0.294). The proportion of participants with TIR >70% at study end was comparable across arms (9%, 16%, 20% and 17%; p = 0.594).
Other blinded CGM-derived endpoints and HbA1c
Compared with usual care, glycaemic variability (% difference of 11 [1, 20]) and glucose CV (% difference of 11 [4, 17]) decreased with ABC+isCGM, but did not change in the two other intervention arms compared with usual care. Mean glucose did not change in the intervention arms compared with usual care (p ≥ 0.314).
Compared with usual care, HbA1c decreased in the ABC+isCGM arm (4 [1, 8] mmol/mol) (0.4 [0.1, 0.7] %-point), but not among ABC or isCGM only participants (p ≥ 0.184). Similar results were found in the per-protocol analysis. The HbA1c reduction among all participants using isCGM (isCGM or ABC+isCGM) was greater than among those in the usual care arm (3 [0, 6] mmol/mol; 0.3 [0.0, 0.6] %-point; p = 0.049).
Patient-reported outcomes
Compared with usual care, treatment satisfaction (DTSQs, DTSQc) improved in the isCGM (5, 6 units; p < 0.001) and ABC+isCGM (4, 4 units; p < 0.012) arms. Psychosocial self-efficacy (0.3, p = 0.014), ADDQoL present quality of life (0.5, p = 0.018) and DTSQc-perceived frequency of hyperglycaemia (−1.2, p = 0.003) improved among ABC+isCGM vs usual care participants. The total ADDQoL score worsened in the isCGM arm (−0.5, p = 0.017), and, when solely analysing the ‘dietary freedom’ domain, the figure for change in impact on quality of life in the isCGM arm was −1.1 (p = 0.066) vs usual care.
Within-group changes
There were several significant within-group improvements from baseline to study end in glycaemic and patient-reported outcomes (Table 2, Fig. 2, ESM Table 1).
Adjustment for multiple testing
For the secondary outcomes, statistical significance of the intervention was maintained after adjustment for multiple testing with regard to glucose CV, treatment satisfaction (DTSQs and DTSQc) and DTSQc-perceived frequency of hyperglycaemia among ABC+isCGM participants, and treatment satisfaction (DTSQc) was maintained among isCGM only participants, compared with usual care. For other predefined secondary outcomes and additional analyses, statistical significance was lost when adjusting for multiple testing.
Safety
The change in mild hypoglycaemia episodes from baseline to study end in the intervention arms was comparable to usual care. The numbers of reported mild symptomatic hypoglycaemic weekly events were 2 (1–4) (usual care), 2 (1–4) (ABC), 5 (2.5–8) (isCGM) and 3 (1–7) (ABC+isCGM) (p < 0.001) 2 weeks after intervention initiation. The corresponding values for the whole cohort midway and at study end were 2 (1–3.5) and 3 (1–4), without differences across arms. There were six reported severe hypoglycaemic events (one event occurred before initiation of intervention) in five participants allocated usual care (n = 1), ABC (n = 2) and isCGM (n = 2). All participants fully recovered. There were no cases of ketoacidosis.
Discussion
In this large, randomised, controlled, multicentre trial, we approached MDI-treated people with type 1 diabetes, with suboptimal glycaemic control and naive to insulin pumps, CGM and advanced carbohydrate counting. Three different treatment approaches were evaluated compared with usual care: ABC, isCGM and the two combined. None of the treatment approaches led to a significant increase in TIR compared with usual care as measured by blinded CGM for 2 weeks at baseline and study end. Combined intervention with ABC+isCGM improved glucose CV and treatment satisfaction compared with usual care and resulted in a clinically relevant HbA1c reduction over time.
The main strength of our study is the robust design with large numbers, as well as the multicentre approach. Multicomponent interventions are probably pivotal in striving for glycaemic control, especially among those far from target. To our knowledge, the effect of a possible additive effect of ABC on isCGM has not yet been examined by others. However, the four-arm set-up with single and combined interventions may lead to interpretation difficulties. The main limitations are the recruitment difficulties and the higher than anticipated discontinuation rate in the ABC arm, resulting in a slightly underpowered arm. Other important weaknesses are the lack of TIR or HbA1c stratification at inclusion, resulting in somewhat unbalanced groups. Further, blinded CGM sampling for 2 weeks two times probably does not truly reflect blood glucose levels during the entire participation period. Whether participants’ behaviour was affected by blinded CGM wear, as indicated by a somewhat lower mean glucose than reflected by HbA1c, remains speculative.
We could not demonstrate any improvement in TIR by using isCGM vs usual care. To our knowledge, there are no other RCTs that have obtained blinded CGM data in their evaluation of isCGM efficiency. In our study, HbA1c was reduced by 3 mmol/mol (0.3%-point) among isCGM users. Recently, nationwide register-based studies from Sweden [33, 34], the Netherlands [19], Belgium [35] and the UK [18] including 1050–14,450 people with type 1 diabetes and baseline HbA1c of 62–68 mmol/mol (7.8–8.4%) were published. HbA1c reductions were found 8–24 months after isCGM initiation in four studies; however, the actual figures were rather modest: 1 [34], 3 [33], 5 [18] and 3 [19] mmol/mol (0.1, 0.3, 0.5 and 0.3%-point), and in the fifth study HbA1c was maintained [35]. Since we had provided an educational programme with structured group courses and close follow-up by an experienced team, we had probably expected greater HbA1c reduction by using isCGM. Our study population is characterised by people not obtaining the recommended HbA1c target but not necessarily with pronounced dysregulation, with whom one expects greater HbA1c reductions. In similar populations, greater effects on glycaemic variables have been found with real-time CGM with alarms [2, 3]. In our population, the HbA1c reduction in all isCGM users was greater than in the usual care arm, but then the effect of ABC was not taken into consideration, and the result lost its significance after multiple testing adjustments. Also, the recommended number and the actual figures of daily scans in the isCGM arms were lower than recent studies have demonstrated are needed to obtain target HbA1c [18, 21, 36]. Our finding of a reduced scanning frequency over time highlights the importance of adequate support to persistent compliance.
Existing data on severe hypoglycaemia and hospitalisations after initiation of isCGM are somewhat inconsistent, since there are reports on reductions [34, 35] but also a fourfold increase in hypoglycaemia admissions [16]. A newly published RCT in youth with poor glycaemic control reports increased glucose testing and treatment satisfaction with isCGM compared with BGM, but no translation to glycaemic improvement [37]. Others report moderate increases in mild hypoglycaemic events [14, 15], but fewer diabetic [2, 3] and ketoacidosis admissions [14], improved treatment satisfaction [17, 35, 38] and hypoglycaemia awareness [18], and less work absenteeism [19, 35] and diabetes distress [18] after initiating isCGM.
In our trial, only very few withdrew after isCGM allocation, and treatment satisfaction was markedly improved in the isCGM arm compared with usual care. Therefore, the initial sign of increased burden of diabetes on quality of life in the isCGM arm was surprising, but also lost its significance after multiple testing adjustments. One could speculate on whether the result may reflect a lack of tools to act on glucose data provided by isCGM. It is encouraging, however, that treatment satisfaction, psychosocial self-efficacy and present quality of life were improved in the combined intervention arm compared with usual care, although only treatment satisfaction maintained its significance after multiple testing adjustments. These findings might reflect the convenience of easy access to glucose values simultaneously with carbohydrate counting skills and a device for insulin bolus advice.
Our finding of time in hypoglycaemia of 7–11% at both baseline and study end is higher than recommended by consensus [25]. This is likely to be explained by the relatively limited sensor accuracy in the lower range [39]. This may also explain the relatively high symptomatic hypoglycaemia figures shortly after initiation of isCGM, as the open isCGM consists of the same sensor; we cannot exclude, however, that isCGM reveals otherwise unidentified hypoglycaemia.
Our clinical impression is that far too few in the Danish MDI-treated type 1 diabetes population have been trained in or perform daily bolus calculation based on carbohydrate ratios and insulin sensitivity, even though nutritional therapy has an integral role in diabetes management [40]. We found discontinuation rates lower than expected except among those allocated to ABC only. The workload of carbohydrate counting and the time and technical skills to seek bolus advice before administering rapid-acting insulin several times a day probably influence compliance. On the other hand, those continuing participation in the ABC arm reported improved treatment satisfaction and HbA1c was reduced. Similarly, among those allocated to ABC+isCGM, every third participant discontinued or reduced the use of ABC. Whether the workload of ABC+isCGM could be minimised with integrated tools remains speculative. Furthermore, the chosen ABC device did not hold features allowing bolus adjustments in relation to physical exercise, stress or illness, which are likely to be useful.
The aim of including ABC in this study was to evaluate its effect on top of isCGM and to compare ABC with isCGM. The group course and follow-up content, and how to incorporate isCGM data when adjusting the ABC settings, were developed for this specific study, based on manufacturer guidelines and in line with contemporary publications [41,42,43]. Whereas isCGM enables glucose testing and information on glucose trends and otherwise unrevealed glucose levels, ABC primarily entails focus on postprandial glucose values and may therefore be viewed as a factual insulin intervention. Introducing isCGM typically discloses low glucose values or declining glucose trends during sleep, which may explain the reduction of basal insulin dose and the increase in night-time hyperglycaemia in the isCGM arm, although these findings lost significance after multiple testing adjustments. It also seems as if ABC, at least among those that have the resources to perform it properly over time, has an overall, but especially daytime, glucose stabilising effect compared with usual care. We have not specifically investigated postprandial glucose values across arms in this study.
The study population is likely to be recognised by diabetes healthcare professionals worldwide, as people with type 1 diabetes not practising advanced carbohydrate counting and without access to isCGM/CGM are still common, due to limited resources either among users or within the healthcare system.
The key message is that although no improvement in TIR by isCGM alone was found, treatment satisfaction was markedly improved compared with usual care, and HbA1c was reduced by 3 mmol/mol (0.3%-point) over time, in people with type 1 diabetes with suboptimal HbA1c and naive to CGM and advanced carbohydrate counting. Combined intervention with ABC+isCGM improved glucose CV and treatment satisfaction compared with usual care and HbA1c was reduced by 5 mmol/mol (0.5%-point) over time. Discontinuation was most common among ABC only users. The two different interventions are shown to act on different aspects of blood glucose levels. The combined intervention appears advantageous, even though a considerable number can be expected to show resistance towards ABC.
Individualised preferences are likely to further improve the detected effects, but we still ought to consider other actions to obtain target glucose levels. Some may benefit from motivational counselling whereas others may profit from using insulin pumps, CGMs with alarms, automated devices such as hybrid closed-loop systems or new insulins. Future studies on alternative initiatives for people with long-lasting and MDI-treated type 1 diabetes, without reaching target glucose levels, are warranted.
Data availability
The data that support the findings of this study are available from the corresponding author upon reasonable request.
Change history
15 November 2021
A Correction to this paper has been published: https://doi.org/10.1007/s00125-021-05601-5
Abbreviations
- ABC:
-
Automated bolus calculation
- ADDQoL-19:
-
Diabetes quality of life (scale)
- BGM:
-
Blood glucose monitoring
- CE:
-
Conformité Européenne
- CGM:
-
Continuous glucose monitoring
- DES:
-
Diabetes Empowerment Scale
- DTSQc:
-
Diabetes Treatment Satisfaction Questionnaire change version
- DTSQs:
-
Diabetes Treatment Satisfaction Questionnaire status version
- FDA:
-
Food and Drug Administration
- isCGM:
-
Intermittently scanned continuous glucose monitoring
- MDI:
-
Multiple daily insulin injections
- TIR:
-
Time in range
References
Beck RW, Bergenstal RM, Laffel LM, Pickup JC (2019) Advances in technology for management of type 1 diabetes. Lancet 394(10205):1265–1273. https://doi.org/10.1016/S0140-6736(19)31142-0
Beck RW, Riddlesworth T, Ruedy K et al (2017) Effect of continuous glucose monitoring on glycemic control in adults with type 1 diabetes using insulin injections. JAMA 317(4):371–378. https://doi.org/10.1001/jama.2016.19975
Lind M, Polonsky W, Hirsch IB et al (2017) Continuous glucose monitoring vs conventional therapy for glycemic control in adults with type 1 diabetes treated with multiple daily insulin injections. JAMA 317(4):379–387. https://doi.org/10.1001/jama.2016.19976
American Diabetes Association (2021) Diabetes technology: standards of medical care in diabetes−2021. Diabetes Care 44:S85–S99. https://doi.org/10.2337/dc21-S007
American Diabetes Association (2021) Glycemic targets: standards of medical care in diabetes−2021. Diabetes Care 44:S73–S84. https://doi.org/10.2337/dc21-S006
Schmidt S, Norgaard K (2014) Bolus calculators. J Diabetes Sci Technol 8:1035–1041. https://doi.org/10.1177/1932296814532906
Schmidt S, Schelde B, Nørgaard K (2014) Effects of advanced carbohydrate counting in patients with type 1 diabetes: a systematic review. Diabet Med 31(8):886–896. https://doi.org/10.1111/dme.12446
Schmidt S, Meldgaard M, Serifovski N et al (2012) Use of an automated bolus calculator in MDI-treated type 1 diabetes: the BolusCal study, a randomized controlled pilot study. Diabetes Care 35(5):984–990. https://doi.org/10.2337/dc11-2044
Bell KJ, Barclay AW, Petocz P, Colagiuri S, Brand-Miller JC (2014) Efficacy of carbohydrate counting in type 1 diabetes: a systematic review and meta-analysis. Lancet Diabetes Endocrinol 2(2):133–140. https://doi.org/10.1016/S2213-8587(13)70144-X
Hommel E, Schmidt S, Vistisen D et al (2017) Effects of advanced carbohydrate counting guided by an automated bolus calculator in type 1 diabetes mellitus (StenoABC): a 12-month, randomized clinical trial. Diabet Med 34(5):708–715. https://doi.org/10.1111/dme.13275
Meldgaard M, Damm-Frydenberg C, Vesth U, Nørgaard K, Schmidt S (2015) Use of advanced carbohydrate counting and an automated bolus calculator in clinical practice: the BolusCal ® training concept. Int Diabetes Nurs 12(1):8–13. https://doi.org/10.1179/2057331615z.0000000002
Cowart K, Updike W, Bullers K (2020) Systematic review of randomized controlled trials evaluating glycemic efficacy and patient satisfaction of intermittent-scanned continuous glucose monitoring in patients with diabetes. Diabetes Technol Ther 22:337–345. https://doi.org/10.1089/dia.2019.0345
Leelarathna L, Wilmot EG (2018) Flash forward: a review of flash glucose monitoring. Diabet Med 35(4):472–482. https://doi.org/10.1111/dme.13584
Tyndall V, Stimson RH, Zammitt NN et al (2019) Marked improvement in HbA1c following commencement of flash glucose monitoring in people with type 1 diabetes. Diabetologia 62(8):1349–1356. https://doi.org/10.1007/s00125-019-4894-1
Paris I, Henry C, Pirard F, Gérard A-C, Colin IM (2018) The new FreeStyle libre flash glucose monitoring system improves the glycaemic control in a cohort of people with type 1 diabetes followed in real-life conditions over a period of one year. Endocrinol Diabetes Metab 1(3):e00023. https://doi.org/10.1002/edm2.23
Stimson RH, Dover AR, Ritchie SA et al (2020) HbA1c response and hospital admissions following commencement of flash glucose monitoring in adults with type 1 diabetes. BMJ Open Diabetes Res Care 8(1):e001292. https://doi.org/10.1136/bmjdrc-2020-001292
Bolinder J, Antuna R, Geelhoed-Duijvestijn P, Kröger J, Weitgasser R (2016) Novel glucose-sensing technology and hypoglycaemia in type 1 diabetes: a multicentre, non-masked, randomised controlled trial. Lancet 388(10057):2254–2263. https://doi.org/10.1016/S0140-6736(16)31535-5
Deshmukh H, Wilmot EG, Gregory R et al (2020) Effect of flash glucose monitoring on glycemic control, hypoglycemia, diabetes-related distress, and resource utilization in the Association of British Clinical Diabetologists (ABCD) nationwide audit. Diabetes Care 43(9):2153–2160. https://doi.org/10.2337/dc20-0738
Fokkert M, Van Dijk P, Edens M et al (2019) Improved well-being and decreased disease burden after 1-year use of flash glucose monitoring (FLARE-NL4). BMJ Open Diabetes Res Care 7:e000809. https://doi.org/10.1136/bmjdrc-2019-000809
Rouhard S, Buysschaert M, Alexopoulou O, Preumont V (2020) Impact of flash glucose monitoring on glycaemic control and quality of life in patients with type 1 diabetes: a 18-month follow-up in real life. Diabetes Metab Syndr Clin Res Rev 14(2):65–69. https://doi.org/10.1016/j.dsx.2019.12.007
Laurenzi A, Caretto A, Barrasso M et al (2020) Frequency of flash glucose monitoring readings, hemoglobin A1c and time in range: a real life study in adults with type 1 diabetes. Acta Diabetol 57(11):1395–1397. https://doi.org/10.1007/s00592-020-01577-7
Foster NC, Beck RW, Miller KM et al (2019) State of type 1 diabetes management and outcomes from the T1D exchange in 2016-2018. Diabetes Technol Ther 21:66–72. https://doi.org/10.1089/dia.2018.0384
Secher AL, Pedersen-Bjergaard U, Svendsen OL et al (2020) Study protocol for optimising glycaemic control in type 1 diabetes treated with multiple daily insulin injections: intermittently scanned continuous glucose monitoring, carbohydrate counting with automated bolus calculation, or both? A randomised controlled trial. BMJ Open 10(4):e036474. https://doi.org/10.1136/bmjopen-2019-036474
Danish Endocrine Society. Type 1 diabetes. Available from https://endocrinology.dk/nbv/diabetes-melitus/type-1-diabetes-mellitus/. Accessed 6 Apr 2021[website in Danish]
Battelino T, Danne T, Bergenstal RM et al (2019) Clinical targets for continuous glucose monitoring data interpretation: recommendations from the international consensus on time in range. Diabetes Care 42(8):1593–1603. https://doi.org/10.2337/dci19-0028
Bradley C, Plowright R, Stewart J, Valentine J, Witthaus E (2007) The Diabetes Treatment Satisfaction Questionnaire change version (DTSQc) evaluated in insulin glargine trials shows greater responsiveness to improvements than the original DTSQ. Health Qual Life Outcomes 5(1):57. https://doi.org/10.1186/1477-7525-5-57
Bradley C, Todd C, Gorton T, Symonds E, Martin A, Plowright R (1999) The development of an individualized questionnaire measure of perceived impact of diabetes on quality of life: the ADDQoL. Qual Life Res 8(1–2):79–91. https://doi.org/10.1023/A:1026485130100
Wee HL, Tan CE, Goh SY, Li SC (2006) Usefulness of the Audit of Diabetes-Dependent Quality-of-Life (ADDQoL) questionnaire in patients with diabetes in a multi-ethnic Asian country. Pharmacoeconomics 24(7):673–682. https://doi.org/10.2165/00019053-200624070-00006
Høi-Hansen T, Pedersen-Bjergaard U, Thorsteinsson B (2010) Classification of hypoglycemia awareness in people with type 1 diabetes in clinical practice. J Diabetes Complicat 24(6):392–397. https://doi.org/10.1016/j.jdiacomp.2009.07.006
Pedersen-Bjergaard U, Pramming S, Thorsteinsson B (2003) Recall of severe hypoglycaemia and self-estimated state of awareness in type 1 diabetes. Diabetes Metab Res Rev 19(3):232–240. https://doi.org/10.1002/dmrr.377
Danne T, Nimri R, Battelino T et al (2017) International consensus on use of continuous glucose monitoring. Diabetes Care 40(12):1631–1640. https://doi.org/10.2337/dc17-1600
Benjamini Y, Hochberg Y (1995) Controlling the false discovery rate: a practical and powerful approach to multiple testing. J R Stat Soc Ser B Methodol 57:289–300
Eeg-Olofsson K, Svensson A-M, Franzén S, Ismail H, Levrat-Guillen F (2020) Sustainable HbA1c decrease at 12 months for adults with type 1 and type 2 diabetes uring the FreeStyle libre system: a study within the National Diabetes Register in Sweden. Diabetes 69(Suppl 1):74–LB. https://doi.org/10.2337/db20-74-LB
Nathanson D, Svensson A-M, Miftaraj M (2021) Effect of flash glucose monitoring in adults with type 1 diabetes: a nationwide, longitudinal observational study of 14,372 flash users compared with 7691 glucose sensor naive controls. Diabetologia 64(7):1595–1603. https://doi.org/10.1007/s00125-021-05437-z
Charleer S, De Block C, Van Huffel L et al (2020) Quality of life and glucose control after 1 year of nationwide reimbursement of intermittently scanned continuous glucose monitoring in adults living with type 1 diabetes (FUTURE): a prospective observational real-world cohort study. Diabetes Care 43(2):389–397. https://doi.org/10.2337/dc19-1610
Dunn TC, Xu Y, Hayter G, Ajjan RA (2018) Real-world flash glucose monitoring patterns and associations between self-monitoring frequency and glycaemic measures: a European analysis of over 60 million glucose tests. Diabetes Res Clin Pract 137:37–46. https://doi.org/10.1016/j.diabres.2017.12.015
Boucher SE, Gray AR, Wiltshire EJ et al (2020) Effect of 6 months of flash glucose monitoring in youth with type 1 diabetes and high-risk glycemic control: a randomized controlled trial. Diabetes Care 43(10):2388–2395. https://doi.org/10.2337/dc20-0613
Olafsdattir AF, Attvall S, Sandgren U et al (2017) A clinical trial of the accuracy and treatment experience of the flash glucose monitor FreeStyle Libre in adults with type 1 diabetes. Diabetes Technol Ther 19:164–172. https://doi.org/10.1089/dia.2016.0392
Jafri RZ, Balliro CA, El-Khatib F et al (2020) A three-way accuracy comparison of the Dexcom G5, Abbott Freestyle Libre Pro, and Senseonics Eversense continuous glucose monitoring devices in a home-use study of subjects with type 1 diabetes. Diabetes Technol Ther 22(11):846–852. https://doi.org/10.1089/dia.2019.0449
American Diabetes Association (2021) Facilitating behavior change and well-being to improve health outcomes: standards of medical care in diabetes−2021. Diabetes Care 44:S53–S72. https://doi.org/10.2337/dc21-S005
Bianchi C, Aragona M, Rodia C et al (2019) Freestyle Libre trend arrows for the management of adults with insulin-treated diabetes: a practical approach. J Diabetes Complicat 33:6–12. https://doi.org/10.1016/j.jdiacomp.2018.10.012
Kudva YC, Ahmann AJ, Bergenstal RM et al (2018) Approach to using trend arrows in the FreeStyle Libre flash glucose monitoring systems in adults. J Endocr Soc 2(12):1320–1337. https://doi.org/10.1210/js.2018-00294
Borot S, Benhamou PY, Atlan C et al (2018) Practical implementation, education and interpretation guidelines for continuous glucose monitoring: a French position statement. Diabetes Metab 44(1):61–72. https://doi.org/10.1016/j.diabet.2017.10.009
Acknowledgements
We are deeply thankful for the participants’ willingness to spend the time needed on data contribution. We also thank all study nurses and dietitians, especially A. G. Skouboe, M. M. Andersen, H. H. Nørgaard, A. Frederiksen and A. E. K. Linddal; the referring clinicians; and M. A. Campbell and K. Panduro for their help on the data collection procedure.
Authors’ relationships and activities
ALS, OLS, BG-R and LHR declare that there are no relationships or activities that might bias, or be perceived to bias, their work. UP-B has received research grants from Novo Nordisk and personal fees as adviser or consultant from Abbott, Novo Nordisk, Sanofi and Zealand Pharma, and is a member of the Editorial Board of Diabetologia. DV holds stocks in Novo Nordisk. TA holds stocks in Novo Nordisk. KN serves as an adviser to Medtronic, Abbott and Novo Nordisk; owns shares in Novo Nordisk; has received research grants from Novo Nordisk, Zealand Pharma, Medtronic and Roche; and has received fees for speaking from Medtronic, Roche, Zealand Pharma, Novo Nordisk and Dexcom.
Funding
The study is investigator initiated and financed by the Capital Region of Denmark. This research received no specific grant from any funding agency in the public, commercial or not-for-profit sectors. None of the investigators have personal financial interest in the conduct or the outcome of the project.
Author information
Authors and Affiliations
Contributions
KN is the primary investigator of the study. All authors were involved in the design of the study, the literature research, and the acquisition and/or analysis of data. All authors also contributed to this manuscript with critical revision and approved the final version. KN and ALS had full access to the data and take responsibility for the decision to submit for publication.
Corresponding author
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
The original online version of this article was revised: The footnote for Table 2 incorrectly stated the data were median (IQR), but should have reported the data as mean.
Supplementary Information
ESM 1
(PDF 96 kb)
Rights and permissions
About this article
Cite this article
Secher, A.L., Pedersen-Bjergaard, U., Svendsen, O.L. et al. Flash glucose monitoring and automated bolus calculation in type 1 diabetes treated with multiple daily insulin injections: a 26 week randomised, controlled, multicentre trial. Diabetologia 64, 2713–2724 (2021). https://doi.org/10.1007/s00125-021-05555-8
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00125-021-05555-8