Abstract
We evaluated the real-time diagnostic capability of a new tool enhancing early diagnosis of pump thrombosis (PT) of the HeartWare left ventricular assist device via time-frequency analysis (TFA) of the log files. We analyzed 173 log files, including 24 (14%) associated with a clinical diagnosis of PT and 149 (86%) controls. The 30-day log file records were discretized into consecutive windows of a 24-h duration, which were iteratively acquired and processed via TFA. This way, we simulated longitudinal acquisition of pump parameters and provided real-time analysis of consecutive data, thus resembling the clinical scenario. Sensitivity and specificity of the tool were 79% and 84%, respectively. Sensitivity against PT events with progressive “build-up” thrombus increased up to 95%, and early signs of a forthcoming PT were identified 10±8 days prior to its clinical manifestation. This study demonstrates high reliability and the potential for effective clinical translation of this prognostic tool.
Similar content being viewed by others
Abbreviations
- INTERMACS:
-
Interagency Registry of Mechanically Assisted Circulatory Support
- LDH:
-
Lactate dehydrogenase
- LVAD:
-
Left ventricular assist device
- PT:
-
Pump thrombosis
- TFA:
-
Time-frequency analysis
References
Zimpfer, D., Fiane, A. E., Larbalestier, R., Tsui, S., Jansz, P., Simon, A., Schueler, S., Strueber, M., & Schmitto, J. D. (2020). Long-term survival of patients with advanced heart failure receiving an left ventricular assist device intended as a bridge to transplantation: The registry to evaluate the HeartWare Left ventricular assist system. Circulation. Heart Failure, 13(3), e006252. https://doi.org/10.1161/CIRCHEARTFAILURE.119.006252.
Mehra, M. R., Uriel, N., Naka, Y., Cleveland Jr., J. C., Yuzefpolskaya, M., Salerno, C. T., Walsh, M. N., Milano, C. A., Patel, C. B., Hutchins, S. W., Ransom, J., Ewald, G. A., Itoh, A., Raval, N. Y., Silvestry, S. C., Cogswell, R., John, R., Bhimaraj, A., Bruckner, B. A., Lowes, B. D., Um, J. Y., Jeevanandam, V., Sayer, G., Mangi, A. A., Molina, E. J., Sheikh, F., Aaronson, K., Pagani, F. D., Cotts, W. G., Tatooles, A. J., Babu, A., Chomsky, D., Katz, J. N., Tessmann, P. B., Dean, D., Krishnamoorthy, A., Chuang, J., Topuria, I., Sood, P., Goldstein, D. J., & MOMENTUM 3 Investigators. (2019). A fully magnetically levitated left ventricular assist device - Final report. The New England Journal of Medicine, 380(17), 1618–1627. https://doi.org/10.1056/NEJMoa1900486.
Chorpenning, K., Brown, M. C., Voskoboynikov, N., Reyes, C., Dierlam, A. E., & Tamez, D. (2014). HeartWare controller logs a diagnostic tool and clinical management aid for the HVAD pump. ASAIO Journal, 60(1), 115–118. https://doi.org/10.1097/MAT.0000000000000022.
https://www.medtronic.com/content/dam/medtronic-com/products/cardiac-rhythm/ventricular-assist-device/documents/autologs1-2/UC201909382-EN-Autologs-1-2-Atlas.pdf. Accessed Feb 16th 2020.
Consolo, F., Esposti, F., Gustar, A., De Bonis, M., & Pappalardo, F. (2019). Log files analysis and evaluation of circadian patterns for the early diagnosis of pump thrombosis with a centrifugal continuous-flow left ventricular assist device. The Journal of Heart and Lung Transplantation, 38(10), 1077–1086. https://doi.org/10.1016/j.healun.2019.04.008.
Pappalardo, F., Bertoldi, L. F., Sanvito, F., Marini, C., & Consolo, F. (2021). Inflow cannula obstruction of the HeartWare left ventricular assist device: What do we really know? Cardiovascular Pathology, 50, 107299. https://doi.org/10.1016/j.carpath.2020.107299.
Chui, C. K. (1992). An introduction to wavelets. In Wavelet analysis and its applications (Vol. 6, 1st ed.). Academic Press ISBN0121745848.
Jorde, U. P., Aaronson, K. D., Najjar, S. S., Pagani, F. D., Hayward, C., Zimpfer, D., et al. (2015). Identification and management of pump thrombus in the HeartWare left ventricular assist device system: A novel approach using log file analysis. JACC Heart Failure, 3(11), 849–856. https://doi.org/10.1016/j.jchf.2015.06.015.
Slaughter, M. S., Ising, M. S., Tamez, D., O’Driscoll, G., Voskoboynikov, N., Bartoli, C. R., et al. (2010). Increase in circadian variation after continuous-flow ventricular assist device implantation. The Journal of Heart and Lung Transplantation, 29(6), 695–697. https://doi.org/10.1016/j.healun.2009.12.006.
Cervi, F., Vismara, E., Pappalardo, F., & Consolo, F. (2020). Log files of continuous-flow left ventricular assist devices reveal diurnal changes of pump parameters beyond circadian variations. ASAIO Journal, 67(2), e62–e63. https://doi.org/10.1097/MAT.0000000000001201.
Grabska, J., Schlöglhofer, T., Gross, C., Maw, M., Dimitrov, K., Wiedemann, D., et al. (2020). Early detection of pump thrombosis in patients with left ventricular assist device. ASAIO Journal, 66(4), 348–354. https://doi.org/10.1097/MAT.0000000000001015.
Gyoten, T., Morshuis, M., Rojas, S. V., Deutsch, M. A., Schramm, R., Gummert, J. F., & Fox, H. (2020). Identification of characteristics, risk factors, and predictors of recurrent LVAD thrombosis: Conditions in HeartWare devices. Journal of Artificial Organs, 24(2), 173–181. https://doi.org/10.1007/s10047-020-01228-2.
Brown, M. C., Ramos, V., & Voskoboynikov, N. (2020). Evolution and clinical adoption of the Autologs system: An automated analysis for enhanced patient management in MCS. Artificial Organs, 44(9), 935–938. https://doi.org/10.1111/aor.13739.
Author information
Authors and Affiliations
Contributions
All authors contributed to the study conception and design, material preparation, data collection, and analysis. The first draft of the manuscript was written by Filippo Consolo and all authors commented on previous versions of the manuscript. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics Approval
This study was performed in line with the principles of the Declaration of Helsinki and was approved by the institutional review board of San Raffaele Scientific Institute (June 2017). All patients provided written informed consent. No animal studies were carried out by the authors for this article.
Conflict of Interest
FC and FP are co-inventors of the algorithm used in the study that is the object of an international patent application.
Additional information
Associate Editor Navin Kumar Kapur oversaw the review of this article
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Clinical registration number: https://clinicaltrials.gov/ct2/show/NCT03255928
ClinicalTrials.govIdentifier: NCT03255928
Rights and permissions
About this article
Cite this article
Consolo, F., Pappalardo, F. Real-Time Analysis of the Log Files of the HeartWare Continuous-Flow Left Ventricular Assist Device for the Early Diagnosis of Pump Thrombosis: a Step Forward Toward Clinical Translation. J. of Cardiovasc. Trans. Res. 15, 408–415 (2022). https://doi.org/10.1007/s12265-021-10157-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12265-021-10157-1