Skip to main content
  • Original research
  • Open access
  • Published:

Standardisation of conventional and advanced iterative reconstruction methods for Gallium-68 multi-centre PET-CT trials

Abstract

Purpose

To assess the applicability of the Fluorine-18 performance specifications defined by EANM Research Ltd (EARL), in Gallium-68 multi-centre PET-CT trials using conventional (ordered subset expectation maximisation, OSEM) and advanced iterative reconstructions which include the systems’ point spread function (PSF) and a Bayesian penalised likelihood algorithm (BPL) commercially known as Q.CLEAR. The possibility of standardising the two advanced reconstruction methods was examined.

Methods

The NEMA image quality phantom was filled with Gallium-68 and scanned on a GE PET-CT system. PSF and BPL with varying post-reconstruction Gaussian filter width (2–6.4 mm) and penalisation factor (200–1200), respectively, were applied. The average peak-to-valley ratio from six profiles across each sphere was estimated to inspect any edge artefacts. Image noise was assessed using background variability and image roughness. Six GE and Siemens PET-CT scanners provided Gallium-68 images of the NEMA phantom using both conventional and advanced reconstructions from which the maximum, mean and peak recoveries were drawn. Fourteen patients underwent 68Ga-PSMA PET-CT imaging. BPL (200-1200) reconstructions of the data were compared against PSF smoothed with a 6.4-mm Gaussian filter.

Results

A Gaussian filter width of approximately 6 mm for PSF and a penalisation factor of 800 for BPL were needed to suppress the edge artefacts. In addition, those reconstructions provided the closest agreement between the two advanced iterative reconstructions and low noise levels with the background variability and the image roughness being lower than 7.5% and 11.5%, respectively. The recoveries for all methods generally performed at the lower limits of the EARL specifications, especially for the 13- and 10-mm spheres for which up to 27% (conventional) and 41% (advanced reconstructions) lower limits are suggested. The lesion standardised uptake values from the clinical data were significantly different between BPL and PSF smoothed with a Gaussian filter of 6.4 mm wide for all penalisation factors except for 800 and 1000.

Conclusion

It is possible to standardise the advanced reconstruction methods with the reconstruction parameters being also sufficient for minimising the edge artefacts and noise in the images. For both conventional and advanced reconstructions, Gallium-68 specific recovery coefficient limits were required, especially for the smallest phantom spheres.

Introduction

Standardisation is a key aspect when conducting PET multi-centre clinical trials in order to minimise the sources of variation between participating sites. Moreover, it improves the robustness of the performed analyses, especially when quantification is of importance. Unaccounted-for inter-scanner variations could lead to significant differences in image noise and the accuracy of standardised uptake values (SUVs) which could in turn significantly reduce the power of the conducted studies [1, 2].

The importance of standardisation has been acknowledged by the scientific community especially in trials using 18F-fluoro-2-deoxy-d-glucose (FDG). This is mainly due to the radiotracer’s well-established role in oncology and its availability in most PET centres, making it logistically easier to conduct multi-centre trials. Gallium-68-based radiotracers are increasingly finding their way into clinical PET applications owing to the introduction of Germanium-68/Gallium-68 generators, their fast target localisation, their fast blood clearance and the short physical half-life of Gallium-68 (T1/2 = 68 min) [3]. The most commonly used, 68Ga-prostate specific-membrane antigen (PSMA), has exhibited high specificity in prostate cancer with high proportion of changes in management (21%) of patients undergoing the scan [4, 5]. Similarly, 68Ga-tetraazacyclododecanetetraacetic acid-DPhe1-Tyr3-octreotate (DOTATATE) has shown advantages in the management of patients, especially in well-differentiated neuroendocrine tumours compared to 18F-FDG [6]. Published guidelines for scanning protocols with those tracers recommend an injected activity of up to 200MBq and acquisition time of 2–4 min per bed position [7, 8]. Taking also into account the distinctly different physical characteristics of the two radionuclides (approximately 5 times longer positron range, with approximate mean ranges of 0.6 mm for Fluorine-18 and 2.9 mm for Gallium-68 in water [9], half the physical half-life and 8% lower positron yield for Gallium-68 compared to Fluorine-18), it becomes evident that the Fluorine-18 based guidelines from EANM Research Ltd. (EARL) might not be directly applicable for [68Ga]-labelled tracers [10]. Huizing et al. reported greater variation between sites for Gallium-68 and an 11% decrease in recovery coefficient compared to Fluorine-18 when scanning the NEMA Image Quality phantom as per the EARL guidelines [11]. In their study, the main source of difference between the two tracers was due to inaccuracies in the scaling factor for Gallium-68 in the dose calibrators as set by the manufacturers and was then partly alleviated by scaling the recovery coefficients by the measured discrepancy [12]. Soderlund et al. also reported an inverse relationship between recovery and positron range when comparing phantom acquisitions reconstructed with the same parameters, indicating the need for tracer specific limits [13].

Incorporation of point spread function (PSF) in iterative reconstruction is reported to lead to images with significantly higher contrast compared to standard iterative algorithms, especially for small lesions, but resulting in the appearance of edge artefacts in lesions [14, 15]. PSF reconstruction is now offered by most PET-CT systems. As in non-PSF iterative reconstruction, three of the main user defined parameters on the scanner are the number of subsets, number of iterations and the post-reconstruction Gaussian filter. The iterations and the subsets define the agreement between measured and estimated projection data and their product (effective iterations) is usually large enough (typically between 32 and 48) to ensure convergence of the radioactivity concentration in the region of interest. However, the improvement in accuracy by increasing the number of effective iterations comes at the cost of increased noise, which adds uncertainty in identifying small lesions. The most commonly used method for suppressing this noise is the application of a post-reconstruction Gaussian filter which performs a weighted averaging in neighbouring pixels and is typically 3–10 mm wide. The downside of this approach is the blurring caused on the edges of regions and apparent decreased uptake in metabolically active lesions which could potentially counter the benefit of using PSF reconstruction in the first place. Consequently, there is no consensus in the literature on the input values that should be applied, especially for the width of Gaussian filter, which could lead to significant variations in SUVs [16, 17].

GE Healthcare has recently further expanded on the PSF reconstruction algorithm by including a Bayesian penalised likelihood (BPL) reconstruction method (Q.CLEAR), which is claimed to increase image accuracy as it iterates the image until full convergence has been achieved. The only input parameter is the penalisation factor (β) which regulates the level of noise in the images [18]. Recent qualitative studies using Fluorine-18 labelled tracers in clinical data [19,20,21], reported an optimum value of β=300–400. Quantitative assessments [19, 21] also revealed that this value provided a higher signal-to-noise ratio when compared to standard iterative reconstruction. For 68Ga-PSMA clinical scans, ter Voert et al. reported an optimum SNR for β values between 400 and 550 [22]. Effects such as edge artefacts and effect of noise and contrast recovery in regions of different sizes can be difficult to assess qualitatively and, to our knowledge, a quantitative analysis on phantom data for the investigation of the optimum β value for Gallium-68 labelled tracers still needs to be conducted. Moreover, as Q.CLEAR and PSF reconstructions become the standard of care for many sites, standardisation procedures for Gallium-68 multi-centre trials need to be established when using those methods.

In this study, we initially investigated the optimum level of Gaussian filtering for traditional PSF reconstruction and the optimum penalisation factor for Q.CLEAR in terms of edge artefact suppression. The applicability of the EARL Fluorine-18 defined accreditation limits for the standardisation of six UK-based scanners planning to participate in Gallium-68 multi-centre trials was reviewed for the most commonly used ordered subset expectation maximisation reconstruction (OSEM) while standardisation of the advanced reconstruction techniques was also attempted based both on their recovery coefficients and noise characteristics. As it can be argued that the NEMA phantom might be an unrealistic approach for evaluating the differences between two reconstruction methods in real clinical data with lesions of various metabolic activities, sizes and neighbouring tissues, the agreement between PSF and Q.CLEAR reconstruction was also evaluated using clinical scans of patients scanned with 68Ga-PSMA.

Methods

Phantom data

Optimisation of advanced reconstruction methods

For the investigation of the performance of Q.CLEAR and its harmonisation with PSF reconstruction, the NEMA IEC PET Body Phantom™ (Data Spectrum) was scanned on the GE Discovery 710 PET-CT scanner. Gallium-68 was inserted in all six spheres (37, 28, 22, 17, 13, and 10 mm) and the background region, achieving radioactivity concentrations of 19.9 kBq/ml and 2.4 kBq/ml, respectively. The lung insert packed with styrofoam beads was also included in the background. The raw data were reconstructed with Q.CLEAR with varying β values (200, 400, 600, 800, 1000 and 1200) and with PSF using 2 iterations and 24 subsets and varying Gaussian post-reconstruction filter width (2, 4, 5 and 6.4 mm). Both methods use the OSEM algorithm, including time of flight (TOF) and PSF but with the addition of a BPL method in Q.CLEAR. A straightforward nomenclature to highlight the main differences of the two reconstructions would be OSEM+TOF+PSFw with w being the filter width in mm for the traditional PSF reconstruction and OSEM+TOF+PSF+BPLβ with β being the penalisation factor for the Q.CLEAR method. For better legibility though, the two methods will be referred to simply as PSFw and BPLβ. The figures for all analyses were generated using Matlab R2020b.

Multi-centre Fluorine-18 and Gallium-68 acquisition

All scanners included in this study had been accredited by the UK PET Core Lab (http://www.ncri-pet.org.uk/) for participation in Fluorine-18-based PET-CT multi-centre trials. For the accreditation process, the same specification limits as defined by EARL standard 1 were used.

The NEMA IEC PET Body Phantom™ (Data Spectrum) was scanned on 12 Siemens and GE scanners with Fluorine-18. Local sites were provided with a standardised filling procedure and asked to acquire the phantom using local clinical protocols for FDG and a sphere-to-background ratio of 4.8:1 to approximate the average tumour-to-background ratio as observed in clinical FDG studies [23, 24]. A summary of the scanner models and reconstruction parameters are shown in Table 1. The sphere-to-background ratio was 4.61 ± 0.42 (average ± standard deviation) with all six spheres containing radioactivity and the lung insert included in the background region. All images were reconstructed both with OSEM+TOFw and PSFw. The number of effective iterations was maintained the same between OSEM+TOFw and PSFw as it has been previously shown that the convergence of the recovery coefficient is similar between the two for that level of post-reconstruction smoothing [25].

Table 1 Reconstruction and acquisition parameters for all NEMA IQ scans using Fluorine-18 and Gallium-68

Two Siemens and four GE scanners provided acquisitions of the NEMA IQ phantom with Gallium-68 again using centrally distributed procedures and with 8.22 ± 0.53 sphere-to-background ratio. The higher sphere-to-background ratio compared to the FDG scans was selected in order to take into account the higher tumour-to-background ratio observed in 68Ga-PSMA scans. A conservative approach was selected though compared to the published contrast in clinical images, which also enables the results from this study to be compared with the existing literature [16, 23, 26, 27]. The reconstruction parameters were based on the local clinical protocols for whole-body 68Ga-PSMA scans. One Siemens scanner (Siemens mCT Flow) provided both OSEM+TOF5 and PSF5 reconstructions but for the second one (Siemens TrueV), in which these options were not available, only the OSEM5 reconstruction was applied. All GE scanners reconstructed the data using OSEM+TOF6.4. Three scanners (two Discovery 710 and Discovery 690) also provided a PSF6.4 reconstruction and three sites (two Discovery 710 and Discovery MIDR) a BPL800 reconstruction.

Analysis of phantom data

All sphere regions were semi-automatically drawn using a 3D isocontour at 50% of the maximum value within each sphere [10, 16] using MIM Software Inc. 6.9.3. Also, six concentric, circular regions were drawn in the background with diameters equal to the nominal diameter of the hot sphere regions. Those were replicated 12 times and on five different slices (360 in total) as described by Tong et al. [25]. To assess the compliance of the spheres with the EARL specifications (http://earl.eanm.org/ accessed on January 22, 2021), the recovery coefficient RC was used:

$$ RC=\frac{C_m}{C_{dc}} $$
(1)

Cm is the measured radioactivity concentration in the images and Cdc is the radioactivity concentration inserted in the spheres as measured from the radionuclide calibrator, corrected for decay and residual. For the sphere regions, the RC was measured for the voxel with the maximum reconstructed activity concentration (RCmax), for the mean activity (RCmean) and for a small (1cm3) spherical region of interest in the area of the highest reconstructed activity concentration (RCpeak) which was allowed to extend outside the boundaries of the delineated sphere region [28].

For comparison between scanners and acquisitions, the contrast recovery coefficient CRC was used in order to account for variations between sphere-to-background ratios and investigate the effect of varying region size [29].

$$ CRC=\frac{C_m-{C}_{bkg.m}}{C_{dc}-{C}_{bkg. dc}} $$
(2)

Cbkg.m is the measured radioactivity concentration in the background from the images and Cbkg.dc the inserted radioactivity concentration in the background as measured from the radionuclide calibrator, corrected for decay and residual. The CRCmax, CRCmean and CRCpeak were calculated.

The noise in the images was assessed by measuring background variability (BV) and image roughness (IR) [25]:

$$ {BV}_r=\frac{\sqrt{\frac{1}{K-1}\sum \limits_{k=1}^K{\left({C}_{r,k}-\overline{C_r}\right)}^2}}{\ \overline{C_r}} $$
(3)
$$ {IR}_{r,k}=\frac{\sqrt{\frac{1}{I-1}\sum \limits_{i\epsilon {ROI}_{r,k}}{\left({C}_{i,r}-{C}_{r,k}\right)}^2}}{C_{r,k}} $$
(4)

K is the number of regions (60 for each ROI size), r indicates the region size for the six different sizes used, Cr, k the average radioactivity concentration for ROI k, \( \overline{C_r} \) the average background radioactivity concentration for all ROIs of the same size, I the number of voxels in each ROI, and Ci, r the radioactivity concentration of voxel i in region r.

In order to investigate and quantify the presence of edge artefacts, for each sphere, the axial slice with the largest diameter was selected and the surface areas of the spheres were plotted. Six 1D profiles were also drawn for each sphere on the same slice used for the surface plots, and the Peak-To-Valley (PTV) ratio was estimated with peak being the highest value along the profile and the valley the lowest within the defined regions [30].

Clinical data

Fourteen patients aged between 51 and 79 years old with prostate cancer were injected with 163 ± 23 MBq (mean ± standard deviation) 68Ga-PSMA and underwent whole body PET-CT scanning approximately 1 h later, on the GE Discovery 710 scanner at the PET Centre, St Thomas’ Hospital, London, UK, between 2017 and 2018. The data were reconstructed with OSEM+TOF and PSF both smoothed with a 6.4 mm Gaussian filter, and BPL with β = 200, 400, 600, 800, 1000 and 1200.

Analysis of clinical data

Thirteen prostate, nine lymph nodes and two bone lesions (24 lesions in total) were manually delineated on MIM Software Inc. 6.9.3 and checked by an experienced nuclear medicine physician. The log transformed mean, maximum and peak standard uptake values (SUVmean, SUVmax and SUVpeak) were calculated, and Bland-Altman plots were generated to assess the agreement between PSF and BPL reconstructions [31].

Statistical analyses

Welch’s t test at p<0.05 was used to compare the CRC between Fluorine-18 and Gallium-68 for the phantom data and paired t test at p<0.05 to compare the SUV values as measured from the different reconstructions for the clinical data.

Results

Optimisation of advanced reconstruction methods for Gallium-68

Edge artefacts

Edge artefacts were observed in both PSF and BPL reconstructions for low levels of smoothing as shown in Fig. 1. These were mainly prominent in the 37-, 28-, and 22-mm spheres while PTV could not be estimated for the rest as the surface had a “conical” rather than a “concave” shape and a PTV could not be measured. Moreover, the limited spatial resolution did not allow the drawing of more than 2–4 different profiles. For all profiles, and in the three largest spheres, the valley was identified in the centre of the region and the peak at the edges. PTV was decreased with the increase of β and Gaussian filter width. The optimum β was region dependent with 22- and 28-mm spheres requiring a value between 600 and 800 in order to achieve PTV = 1 ± 0.02 while β=1200 was required for the 37-mm sphere. Similarly, a 6.4-mm filter was required for PSF reconstruction for the 37- and 28-mm spheres to achieve PTV equal to 1.7 and 1.02, respectively, and 5-mm for the 22-mm sphere to achieve PTV=1.

Fig. 1
figure 1

Surface plots for the 37-mm sphere of the NEMA IQ phantom taken from the slice with the largest diameter. The top row refers to BPL reconstructions with increasing penalisation factor (left to right) and the bottom row to PSF reconstructions with increasing width (left to right) of Gaussian post-reconstruction filter. The averaged peak-to-valley ratios from six profiles are also shown on the right for BPL (top) and PSF (bottom) reconstructions for the 37-, 28-, and 22-mm spheres. The error bars represent the standard deviations.

Recovery coefficients

RCs decreased with increasing Gaussian filter for PSF as shown in Fig. 2. For RCmax, minimal effect was observed for filters up to 4-mm wide (less than 2.9% change for any of the spheres). When a 5-mm filter was applied though, the decrease in RCmax ranged between 4 and 26%. A more conservative decrease with wider smoothing filter was noticed in RCmean and RCpeak where a maximum of 12% decrease was noticed between the 2 and 6.4 mm for both RCmean and RCpeak. The largest differences for all recovery metrics were noticed for the 13-mm sphere which exhibited an overshoot in RCmax, i.e. a sudden increase in the curves with recoveries >1, for filter widths smaller than 5 mm.

Fig. 2
figure 2

RCmax (left), RCmean (middle) and RCpeak (right) for Gallium-68 measured from the NEMA IQ phantom scanned on a GE Discovery 710 scanner and reconstructed using BPL with varying penalisation factor (top row) and PSF with varying post-reconstruction Gaussian filter width (bottom row).

A similar pattern was observed for the BPL reconstruction with increasing penalisation factor. The RCpeak was the least sensitive metric for regions larger than 22 mm for which, changes smaller than 5% between BPL200 and BPL1200 were observed. RCmax did not change by more than 3.5% for spheres >17 mm when increasing β to values higher than 600 while differences for the smallest spheres were 4 times higher and the curves became smoother.

Recovery coefficient for Gallium-68—Standard iterative reconstruction

The Gallium-68 RC curves for each of the six sites along with the EARL standard 1 and the comparison of the average CRC with Fluorine-18 are shown in Figs. 3 and 4, respectively. The RCmean, RCmax and RCpeak for Gallium-68 were at the lower limit of the EARL specifications, especially for the three smallest spheres. The RCmeans for the four largest spheres for three sites were between 1 and 7% lower than the lower EARL limit specified for Fluorine-18, while the 10- and 13-mm spheres were between 1 and 19% lower for five out of six sites. RCmax was within limits for all sites.

Fig. 3
figure 3

RCmax (left), RCmean (middle) and RCpeak (right) for Gallium-68 from all six centres measured from the NEMA IQ phantom reconstructed with standard OSEM. The RCs (dashed black lines) for the EARL 1 FLuorine-18 standards are also displayed as opposed to the limits proposed in this study for Gallium-68 (black solid lines)

Fig. 4
figure 4

Average CRCmax (left), CRCmean (middle) and CRCpeak (right) for all 6 Gallium-68 (blue curve) and the 12 Fluorine-18 (red curve) acquisitions reconstructed with standard OSEM. The error bars represent the standard deviation between sites. Significant difference was noticed for the 13-mm sphere at the 5% significance level for CRCmax and CRCpeak

No significant difference was found between the average CRCmean of all six Gallium-68 scans and the 12 Fluorine-18 scans (all of which were within the EARL specifications) as shown in Fig. 4. CRCmax and CRCpeak were comparable for the three largest spheres but consistently lower for Gallium-68. Welch’s t test indicated statistically significant differences between Gallium-68 and Fluorine-18 for CRCmax and the CRCpeak for the 13-mm sphere.

Standardisation of BPL and PSF reconstruction for Gallium-68

Investigation of recovery coefficients

Since all scanners have the ability to perform PSF reconstructions while only the GE scanners have the BPL functionality, BPL was standardised to PSF. Therefore, the various BPLβ reconstructions from GE 710 (1) were compared against PSF6.4 from the same scanner. As shown in Fig. 5, BPL800 was the reconstruction which closely agreed to PSF6.4 for the RCmax with a maximum of 8% difference being observed for the 13-mm sphere. Higher penalisation factors seemed to produce lower maximum values for the BPL compared to PSF6.4 reconstruction, especially for the two smallest spheres. Similarly, RCmean and RCpeak agreement were region-related with BPL1200 having the best agreement (<5%) for all but the smallest sphere where differences of 10% and 7% were observed for RCmean and RCpeak, respectively. BPL800 exhibited the best agreement for the 10-mm sphere (1.7%) with the differences in the rest ranging between 2.6 and 13.8% for RCmean and 2.4 and 11.9% for RCpeak.

Fig. 5
figure 5

Percentage difference of RCmax (left), RCmean (middle) and RCpeak (right) between BPLβ and PSF6.4 reconstruction [(\( {RC}_{BPL_{\beta }}-{RC}_{PSF_{6.4}}\left)/{RC}_{PSF_{6.4}}\right] \) for all spheres. Each group of bars corresponds to the comparison with a single β value (200, 400, 600, 800, 1000 and 1200)

The RCs in Fig. 6 did not show any systematic or noticeable differences between BPL800 and PSF5/6.4 (PSF5 and PSF6.4) reconstructions. The BPL800 from the GE MI DR scanner provided higher RC values compared to the rest, with differences of up to 21% for RCmax and RCmean and up to 29% for RCmean when compared to the PSF6.4 from the GE 690 which in general provided the lower values. Nevertheless, the ranges of the RC were comparable to the width of the EARL Fluorine-18 standard 1 and 2 specifications, depending on sphere size, with the later defined for advanced reconstruction methods.

Fig. 6
figure 6

RCmax (left), RCmean (middle) and RCpeak (right) for Gallium-68 from all five centres measured from the NEMA IQ phantom reconstructed with BPL800 (solid lines) and PSF6.4 for the GE scanners and PSF5 for the Siemens scanner (dashed lines). The 2 RCs (dashed black lines) for the EARL 2 Fluorine-18 standards are also displayed as opposed to the limits proposed in this study for Gallium-68 (black solid lines)

The average CRC values from all advanced reconstruction methods for Gallium-68 were also comparable to the ones for Fluorine-18 averaged from 12 different sites with the images reconstructed using PSF5/6.4 (Fig. 7). A significant difference was found for CRCmean in the 10-mm sphere.

Fig. 7
figure 7

Average CRCmax (left), CRCmean (middle) and CRCpeak (right) for all 6 Gallium-68 (blue curve) and the 12 Fluorine-18 (red curve) acquisitions. All Fluorine-18 acquisitions have been reconstructed with PSF6.4/5, and the Gallium-68 includes all PSF6.4/5 and BPL800 reconstructions. The error bars represent the standard deviation between sites. Significant difference was noticed for the 10-mm sphere at the 5% significance level for CRCmean

Investigation of noise characteristics

Both BV and IR were at similar levels for both standard OSEM+TOF and PSF when smoothed with a 6.4 mm Gaussian filter as shown in Fig. 8. Both noise metrics decreased with increasing β for BPL. A β value of >400 was required in order for the images to have a similar or lower noise level compared to OSEM+TOF and PSF with BV ranging between 7.5 and 2.8% and IR between 9.1 and 11.5%. A sharp decrease of 31.6–48.1% and 61.0–69.6%, depending on the region size, was observed for BV and IR, respectively, when using BPL800 compared to BPL200 while additional increase in β resulted in more moderate improvements.

Fig. 8
figure 8

Background variability (top row) and image roughness (bottom row) for Gallium-68 acquisitions of the NEMA IQ phantom. The left column displays the metrics for varying β values of the BPL reconstruction including OSEM+TOF6.4 and PSF6.4, the middle column the OSEM and OSEM+TOF reconstructions for six different sites and the right figure for the advanced reconstructions methods

The images generated with 2D reconstruction from the Siemens TrueV exhibited the highest values (~15% IR and 13% BV for the small sphere) as illustrated in the middle row of Fig. 8. For the rest of the sites BV did not exceed 8% and IR 13% for any of the spheres when reconstructed with standard OSEM+TOF. Comparable noise levels were measured for the PSF6.4 reconstructions while BPL800 exhibited ~40% lower IR and between 6.6 and 27.8% lower BV (depending on the region size) compared to their corresponding PSF reconstructions.

Definition of standardisation specifications

The proposed limits for Gallium-68 trials for both OSEM and advanced reconstruction methods were defined by taking ±2 standard deviations from the mean RCs of all submitted images [32] and are shown in Table 2. The proposed limits for RCmax were comparable with the ones proposed by EARL Fluorine-18 standard 1 for the 4 largest spheres with the ones proposed in this study being up to 8% lower. For the 13-mm and 10-mm spheres though, the upper limit of RCmax was 23% and 14% lower, respectively. The proposed limits for RCmean were up to 12% lower for the four largest spheres when compared to EARL Fluorine-18 standard 1 while for the two smaller spheres, up to 27% lower.

Table 2 Proposed specification limits for Gallium-68 multi-centre trials

Similarly, when compared to EARL Fluorine-18 standard 2, both the upper and lower proposed RCmax limits for advanced reconstruction methods, were up to 16% lower for the four largest spheres but for the two smallest spheres were between 19 and 33% lower. Lower limits are also proposed here for RCmean, with the decrease ranging between 11 and 25% for the lower and 16–41% for the upper limit compared to EARL Fluorine-18 standard 2. When compared to the proposed limits for conventional OSEM reconstruction, the three largest spheres had comparable RCmax and RCmean specifications but considerably higher for the rest with RCmax being 13–28% and 19–51% higher and RCmean 19–33 and 3–27% higher for the upper and lower limit, respectively. RCpeak was also 22% and 24% higher for spheres 17 and 13 mm when an advanced reconstruction was performed but less than 11% for the rest of the limits.

Assessment of clinical data

Consistent with the phantom data, visual inspection of the reconstructed clinical images, indicated reduced noise for images reconstructed with BPL800 compared to the rest of the images (Fig. 9). Minimal differences in terms of noise but reduced contrast were observed with increasing β.

Fig. 9
figure 9

Coronal views of prostate cancer patient, scanned with 68Ga-PSMA with the data reconstructed using OSEM+TOF6.4, PSF6.4, BPL200, BPL800 and BPL1200 from left to right

The Bland-Altman plots in Fig. 10 indicated that all but one lesion for SUVpeak were within the 95% confidence intervals when comparing SUVs between PSF6.4 and BPL. Similar results were observed for SUVmax and SUVmean with a maximum of 2 lesions being outside the 95% confidence intervals. The only BPL reconstructions for which the SUVs were not significantly different from PSF6.4 were BPL800 and BPL1000. When comparing PSF6.4 to OSEM+TOF6.4, the difference in SUVmax ranged between −19 to 29%. SUVmean and SUVpeak were more comparable between reconstructions with PSF6.4 being up to 14% higher.

Fig. 10
figure 10

Bland-Altman plots for SUVmax (left column), SUVmean (middle column) and SUVpeak (right column) of all 22 lesions with the black solid line being the average difference of the log-transformed SUVs between PSF6.4 and BPL600 (first row), BPL800 (middle row) and BPL1000 (bottom row) and the dashed lines 1.96 x the standard deviation. The p value from the paired t test is also quoted on each figure with * indicating significant differences at the 5% significance level

Discussion

In this study, it was shown that multi-centre trials using Gallium-68 based tracers are feasible for both conventional iterative and advanced reconstruction methods which include PSF with or without a Bayesian penalisation likelihood algorithm (Q.CLEAR) by adhering to the adjusted specifications. Those specifications mainly differ from the ones defined by EARL Fluorine-18 standards 1 and 2 for the smallest spheres of the NEMA IQ phantom. Still, two different sets of limits were needed for conventional and advanced reconstructions, similar to EARL for Fluorine-18, as the two methods cannot be used interchangeably. A PSF reconstruction using a Gaussian post-reconstruction filter of approximately 6 mm and an BPL reconstruction with penalisation value of 800–1000 were needed for the two methods to provide comparable results, suppress the edge artefacts and for BPL to provide images with lower noise levels when compared to PSF. The results from the phantom study regarding the agreement of the two advanced reconstruction methods in terms of SUVs were also in agreement with the results from clinical analysis.

Consistent with the literature, the RCs for Gallium-68 were at the lower limits of the EARL Fluorine-18 standard 1 for conventional iterative reconstruction [11, 33,34,35]. No additional correction factors were needed in this study to account for well counter variability [11] as they all had traceability to the National Physics Laboratory. The calibration factor used for Gallium-68 was therefore locally defined if it was deemed necessary rather than using the one specified from the manufacturer. Moreover, all scanners had to provide 2-week results from their daily quality assurance scans as part of their accreditation procedure for the UK PET Core Lab (http://www.ncri-pet.org.uk/) to assess their stability and accuracy. Therefore, no additional correction for system calibration bias was needed either [32]. The observed decrease was related to blurring due to the higher positron range of Gallium-68 when compared to Fluorine-18 and was, as expected, mainly apparent in the smaller spheres of the NEMA IQ phantom [13]. In the corresponding comparison when PSF was included, there seemed to be an even better agreement between the CRCs of the two isotopes. This could be attributed to the fact that blurring due to positron range was partially accounted for.

The proposed specification limits in Table 2 depict the aforementioned results with the limits for the OSEM reconstruction being similar to the ones proposed by EARL Fluorine-18 standard 1 for the four larger spheres and lower for the smaller spheres in which the blurring due to positron range is more prominent. Despite the consensus of decreased recovery in Gallium-68 images, it has been proposed that the EARL specifications can still be applied as the curves are not noticeably different [11, 33]. Although this approach might be feasible, it could be difficult to manage as Gallium-68-based multi-centre trials become more popular and a large number of perfectly functional scanners would be prohibited from participating. For example, in Fig. 3, five out of six scanners have at least one sphere with RC below the lower limit of EARL Fluorine-18 standard 1. Similarly, in the multi-centre study performed by Huizing et al. [11], at least 5 out of 13 scanners did not meet the EARL criteria, even after the proposed correction for variances in the well counter, while valid results presented in other studies would also render the scanners ineligible from participating in multi-centre trials [33, 35]. The proposed limits in this study are closer to previously published results [11, 33,34,35] and consistent with the observation by Huizing et al. of higher variation between PET-CT systems [11]. Despite the fact that all scanners performed similarly in this study in terms of recovery, it needs to be taken into account that only six were used for the definition of the specification limits which is a limitation of this study, including scanners of different generations such as the Siemens Biograph TrueV and the GE MI DR. Since according to the EARL specifications the results from the TrueV scanner would be acceptable for participation in multi-centre trials with a number of those scanners still operational and used in research, it was decided to include it in this study but highlight the effect of the noise in comparison to the rest of the scanners. Moreover, this scanner complies with the EARL standard 1 which includes scanner without TOF capability. In addition, no Philips scanners participated in this study although the limits proposed in here are expected to be applicable for those scanners as well based on previous studies which did not notice any systematic differences between manufacturers [11, 32]. Nevertheless, it might be useful to revisit those limits in the future when most of the old scanners will be decommissioned and to evaluate them in studies were Philips scanners also participate.

A potential limitation when comparing the CRCs between Gallium-68 and Fluorine-18 scans is the difference in the sphere-to-background ratio between the different acquisitions. To investigate this effect, the NEMA IQ phantom was scanned on the GE-710 (1) scanner with sphere-to-background ratios ranging from 2:1 to 15:1, reconstructed using OSEM with time-of-flight and smoothed with a post-reconstruction Gaussian filter of 6.4 mm. The results displayed in Suppl. Figure 1 indicated that all coefficients were comparable for ratios between 5:1 and 10:1. Partial volume effects became more prominent for smaller ratios for all spheres while for larger ratios, the smaller sphere was more accurately measured. The range of ratios for the phantoms included in this study though was within the acceptable limits. Moreover, the level of the post-reconstruction smoothing is expected to have suppressed any edge artefacts within this range according to the results of this study, even in the case of PSF reconstruction.

Another point of consideration in this study is the difference in the reconstruction parameters especially in terms of voxel size between scanners. The recovery curves however were comparable and within the variation expected due to reproducibility between the images from scanner with the largest voxel size (GE Discovery 690) and the scanner with the smallest size (GE Discovery 710) for both conventional and advanced reconstructions. As expected, the effect of the largest voxel is highlighted in the lower noise for that acquisition. Similar results have also been reported in the literature where images reconstructed with a comparable range of voxels sizes have been included [11, 32].

As expected, edge artefacts were apparent in both PSF and BPL reconstructions which were mitigated with increased Gaussian filtering and β value, respectively. Consistent with the literature, the characteristic increase of the RCmax in the 13-mm sphere was noticed in both reconstructions due to the superimposition of the edges for this region which are heavily influenced by the edge artefacts [14, 17, 32]. For the 10-mm sphere, this effect is countered by the partial volume effects. Assessment of the recovery curves on their own would have made it difficult to evaluate the level of suppression of the edge artefacts as only the 13-mm sphere directly indicated the problem in the images. The surface curves demonstrated the level each region was affected and their respective smoothing requirements. This could be useful for a study interested in protocol optimisation when the approximate dimensions of the region of interest are known. Moreover, in studies where metrics such as the SUVmax and SUVpeak are applied, it is important for a similar assessment to be followed in order to avoid usage of voxels affected by the artefacts.

The required level of Gaussian filtering necessary to partially correct the edge artefacts in PSF reconstruction, also resulted in images with similar noise levels in terms of image roughness and background variability as the standard OSEM+TOF reconstruction when smoothed with the same filter, in accordance to the results from Tong et al. [25]. Even though a β value of 400 also resulted in images with similar noise properties, it did not seem sufficient for addressing the edge artefacts and a higher value was required which resulted in even lower noise levels but at the expense of recovery.

The optimum β value needed for standardising the recovery of BPL with the PSF reconstruction was region-dependant but the best agreement on average was observed for β=800. For larger β values, the recovery from the two smaller spheres, which were the ones mainly benefited from PSF and BPL reconstructions, was approximately 20% lower compared to PSF. This was in agreement with the clinical results for which all BPL reconstructions except BPL800 and BPL1000 were found to be significantly different compared to PSF6.4. Moreover, visual assessment of the images indicated minimal effects in image quality when increasing β to values higher than 800. Therefore, BPL800 is proposed for standardising Gallium-68 based PET acquisitions reconstructed with PSF and BPL.

The focus of this study was to harmonise the two types of reconstruction rather than optimising them. Depending on the tumour size, it could be argued that for a single-centre study focused on BPL reconstruction a slightly smaller β value such as 600 would provide higher contrast with acceptable noise levels. In such an approach, the assessment of the noise characteristics would need to be part of the optimisation process. In addition, better results both in terms of accuracy and noise could be expected if an isotope-specific PSF was implemented [36]. For this study, the data were reconstructed on the scanner though, using the manufacturer’s PSF as that would be the case for any prospective clinical trial. Ideally, taking into account the considerable difference in positron range between Fluorine-18 and Gallium-68, the advanced reconstruction methods would have been reconstructed with a Gallium-68specific PSF. Nonetheless, the selected level of Gaussian smoothing and penalisation value for standardising the PSF and BPL reconstruction, retained the improved recovery, especially in the smaller regions, and lower noise when compared to OSEM+TOF while they were found to be the minimum values for minimising the edge artefacts.

This study also indicates that isotope-specific limits might be necessary for performing multi-centre trials especially for isotopes with distinctly different positron ranges. As more and more isotopes find their way into the clinic though, this might prove logistically difficult as it could result in multiple standards that need to be regularly reviewed and updated. A method to move this field forward might be to establish a model for defining the limits based on the isotopes’ characteristics [37]. An alternative approach could be to utilise newly developed deep learning approaches to optimise the images [38]. The latter could potentially convert the images from each scanner to the samestandard and improve the quality in a more automated process, allowing direct image comparison between scanners. For the time being though, where the number of isotopes used in multi-centre trials is limited, careful assessment of the specification limits is needed.

Conclusion

Consistent with the literature, the recovery coefficients of Gallium-68 PET-CT performed near and below the lower levels of the EARL1 performance specifications for Fluorine-18 and well below the recently defined levels for the smaller spheres in the NEMA IQ phantom. Therefore, new Gallium-68 specific levels are proposed for multi-centre trials. Gaussian filtering of 6 mm (or more) and a penalisation value of 800 (or more) were required to minimise edge artefacts from the majority of the regions from the phantom in reconstructions using PSF and PSF with Bayesian penalisation likelihood, respectively. Those two reconstructions also provided similar recovery coefficient results for the phantom and similar SUVs in a clinical assessment of patients with prostate cancer. Moreover, they resulted in similar or lower noise levels and higher recovery coefficients when compared to standard iterative reconstruction and separate performance specifications for advanced reconstruction methods were defined.

Availability of data and materials

The datasets used and analysed during the current study are available from the corresponding author on reasonable request.

References

  1. Westerterp M, Pruim J, Oyen W, Hoekstra O, Paans A, Visser E, et al. Quantification of FDG PET studies using standardised uptake values in multi-centre trials: effects of image reconstruction, resolution and ROI definition parameters. Eur J Nucl Med Mol Imaging. 2007;34(3):392–404. https://doi.org/10.1007/s00259-006-0224-1.

    Article  PubMed  Google Scholar 

  2. Boellaard R, Krak NC, Hoekstra OS, Lammertsma AA. Effects of noise, image resolution, and ROI definition on the accuracy of standard uptake values: a simulation study. J Nucl Med. 2004;45(9):1519–27.

    PubMed  Google Scholar 

  3. Banerjee SR, Pomper MG. Clinical applications of Gallium-68. Appl Radiat Isot. 2013;76:2–13. https://doi.org/10.1016/j.apradiso.2013.01.039.

    Article  CAS  PubMed  Google Scholar 

  4. Roach PJ, Francis R, Emmett L, Hsiao E, Kneebone A, Hruby G, et al. The impact of (68)Ga-PSMA PET/CT on management intent in prostate cancer: results of an australian prospective multicenter study. J Nucl Med. 2018;59(1):82–8. https://doi.org/10.2967/jnumed.117.197160.

    Article  CAS  PubMed  Google Scholar 

  5. Cook GJR, Kulkarni M, Warbey VS. PSMA PET/CT imaging for primary staging of intermediate and high-risk prostate cancer. BJU Int. 2019;124(3):357–8. https://doi.org/10.1111/bju.14856.

    Article  PubMed  Google Scholar 

  6. Johnbeck CB, Knigge U, Kjær A. PET tracers for somatostatin receptor imaging of neuroendocrine tumors: current status and review of the literature. Future Oncol. 2014;10(14):2259–77. https://doi.org/10.2217/fon.14.139.

    Article  CAS  PubMed  Google Scholar 

  7. Fendler WP, Eiber M, Beheshti M, Bomanji J, Ceci F, Cho S, et al. Ga-PSMA PET/CT: Joint EANM and SNMMI procedure guideline for prostate cancer imaging: version 1.0. Eur J Nucl Med Mol Imaging. 2017;44:1014–24. https://doi.org/10.1007/s00259-017-3670-z.

    Article  PubMed  Google Scholar 

  8. Subramaniam RM, Bradshaw ML, Lewis K, Pinho D, Shah C, Walker RC. ACR practice parameter for the performance of Gallium-68 DOTATATE PET/CT for neuroendocrine tumors. Clin Nucl Med. 2018;43(12):899–908. https://doi.org/10.1097/RLU.0000000000002309.

    Article  PubMed  Google Scholar 

  9. Jødal L, Le Loirec C, Champion C. Positron range in PET imaging: an alternative approach for assessing and correcting the blurring. Phys Med Biol. 2012;57(12):3931–43. https://doi.org/10.1088/0031-9155/57/12/3931.

    Article  CAS  PubMed  Google Scholar 

  10. Kaalep A, Sera T, Oyen W, Krause BJ, Chiti A, Liu Y, et al. EANM/EARL FDG-PET/CT accreditation - summary results from the first 200 accredited imaging systems. Eur J Nucl Med Mol Imaging. 2018;45(3):412–22. https://doi.org/10.1007/s00259-017-3853-7.

    Article  CAS  PubMed  Google Scholar 

  11. Huizing DMV, Koopman D, van Dalen JA, Gotthardt M, Boellaard R, Sera T, et al. Multicentre quantitative (68)Ga PET/CT performance harmonisation. EJNMMI Phys. 2019;6(1):19. https://doi.org/10.1186/s40658-019-0253-z.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Sanderson T, Solomon J, Nottage C, Dickson J. Underestimation of 68Ga PET/CT SUV caused by activity overestimation using default calibrator settings. Phys Med. 2019;59:158–62. https://doi.org/10.1016/j.ejmp.2019.03.006.

    Article  PubMed  Google Scholar 

  13. Soderlund AT, Chaal J, Tjio G, Totman JJ, Conti M, Townsend DW. Beyond 18F-FDG: characterization of PET/CT and PET/MR scanners for a comprehensive set of positron emitters of growing application--18F, 11C, 89Zr, 124I, 68Ga, and 90Y. J Nucl Med. 2015;56(8):1285–91. https://doi.org/10.2967/jnumed.115.156711.

    Article  CAS  PubMed  Google Scholar 

  14. Rahmim A, Qi J, Sossi V. Resolution modeling in PET imaging: theory, practice, benefits, and pitfalls. Med Phys. 2013;40(6Part1):064301. https://doi.org/10.1118/1.4800806.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Akamatsu G, Ishikawa K, Mitsumoto K, Taniguchi T, Ohya N, Baba S, et al. Improvement in PET/CT image quality with a combination of point-spread function and time-of-flight in relation to reconstruction parameters. J Nucl Med. 2012;53(11):1716–22. https://doi.org/10.2967/jnumed.112.103861.

    Article  PubMed  Google Scholar 

  16. Boellaard R, Oyen WJ, Hoekstra CJ, Hoekstra OS, Visser EP, Willemsen AT, et al. The Netherlands protocol for standardisation and quantification of FDG whole body PET studies in multi-centre trials. Eur J Nucl Med Mol Imaging. 2008;35(12):2320–33. https://doi.org/10.1007/s00259-008-0874-2.

    Article  PubMed  Google Scholar 

  17. Lasnon C, Desmonts C, Quak E, Gervais R, Do P, Dubos-Arvis C, et al. Harmonizing SUVs in multicentre trials when using different generation PET systems: prospective validation in non-small cell lung cancer patients. Eur J Nucl Med Mol Imaging. 2013;40(7):985–96. https://doi.org/10.1007/s00259-013-2391-1.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  18. Steve R. “Q. clear.” GE Healthcare, White Paper. 2014. p. 1-9.

  19. Teoh EJ, McGowan DR, Macpherson RE, Bradley KM, Gleeson FV. Phantom and clinical evaluation of the Bayesian penalized likelihood reconstruction algorithm Q.Clear on an LYSO PET/CT System. J Nucl Med. 2015;56:1447–52. https://doi.org/10.2967/jnumed.115.159301.

    Article  CAS  PubMed  Google Scholar 

  20. Teoh EJ, McGowan DR, Schuster DM, Tsakok MT, Gleeson FV, Bradley KM. Bayesian penalised likelihood reconstruction (Q.Clear) of 18F-fluciclovine PET for imaging of recurrent prostate cancer: semi-quantitative and clinical evaluation. Br J Radiol. 2018;91:20170727. https://doi.org/10.1259/bjr.20170727.

    Article  PubMed  PubMed Central  Google Scholar 

  21. Rogasch JM, Suleiman S, Hofheinz F, Bluemel S, Lukas M, Amthauer H, et al. Reconstructed spatial resolution and contrast recovery with Bayesian penalized likelihood reconstruction (Q.Clear) for FDG-PET compared to time-of-flight (TOF) with point spread function (PSF). EJNMMI Phys. 2020;7:2. https://doi.org/10.1186/s40658-020-0270-y.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Ter Voert EEGW, Muehlematter UJ, Delso G, Pizzuto DA, Müller J, Nagel HW, et al. Quantitative performance and optimal regularization parameter in block sequential regularized expectation maximization reconstructions in clinical 68Ga-PSMA PET/MRI. EJNMMI Res. 2018;8(1):70. https://doi.org/10.1186/s13550-018-0414-4.

    Article  PubMed  PubMed Central  Google Scholar 

  23. Øen SK, Aasheim LB, Eikenes L, Karlberg AM. Image quality and detectability in Siemens Biograph PET/MRI and PET/CT systems-a phantom study. EJNMMI Phys. 2019;6(1):16. https://doi.org/10.1186/s40658-019-0251-1.

    Article  PubMed  PubMed Central  Google Scholar 

  24. Rogasch JM, Steffen IG, Hofheinz F, Großer OS, Furth C, Mohnike K, et al. The association of tumor-to-background ratios and SUVmax deviations related to point spread function and time-of-flight F18-FDG-PET/CT reconstruction in colorectal liver metastases. EJNMMI Res. 2015;5(1):31. https://doi.org/10.1186/s13550-015-0111-5.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  25. Tong S, Alessio AM, Kinahan PE. Noise and signal properties in PSF-based fully 3D PET image reconstruction: an experimental evaluation. Phys Med Biol. 2010;55(5):1453–73. https://doi.org/10.1088/0031-9155/55/5/013.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  26. Daube-Witherspoon ME, Karp JS, Casey ME, DiFilippo FP, Hines H, Muehllehner G, et al. PET performance measurements using the NEMA NU 2-2001 standard. J Nucl Med. 2002;43(10):1398–409.

    PubMed  Google Scholar 

  27. Afshar-Oromieh A, Malcher A, Eder M, Eisenhut M, Linhart HG, Hadaschik BA, et al. PET imaging with a [68Ga]gallium-labelled PSMA ligand for the diagnosis of prostate cancer: biodistribution in humans and first evaluation of tumour lesions. Eur J Nucl Med Mol Imaging. 2013;40(4):486–95. https://doi.org/10.1007/s00259-012-2298-2.

    Article  CAS  PubMed  Google Scholar 

  28. Wahl RL, Jacene H, Kasamon Y, Lodge MA. From RECIST to PERCIST: evolving considerations for PET response criteria in solid tumors. J Nucl Med. 2009;50(Suppl 1):122S–50S. https://doi.org/10.2967/jnumed.108.057307.

    Article  CAS  PubMed  Google Scholar 

  29. Kessler RM, Ellis JR Jr, Eden M. Analysis of emission tomographic scan data: limitations imposed by resolution and background. J Comput Assist Tomogr. 1984;8(3):514–22. https://doi.org/10.1097/00004728-198406000-00028.

    Article  CAS  PubMed  Google Scholar 

  30. Zhang J, Olcott PD, Chinn G, Foudray AM, Levine CS. Study of the performance of a novel 1 mm resolution dual-panel PET camera design dedicated to breast cancer imaging using Monte Carlo simulation. Med Phys. 2007;34(2):689–702. https://doi.org/10.1118/1.2409480.

    Article  CAS  PubMed  Google Scholar 

  31. Lodge MA. Repeatability of SUV in Oncologic 18F-FDG PET. J Nucl Med. 2017;58(4):523–32. https://doi.org/10.2967/jnumed.116.186353.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  32. Kaalep A, Sera T, Rijnsdorp S, Yaqub M, Talsma A, Lodge MA, et al. Feasibility of state of the art PET/CT systems performance harmonisation. Eur J Nucl Med Mol Imaging. 2018;45(8):1344–61. https://doi.org/10.1007/s00259-018-3977-4.

    Article  PubMed  PubMed Central  Google Scholar 

  33. Wielaard J, Habraken JBA, Brinks P, Lavalaye J, Boellaard R. Optimization of injected 68Ga-PSMA activity based on list-mode phantom data and clinical validation. EJNMMI Phys. 2020;7(1):20. https://doi.org/10.1186/s40658-020-00289-9.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  34. Jönsson L, Stenvall A, Mattsson E, Larsson E, Sundlöv A, Ohlsson T, et al. Quantitative analysis of phantom studies of 111In and 68Ga imaging for neuroendocrine tumours. EJNMMI Phys. 2018;5(1):5. https://doi.org/10.1186/s40658-018-0204-0.

    Article  PubMed  PubMed Central  Google Scholar 

  35. Sanchez-Crespo A. Comparison of Gallium-68 and Fluorine-18 imaging characteristics in positron emission tomography. Appl Radiat Isot. 2013;76:55–62. https://doi.org/10.1016/j.apradiso.2012.06.034.

    Article  CAS  PubMed  Google Scholar 

  36. Kotasidis FA, Anton-Rodriguez, J. M., Angelis, G. I., Matthews, J. C., Reader, A. J., & Zaidi, H. Isotope specific resolution modelling image reconstruction for high resolution PET imaging. IEEE Nuclear Science Symposium and Medical Imaging Conference (2013 NSS/MIC). Seoul; 2013. p. 1-3.

  37. Prenosil GA, Hentschel M, Fürstner M, Krause T, Weitzel T, Klaeser B. Technical note: transconvolution based equalization of positron energy effects for the use of 68Ge/68Ga phantoms in determining 18F PET recovery. Med Phys. 2017;44(7):3761–6. https://doi.org/10.1002/mp.12330.

    Article  CAS  PubMed  Google Scholar 

  38. Dal Toso L, Pfaehler, E., Boellaard, R., Schnabel, J. A., & Marsden, P. K. Deep learning based approach to quantification of PET tracer uptake in small tumors. International Workshop on Machine Learning for Medical Image Reconstruction, MICCAI 2019. Shenzhen; China: Springer, Cham; 2019. p. 181-92.

Download references

Acknowledgements

We would like to thank Dr. Jose Anton-Rodriguez, Ms. April-Louise Smith, Dr. Nick Bird and Dr. Rebecca Gregory for the provision of data essential for this study and Dr. Joel Dunn for his scientific input.

Code availability

The code used for the analysis of the datasets in this study is available from the corresponding author on reasonable request.

Funding

'This research was funded in part, by the Wellcome Trust/EPSRC Centre for Medical Engineering [WT 203148/Z/16/Z]. For the purpose of open access, the author has applied a CC BY public copyright licence to any Author Accepted Manuscript version arising from this submission. King’s College London and UCL Comprehensive Cancer Imaging Centre are funded by the CRUK and EPSRC in association with the MRC and DoH (England). The research was supported by the National Institute for Health Research (NIHR) Biomedical Research Centre based at Guy’s and St Thomas’ NHS Foundation Trust and King’s College London. Georgios Krokos is funded by the Biomedical Research Centre at Guy’s & St Thomas’ NHS Foundation Trust and King’s College London. The views expressed are those of the authors and not necessarily those of the NHS, the NIHR or the Department of Health.

Author information

Authors and Affiliations

Authors

Contributions

Dr. Georgios Krokos conceived along with Lucy Pike the idea for the project presented in this study. He designed the study, acquired the phantom data presented from the Siemens mCT Flow and one of the GE Discovery 710 scanners, coordinated the acquisition of the phantom data from the rest of the sites, performed the analysis and drafted the manuscript. Ms. Lucy Pike conceived the idea of this study along with Georgios Krokos. She provided historical data available through the UK PET Core Lab of phantoms scanned with Fluorine-18 on PET-CT scanners accredited to participate in multi-centre trials. She also provided scientific input and feedback on the final manuscript. Prof. Gary Cook acquired the necessary approvals for the retrospective analysis of the clinical data. He assisted in the manual delineation of the lesions from the clinical images and provided feedback on the final draft of the manuscript. Prof. Paul Marsden supervised the overall project and provided scientific input and direction during the study. He also provided insight for the effects noticed in the results which was used in the discussion and provided feedback on the final draft of the manuscript. The author(s) read and approved the final manuscript.

Corresponding author

Correspondence to Georgios Krokos.

Ethics declarations

Ethics approval and consent to participate

This service evaluation was granted institutional approval by the Guy’s and St Thomas’ Service Evaluation committee and the requirement for consent for data use that was anonymised before analysis was waived.

Project Number: 9892

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1: Suppl. Figure 1.

Effect of different sphere-to-Background ratios. CRCmax (left), CRCmean (middle), CRCpeak (right) for five PET-CT acquisitions of the NEMA IQ phantom with varying sphere-to-background ratio. For a ratio of 2:1 only the two largest spheres could be defined using the 3D isocontour at 50% of the maximum.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Krokos, G., Pike, L., Cook, G.J.R. et al. Standardisation of conventional and advanced iterative reconstruction methods for Gallium-68 multi-centre PET-CT trials. EJNMMI Phys 8, 52 (2021). https://doi.org/10.1186/s40658-021-00400-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40658-021-00400-8

Keywords