13 February 2021 Transcutaneous monitoring of hemoglobin derivatives during methemoglobinemia in rats using spectral diffuse reflectance
Author Affiliations +
Abstract

Significance: Untreated methemoglobinemia may cause severe hypoxemia and even death when methemoglobin levels in the blood stream exceed 70%. Although CO-oximetry can be used to monitor the response to treatment for methemoglobinemia, it is costly and requires an invasive procedure for collecting blood samples from patients. A pulse CO-oximeter with a contact probe can be used to continuously and non-invasively measure the percentage of methemoglobin, as well as the percutaneous oxygen saturation. In terms of the prevention of infectious diseases, however, it is desirable to monitor methemoglobin and oxygen saturation levels in a non-contact manner. Diffuse reflectance spectral imaging is promising as a non-contact, non-invasive, and cost-effective clinical diagnostic tool for methemoglobinemia.

Aim: To demonstrate the feasibility of visible spectral diffuse reflectance for in vivo monitoring of hemoglobin derivatives and evaluating methemoglobin production and reduction as well as hypoxemia during methemoglobinemia in rats.

Approach: A new imaging approach based on the multiple regression analysis aided by Monte Carlo simulations for light transport was developed to quantify methemoglobin, oxygenated hemoglobin, and deoxygenated hemoglobin using a hyperspectral imaging system. An in vivo experiment with rats exposed to sodium nitrite (NaNO2) at different doses was performed to confirm the feasibility of the method for evaluating the dynamics of methemoglobin, oxygenated hemoglobin, and deoxygenated hemoglobin during methemoglobinemia. Systemic physiological parameters, including the percutaneous arterial oxygen saturation, heart rate (HR), and pulse distention, were measured by a commercially available pulse oximeter, and the results were compared to those obtained by the proposed method.

Results: Both the methemoglobin concentration and methemoglobin saturation rapidly increased with a half-maximum time of <20  min. They reached their maximal values nearly 60 min after the administration of NaNO2. Tissue oxygen saturation dramatically dropped to a minimum of 33.7  %    ±  0.4  %  , 23.1  %    ±  5.6  %  , 8.8  %    ±  1.7  %  , and 9.7  %    ±  5.1  %   on average for NaNO2 doses of 25, 37.5, 50, and 75 mg/kg, respectively. Changes in methemoglobin concentration and tissue oxygen saturation are indicative of the temporary production of methemoglobin and severe hypoxemia during methemoglobinemia. Profound increases in the HR and pulse distention implied an elevated cardiac output caused by tachycardia and the resultant increase in peripheral blood volume to compensate for the hypoxia and hypoxemia during methemoglobinemia. This was in agreement with the time course of the peripheral hemoglobin volume concentration obtained by the proposed method.

Conclusions: The proposed method is capable of the in vivo non-contact simultaneous evaluation of methemoglobin levels and hypoxemia during methemoglobinemia, and that it has potential as a tool for the diagnosis and management of methemoglobinemia.

1.

Introduction

The spectral features of human skin are often affected by various skin diseases, which cause altered attenuation of light within multiple layers of the skin. In the epidermis and dermis, similar changes are associated with fluctuations in the concentration of melanin and blood-borne pigments, such as hemoglobin.1 Each chromophore has its own characteristic absorption spectrum in the visible to near-infrared wavelength regions, which change with wavelength and can function as a fingerprint for that molecule. The binding of other molecules to the functional chromophore causes variations in the structure of a molecule, resulting in an altered absorption spectrum for each derivative of a molecule that is formed. Figure 1 shows the extinction coefficient spectra of typical hemoglobin derivatives and melanin at visible wavelengths between 500 and 700 nm.24 The actual absorption spectra of hemoglobin depend primarily upon whether or not the blood is oxygenated, and there are thus two main derivatives, i.e., oxyhemoglobin in oxygenate blood and deoxyhemoglobin in deoxygenated blood. There are also hemoglobin derivatives that are incapable of transporting oxygen molecules, and they are, therefore, collectively known as dyshemoglobins; these include methemoglobin, carboxyhemoglobin, and sulfhemoglobin. The absorption can differ drastically between dysfunctional hemoglobin and functional hemoglobin.5,6

Fig. 1

Extinction coefficient spectra of typical hemoglobin derivatives (oxygenated hemoglobin HbO,2 deoxygenated hemoglobin HbR,2 and methemoglobin3) and melanin4 in the visible wavelengths between 500 and 700 nm.

JBO_26_3_033708_f001.png

Methemoglobin is a dysfunctional form of hemoglobin, in which the ferrous (Fe2+) irons of heme are oxidized to the ferric (Fe3+) state and cannot bind oxygen molecules. Therefore, methemoglobin is incapable of transporting oxygen molecules to peripheral tissues.7 As a consequence, a large amount of methemoglobin causes methemoglobinemia, which leads to tissue hypoxia.8 Under normal physiological states, only a small amount (<2%) of hemoglobin exists in the form of methemoglobin.6 However, a higher percentage is found in individuals who suffer from methemoglobinemia due to genetic diseases912 or exposure to certain chemicals and drugs.1317 Severe untreated methemoglobinemia is life-threatening and may cause death when methemoglobin levels in the blood stream exceed 70%.6,18 There are many case reports of potentially serious methemoglobin levels induced by topical anesthesia with lidocaine and benzocaine.15,16 As with other potentially life-threatening sicknesses, early assessment of methemoglobinemia is essential for effective monitoring and treatment.

The conventional methods for evaluating methemoglobin in hypoxemic patients are arterial blood gas analysis, pulse oximetry, and CO-oximetry. Arterial blood gas analysis results could show a normal oxygen partial pressure (pO2) even when there is a high concentration of methemoglobin in the blood. This is because arterial blood gas analysis calculates only the dissolved oxygen and not the actual amount of oxygen molecules bound to hemoglobin in blood.15 Pulse oximetry measures tissue oxygen saturation at two wavelengths (typically 660 and 940 nm), and reliable results are obtained only in the presence of oxyhemoglobin and deoxyhemoglobin. Methemoglobin absorbs light equally at both wavelengths, resulting in an absorbance ratio of one, which corresponds to an oxygen saturation of 85%.19 However, when the level of methemoglobin is high, the result is unreliable, because the pulse oximeter readings remain stable. 1922 CO-oximetry is a bench-top analysis method for the evaluation of methemoglobin, and it has been used as the traditional detection tool for methemoglobinemia. Although CO-oximetry can be used to monitor the response to treatment for methemoglobinemia, it is costly and requires an invasive procedure for collecting blood samples from patients. Moreover, the testing of arterial blood gas samples by CO-oximetry requires an order from the physician. Thus the use of CO-oximetry for the measurement of methemoglobin concentrations is limited by the frequency of orders from physicians for a sequential blood gas analysis. A more rapid, simple, and cost-effective noninvasive monitoring method may allow clinicians to perform accurate diagnostic tests for methemoglobinemia.23 A pulse CO-oximeter24 with a contact probe can measure the percentage of methemoglobin continuously and non-invasively, as well as the percutaneous oxygen saturation. The Masimo Pulse CO-Oximeter (Radical-7, Masimo Corp., Irvine, CA) and The Masimo Pulse oximeter (Radical-5, Masimo Corp., Irvine, CA) have been widely used to measure the percentage of methemoglobin and the percutaneous arterial oxygen saturation (SpO2), respectively. It has been reported that the Radical-7 had a sensitivity of 97.1% and a specificity of 92.4% for detecting the percentage of methemoglobin 10% when the arterial oxygen saturation (SaO2) is >95%,25 whereas the SpO2 reading by the Radical-5 showed a sensitivity of 100% and a specificity of 93%.26 In terms of the prevention of infectious diseases, it is desirable to monitor methemoglobin and oxygen saturation levels in a non-contact manner. The chemical reactions in the heme group can result in remarkable variations in the hemoglobin absorption spectrum in the visible region. Following the oxidative reaction of hemoglobin, a significant increase in absorption occurs in the red zone of the visible spectrum, and methemoglobin has a characteristic absorption peak at 630  nm.3 Changes in the absorption spectrum due to methemoglobin formation are useful for diagnosis of methemoglobinemia and burn wound staging. Some researchers have demonstrated that optical methods have great potential for the non-invasive in vivo monitoring of methemoglobin absorption properties; these methods include broadband diffuse optical spectroscopy2730 and photoacoustic imaging.31,32 Since diffuse reflectance spectroscopy (DRS) can be achieved using simple optical components and apparatuses, it is promising as a non-invasive clinical diagnostic technique. DRS can accurately and simultaneously quantify the in vivo concentrations of melanin, oxygenated hemoglobin, and deoxygenated hemoglobin.3335 Multispectral and hyperspectral imaging techniques can be used to collect spectral information from each pixel in images, which enables the identification of various physiological conditions in living tissues.3638

In the existing literature, a method has been proposed for the measurements of melanin, oxygenated hemoglobin, and deoxygenated hemoglobin in skin tissues,35 and it has been applied to multispectral imaging for visualizing spatiotemporal changes in peripheral hemodynamics and pigmentations in response to physiological stimuli.39 In this approach, multiple regression analysis is performed using the absorbance spectrum of human skin between 500 and 600 nm as a response variable and the known extinction coefficient spectra of melanin, oxygenated hemoglobin, and deoxygenated hemoglobin as predictor variables to provide multiple regression coefficients. The concentrations of melanin and hemoglobin are then determined from the regression coefficients using empirical formulas that are deduced numerically in advance. A Monte Carlo simulation (MCS) of light transport in a human skin model is carried out to numerically establish the empirical formulas. We leveraged this approach in our study and extended it to the simultaneous quantification of methemoglobin, oxygenated hemoglobin, and deoxygenated hemoglobin. The purpose of our work was to demonstrate the feasibility of visible spectral diffuse reflectance measurement for in vivo monitoring of hemoglobin derivatives and evaluating methemoglobin production and reduction as well as hypoxemia during methemoglobinemia in rats. To confirm the feasibility of the method for quantifying the volume concentrations of methemoglobin, oxygenated hemoglobin, and deoxygenated hemoglobin during methemoglobinemia, we performed in vivo animal experiments with rats exposed to sodium nitrite (NaNO2) at different doses.

2.

Materials and Methods

2.1.

Animal Preparation

Male Wister rats (n=13) weighing 300 to 630 g were used for the animal experiments. All experimental procedures were conducted according to the protocols approved by the Animal Care Committee of Tokyo University of Agriculture and Technology (Approval No. 31-75). Anesthesia of rats was performed with isoflurane and maintained at a depth such that the rat had no response to toe pinching. After the induction of anesthesia, the dorsal region was shaved, and a depilatory agent containing thioglycolic acid was applied to the rat dorsal skin. Thirteen rats were administered a rapid single dose of NaNO2 intraperitoneally at the different dose conditions of 25, 37.5, 50, and 75 mg/kg body weight (2-ml dosing volume). Twelve out of thirteen rats were divided into four different dose groups consisting of three rats, respectively, and used only for spectral diffuse reflectance imaging. One out of thirteen rats was used for simultaneous measurements of pulse oximeter signals and spectral diffuse reflectance images.

2.2.

Experimental Apparatus

Figure 2 shows a schematic illustration of the hyperspectral imaging system used in this study. A halogen lamp light source (LA-150SAE, Hayashi Watch Works Co., Ltd., Tokyo, Japan) was used to illuminate the sample surface via a light guide with a ring-shaped illuminator. Diffusely reflected light was received by a hyperspectral camera (NH-NSD, EBA JAPAN, Tokyo, Japan) with a camera lens to acquire a hyperspectral cube. The hyperspectral camera has an internal optical stage-scanning system consisted of a slit, a collimating lens, a transmission diffraction grating, a relay lens, and a two-dimensional charge-coupled device detector sensor. A line in the scene is imaged onto the slit entrance. After collimation, the transmission diffraction grating splits the light into a series of spectral bands that are then focused onto the detector sensor via the relay lens, where one axis (x axis) records the spatial information (along the slit) and the other axis (z axis) records the spectral information. By scanning the internal optical stage along the direction perpendicular to the slit line, the camera collects two-dimensional images for adjacent lines, creating a hyperspectral cube with two spatial dimensions and one spectral dimension; the first two dimensions were spatial (x and y axes) with 640×480  pixels, whereas the third dimension (z axis) was the wavelength, ranging from 400 to 1000 nm at 10-nm intervals. The total acquisition time for one hyperspectral cube was 13 s. The spectral resolution of the camera is 10 nm. A standard white diffuser with 99% reflectance (SRS-99-020, Labsphere Incorporated, North Sutton, NH, USA) was used to correct for the interinstrument differences in the output of the camera and the spatial non-uniformity of the illumination. A ring-shaped polarizer and an analyzer were set in a crossed Nicols alignment to reduce specular reflection from the skin surface. The hyperspectral image data were then stored on a personal computer and analyzed according to the visualizing process described above. The experiments were performed in the darkroom to prevent the effect of ambient light on the spectral imaging. Therefore, ambient light had no significant effect on the imaging. Simultaneously with the optical imaging of skin tissue, the percutaneous arterial oxygen saturation [SpO2 (%)], heart rate [HR (bpm)], and pulse distention (PD [(μm)] were measured for one rat by a pulse oximeter with a foot sensor (MOUSEOX Pulse Oximeter; Star Life Science, Oakmont, PA, USA). The amplitude of the light absorption signal due to the cardiac pulse, which corresponded to the change in distention of the arterial blood vessels at the sensor location, was taken to be the PD. It can be used as an indicator of local blood flow or peripheral blood volume. After a rest of 5 min, acquisitions of hyperspectral image data were made immediately before the administration of NaNO2 (0 min) and at 5, 10, 20, 30, 60, 120, 40, 300, and 360 min after the administration of NaNO2, whereas recordings of pulse oximeter signals were made for a total of 360 min at 5-s intervals.

Fig. 2

Flow diagram of the process for estimating the concentration of methemoglobin CmetHb, oxygenated hemoglobin CHbO, deoxygenated hemoglobin CHbR, and melanin Cm. (a) Preparation step for establishing the empirical formulas and (b) main process for estimating CmetHb, CHbO, CHbR, and Cm from the measured diffuse reflectance spectrum.

JBO_26_3_033708_f002.png

2.3.

Principle for Quantifying Chromophore Concentrations

2.3.1.

Making empirical formulas for quantifying chromophore concentrations

We modified the approach proposed by Nishidate et al.,35 which was designed to quantify the three major chromophores of oxygenated hemoglobin, deoxygenated hemoglobin, and melanin. We take methemogobin into account in the skin tissue model in addition to oxygenated hemoglobin, deoxygenated hemoglobin, and melanin. Figure 3(a) shows a flow diagram of establishing the empirical formulas for quantifying concentrations of methemoglobin, oxygenated hemoglobin, deoxygenated hemoglobin, and melanin. The attenuation spectrum A(λ) was defined as

Eq. (1)

A(λ)=log101R(λ),
where R(λ) is the diffuse reflectance spectrum normalized by the incident light spectrum. Because attenuation due to light scattering can be treated as pseudochromophore,40,41 the attenuation spectrum A(λ) can be approximated as the sum of attenuations due to absorption and scattering in the skin tissue as follows:

Eq. (2)

A(λ)=CmetHbld(λ,CmetHb,CHbO,CHbR,μs)εmetHb(λ)+CHbOld(λ,CmetHb,CHbO,CHbR,μs)εHbO(λ)+CHbRld(λ,CmetHb,CHbO,CHbR,μs)εHbR(λ)+Cmle(λ,Cm,μs)εm(λ)+S(λ,μs),
where C (M) is the molar concentration, l (cm) is the mean path length, ε (λ) (cm1M1) is the molar extinction coefficient, and S(λ,μs) indicates attenuation due to light scattering in the tissue. The subscripts metHb, HbO, HbR, m, d, and e indicate methemoglobin, oxygenated hemoglobin, deoxygenated hemoglobin, melanin, dermis, and epidermis, respectively. Using A(λ) at λ=550 to 680 nm at 10-nm intervals as the response variable and ε(λ) at the same wavelength range as the predictor variables, the multiple regression analysis can be performed as

Eq. (3)

A(λ)=ametHbεmetHb(λ)+aHbOεHbO(λ)+aHbRεHbR(λ)+amεm(λ)+a0,
where ametHb, aHbO, aHbR, am, and a0 are the regression coefficients. We refer to this multiple regression analysis as MRA1. In the wavelength range 550 to 680 nm, oxygenated hemoglobin and deoxygenated hemoglobin have isosbestic points at 570 and 585 nm. Oxygenated hemoglobin and methemoglobin have an isosbestic point at 590 nm. Deoxygenated hemoglobin and methemoglobin have isosbestic points around 590 and 660 nm. Reflectance at 560 nm is sensitive to the oxygen state of hemoglobin. Therefore, we chose the wavelength range 550 to 680 nm to use for the multiple regression analysis. As described in Sec. 2.4, we used a hyperspectral camera with a 10-nm spectral resolution in this study. Therefore, we used a spectrum with 10-nm-intervals for this multiple regression analysis. The multiple regression coefficients ametHb, aHbO, aHbR, and am describe the degree of contribution of εmetHb(λ), εHbO(λ), εHbR(λ), and εm(λ), respectively, to A(λ). Therefore, the values of ametHb, aHbO, aHbR, and am are related to the products of chromophore concentrations and mean optical path lengths. We used the multiple regression coefficients ametHb, aHbO, aHbR, am, and a0 to estimate the concentrations of CmetHb, CHbO, CHbR, and Cm. For this purpose, we assumed the empirical formulas for CmetHb, CHbO, CHbR, and Cm as

Eq. (4)

CmetHb=bmetHb·a,

Eq. (5)

CHbO=bHbO·a,

Eq. (6)

CHbR=bHbR·a,

Eq. (7)

Cm=bm·a,
where

Eq. (8)

a=[1,ametHb,aHbO,aHbR,am,a0]T,

Eq. (9)

bmetHb=[bmetHb,0,bmetHb,1,bmetHb,2,bmetHb,3,bmetHb,4,bmetHb,5],

Eq. (10)

bHbO=[bHbO,0,bHbO,1,bHbO,2,bHbO,3,bHbO,4,bHbO,5],

Eq. (11)

bHbR=[bHbR,0,bHbR,1,bHbR,2,bHbR,3,bHbR,4,bHbR,5],

Eq. (12)

bm=[bm,0,bm,1,bm,2,bm,3,bm,4,bm,5].

Fig. 3

Schematic illustration of the hyperspectral imaging system.

JBO_26_3_033708_f003.png

The symbol []T represents the transposition of a vector. The coefficient vectors bmetHb, bHbO, bHbR, and bm are unknown and must be determined before estimating CmetHb, CHbO, CHbR, and Cm. We conducted further multiple regression analyses to establish the vectors bmetHb, bHbO, bHbR, and bm. In this second multiple regression analysis, CmetHb, CHbO, CHbR, and Cm were regarded as response variables, and the five regression coefficients (ametHb, aHbO, aHbR, am, and a0) that were obtained from MRA1 were regarded as predictor variables to determine the empirical formulas for CmetHb, CHbO, CHbR, and Cm. We refer to this analysis as MRA2.

We used the MCML software developed by Wang et al.42 for the MCS to simulate the diffuse reflectance spectra of skin tissues. The simulated spectra were used to derive the datasets of chromophore concentrations and the multiple regression coefficients of ametHb, aHbO, aHbR, am, and a0. The simulation model consisted of two layers representing the epidermis and dermis. 5,000,000 photon packets were launched in a single simulation of diffuse reflectance at each wavelength. The absorption coefficients of oxygenated hemoglobin μa,HbO(λ) (cm1), deoxygenated hemoglobin μa,HbR(λ) (cm1), methemoglobin μa,metHb(λ) (cm1), and melanin μa,m(λ) (cm1) were obtained from the values of εHbO(λ) (cm1M1), εHbR(λ) (cm1M1), εmetHb(λ) (cm1M1), and εmel(λ) (cm1M1) as shown in Fig. 1. The value of μa (λ) (cm1) was derived from the product of the molar concentration C (M) and the molar extinction coefficient ε(λ) (cm1M1) as, μa (λ)=2.303Cε(λ). The absorption coefficients μa,m(λ) for 10 different melanin concentrations were input for the epidermis layer in the MCS as Cm=1 to 10 vol. % at 1-vol. % intervals. We assumed that the whole blood with 2.32 mM of hemoglobin is 100% volume concentration of total hemoglobin (CHbT=100 vol. %). The sum of the absorption coefficients of oxygenated hemoglobin μa,HbO(λ), deoxygenated hemoglobin μa,HbR(λ), and methemoglobin μa,metHb(λ) represents the absorption coefficients of total hemoglobin μa,HbT(λ) (cm1). The absorption coefficients for total hemoglobin μa,HbT(λ) for the values of CHbT=0.2 to 1.0 vol. % at 0.2-vol. % intervals were input for the dermis layer in the MCS. Tissue oxygen saturation (StO2) and methemoglobin saturation (StMet) were determined by μa,HbO(λ)/μa,HbT(λ) and μa,metHb(λ)/μa,HbT(λ), respectively, and values ranged from 0% to 100% were used for the simulation. The refractive index of the epidermis and dermis layers was assumed to be same and fixed at 1.4. The thicknesses of epidermis and dermis layers were set to 0.06 and 4.94 mm, respectively. The reduced scattering coefficient μs(λ)(cm1) calculated from the typical values for the scattering coefficient35 μs(λ) (cm1) and anisotropy factor35 g(λ) were used for both epidermis and dermis layers. The value of μs at 550 nm was 31.62  cm1. In total, 1550 diffuse reflectance spectra at λ=550 to 680 nm at 10-nm intervals were simulated under the various combinations of CmetHb vol. %, CHbO vol. %, CHbR vol. %, and Cm vol. %. The MRA1 analysis for each simulated spectrum generated 1550 sets of vector a and concentrations of CmetHb vol. %, CHbO vol. %, CHbR vol. %, and Cm vol. %. The coefficient vectors bmetHb, bHbO, bHbR, and bm were determined statistically by performing MRA2. In this way, we used both multiple regression analysis and the MCS together to establish the empirical model.

2.3.2.

Estimating chromophore concentrations using the established empirical formulas

Figure 3(b) shows the main process for estimating concentrations of methemoglobin, oxygenated hemoglobin, deoxygenated hemoglobin, and melanin from the measured diffuse reflectance spectrum using the established empirical formulas. The measured diffuse reflectance spectrum R(λ) is converted into the attenuation spectrum A(λ). MRA1 is then performed using the measured A(λ) at λ=550 to 680 nm at 10-nm intervals as the response variable and ε(λ) at the same wavelength range as the predictor variables. The resultant multiple regression coefficients ametHb, aHbO, aHbR, am, and a0 are substituted into the empirical formulas established in the preparation step described in Sec 2.3.1 and finally the estimated concentrations of methemoglobin CmetHb vol. %, oxygenated hemoglobin CHbO vol. %, deoxygenated hemoglobin CHbR vol. %, and melanin Cm vol. % are obtained. The above procedure was done for each pixel.

2.4.

In Silico Experiments

We performed in silico experiments with diffuse reflectance samples generated by the MCS to validate the accuracy of the proposed method. For test samples, the values of CHbT were set to 0.3, 0.5, and 0.7 vol. %, whereas those of Cm were set to 1.5, 3.5, and 5.5 vol. % The values of StO2 were ranged from 2% to 72%, whereas those of StMet were set to 20%, 50%, and 80%. Other conditions for the MCS were fixed to be the same as those in Sec. 2.3.1. In total, 81 diffuse reflectance spectra at λ=550 to 680 nm at 10-nm intervals were generated under the combinations of CHbT, Cm, StO2, and StMet.

2.5.

Statistical Analysis

To assess the results from in silico experiments, Pearson’s correlation coefficient and root-mean-square error (RMSE) were calculated using the ground truth values and estimated values by the proposed method. A probability value of p<0.05 indicates statistical significance. We used only one area of each rat for data analysis of hyperspectral images. Region of interest (ROI) of 300×300  pixels was set in each image and the mean and the standard deviation (SD) over the ROI were calculated for the analysis of time courses in CmetHb CHbO, CHbR, CHbT Cm, StO2, and StMet. Therefore, data are expressed as mean ± SD. In order to compare the difference in each parameter across the dose conditions, the mean over the three rats in each group was also calculated.

3.

Results and Discussion

We first validated the accuracy of the proposed method by in silico experiments with diffuse reflectance samples generated by the MCS (Fig. 4). The estimated and ground truth values for total hemoglobin concentration CHbT, melanin concentration Cm, tissue oxygen saturation StO2, methemoglobin saturation StMet, methemoglobin concentration CmetHb, oxygenated hemoglobin concentration CHbO, and deoxygenated hemoglobin concentration CHbR are shown in Figs. 4(a)–(g), respectively. The estimated values are well correlated with the ground truth values for given ranges of each parameter. The value of Pearson’s correlation coefficient between the estimated and ground truth values of CHbT, Cm, StO2, StMet, CmetHb, CHbO, and CHbR were calculated to be 0.997 (p<0.0001), 0.999 (p<0.0001), 0.997 (p<0.0001), 0.998 (p<0.0001), 0.996 (p<0.0001), 0.998 (p<0.0001), and 0.997 (p<0.0001), respectively. The RMSEs of CHbT, Cm, StO2, StMet, CmetHb, CHbO, and CHbR were 0.39%, 0.18%, 3.06%, 0.45%, 0.70%, 3.03%, and 2.21%, respectively.

Fig. 4

Comparisons between the estimated and ground truth values for (a) total hemoglobin concentration CHbT; (b) melanin concentration Cm; (c) tissue oxygen saturation StO2; (d) methemoglobin saturation StMet; (e) methemoglobin concentration CmetHb; (f) oxygenated hemoglobin concentration CHbO; and (g) deoxygenated hemoglobin concentration CHbR, obtained from the in silico experiments.

JBO_26_3_033708_f004.png

We examined the spectral diffuse reflectance images obtained from rat dorsal skin at 0, 120, and 360 min after the administration of NaNO2 to confirm the impact of light absorption by methemoglobin on the spectrum. As expected from the absorption spectrum of methemoglobin shown in Fig. 1, the diffuse reflectance values in the range from 610 to 650 nm were clearly decreased at 120 min after the administration of NaNO2 (Fig. 5). We found that an increase in metHb caused by NaNO2 administration would lead to a maximum percentage decrease of 17% in diffuse reflectance at 630 nm [Fig. 6 (c)]. The diffuse reflectance at 520 nm showed that there was an initial slight increase of 2% at 5 min, followed by a decrease of 5% with a minimum value at 120 min, then it returned to its normal level at 360 min [Fig. 6 (c)]. A decrease in R560/R570, i.e., the local peak of the diffuse reflectance at 560 nm relative to that at 570 nm at 10 to 60 min after NaNO2 administration reflected the deoxygenation of hemoglobin [Fig. 6 (d)]. Methemoglobin, oxygenated hemoglobin, and deoxygenated hemoglobin have almost the same molar extinction coefficient at 520 nm, as shown in Fig. 1. Therefore, the change in diffuse reflectance at 520 nm implied a temporary increase in the volume concentration of total hemoglobin.

Fig. 5

Typical spectral diffuse reflectance images of rat dorsal skin measured by the hyperspectral camera at 0, 120, and 360 min after the administration of NaNO2. Three representative time points are illustrated to show the largest differences, i.e., at 120 min the diffuse reflectance at 610 to 650 nm was the lowest.

JBO_26_3_033708_f005.png

Fig. 6

Typical time course of diffuse reflectance spectra obtained from rat dorsal skin before and after the administration of NaNO2 for (a) 0 to 60 min and (b) 120 to 360 min. Each spectrum is the average value over the ROI in the spectral diffuse reflectance image measured by the hyper spectral camera. (c) The percent change in diffuse reflectance relative to the value at 0 min for 520, 560, 570, and 630 nm. (d) Time course of the ratio of diffuse reflectance at 560 nm and that at 570 nm.

JBO_26_3_033708_f006.png

We applied the proposed method to the spectral diffuse reflectance images obtained from rat dorsal skin before and after the administration of NaNO2 at a dose of 50 mg/kg to evaluate quantitatively changes in chromophore concentrations. Figure 7 shows typical sequential images of (a) CmetHb, (b) CHbO, (c) CHbR, (d) StMet, (e) StO2, (f) CHbT, and (g) Cm obtained from rat dorsal skin before and after the administration of NaNO2 at a dose of 50 mg/kg. Figure 8 shows the typical time courses of (a) CmetHb, (b) CHbO, (c) CHbR, (d) StMet, (e) StO2, (f) CHbT, and (g) Cm averaged over the entire region of each corresponding image shown in Fig. 7. Both CmetHb and StMet increased after the administration of NaNO2, whereas CHbO and CHbR decreased and increased, respectively, which was indicative of temporary methemoglobinemia. At 360 min after the administration of NaNO2, both CmetHb and StMet returned to their normal levels due to the self-reductive systems of the rat. The value of StO2 decreased after the administration of NaNO2, indicating temporary hypoxemia caused by methemoglobinemia as expected (Figs. 7 and 8). The value of CHbT showed an initial decrease of 0.03 vol. %, followed by a profound increase of 0.16 vol. % after the administration of NaNO2. The first phase of total hemoglobin change coincided with the decrease in both CHbO and CHbR (Fig. 8).

Fig. 7

Typical sequential images of (a) methemoglobin concentration CmetHb; (b) oxygenated hemoglobin concentration CHbO; (c) deoxygenated hemoglobin concentration CHbR; (d) methemoglobin saturation StMet; (e) tissue oxygen saturation StO2; (f) total hemoglobin concentration CHbT; and (g) melanin concentration Cm obtained from rat dorsal skin before and after the administration of NaNO2 at a dose of 50 mg/kg. Data were measured by the proposed method.

JBO_26_3_033708_f007.png

Fig. 8

Typical time courses of (a) methemoglobin concentration CmetHb; (b) oxygenated hemoglobin concentration CHbO; (c) deoxygenated hemoglobin concentration CHbR; (d) methemoglobin saturation StMet; (e) tissue oxygen saturation StO2; (f) total hemoglobin concentration CHbT; and (g) melanin concentration Cm averaged over the entire region of each corresponding image shown in Fig. 7.

JBO_26_3_033708_f008.png

To confirm the applicability of the method to imaging applications in monitoring of hemoglobin derivatives, we investigated variations in spatial distributions of CmetHb, CHbO, CHbR, StMet, StO2, and CHbT. Figure 9 shows segmentation results for sequential images shown in Fig. 7. We selected a threshold pixel value for each chromophore image, and segmented all pixels above the threshold value, and looked at how the segmented regions change over time. The average values overall pixels of images at 0 min after the administration of NaNO2 were selected as the threshold pixel values for CHbO, CHbR, and StO2, whereas those at 120 min after the administration of NaNO2 were selected as the threshold pixel values for CmetHb, CHbT, and StMet. The threshold pixel values for CmetHb, CHbO, CHbR, StMet, StO2, and CHbT were 0.24 vol. %, 0.085 vol. %, 0.05 vol. %, 82%, 65%, and 0.3 vol. %, respectively. The region of segmented pixels for CHbT began to increase in the upper half of the image at 30 min after the administration of NaNO2 and spreads to the lower half of the image until 120 min after the administration of NaNO2 [Fig. 9(f)]. The same tendency can be observed in the segmented image of CmetHb [Fig. 9(a)]. On the other hand, the decreases in the regions of segmented pixels for both CHbO and CHbR spread from the upper side to the lower side of the image until 120 min after the administration of NaNO2 [Figs. 9(b) and 9(c)]. The cause of the variations in spatial distribution of CmetHb, CHbO, CHbR, and CHbT is unclear. One possible explanation for those variations is heterogeneous microvascular perfusion distribution. Changes in the regions of segmented pixels for StMet are spatially uniform, which is probably due to that StMet is calculated as the ratio of methemoglobin to total hemoglobin [Fig. 9(d)]. The same applies to the segmentation image of StO2 that is the ratio of oxygenated hemoglobin to total hemoglobin [Fig. 9(e)].

Fig. 9

Segmentation results for sequential images shown in Fig. 7. (a) methemoglobin concentration CmetHb; (b) oxygenated hemoglobin concentration CHbO; (c) deoxygenated hemoglobin concentration CHbR; (d) methemoglobin saturation StMet; (e) tissue oxygen saturation StO2; and (f) total hemoglobin concentration CHbT. Red color and gray color show a pixel above and below a threshold, respectively. The average values overall pixels of images at 0 min were selected as the threshold pixel values for CHbO, CHbR, and StO2, whereas those at 120 min were selected as the threshold pixel values for CmetHb, CHbT, and StMet. The threshold pixel values for CmetHb, CHbO, CHbR, StMet, StO2, and CHbT were 0.24 vol. %, 0.085 vol. %, 0.05 vol. %, 82%, 65%, and 0.3 vol. %, respectively.

JBO_26_3_033708_f009.png

We compared the time courses of (a) CmetHb, (b) CHbT, (c) StMet, and (d) StO2 for NaNO2 doses of 25, 37.5, 50, and 75 mg/kg to investigate the dose-dependent effects of NaNO2 (Fig. 10). The values of CmetHb and StMet rapidly increased with a half-maximum time of <20  min, and the time required to return to the normal levels increased proportionally with the dose [Figs. 10(a) and 10(c)]. On the other hand, at each dose condition, the time course of StO2 dramatically dropped at 60 min after the administration of NaNO2, and then gradually increased again [Fig. 10(d)]. The minimum averaged values of StO2 during methemoglobinemia were 33.7%±0.4%, 23.1%±5.6%, 8.8%±1.7%, and 9.7%±5.1% on average for NaNO2 doses of 25, 37.5, 50, and 75 mg/kg, respectively. The values of StO2 were negatively correlated to those of CmetHb and StMet. All three rats with the dose condition of 75 mg/kg body weight died at 60 to 69 min after the administration of NaNO2, whereas those with the other dose conditions survived. Therefore, the median lethal dose was speculated to be within the dose of 50 to 75 mg/kg. We found that the maximum value of StMet increased with increasing dose up to 50 mg/kg, in which it reached a plateau [Fig. 10 (e)].

Fig. 10

Time courses of (a) methemoglobin concentration CmetHb; (b) total hemoglobin concentration CHbT; (c) methemoglobin saturation StMet; and (d) tissue oxygen saturation StO2 for various doses of NaNO2. (e) Relationship between the maximum value of StMet and the NaNO2 dose. Data were obtained by the proposed method. Solid lines in (a)–(d) are means of three rats for each dose group. Plots and error bars in (e) show means and SDs of three rats for each dose group.

JBO_26_3_033708_f010.png

In order to discuss the results of total hemoglobin concentration CHbT after the administration of NaNO2, we compared the time course of CHbT obtained by the proposed method with HR and pulse distension (PD) measured by the commercial existing pulse oximeter [Figs. 11(a) and 11(b)]. We found that the HR significantly increased to a maximum of 323 bpm at 135 min after the administration of NaNO2, then it gradually decreased for 210 min [Fig. 11(a)]. This increase in HR corresponded to tachycardia, which is a symptom of methemoglobinemia.43 The PD after the administration of NaNO2 initially decreased to 9.3  μm, but then it significantly increased to a maximum of 76.2  μm at 63 min, then subsequently returned to normal levels [Fig. 11(b)]. The profound increase in PD indicated an elevated cardiac output caused by tachycardia and the resultant increase in peripheral blood volume to compensate for the hypoxia and hypoxemia during methemoglobinemia; this was in agreement with the time course of CHbT obtained by the proposed method.

Fig. 11

Time courses of (a) HR and total hemoglobin concentration CHbT; (b) PD and total hemoglobin concentration CHbT; and (c) percutaneous arterial oxygen saturation SpO2 and tissue oxygen saturation StO2 for rat after the administration of NaNO2 at a dose of 37.5 mg/kg. HR, PD, and SpO2 were measured by the commercial existing pulse oximeter, whereas CHbT and StO2 were obtained by the proposed method.

JBO_26_3_033708_f011.png

We also compared the time course of StO2 obtained by the proposed method with percutaneous arterial oxygen saturation SpO2 measured by the commercial existing pulse oximeter. We found that the SpO2 significantly dropped after the administration of NaNO2, then remained around 82.5% for 160 min [Fig. 11(c)]. This observation was in agreement with several reports of SpO2 measured by conventional pulse oximetry during methemoglobinemia.1922 It has been reported that methemoglobinemia can cause a discrepancy between the values of SpO2 measured by a conventional pulse oximeter and the actual arterial oxygen saturation (SaO2) calculated by arterial blood gas measurement.1922 The values of SpO2 are not proportional to the actual SaO2 during methemoglobinemia, since elevated methemoglobin levels (up to 60%) force the SpO2 value to be around 85%, whereas SaO2 reaches 35%.19 When methemoglobin is elevated, the SpO2 values measured by conventional pulse oximetry are inaccurate and can falsely diagnose the actual status of patients.

In contrast, the values of StO2 obtained by the proposed method were dramatically decreased with a minimum of 19.9% as shown in Fig. 11(c). Since the value of StO2 estimated by the proposed method represents the oxygen saturation for a mixture of SaO2 and SvO2, it should be calculated to be lower than the SaO2 value. Considering that the SaO2 value has been reported to be 35%during methemoglobinemia,19 it is reasonable that the minimum averaged values of StO2 during methemoglobinemia showed 33.7%±0.4%, 23.1%±5.6%, 8.8%±1.7%, and 9.7%±5.1% on average for NaNO2 doses of 25, 37.5, 50, and 75 mg/kg, respectively, as shown in Fig. 10(d). The results for StO2 indicated that the proposed method can be used to evaluate hypoxemia during methemoglobinemia.

The advantage of the diffuse reflectance imaging over the DRS is the ability to visualize the spatial distribution of chromophores. For example, the spatial distributions of oxygenated hemoglobin concentration, deoxygenated hemoglobin concentration, methemoglobin concentration, and tissue oxygen saturation are important for evaluating the depth of burn injury to classify the severity of burn injury. This issue should be studied further in the future. Near-infrared spectroscopy measurements2932 also considered lipids and water as major chromophores in addition to hemoglobin derivatives because lipids and water have distinct absorption peaks in the near-infrared wavelength region. On the other hand, there is no significant absorption by lipids or water in the wavelength range from 550 to 680 nm we used. Therefore, we did not consider lipids and water as major chromophores in this study. The current empirical formulas for chromophores were derived from the MCS with a typical reduced scattering coefficient spectrum. However, the reduced scattering coefficient spectrum usually differs among body parts and may vary with the age of the subjects. Correct estimation of the scattering properties or consideration of the variations in the reduced scattering coefficient spectrum is essential to precisely estimate the chromophore concentrations. We assumed the typical thicknesses of epidermis and the dermis in the MCS to establish the empirical model. The assumed epidermis and dermis thicknesses could have an impact to the results. The probability that light is absorbed by melanin in the epidermis will be higher than the probability that light is absorbed by hemoglobin in the dermis as epidermis becomes thicker. Therefore, if the thickness of epidermis is larger than the typical value, melanin concentration in epidermis or hemoglobin derivatives concentrations may be under- or overestimated, respectively. The full thickness of the skin tissue model was set to 5 mm in the MCS. The simulated diffuse reflectance could be increased if the total thickness of skin tissue model is larger than 5 mm. Hence, the MCS on a thicker tissue model with a semi-infinite geometry could change the results. However, the possibility of relative measurements of oxygenated hemoglobin, deoxygenated hemoglobin, methemoglobin, and melanin in the skin tissue could still prove valuable. Although we did not consider the polarized photons in the MCS, the diffuse reflectance was calculated separately from the specular reflectance in the simulation and used for in a data set for the empirical model. The quantification accuracy may be improved by taking into account the polarization effects in the MCS. We performed the MCS using the MCML software. The MCML software basically output the physical quantities of photon absorption, fluence, diffuse reflectance, specular reflectance, and total transmittance. It could be possible to estimate the mean path length if we modified the MCML source code. The estimated mean path length should be compared with the values of a and b in the future. The present method lacks depth resolution, because it relies on the integration of all diffusing reflection information along the depth direction. Moreover, the present method is based on the MCS model, which assumes uniformly distributed scattering and absorption properties. The current algorithm performs the multiple regression analysis using each pixel in the spectral image to calculate the multiple regression coefficients. This process is relatively time-consuming, and the computational load increases when the number of pixels in an image for analysis increases. Therefore, this method may not be suitable for real-time high temporal and spatial resolution imaging of chromophores in vivo. Although hyperspectral imaging techniques have been widely used for scientific purposes in the laboratory environment, it is impractical for clinical use due to the large size and high cost of commercial setups. It is advantageous that the proposed algorithm can also be combined with the RGB camera-based spectral image reconstruction technique based on the Wiener estimation method.44 This is promising for the establishment of a compact and affordable imaging system that can simultaneously evaluate the spatiotemporal distributions of methemoglobin, oxygenated hemoglobin, and deoxygenated hemoglobin in clinical point-of-care testing. Such a device would have an enormous impact on the health of people of all ages worldwide, and this issue should be studied further in the future.

4.

Conclusions

In summary, this study demonstrated the usefulness of a new non-contact measuring method based on DRS for the simultaneous evaluations of the precutaneous volume concentrations of methemoglobin, oxygenated hemoglobin, deoxygenated hemoglobin, and melanin as well as tissue oxygen saturation and methemoglobin saturation. In vivo experiments with rat dorsal skin before and after the administration of NaNO2 at doses of 25, 37.5, 50, and 75 mg/kg were performed to induce methemoglobinemia. Both the methemoglobin concentration and methemoglobin saturation rapidly increased with a half-maximum time of <20  min. They reached their maximal values nearly 60 min after the administration of NaNO2, and the time required to return to the normal levels increased proportionally with the dose, which implied that the methemoglobinemia was temporary. The values of tissue oxygen saturation dramatically dropped after the administration of NaNO2, and then they gradually increased for each dose condition, indicating temporary hypoxemia caused by methemoglobinemia. The time course of the peripheral volume concentration of total hemoglobin during methemoglobinemia obtained by the proposed method reflects both the HR and PD that are measured by the commercially available pulse oximeter. The results of this study showed that the proposed method has potential for the simultaneous evaluations of the methemoglobin level, regional peripheral blood volume, and hypoxemia during methemoglobinemia.

Disclosures

The authors declare no conflicts of interest.

Acknowledgments

This manuscript is a revised version of a report previously published in an SPIE conference proceedings Volume 11521, 115210G (2020). Part of this work was supported by JSPS KAKENHI Grant No. 17H06102.

References

1. 

A. N. Yaroslavsky et al., “Optics of Blood,,” Handbook of Optical Biomedical Diagnostics, SPIE Press, Bellingham, Washington (2002). Google Scholar

2. 

S. A. Prahl, “Tabulated molar extinction coefficient for hemoglobin in water,” (1999). http://omlc.ogi.edu/spectra/hemoglobin/summary.html Google Scholar

3. 

W. G. Zijlstra and A. Buursma, “Spectrophotometry of hemoglobin: absorption spectra of bovine oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin, and methemoglobin,” Comp. Biochem. Physiol., 118 (4), 743 –749 (1997). https://doi.org/10.1016/S0305-0491(97)00230-7 Google Scholar

4. 

S. L. Jacques, R. D. Glickman and J. A. Schwartz, “Internal absorption coefficient and threshold for pulsed laser disruption of melanosomes isolated from retinal pigment epithelium,” Proc. SPIE, 2681 468 –477 (1996). https://doi.org/10.1117/12.239608 PSISDG 0277-786X Google Scholar

5. 

G. V. G. Baranoski et al., “On the noninvasive optical monitoring and differentiation of methemoglobinemia and sulfhemoglobinemia,” J. Biomed. Opt., 17 (9), 097005 (2012). https://doi.org/10.1117/1.JBO.17.9.097005 JBOPFO 1083-3668 Google Scholar

6. 

S. Haymond et al, “Laboratory assessment of oxygenation in methemoglobinemia,” Clin. Chem., 51 (2), 434 –444 (2005). https://doi.org/10.1373/clinchem.2004.035154 Google Scholar

7. 

R. K. Murray, D. K. Granner and V. W. Rodwell, Harper’s Illustrated Biochemistry, 27th ed.McGraw Hill Companies, Asia (2006). Google Scholar

8. 

M. C. Kohn et al., “Pharmacokinetics of sodium nitrite-induced methemoglobinemia in the rat,” Drug. Metab. Dispos., 30 676 –683 (2002). https://doi.org/10.1124/dmd.30.6.676 DMDSAI 0090-9556 Google Scholar

9. 

S. S. Da-Silva, I. S. Sajan and J. P. Underwood, “Congenital methemoglobinemia: a rare cause of cyanosis in the newborn—a case report,” Pediatrics, 112 e158 –e161 (2003). https://doi.org/10.1542/peds.112.2.e158 PEDIAU 0031-4005 Google Scholar

10. 

R. B. May, “An infant with sepsis and methemoglobinemia,” J. Emerg. Med., 3 261 –264 (1985). https://doi.org/10.1016/0736-4679(85)90428-7 JEMMEP 1351-0622 Google Scholar

11. 

K. Ohashi et al., “Elevated methemoglobin in patients with sepsis,” Acta. Anaesthesiol. Scand., 42 713 –716 (1998). https://doi.org/10.1111/j.1399-6576.1998.tb05306.x Google Scholar

12. 

R. O. Wright, W. J. Lewander and A. D. Woolf, “Methemoglobinemia: etiology, pharmacology, and clinical management,” Ann. Emerg. Med., 34 646 –656 (1999). https://doi.org/10.1016/S0196-0644(99)70167-8 AEMED3 0196-0644 Google Scholar

13. 

World Health Organization, “Nitrate and nitrite in drinking-water,,” WHO/FWC/WSH/16.52(2016). Google Scholar

14. 

G. F. Craun, D. G. Greathouse and D. H. Gunderson, “Methemoglobin levels in young children consuming high nitrate well water in the United States,” Int. J. Epidem., 10 309 –317 (1981). https://doi.org/10.1093/ije/10.4.309 IJEPBF 0300-5771 Google Scholar

15. 

K. Kern, P. B. Langevin and B. M. Dunn, “Methemoglobinemia after topical anesthesia with lidocaine and benzocaine for a difficult intubation,” J. Clin. Anesth., 12 167 –172 (2000). https://doi.org/10.1016/S0952-8180(00)00113-6 JCLBE7 Google Scholar

16. 

R. K. Singh et al., “Benzocaine induced methemoglobinemia accompanying adult respiratory distress syndrome and sepsis syndrome: case report,” J. Trauma, 50 1153 –1157 (2001). https://doi.org/10.1097/00005373-200106000-00030 Google Scholar

17. 

C. Udeh, J. Bittikofer and S. T. Sum-Ping, “Severe methemoglobinemia on reexposure to benzocaine,” J. Clin. Anesth., 13 128 –130 (2001). https://doi.org/10.1016/S0952-8180(01)00222-7 JCLBE7 Google Scholar

18. 

H. U. Rehman, “Methemoglobinemia,” West. J. Med., 175 (3), 193 –196 (2001). https://doi.org/10.1136/ewjm.175.3.193 Google Scholar

19. 

S. J. Barker, K. K. Tremper and J. Hyatt, “Effects of methemoglobinemia on pulse oximetry and mixed venous oximetry,” Anesthesiology, 70 112 –117 (1989). https://doi.org/10.1097/00000542-198901000-00021 ANESAV 0003-3022 Google Scholar

20. 

C. M. Alexander, L. E. Teller and J. B. Gross, “Principles of pulse oximetry: theoretical and practical considerations,” Anesth. Analg., 68 368 –376 (1989). https://doi.org/10.1213/00000539-198903000-00035 Google Scholar

21. 

L. Delwood et al., “Methemoglobinemia and its effect on pulse oximetry,” Crit. Care. Med., 19 988 (1991). https://doi.org/10.1097/00003246-199107000-00030 CCMDC7 0090-3493 Google Scholar

22. 

A. S. O. Tang et al., “The mystery of ‘saturation gap’: a case of dapsone-induced methaemoglobinemia in a pregnant mother with leprosy,” Oxf. Med. Case Rep., 2019 (1), 8 –10 (2019). https://doi.org/10.1093/omcr/omy111 Google Scholar

23. 

R. Ash-Bernal, R. Wise and S. M. Wright, “Acquired methemoglobinemia: a retrospective series of 138 cases at 2 teaching hospitals,” Medicine, 83 (5), 265 –273 (2004). https://doi.org/10.1097/01.md.0000141096.00377.3f MEDIAV 0025-7974 Google Scholar

24. 

E. H. Annabi and S. J. Barker, “Severe methemoglobinemia detected by pulse oximetry,” Anesth. Analg., 108 (3), 898 –899 (2009). https://doi.org/10.1213/ane.0b013e318172af73 Google Scholar

25. 

J. R. Feiner and P. E. Bickler, “Improved accuracy of methemoglobin detection by pulse CO-oximetry during hypoxia,” Anesth. Analg., 111 (5), 1160 –1167 (2010). https://doi.org/10.1213/ANE.0b013e3181f46da8 Google Scholar

26. 

N. Shah et al., “Performance of three new-generation pulse oximeters during motion and low perfusion in volunteers,” J. Clin. Anesth., 24 (5), 385 –391 (2012). https://doi.org/10.1016/j.jclinane.2011.10.012 JCLBE7 Google Scholar

27. 

J. Lee et al., “Noninvasive in vivo monitoring of methemoglobin formation and reduction with broadband diffuse optical spectroscopy,” J. Appl. Physiol., 100 (2), 615 –622 (2006). https://doi.org/10.1152/japplphysiol.00424.2004 Google Scholar

28. 

J. B. Fishkin et al., “Frequency-domain method for measuring spectral properties in multiple-scattering media: methemoglobin absorption spectrum in a tissuelike phantom,” Appl. Opt., 34 1143 –1155 (1995). https://doi.org/10.1364/AO.34.001143 APOPAI 0003-6935 Google Scholar

29. 

R. B. Saager et al., “Impact of hemoglobin breakdown products in the spectral analysis of burn wounds using spatial frequency domain spectroscopy,” J. Biomed. Opt., 24 (2), 020501 (2019). https://doi.org/10.1117/1.JBO.24.2.020501 JBOPFO 1083-3668 Google Scholar

30. 

G. Leung et al., “Rapid tissue viability evaluation using methemoglobin as a biomarker in burns,” Int. J. Burns Trauma, 8 (5), 126 –134 (2018). Google Scholar

31. 

M. Tang et al., “Noninvasive photoacoustic microscopy of methemoglobin in vivo,” J. Biomed. Opt., 20 (3), 036007 (2015). https://doi.org/10.1117/1.JBO.20.3.036007 JBOPFO 1083-3668 Google Scholar

32. 

K. Aizawa et al., “In vivo photoacoustic spectroscopic imaging of hemoglobin derivatives in thermally damaged tissue,” Jpn. J. Appl. Phys., 48 062302 (2009). https://doi.org/10.1143/JJAP.48.062302 Google Scholar

33. 

G. Zonios, J. Bykowski and N. Kollias, “Skin melanin, hemoglobin, and light scattering properties can be quantitatively assessed in vivo using diffuse reflectance spectroscopy,” J. Invest. Dermatol., 117 1452 –1457 (2001). https://doi.org/10.1046/j.0022-202x.2001.01577.x JIDEAE 0022-202X Google Scholar

34. 

S.-H. Tseng et al., “Chromophore concentrations, absorption and scattering properties of human skin in vivo,” Opt. Expess, 17 14599 –14617 (2009). https://doi.org/10.1364/OE.17.014599 Google Scholar

35. 

I. Nishidate, Y. Aizu and H. Mishina, “Estimation of melanin and hemoglobin in skin tissue using multiple regression analysis aided by Monte Carlo simulation,” J. Biomed. Opt., 9 (4), 700 –710 (2004). https://doi.org/10.1117/1.1756918 JBOPFO 1083-3668 Google Scholar

36. 

A. Nkengne et al., “‘SpectraCam®: a new polarized hyperspectral imaging system for repeatable and reproducible in vivo skin quantification of melanin, total hemoglobin, and oxygen saturation,” Skin Res. Technol., 24 99 –107 (2018). https://doi.org/10.1111/srt.12396 Google Scholar

37. 

S. F. Bish et al., “Handheld diffuse reflectance spectral imaging (DRSi) for in-vivo characterization of skin,” Biomed. Opt. Express, 5 573 –586 (2014). https://doi.org/10.1364/BOE.5.000573 BOEICL 2156-7085 Google Scholar

38. 

B. S. Sorg et al., “Hyperspectral imaging of hemoglobin saturation in tumor microvasculature and tumor hypoxia development,” J. Biomed. Opt., 10 (4), 044004 (2005). https://doi.org/10.1117/1.2003369 JBOPFO 1083-3668 Google Scholar

39. 

I. Nishidate et al., “Non-invasive spectral imaging of skin chromophores based on multiple regression analysis aided by Monte Carlo simulation,” Opt. Lett., 36 3239 –3241 (2011). https://doi.org/10.1364/OL.36.003239 OPLEDP 0146-9592 Google Scholar

40. 

M. Kohl et al., “Physical model for the spectroscopic analysis of cortical intrinsic optical signals,” Phys. Med. Biol., 45 (12), 3749 –3764 (2000). https://doi.org/10.1088/0031-9155/45/12/317 PHMBA7 0031-9155 Google Scholar

41. 

P. B. Jones et al., “Simultaneous multispectral reflectance imaging and laser speckle flowmetry of cerebral blood flow and oxygen metabolism in focal cerebral ischemia,” J. Biomed. Opt., 13 (4), 044007 (2008). https://doi.org/10.1117/1.2950312 JBOPFO 1083-3668 Google Scholar

42. 

L. Wang, S. L. Jacques and L. Zheng, “MCML: Monte Carlo modeling of light transport in multi-layered tissues,” Comput. Methods Prog. Biomed., 47 131 –146 (1995). https://doi.org/10.1016/0169-2607(95)01640-F Google Scholar

43. 

“FDA Drug Safety Communication: FDA continues to receive reports of a rare, but serious and potentially fatal adverse effect with the use of benzocaine sprays for medical procedures,” (2011) https://www.fda.gov/drugs/drug-safety-and-availability/fda-drug-safety-communication-fda-continues-receive-reports-rare-serious-and-potentially-fatal Google Scholar

44. 

I. Nishidate et al., “Estimation of melanin and hemoglobin using spectral reflectance images reconstructed from a digital RGB image by the Wiener Estimation Method,” Sensors, 13 7902 –7915 (2013). https://doi.org/10.3390/s130607902 SNSRES 0746-9462 Google Scholar

Biography

Fahima Khatun received her DVM and MS degrees in parasitology from Bangladesh Agricultural University in 2010 and 2012, respectively. She is a PhD student at Tokyo University of Agriculture and Technology. Her current research interest is biomedical optics for disease diagnosis of human and animal beings.

Yoshihisa Aizu is a professor at Muroran Institute of Technology, Japan. He received his BE degree in electronics from Musashi Institute of Technology in 1980, ME and PhD degrees in electronics from Hokkaido University in 1982 and 1985, respectively. From 1992 to 1993, he was a visiting researcher in University Erlangen, Germany. His current research activities are in applications of optical imaging, spectroscopy, colorimetry, and laser speckle imaging to human and biological objects.

Izumi Nishidate received his MS and PhD degrees from Muroran Institute of Technology, Japan. He is an associate professor in the Graduate School of Bio-Applications and Systems Engineering, Tokyo University of Agriculture and Technology. His research interests include diffuse reflectance spectroscopy, light transport in biological tissues, multispectral imaging, in vivo functional imaging of living tissues, and photodynamic therapy.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Fahima Khatun, Yoshihisa Aizu, and Izumi Nishidate "Transcutaneous monitoring of hemoglobin derivatives during methemoglobinemia in rats using spectral diffuse reflectance," Journal of Biomedical Optics 26(3), 033708 (13 February 2021). https://doi.org/10.1117/1.JBO.26.3.033708
Received: 31 August 2020; Accepted: 21 January 2021; Published: 13 February 2021
Lens.org Logo
CITATIONS
Cited by 7 scholarly publications and 1 patent.
Advertisement
Advertisement
KEYWORDS
Diffuse reflectance spectroscopy

Oxygen

Tissues

Blood

Absorption

Skin

Image segmentation

Back to Top