Safety and Efficacy of the Neuroform Atlas Stent for Treatment of Intracranial Aneurysms

A Systematic Review and Meta-Analysis

  • Original Article
  • Published:
Clinical Neuroradiology Aims and scope Submit manuscript

Abstract

Objective

The Neuroform Atlas (Stryker Neurovascular, Fremont, CA, USA) is a low-profile laser cut self-expanding nitinol stent designed to provide coil support and wall apposition during aneurysm embolisation. In this study, we performed a meta-analysis of outcomes after treatment with the Neuroform Atlas stent for the purpose of coil embolisation.

Methods

The primary objectives of this meta-analysis were to define the safety (treatment-related complications, neurologic outcomes, mortality rate) and the efficacy (aneurysm occlusion rate) of the treatment of intracranial aneurysms with the Neuroform Atlas stent. A systematic review and meta-analysis was performed by searching PubMed, EMBASE, and the Cochrane CENTRAL Library for all published studies on the treatment of intracranial aneurysms with the Neuroform Atlas device up to 6 April 2020. The review was prepared in accordance with the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines.

Results

A total of 14 studies were analysed (577 patients with 593 intracranial aneurysms). The mean age was 58.2 years and 35.6% were male. Technical success of the procedure was 100%. RROC1/RROC2 (Raymond-Roy occlusion classification (RROC) 1/2) (total occlusion/neck remnant) at a mean follow-up of 8.9 months was achieved in 94.8%. RROC3 was 4.9%. All-cause mortality was 1.8% and permanent residual neurological deficit or disability was 2.7%. Overall complications at follow-up were 6.2%.

Conclusion

Our analysis demonstrated good rates of occlusion at follow-up for aneurysms treated with the Atlas device at follow-up. The safety profile appears similar to other low-profile intracranial stents.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  1. McLaughlin N, McArthur DL, Martin NA. Use of stent-assisted coil embolization for the treatment of wide-necked aneurysms: A systematic review. Surg Neurol Int. 2013;4:43.

    Article  Google Scholar 

  2. Fiorella D, Albuquerque FC, Han P, McDougall CG. Preliminary experience using the Neuroform stent for the treatment of cerebral aneurysms. Neurosurgery. 2004;54:6–16; discussion 16–7.

    Article  Google Scholar 

  3. Howington JU, Hanel RA, Harrigan MR, Levy EI, Guterman LR, Hopkins LN. The Neuroform stent, the first microcatheter-delivered stent for use in the intracranial circulation. Neurosurgery. 2004;54:2–5.

    Article  Google Scholar 

  4. Shapiro M, Becske T, Sahlein D, Babb J, Nelson PK. Stent-supported aneurysm coiling: a literature survey of treatment and follow-up. AJNR Am J Neuroradiol. 2012;33:159–63.

    Article  CAS  Google Scholar 

  5. Lopes DK, Johnson AK, Kellogg RG, Heiferman DM, Keigher KM. Long-term radiographic results of stent-assisted embolization of cerebral aneurysms. Neurosurgery. 2014;74:286–91.

    Article  Google Scholar 

  6. Chalouhi N, Starke RM, Koltz MT, Jabbour PM, Tjoumakaris SI, Dumont AS, Rosenwasser RH, Singhal S, Gonzalez LF. Stent-assisted coiling versus balloon remodeling of wide-neck aneurysms: comparison of angiographic outcomes. AJNR Am J Neuroradiol. 2013;34:1987–92.

    Article  CAS  Google Scholar 

  7. Lopes D, Sani S. Histological postmortem study of an internal carotid artery aneurysm treated with the Neuroform stent. Neurosurgery. 2005;56:E416; discussion E416.

  8. Aydin K, Arat A, Sencer S, Barburoglu M, Men S. Stent-Assisted Coiling of Wide-Neck Intracranial Aneurysms Using Low-Profile LEO Baby Stents: Initial and Midterm Results. AJNR Am J Neuroradiol. 2015;36:1934–41.

    Article  CAS  Google Scholar 

  9. Möhlenbruch M, Herweh C, Behrens L, Jestaedt L, Amiri H, Ringleb PA, Bendszus M, Pham M. The LVIS Jr. microstent to assist coil embolization of wide-neck intracranial aneurysms: clinical study to assess safety and efficacy. Neuroradiology. 2014;56:389–95.

    Article  Google Scholar 

  10. Park SY, Oh JS, Oh HJ, Yoon SM, Bae HG. Safety and Efficacy of Low-Profile, Self-Expandable Stents for Treatment of Intracranial Aneurysms: Initial and Midterm Results - A Systematic Review and Meta-Analysis. Interv Neurol. 2017;6:170–82.

    Article  Google Scholar 

  11. Jankowitz BT, Hanel R, Jadhav AP, Loy DN, Frei D, Siddiqui AH, Puri AS, Khaldi A, Turk AS, Malek AM, Sauvageau E, Hetts SW, Zaidat OO. Neuroform Atlas Stent System for the treatment of intracranial aneurysm: primary results of the Atlas Humanitarian Device Exemption cohort. J Neurointerv Surg. 2019;11:801–6.

    Article  Google Scholar 

  12. Moher D, Liberati A, Tetzlaff J, Altman DG; PRISMA Group. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. Int J Surg. 2010;8:336–41. Erratum in: Int J Surg. 2010;8:658.

    Article  Google Scholar 

  13. Roy D, Milot G, Raymond J. Endovascular treatment of unruptured aneurysms. Stroke. 2001;32:1998–2004.

    Article  CAS  Google Scholar 

  14. Stang A. Critical evaluation of the Newcastle-Ottawa scale for the assessment of the quality of nonrandomized studies in meta-analyses. Eur J Epidemiol. 2010;25:603–5.

    Article  Google Scholar 

  15. Ciccio G, Robert T, Smajda S, Fahed R, Desilles JP, Redjem H, Escalard S, Mazighi M, Blanc R, Piotin M. Double stent assisted coiling of intracranial bifurcation aneurysms in Y and X configurations with the Neuroform ATLAS stent: immediate and mid term angiographic and clinical follow-up. J Neurointerv Surg. 2019;11:1239–42.

    Article  Google Scholar 

  16. Aydin K, Balci S, Sencer S, Barburoglu M, Umutlu MR, Arat A. Y-Stent-Assisted Coiling With Low-Profile Neuroform Atlas Stents for Endovascular Treatment of Wide-Necked Complex Intracranial Bifurcation Aneurysms. Neurosurgery. 2020;87:744–53. https://doi.org/10.1093/neuros/nyz516.

    Article  PubMed  Google Scholar 

  17. Baek JW, Jin SC, Kim JH, Yoo MW, Jeong HW, Seo JH, Han JY, Heo YJ, Kim ST. Initial multicentre experience using the neuroform atlas stent for the treatment of un-ruptured saccular cerebral aneurysms. Br J Neurosurg. 2020;34:333–8.

    Google Scholar 

  18. Pan S, Kannath SK, Rajan JE. Low-Profile Neuroform Atlas Stent-Assisted Coiling of Intracranial Aneurysms: Single-Center Initial Experience. J Clin Interv Radiol Isvir. 2019;03:117–21.

    Article  Google Scholar 

  19. Park KY, Jang CK, Lee JW, Kim DJ, Kim BM, Chung J. Preliminary experience of stent-assisted coiling of wide-necked intracranial aneurysms with a single microcatheter. BMC Neurol. 2019;19:245.

    Article  Google Scholar 

  20. Zhang X, Zhong J, Gao H, Xu F, Bambakidis NC. Endovascular treatment of intracranial aneurysms with the LVIS device: a systematic review. J Neurointerv Surg. 2017;9:553–7.

    Article  Google Scholar 

  21. Cagnazzo F, Cappucci M, Lefevre PH, Dargazanli C, Gascou G, Morganti R, Mazzotti V, di Carlo D, Perrini P, Mantilla D, Riquelme C, Bonafe A, Costalat V. Treatment of Intracranial Aneurysms with Self-Expandable Braided Stents: A Systematic Review and Meta-Analysis. AJNR Am J Neuroradiol. 2018;39:2064–9.

    Article  CAS  Google Scholar 

  22. Cagnazzo F, Limbucci N, Nappini S, Renieri L, Rosi A, Laiso A, Tiziano di Carlo D, Perrini P, Mangiafico S. Y-Stent-Assisted Coiling of Wide-Neck Bifurcation Intracranial Aneurysms: A Meta-Analysis. AJNR Am J Neuroradiol. 2019;40:122–8.

    Article  CAS  Google Scholar 

  23. Phan K, Huo YR, Jia F, Phan S, Rao PJ, Mobbs RJ, Mortimer AM. Meta-analysis of stent-assisted coiling versus coiling-only for the treatment of intracranial aneurysms. J Clin Neurosci. 2016;31:15–22.

    Article  Google Scholar 

  24. Hong Y, Wang YJ, Deng Z, Wu Q, Zhang JM. Stent-assisted coiling versus coiling in treatment of intracranial aneurysm: a systematic review and meta-analysis. PLoS One. 2014;9:e82311.

    Article  CAS  Google Scholar 

  25. King B, Vaziri S, Singla A, Fargen KM, Mocco J. Clinical and angiographic outcomes after stent-assisted coiling of cerebral aneurysms with Enterprise and Neuroform stents: a comparative analysis of the literature. J Neurointerv Surg. 2015;7:905–9.

    Article  Google Scholar 

  26. Gross BA, Ares WJ, Ducruet AF, Jadhav AP, Jovin TG, Jankowitz BT. A clinical comparison of Atlas and LVIS Jr stent-assisted aneurysm coiling. J Neurointerv Surg. 2019;11:171–4.

    Article  Google Scholar 

  27. Kim CH, Kim YH, Sung SK, Son DW, Song GS, Lee SW. Clinical Safety and Effectiveness of Stent-Assisted Coil Embolization with Neuroform Atlas Stent in Intracranial Aneurysm. J Korean Neurosurg Soc. 2020;63:80–8.

    Article  Google Scholar 

  28. Tsai JP, Hardman J, Moore NZ, Hussain MS, Bain MD, Rasmussen PA, Masaryk TJ, Elgabaly MH, Sheikhi L, Toth G. Early post-Humanitarian Device Exemption experience with the Neuroform Atlas stent. J Neurointerv Surg. 2019;11:1141–4.

    Article  Google Scholar 

  29. Quintana EM, Valdes PV, Deza EM, García AG, Rodríguez MC, Pérez JMJ, Chaviano J, Morales JCG, Batista KP, García FA. Initial experience and one-year follow-up with Neuroform Atlas Stent System for the treatment of brain aneurysms. Interv Neuroradiol. 2019;25:521–9.

    Article  Google Scholar 

  30. Ten Brinck MFM, de Vries J, Bartels RHMA, Grotenhuis JA, Boogaarts HD. NeuroForm Atlas Stent-Assisted Coiling: Preliminary Results. Neurosurgery. 2019;84:179–89.

    Article  Google Scholar 

  31. Goertz L, Dorn F, Siebert E, Herzberg M, Borggrefe J, Schlamann M, Krischek B, Stavrinou P, Mpotsaris A, Bohner G, Liebig T, Kabbasch C. Safety and efficacy of the Neuroform Atlas for stent-assisted coiling of intracranial aneurysms: A multicenter experience. J Clin Neurosci. 2019;68:86–91.

    Article  Google Scholar 

  32. Cay F, Peker A, Arat A. Stent-assisted coiling of cerebral aneurysms with the Neuroform Atlas stent. Interv Neuroradiol. 2018;24:263–9.

    Article  Google Scholar 

  33. Caragliano AA, Papa R, Pitrone A, Limbucci N, Nappini S, Ruggiero M, Visconti E, Alexandre A, Menozzi R, Lauretti D, Cavasin N, Alibrandi A, Tessitore A, Longo M, Vinci SL. The low-profile Neuroform Atlas stent in the treatment of wide-necked intracranial aneurysms - immediate and midterm results: An Italian multicenter registry. J Neuroradiol. 2020;47:421–7. https://doi.org/10.1016/j.neurad.2019.03.005.

    Article  PubMed  Google Scholar 

  34. Ulfert C, Pham M, Sonnberger M, Amaya F, Trenkler J, Bendszus M, Möhlenbruch MA. The Neuroform Atlas stent to assist coil embolization of intracranial aneurysms: a multicentre experience. J Neurointerv Surg. 2018;10:1192–6.

    Article  Google Scholar 

Download references

Funding

No funding was received for the performance of this meta-analysis.

Author information

Authors and Affiliations

Authors

Contributions

JL and SD design and conception of the work. JL, JS, and SS data collection. All authors: critical review and revision of the manuscript, interpretation of data.

Corresponding author

Correspondence to Jeremy Lynch.

Ethics declarations

Conflict of interest

J. Lynch, S. Sciacca, J. Siddiqui and L. Kanagarajah declare that they have no competing interests. S. Derakhshani has worked as a consultant for Stryker Neurovascular.

Ethical standards

No novel patient data were obtained for this study. No human or animal studies were undertaken in this study.

Additional information

Availability of Data and Material

All collected data are included in the manuscript and accompanying tables.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Lynch, J., Sciacca, S., Siddiqui, J. et al. Safety and Efficacy of the Neuroform Atlas Stent for Treatment of Intracranial Aneurysms. Clin Neuroradiol 31, 1167–1180 (2021). https://doi.org/10.1007/s00062-020-00979-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00062-020-00979-y

Keywords

Navigation