Abstract
This paper describes a metrological approach to evaluate the measurement capability of laboratories participating in a then Asia Pacific Laboratory Accreditation Cooperation [APLAC, now Asia Pacific Accreditation Cooperation] proficiency testing (PT) programme on mass fraction of acesulfame potassium and sucralose in cake mix flour. The programme was jointly organised by the Chemical Metrology Laboratory of the Health Sciences Authority, Singapore, and the Singapore Accreditation Council. Through the extensive APLAC network of accreditation bodies, 41 nominated laboratories from 32 economies registered, of which 40 laboratories submitted their results. Reference values of artificial sweeteners in the programme for performance evaluation were obtained from a high-accuracy liquid chromatography isotope dilution mass spectrometry (LC-IDMS) method. One of the unique features of this PT programme was the provision of pure substance certified reference materials (CRMs) of acesulfame potassium and sucralose to the participating laboratories as calibration standards to ensure comparability and metrological traceability of results. The participating laboratories’ performance, based on z-scores, were evaluated using the assigned reference value determined from IDMS. Five and one of the laboratories’ results would be incorrectly evaluated using consensus value calculated from the laboratories’ results for performance evaluation and using commercial standard as calibrant, respectively. This PT programme demonstrates the importance of the use of metrologically traceable reference values as the assigned values for appropriate evaluation of accuracy and reliable assessment. The provision of CRMs also heightened the testing laboratories’ awareness on the importance of using CRMs as calibrants for accurate measurements.
Similar content being viewed by others
References
Zimmet PZ, Magliano DJ, Herman WH, Shaw JE (2014) Diabetes: a 21st century challenge. Lancet Diabetes Endocrinol 2:56–64
Bruyere O, Ahmed SH, Altan C, Belegaud J, Bortolotti M, Canivenc-Lavier MC, Charriere S, Girardet JP, Houdart S, Kalonki E, Nadaud P, Rajas F, Slama G, Margaritis I (2015) Review of the nutritional benefits and risks related to intense sweeteners. Arch Public Health 73:41
Zygler A, Wasik A, Namiesnik J (2009) Analytical methodologies for determination of artificial sweeteners in foodstuffs. Trends Anal Chem 28:1082–1102
Kroger M, Meister K, Kava R (2006) Low calorie sweeteners and other sugar substitutes: a review of the safety issues. Compr Rev Food Sci Food Saf 5:35–47
Garcia-Almeida JM, Casado Fdez GM, Garcia Aleman J (2013) A current and global review of sweeteners. Regulatory aspects. Nutr Hosp 28:17–31
Wasik A, McCourt J, Buchgraber M (2007) Simultaneous determination of nine intense sweeteners in foodstuffs by high performance liquid chromatography and evaporative light scattering detection—development and single-laboratory validation. J Chromatogr A 1157:187–196
ISO/IEC 17025 (2017) General requirements for the competency of testing and calibration laboratories, International Standards Organisation, Geneva, Switzerland
Thompson M, Ellison S, Wood R (2006) The international harmonised protocol for the proficiency testing of analytical chemistry laboratories. Pure Appl Chem 78:145–196
ISO 13528 (2015) Statistical methods for use in proficiency testing by interlaboratory comparisons, International Standards Organisation, Geneva, Switzerland
Taverniers I, Loose MD, Bockstaele EV (2004) Trends in quality in the analytical laboratory. II. Analytical method validation and quality assurance. TrAC Trends Anal Chem 23:535–552
Website of the Asia Pacific Accreditation Cooperation (APAC). apac-accrediation.org. Accessed 2 Mar 2020
List of APAC PT Programmes. APAC PT Masterlist-20180919. apac-accreditation.org/publications/tec2-series/apac-pt-masterlist-20180919/. Accessed 2 Mar 2020
Kaarls R, Mackay L, Samuel A, Sin DW, Mok CS, Wong YL, Yip YC (2017) Laboratory capacity building through the use of metrologically traceable reference values in proficiency testing programmes. Accred Qual Assur 22:321–334
Yarita T, Otake T, Aoyagi Y, Kuroiwa T, Numata M, Takatsu A (2015) Proficiency testing for determination of pesticide residues in soybean: comparison of assigned values from participants’ results and isotope-dilution mass spectrometric determination. Talanta 132:269–277
Dayarathna D, Thirimanna C, Mubarak A, Mackay L, Rogerson J, Tang H, Feng J, Sin DW, Wong YC (2013) A multinational joint project on the evaluation of residual pesticide analysis in tea in the Asia Pacific region. Food Res Int 53:931–937
Heuillet M, Lalere B, Peignaux M, Graeve JD, Vaslin-Reimann S, Barros JP, Gambert P, Duvillard L, Delatour V (2013) Validation of a reference method for total cholesterol measurement in human serum and assignation of reference values to proficiency testing samples. Clin Biochem 46:359–364
Bievre P (2012) A reference value in a proficiency testing scheme should be as metrologically credible as possible. Accred Qual Assur 17:561–562
Narukawa T, Inagaki K, Naito S, Zhu Y, Miyashita S, Kuroiwa T, Hioki A, Fujimoto T, Chiba K (2016) Assessment of technical problems in the analysis of inorganic elements in squid through proficiency testing. TrAC Trends Anal Chem 76:216–226
Sin DW, Wong YC (2015) Use of assigned reference values: revisiting a small-scale inter-laboratory comparison for residual pesticides in tea. Accred Qual Assur 20:495–500
Cheng S, Mesquida C, Gui EM, Cheow PS, Lee TK, Teo TL (2014) Providing a traceable assigned value in a proficiency testing programme on the determination of benzoic acid in orange juice. Accred Qual Assur 19:439–443
Otake T, Yarita T, Aoyagi Y, Numata M, Takatsu A (2014) Evaluation of the performance of 57 Japanese participating laboratories by two types of z-scores in proficiency test for the quantification of pesticide residues in brown rice. Anal Bioanal Chem 406:7337–7344
Kardash E, Kuselman I, Pankratov I, Elhanan S (2013) Proficiency testing of pH and electrolytic conductivity measurements of groundwater: a case study of the difference between consensus and metrologically traceable values. Accred Qual Assur 18:373–381
Hon PYT, Wan MY, Sin DW, Wong YC (2013) Performance evaluation of polycyclic aromatic hydrocarbons analysis in sediment from proficiency testing with metrological reference values. Anal Bioanal Chem 405:7241–7250
Yarita T, Otake T, Aoyagi Y, Numata M, Takatsu A (2016) Difference between consensus value of participants’ results and isotope-dilution mass spectrometric results in proficiency testing for pesticide residues in husked wheat. Anal Sci 32:557–580
Zhu Y, Kuroiwa T, Narukawa T, Inagaki K, Chiba K (2012) Proficiency test in Japan for the elements in tea-leaf powder. TrAC Trends Anal Chem 34:152–160
Chan KM, Cheung STC, Wong YL, Cheng ALS, Mok CS, Wong YC, Wong WW, Tholen DW (2010) Proficiency tests for contaminants in food and herbal medicine in the Asia Pacific region. TrAC Trends Anal Chem 29:562–576
Kim SH, Heo SW, Cho H, Yim YH, Lim Y, Lee K, Hwang E (2016) Asia-Pacific Metrology Program (APMP) and Asia-Pacific Laboratory Accreditation Cooperation (APLAC) joint proficiency testing with metrological reference values for hazardous elements in cabbage. TrAC Trends Anal Chem 85:98–106
Kim B, Ahn S, Mitani Y (2011) Interlaboratory comparison for the determination of pesticide residues in Chinese cabbage. Accred Qual Assur 16:499–505
Kong MF, Chan S, Wong YC, Wong SK, Sin DW (2007) Interlaboratory comparison for the determination of five residual organochlorine pesticides in ginseng root samples by gas chromatography. J AOAC Int 90:1133–1141
Beltrami MC, Doring T, Lindner JD (2018) Sweeteners and sweet taste enhancers in the food industry. Food Sci Technol 38:181–187
ISO/IEC 17043 (2010) Conformity assessment—general requirements for proficiency testing. International Organisation for Standardisation, Geneva, Switzerland
Mackay LG, Taylor CP, Myors RB, Hearn R, King B (2003) High accuracy analysis by isotope dilution mass spectrometry using an iterative exact matching technique. Accred Qual Assur 8:191–194
De Bievre P, Peiser HS (1997) Basic equations and uncertainties in isotope-dilution mass spectrometry for traceability to SI of values obtained by this primary method. Fresenius J Anal Chem 359:523–525
Sargent M, Harte R, Harrington C (2002) Guidelines for achieving high accuracy in isotope dilution mass spectrometry (IDMS). The Royal Society of Chemistry, Cambridge
Gui EM, Lu T, Teo TL, Cheow PS, Lee TK (2017) Optimisation of extraction methods and quantification of benzo[a]pyrene and benz[a]anthracene in yerba mate tea by isotope dilution mass spectrometry. Anal Bioanal Chem 26:6069–6080
Teo TL, Quan C, Cheow PS, Huang T, Lee TK, Li H, Zhang W (2012) A bilateral study for the establishment of comparability in the measurement of ethanol in bioethanol fuel using isotope dilution mass spectrometry. Accred Qual Assur 17:315–321
Milton MJT, Quinn TJ (2001) Primary methods for the measurement of amount of substance. Metrologia 38:289–296
ISO Guide 35 (2017) Reference materials—guidance for characterisation and assessment of homogeneity and stability. International Organisation for Standardisation, Geneva, Switzerland
ISO/IEC Guide 98-3:2008 Uncertainty of measurement—part 3: guide to the expression of uncertainty in measurement (GUM:1995)
Mann I, Brookman B (2011) Eurachem/CITAC guide: selection, use and interpretation of proficiency testing (PT) schemes, 2nd edn. www.eurachem.org
Ellison SLR, William A (eds) (2019) Eurachem/CITAC guide: metrological traceability in analytical measurement, 2nd edn. ISBN: 978-0-948926-34-1. www.eurachem.org
Linsinger TPJ (2008) Use of recovery and bias information in analytical chemistry and estimation of its uncertainty contribution. Trends Anal Chem 27:916–923
Thompson M, Wood R (1995) Harmonised guidelines for internal quality control in analytical chemistry laboratories. Pure Appl Chem 67:649–666
Da Silva R, Camões M (2010) Comparability of measurement results for pesticide residues in foodstuffs: an open issue. Accred Qual Assur 15:691–704
Acknowledgements
We would like to express our sincere thanks to colleagues (past and present) in HSA and SAC, all the participating accreditation bodies and laboratories for their contributions to the APLAC PT programme, as well as APLAC for the partial financial support to this PT programme.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors have declared no conflict of interest.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Cheow, P.S., Teo, T.L., Lu, T. et al. An accuracy-based proficiency testing programme with metrological reference values for the mass fraction of acesulfame potassium and sucralose in cake mix flour. Accred Qual Assur 25, 355–366 (2020). https://doi.org/10.1007/s00769-020-01449-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00769-020-01449-1