Silver-coated three-core fiber Michelson interferometer for liquid-level measurement

  • Chen Zheng , Wenlin Feng EMAIL logo , Xiaozhan Yang , Bangxing Li and Zhi Chen EMAIL logo

Abstract

The Michelson liquid-level sensor based on silver coated the end face of the three-core fiber reflection structure has been proposed to measure continuous or discrete liquid level. The Michelson interference structure can be obtained by the combination of the single-mode optical fiber and the three-core optical fiber with the silver film coated on the other end face of it. The inter-mode interference can be obtained by the fiber-core mismatch at the fusion joint. The liquid level can be measured by monitoring the dip wavelength shift of the interference spectrum. The results indicate that the sensitivity of the liquid-level sensor will decrease with the increasing sensing length. The sensing length of 30 mm is selected to investigate the performance of the sensor. The sensitivity of water level is reached as high as 392.83 pm/mm with an excellent linearity of 0.99946. Interestingly, the sensitivity of the sensor increases with the liquid refractive index and the sensitivity of the refractive index for NaCl solution is 4410.74 pm/mm/RIU. The performance of the sensor is very stable in the range of 20–90 °C. The maximum drift for temperature is 0.3001 nm. The sensor can be applied to the measurement of the liquid level in different environments.

1 Introduction

Liquid-level sensor has been widely used in many fields, such as petrochemical, biopharmaceutical and tidal lake water-level monitoring. The sensor often work in the complex and changeable ambient conditions, for example, the strong electromagnetic interference, inflammable, explosive, corrosive and low temperature, etc. [1], [2], [3], [4], [5]. The traditional sensors based on capacitance have to work under the electric power, which easily leads to security risks by the leakage of electric power [2], [3], [4], [5]. Moreover, the traditional sensors based on electrical sensing technology cannot work accurately under the conditions of dynamics, large temperature variation and harsh environment [6]. Compared with other electrochemical liquid-level sensors, optic-fiber liquid-level sensors have the advantages of safety, low cost and variety application environments [7], [8], [9], [10]. Therefore, the research on optical fiber liquid-level sensors has great practical application value in daily life, military aerospace and other fields [11], [12], [13], [14].

The general mechanism of optic-fiber Michelson sensor is the interference between two optical signals with the same wavelength and a relative phase difference. This phase difference is caused by the transmission of light in different media or along different path lengths [15]. This structure has the advantages of compact and simple manufacturing process. In recent years, there were many reports of optical fiber liquid-level sensors based on the utilization of Michelson interference structure. Rong et al. proposed a fiber-optic quasi-Michelson liquid-level sensor composed of a small-core fiber and single-mode fiber (SMF) coated with a silver reflective film. This sensor had a simple structure with low sensitivity of −68.3 pm/mm [16]. Li, et al. proposed a Michelson liquid-level sensor based on an optical fiber with core-offset structure [17]. Although the sensitivity of liquid-level sensor was increased to 77 pm/mm, the complexity of fusion process of the core offset structure hindered the repeatability of the experiment. They also did not study the temperature stability of the sensor. Liang, et al. used a self-made dual-mode elliptical multilayer core fiber (EMCF) and a SMF coated with a silver film to obtain a fiber-optic Michelson sensor [18], but the sensitivity to liquid level is not very high, and the highest sensitivity is only 48.93 pm/mm.

Here, a simple Michelson liquid-level sensor based on a three-core fiber reflection structure is presented. The sensor is composed of a SMF and three-core fiber with a silver reflective film coated on the end face. The sensitivity and stability of the sensor have been investigated in detail and the results are discussed.

2 Preparation and principle of sensor

First, the three-core fiber (TCF, Wuhan Senhui Optoelectronic Technology Co., Ltd., China) was cut into the required length, and the two end faces were cut flat. Second, a layer of silver film was coated on the outer surface of TCF by silver mirror reaction. Finally, the other end face of TCF was fused with the SMF with the fusion procedure of the optical fiber fusion splicer. So the required sensing element was obtained. The sensing structure of the liquid-level sensor is shown in Figure 1(a). The cross-section of the TCF is shown in Figure 1(b). The three cores form a shape of equilateral triangle. The diameters of the cladding and cores of TCF are 125 and 8.6 μm, respectively. The distance between the two cores is about 41.5 μm. The mode of light in SMF will be converted into core mode and cladding mode at the fusion point between the SMF and TCF because of the mode mismatch. The two modes will be reflected by silver film on the end surface of TCF, then propagated through the TCF and eventually coupled into SMF.

Figure 1: (a) Sensing structure and (b) the end face of TCF.
Figure 1:

(a) Sensing structure and (b) the end face of TCF.

The propagation process of optical power from SMF to TCF has been simulated by beam propagation method (BPM). In this model, the lengths of SMF and TCF are 1 and 30 mm, respectively. Figure 2(a) shows the optical power distribution of the cross section near the interface between SMF and TCF. The optical power is first transmitted from SMF to the cladding of TCF to excite the cladding mode. The light distribution of the cross section at the position of 12.5 μm near the fusion point in TCF is shown in Figure 2(b), which illustrates that the light beam gradually enter the three cores at this position to stimulate the fiber-core mode. The light distribution at the end surface of TCF is shown in Figure 2(c), which indicates stronger optical power in cladding, compared with the three cores. As shown in Figure 2(d), there are more light distribution in the cladding than that of the core.

Figure 2: (a) Z-direction cross-sectional light distribution near the interface of SMF and TCF, (b) Z-direction cross-sectional light distribution at 0.125 mm of TCF, (c) Z-direction cross-sectional light distribution at the end of the sensor and (d) XZ cross-sectional light distribution.
Figure 2:

(a) Z-direction cross-sectional light distribution near the interface of SMF and TCF, (b) Z-direction cross-sectional light distribution at 0.125 mm of TCF, (c) Z-direction cross-sectional light distribution at the end of the sensor and (d) XZ cross-sectional light distribution.

The interference pattern is formed by the interference between core mode and cladding mode, and its intensity can be expressed as [16],

(1)I=Icore+Icladm+2mIcoreIcladmcosϕm

where Icore and Icladm are the total light intensities of core mode in the three cores and the light intensity of the mth-cladding mode, respectively. Φm is the phase difference between the core and the mth-cladding mode, which is proportional to the length of the TCF and the effective refractive index difference between the core and the mth-cladding mode, and it can be expressed as [19],

(2)ϕm=4π(neffcoreneffclad,m)Lλ=4πΔneffmLλ

where neffcore and neffclad,m are, respectively, the effective refractive indices of the core mode and the mth-cladding mode. Δneffm is the difference of the effective refractive index between the core mode and the mth-cladding mode. L is the length of the TCF, and λ is the monitoring dip wavelength. According to the interference theory, the attenuation peak of the interference can be expressed as [20],

(3)λm=4ΔneffmL2n+1(n=0,±1,±2,)

During the liquid-level detection, one part sensing length of sensor will be exposed in the air and the other sensing length will be immersed in liquid. The effective refractive index of the cladding will change with the increase of the liquid level, which lead to the change of the interference wavelength. The shift of the interference wavelength can be expressed as [21],

(4)λm=4πΔneffm(LLh)(2n+1)π+4πΔneffmLh(2n+1)π

where Lh and Δneffnm are the length of the three-core fiber immersed in the liquid and the difference of the effective refractive index between the core mode and the mth-cladding mode, respectively.

3 Results and discussion

The schematic diagram of the liquid-level measuring system is shown in Figure 3. The silver mirror reaction was adopted to coat silver film on the end face of TCF. The other end face of TCF was spliced with SMF by the optical-fiber fusion splicer (S178C, Furukawa Electric, Japan). The sensing part was fixed in parallel on the ruler, and placed vertically in the beaker with designed liquid. In order to facilitate the observation of the liquid-level change, the end of the optical fiber is parallel to the zero scale of the ruler, which is regarded as the zero scale of the liquid level. The light beam (wavelength range from 1520 to 1620 nm) first passes through the circulator and transmitted from the SMF to TCF, and reflected by the silver film, then the final reflected light coupled into the SMF at the spliced region and finally output by the circulator to the optical spectrum analyzer (OSA).

Figure 3: Schematic diagram of the liquid-level measuring system.
Figure 3:

Schematic diagram of the liquid-level measuring system.

The effect of the sensing length of TCF on the sensitivity of sensor was investigated. The water level has been tested by the sensor with the sensing length of 23, 30, 40, 50, 60 and 75 mm at room temperature. The sensitivity of sensor varies with the sensing length. In order to compare the difference in sensitivity of the liquid-level sensor with different sensing lengths intuitively, the change in the liquid-level depth detected by the sensor is converted into a percentage of liquid-level change to sensing length. As shown in Figure 4, the change in dip wavelength shift shows a similar trend and maintains good linearity at different sensing lengths with the wavelength shift between 10 and 12 nm according to the fitted data. It can be seen that the length of TCF will affect the sensitivity of the liquid-level sensor, which decrease with the increasing of the sensing length of TCF. Therefore, the appropriate sensing length will be adopted according to the testing environment.

Figure 4: The relationship between the percentage change of the liquid-level depth measured by different TCF lengths vs. the length of the sensing region and the wavelength shift.
Figure 4:

The relationship between the percentage change of the liquid-level depth measured by different TCF lengths vs. the length of the sensing region and the wavelength shift.

As shown in Figure 5(a), the initial spectrum of the sensor with sensing length of 30 mm for TCF in the air is chosen. It can be observed that a wide free spectral region and an obvious monitoring dip appeared under this case. Figure 5(b) shown the wavelength shift of light beam with different water depths at the condition of 30 mm sensing length and 3 mm interval of the water depth. The obvious redshift of the dip wavelength was obtained with the increasing of the water depth. The high sensitivity of 392.83 pm/mm and excellent linearity of 0.99946 are obtained, which are much higher than the sensitivity of −43.7 and 77 pm/mm proposed by Yun et al. and Diaz et al. [12], [13]. Therefore, the sensing length of TCF will be set at 30 mm in the following investigation.

Figure 5: (a) The spectrum of the sensor in the air when the TCF is 30 mm, and (b) the dip wavelength varies with the liquid level.
Figure 5:

(a) The spectrum of the sensor in the air when the TCF is 30 mm, and (b) the dip wavelength varies with the liquid level.

The influence of the sensitivity of presented sensor with refractive index of liquid was also tested. Four different NaCl solutions with concentrations of 0, 5, 10 and 15% were used for liquid-level testing, which corresponding to the refractive index of 1.3333, 1.3424, 1.3510 and 1.3609. As shown in Figure 6(a), the sensitivity of the sensor is increased with the increasing of liquid refractive index though our experiment data, and the different wavelength drift was obtained under different refractive indices too. As shown in Figure 6(b), the response of the liquid-level sensitivity of the sensor to the refractive index is 4410.74 pm/mm/RIU, and the linearity is 0.98398. It shows that the refractive index of the liquid can be tested according to the sensitivity of the sensor at same liquid level, which expands the application range of the sensor.

Figure 6: (a) The sensitivity of the sensor in NaCl solutions with refractive indices of 1.3333, 1.3424, 1.351 and 1.3609, respectively, and (b) the sensitivity of the refractive index.
Figure 6:

(a) The sensitivity of the sensor in NaCl solutions with refractive indices of 1.3333, 1.3424, 1.351 and 1.3609, respectively, and (b) the sensitivity of the refractive index.

In order to explore the effect of temperature on the performance of the sensor, the temperature stability of the sensor was studied. An appropriate amount of deionized water was added to the beaker and turned off the heating device until the water was heated to boiling. Putting the sensor vertically and submerging it, recording the spectral data every 10 °C from 90 to 20 °C. The measured results are shown in Figure 7. The monitoring wavelength appears no drift with temperature, which show maximum shift of 0.3001 nm. It indicates that the sensor can be used at different temperatures without other additional influence.

Figure 7: Temperature sensitivity of the sensor.
Figure 7:

Temperature sensitivity of the sensor.

As shown in Table 1, compared with other fiber-optic Michelson liquid-level sensors with different structures, our sensor has higher sensitivity (392.83 pm/ppm). In addition, the simple sensing structure reduces the influence of temperature.

Table 1:

Performance comparisons of different fiber-optic Michelson liquid-level sensors.

StructureLiquid/water-level sensitivity (pm/mm)Measuring range (mm)References
In-fiber quasi-Michelson interferometer−68.330[16]
Michelson interferometer with the core-offset structure7740[17]
A probe based on MI viadual-mode elliptical multilayer-core fiber−33.4845[18]
Three-core fiber reflection structure392.8330this work

4 Conclusions

In summary, a simple optical fiber liquid-level sensor based on Michelson interference structure was proposed. The sensor has excellent sensitivity and linearity. The change of refractive index around the TCF due to different liquid level will affect the cladding mode of TCF, which in turn lead to redshift of the dip wavelength. The sensing range and sensitivity of the sensor can be controlled by the sensing length of TCF. The response of the liquid-level sensitivity of the sensor to the refractive index of NaCl solution is 4410.74 pm/mm/RIU, which show the sensitivity of sensor increase with the increasing of liquid refractive index. The temperature sensitivity of the sensor is very tiny and can be ignored, which make it can be applied to liquid-level measurement at complex environments.


Corresponding authors: Wenlin Feng, College of Science, Chongqing University of Technology, Chongqing 400054, China; and Chongqing Key Laboratory of Green Energy Materials Technology and Systems, Chongqing 400054, China, E-mail: ; and Zhi Chen, School of Chemistry and Chemical Engineering, Chongqing University of Technology, Chongqing 400054, China, E-mail:

Funding source: Venture & Innovation Support Program for Chongqing Overseas Returnees

Award Identifier / Grant number: cx2019092

Award Identifier / Grant number: 51574054

Funding source: Science and Technology Bureau of Banan District

Award Identifier / Grant number: 2018TJ12

Award Identifier / Grant number: 2019TJ08

Award Identifier / Grant number: KJZD-M201901102

Funding source: Chongqing Science and Technology Bureau

Award Identifier / Grant number: cstc2018jcyjAX0294

Award Identifier / Grant number: CSTCCXLJRC-201905

  1. Author contribution: All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission.

  2. Research funding: The research was supported by the National Natural Science Foundation of China (51574054), Chongqing Municipal Education Commission (KJZD-M201901102), Chongqing Science and Technology Bureau (cstc2018jcyjAX0294, CSTCCXLJRC- 201905), the Venture & Innovation Support Program for Chongqing Overseas Returnees (cx2019092), Science and Technology Bureau of Banan District (2018TJ12, 2019TJ08).

  3. Conflict of interest statement: The authors declare no conflicts of interest.

References

[1] S. Liu, X. Yang, W. Feng, H. Chen, Y. Tao, and Y. Jiang, “Michelson interferometric hydrogen sulfide gas sensor based on NH2-rGO sensitive film,” Z. Naturforch. A, vol. 75, pp. 241–248, 2020, https://doi.org/10.1515/zna-2019-0281.Search in Google Scholar

[2] F. Reverter, X. Li, G. C, and M. Meijer, “Liquid-level measurement system based on a remote grounded capacitive sensor,” Sens. Actuators A: Phys, vol. 138, pp. 1–8, 2007, https://doi.org/10.1016/j.sna.2007.04.027.Search in Google Scholar

[3] B. Kumar, G. Rajita, and N. Mandal, “A review on capacitive-type sensor for measurement of height of liquid level,” Meas. Control, vol. 47, pp. 219–224, 2014, https://doi.org/10.1177/0020294014546943.Search in Google Scholar

[4] K. Chetpattananondha, T. Tapoanoi, P. Phukpattaranont, and N. Jindapetcha, “A self-calibration water level measurement using an interdigital capacitive sensor,” Sens. Actuators A: Phys, vol. 209, pp. 175–182, 2014, https://doi.org/10.1016/j.sna.2014.01.040.Search in Google Scholar

[5] H. Canbolat, “A novel level measurement technique using three capacitive sensors for liquids,” IEEE trans. Instrument. Meas, vol. 58, pp. 3762–3768, 2009, https://doi.org/10.1109/tim.2009.2019715.Search in Google Scholar

[6] E. Terzic, R. Nagarajah, and M. Alamgir, “A neural network approach to fluid quantity measurement in dynamic environments,” Mechatronics, vol. 21, pp. 145–155, 2011, https://doi.org/10.1016/j.mechatronics.2010.09.011.Search in Google Scholar

[7] J. Peng, W. Feng, X. Yang, G. Huang, and S. Liu, “Dual Fabry-Pérot interferometric carbon monoxide sensor based on the PANI/Co3O4 sensitive membrane-coated fibre tip,” Z. Naturforsch. A, vol. 74, pp. 101–107, 2019, https://doi.org/10.1515/zna-2018-0453.Search in Google Scholar

[8] Z. Peng, W. Feng, X. Yang, L. Fang, D. Wei, and X. Liu, “Graphene-based waist-enlarged optical fiber sensor for measurement of sucrose concentration,” Z. Naturforsch. A, vol. 74, pp. 751–756, 2019, https://doi.org/10.1515/zna-2018-0564.Search in Google Scholar

[9] J. E. A. Lopez, J. J. S. Mondragon, P. L. K. Wa, and D. A. M. Arrioja, “Fiber-optic sensor for liquid level measurement,” Opt. Lett., vol. 36, pp. 3425–3427, 2011.10.1364/OL.36.003425Search in Google Scholar PubMed

[10] J. Yu, X. Yang, and W. Feng, “Hydrogen sulfide gas sensor based on copper/graphene oxide composite film-coated tapered single-mode fiber interferometer,” Z. Naturforch. A, vol. 74, pp. 931–936, 2019, https://doi.org/10.1515/zna-2019-0169.Search in Google Scholar

[11] K. R. Sohn, and J. H. Shim, “Liquid-level monitoring sensor systems using fiber Bragg grating embedded in cantilever,” Sens. Actuators A: Phys, vol. 152, pp. 248–251, 2009, https://doi.org/10.1016/j.sna.2009.04.003.Search in Google Scholar

[12] B. F. Yun, N. Chen, and Y. P. Cui, “Highly sensitive liquid-level sensor based on etched fiber Bragg grating,” IEEE Photon. Technol. Lett., vol. 19, pp. 1747–1749, 2007, https://doi.org/10.1109/lpt.2007.905093.Search in Google Scholar

[13] C. A. R. Diaz, A. G. L. Junior, P. S. B. Andre, et al., “Liquid level measurement based on FBG-embedded diaphragms with temperature compensation,” IEEE Sensor. J., vol. 18, pp. 193–200, 2017.10.1109/JSEN.2017.2768510Search in Google Scholar

[14] G. Onorato, G. Persichetti, I. A. Grimaldi, G. Testa, and R. Bernini, “Optical fiber fuel level sensor for aeronautical applications,” Sens. Actuators A: Phys, vol. 260, pp. 1–9, 2017, https://doi.org/10.1016/j.sna.2017.04.015.Search in Google Scholar

[15] P. L. Swart, “Long-period grating Michelson refractometric sensor,” Meas. Sci. Technol., vol. 15, p. 1576, 2004, https://doi.org/10.1088/0957-0233/15/8/025.Search in Google Scholar

[16] Q. Z. Rong, X. G. Qiao, Y. Y. Du, et al.., “In-fiber quasi-Michelson interferometer for liquid level measurement with a core-cladding-modes fiber end-face mirror,” Optic Laser. Eng., vol. 57, pp. 53–57, 2014, https://doi.org/10.1016/j.optlaseng.2013.12.010.Search in Google Scholar

[17] P. F. Li, H. T. Yan, and H. J. Zhang, “Highly sensitive liquid level sensor based on an optical fiber Michelson interferometer with core-offset structure,” Optik, vol. 171, pp. 781–785, 2018, https://doi.org/10.1016/j.ijleo.2018.06.126.Search in Google Scholar

[18] X. Liang, G. B. Ren, Z. B. Liu, H. Wei, and S. S. Jian, “In-fiber liquid-level probe based on Michelson interferometer via dual-mode elliptical multilayer-core fiber,” J. Mod. Optic., vol. 63, no. 13, pp. 1254–1259, 2016.10.1080/09500340.2015.1137367Search in Google Scholar

[19] Z. Y. Li, Y. P. Wang, C. R. Liao, et al.., “Temperature-insensitive refractive index sensor based on in-fiber Michelson interferometer,” Sensor. Actuator. B Chem., vol. 199, pp. 31–35, 2014, https://doi.org/10.1016/j.snb.2014.03.071.Search in Google Scholar

[20] C. Li, T. G. Ning, C. Zhang, et al.., “All-fiber multipath Mach-Zehnder interferometer based on a four-core fiber for sensing applications,” Sens. Actuators A: Phys, vol. 248, pp. 148–154, 2016, https://doi.org/10.1016/j.sna.2016.07.031.Search in Google Scholar

[21] L. C. Li, L. Xia, Z. H. Xie, and D. M. Liu, “All-fiber Mach-Zehnder interferometers for sensing applications,” Optic Express, vol. 20, pp. 11109–11120, 2012, https://doi.org/10.1364/oe.20.011109.Search in Google Scholar

Received: 2020-07-25
Accepted: 2020-10-02
Published Online: 2020-10-27
Published in Print: 2020-11-18

© 2020 Walter de Gruyter GmbH, Berlin/Boston

Downloaded on 23.4.2024 from https://www.degruyter.com/document/doi/10.1515/zna-2020-0201/html
Scroll to top button