The Traceability Improvement and Comparison of Bell Prover as the Indonesian National Standard of Gas Volume Flow Rate

  • Original Paper
  • Published:
MAPAN Aims and scope Submit manuscript

Abstract

Traceability is urgent to guarantee the truth of measurement, particularly for National Standard. Although a bell prover facility has established as the national low pressure gas flow standard of Indonesia since 2008, the measurement capability (15 to 1400 L/min, U95 = 0.20%) of the bell prover has not been traceable to SI Units. Moreover, displacement values provided by the bell prover’s manufacturer by rotary encoder method are still used. The limit detector by optical sensor to measure displacement is worthy to achieve dimensional traceability which can be switched by ultrasonic sensor for reasonable solution. The aim of this study is to improve the traceability of the bell prover by reasonable solution and then validate the gas volume flow rate measurement by inter-laboratory comparison. The Center for Measurement Standards (CMS), Industrial Technology Research Institute (ITRI), was invited to participate in the bilateral comparison, and a gas turbine flow meter was used as the transfer standard. The bell prover has been successfully traceable to SI units, particularly displacement measurement through flying start/stop method by limit of ultrasonic sensor (250 to 1200 L/min, U95 = 0.80% passing assessment report). Despite having larger uncertainty than manufacturer specification, the results are better for more in of layer comparison. Because the effect of turbine drift should be a concern, the gap of measurement time for inter-laboratory comparison is recommended to minimize as many as possible. Furthermore, the measurement range of bell prover should be reduced as reasonability purpose.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6

Similar content being viewed by others

References

  1. Flow Technology Incorporated, Autobell® Bell prover primary standard gas calibration system, Arizona: FTI, (2003).

    Google Scholar 

  2. Flow Technology Incorporated. Report of calibration for Autobell model of FTBP-10-1-2-A012 with serial no. CS07030027. (2007).

  3. J.D. Wright and G.E. Mattingly, NIST calibration services for gas flow meters; Piston prover and bell prover gas flow facilities”, National Institute of Standards and Technology (NIST) Special Publication, 250–49 (1998).

  4. H. Padden, Constant pressure provers from the standards lab to the shop floor, Bios International Corporation, (2003).

  5. National Measurement Laboratory. Report of calibration for turbine flow meter with serial no. 1613794. (2014).

  6. BIPM, The BIPM key comparison database, calibration and measurement capabilities, mass and related quantities, Chinese Taipei, CMS (ITRI Center for Measurement Standards), (February 2018).

  7. J.D. Wright, What is the best transfer standard for gas flow? Proceedings of FLOMEKO. Groningen Netherlands, 12–14th May2003: National Institute of Standards and Technology, (2003).

  8. C.-M. Su, W.-T. Lin and S. Masri, Establishment and verification of a primary low-pressure gas flow standard at NIMT, MAPAN J. Metrol. Soc. India, 26 3 (2011) 173–186

    Google Scholar 

  9. Flow Technology Incorporated, TM-60211 REV. H - Bell prover precision gas prover installation, operation and maintenance manual, Arizona: FTI, (2009).

  10. J. A. Prakosa and L.P. Kozlova, Design and simulation of automatic control valve for gas flow meter calibrator of bell prover, 2018 IEEE conference of Russian young researchers in electrical and electronic engineering (EIConRus), 966–969 (2018).

  11. ISO 9300:2005(en): Measurement of gas flow by means of critical flow Venturi nozzles, (2005).

  12. A. Carullo , P. Marco.An ultrasonic sensor for distance measurement in automotive applications. IEEE Sens. J. 1 2 (2001) 143.

    Article  ADS  Google Scholar 

  13. J.A. Prakosa, A.S. Kukaev, V.A. Parfenov, et al. Simple automatic fluid displacement measurement by time-of-flight laser sensing technology for volume calibrator need. J. Opt., 49 (2020) 69–75. https://doi.org/10.1007/s12596-020-00596-5

  14. R.C. Baker,“Turbine flow meter: II, theoretical and experimental published information”. Flow Measurement and Instrumentation, UK : Butterworth-Heinemann Ltd, 123–44(1993).

  15. Komite Akreditasi Nasional - National Accreditation Body of Indonesia. “Assessment report of gas flow rate on 9–11 August 2016 for Research Center for Metrology LIPI”, Jakarta (2016).

  16. JCGM 100:2008 Evaluation of measurement data guide to the expression of uncertainty in measurement. Paris: BIPM, IEC, IFCC, ILAC, ISO, IUPAC, IUPAP and OIML, (2008).

  17. H.M. Choi, K.-A. Park, Y.K. Oh and Y.M.Choi, Uncertainty evaluation procedure and intercomparison of bellprovers as a calibration system for gas flow meters, KRISS Republic of Korea, (2010).

  18. X. Chunling and S. Yanzuo, Correction For K factor of gas turbine flow meter, FLOMEKO 2003 11th IMEKO TC9 Conference on Flow Measurement. Groningen, The Netherland, (2013)

  19. K. Mattiasson, Key comparisons and the ultimate transfer standard, SP Sveriges Tekniska Forsknings institut AB.

Download references

Acknowledgements

Thank you to management of CMS ITRI, LIPI, and BSN for supporting this research also Ministry of Research and Higher Education of Republic Indonesia for funding through RisetPro Non-Gelar grant.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to J. A. Prakosa.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Prakosa, J.A., Su, C.M., Wang, W.B. et al. The Traceability Improvement and Comparison of Bell Prover as the Indonesian National Standard of Gas Volume Flow Rate. MAPAN 36, 81–87 (2021). https://doi.org/10.1007/s12647-020-00402-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s12647-020-00402-4

Keywords

Navigation