Advances on inorganic scintillator-based optic fiber dosimeters

Abstract

This article presents a new perspective on the development of inorganic scintillator-based fiber dosimeters (IOSFDs) for medical radiotherapy dosimetry (RTD) focusing on real-time in vivo dosimetry. The scintillator-based optical fiber dosimeters (SFD) are compact, free of electromagnetic interference, radiation-resistant, and robust. They have shown great potential for real-time in vivo RTD. Compared with organic scintillators (OSs), inorganic scintillators (IOSs) have larger X-ray absorption and higher light output. Variable IOSs with maximum emission peaks in the red part of the spectrum offer convenient stem effect removal. This article outlines the main advantages and disadvantages of utilizing IOSs for SFD fabrication. IOSFDs with different configurations are presented, and their use for dosimetry in X-ray RT, brachytherapy (BT), proton therapy (PT), and boron neutron capture therapy (BNCT) is reviewed. Challenges including the percentage depth dose (PDD) deviation from the standard ion chamber (IC) measurement, the angular dependence, and the Cherenkov effect are discussed in detail; methods to overcome these problems are also presented.

Introduction

As a fundamental process for RT treatment, RTD ensures that the radiation dose is safely and correctly delivered to the malignant abnormalities. However, current RTD practices have fallen behind the fast rapid scientific and clinical development of RT. Novel techniques applied to RT treatment include intensity-modulated RT (IMR), stereotactic ablative RT (SABR), and magnetic resonance imaging guided linear accelerator (MRI-linac). While these techniques provide better treatment opportunities for cancer patients, they also increase the complexity of the measurement of the dose rate and accumulated dose during treatment and are impacted by the limitations of existing dosimeters. Current safety and quality protocols are well established for traditional RT treatment techniques [1,2,3,4], but these protocols face challenges from aforementioned fast developing RT technologies. There are reports of patients receiving incorrect doses while undergoing radiation treatment utilizing these new RT techniques [5, 6]. This has aroused concerns over the lag of dosimetric equipment development behind the treatment techniques [7,8,9]. Therefore, developing radiation dosimeters capable of real-time in vivo dose monitoring has received significant research attention.

MOSFETs and diode-based devices have been utilized for clinical real-time in vivo dosimetry [10,11,12] owning to their good sensitivity, small size, and real-time readout capabilities. However, MOSFETs are expensive, incident angle-dependent, and of limited durability. Diode dosimeters are sensitive to accumulated dose, temperature, and incident angle. SFD provides an alternative solution for real-time in vivo dosimetry. It is a relatively new type of detector that involves coupling of a scintillator, an optical fiber, and a light-detector. The combination of the scintillator and the optical fiber endows SFD attractive merits such as passive detection, small size, linear response to dose rate, energy independence, immunity to the electromagnetic interference, good mechanical robustness, and capacity for multiplexing. Though the clinical applications of SFD are limited, they have shown great potential for real-time in vivo dosimetry.

To the best of our knowledge, the ideal of a “SFD” first emerged in 1969 when Byfield et al. [13] reported a plastic scintillator-based optical fiber dosimeter for intracavitary dosimetry. The probe comprising a piece of plastic phosphor encapsulated in a stainless-steel tube. The optical signal was collected and transmitted via a Lucite “light pipe”. The detector achieved good accuracy measuring the radiation dose from cobalt60 and megavoltage X-rays in a phantom. Although this probe dosimeter might appear primitive and too large for actual real-time in vivo dosimetry, SFDs later reported/shared the similar design which combines a small piece of scintillator, an optical fiber, and a photodetector.

The fast development of SFD emerged in the 1990s due to seminal and systematic research on plastic scintillator optical fiber dosimeters (PSFDs) [14, 15]. Since then, SFDs have attracted more and more research attention, and numerous articles investigated aspects of SFDs such as scintillator materials [16, 17], configuration optimization [18, 19], and signal processing techniques (e.g., Cherenkov effect removal) [20, 21]. Several featured review papers on SFD development exist in the scientific literature as well [8, 9, 22,23,24,25,26].

The scintillation materials applied in SFDs can be classified into two broad groups. One group is the plastic OS comprising aromatic hydrocarbon molecules. They have similar mass and electron densities as those of water and human tissue. Thus, they have photon/electron interaction properties similar to water and human tissue (i.e., “water-equivalent” property). Another group includes the IOSs, most of which are alkali halide (e.g., NaI:Tl, CsI:Tl, and CsI:Na) or oxide (e.g., Y2O3:Eu and Lu2O3:Eu) crystals doped with lanthanide ions. Though IOSs have been widely utilized for X-ray detection and imaging, their application in SFD is still quite limited partially due to their non-linear response to X-ray energies under 100 keV and non-water equivalency.

However, research on the IOS-based SFDs (IOSFDs) shows that this particular type of SFD has promise in RT dosimetry applications due to their high sensitivity to low radiation dose rate and overwhelming signal intensity compared with the Cherenkov effect. This paper reviews the state-of-the-art development of IOSFDs for radiation dosimetry in oncology, focusing on X-ray RT. We first evaluate the fundamentals of IOSFD in the “Fundamentals” section. The scintillation mechanism of scintillators under ionizing photon radiation will be briefly introduced—this section evaluates the properties of the scintillators that we considered important for RT dosimetry (RTD). Second, we look at the development of IOSFDs using different scintillators and designs in the “Development of IOSFDs” section including their applications for dosimetry in RT such as BT, PT, and BNCT. Third, we discuss some challenges facing IOSFD for RT dosimetry application in the “Challenges of IOSFD for RTD application” section including the measured PDD discrepancy in comparison with the standard IC measurement and the angular dependence. The Cherenkov removal techniques are summarized in the “Cherenkov effect removal techniques” section. Finally, we conclude the article and provide a future outlook of ISFD for RTD.

Fundamentals

The regular SFD configuration involves the coupling of a piece of scintillation material to an optical fiber. Figure 1 gives a simple schematic of how SFD works under the ionizing radiation. The scintillator undergoes a radio-luminescence process when irradiated by a high-energy ionizing beam (Fig. 1). The light photons generated in the scintillator are collected by an optical fiber and remotely transmitted to the photon detector (e.g., photomultiplier, CCD camera, and spectrometer). The output can be analyzed via a remote computer terminal. Theoretically, the scintillation light emission is proportional to the dose rate.

Fig. 1
figure 1

A simple schematic of the radiation detection process using SFD

Overall, the performance of the SFD for oncology radiation measurement is determined by the scintillation light collection efficiency and the response of the scintillator to the incident radiation particles. The scintillation light collection efficiency resembles the light extraction efficiency for LED. It describes how efficiently the scintillation light is extracted and transported by the optical fiber coupled to the scintillator. The scintillation light photons are emitted towards arbitrary directions. To collect these photons as many as possible, it is key to achieve an efficient coupling between the scintillation domain and the optical fiber with a limited numerical aperture. Such coupling required a delicate design of the configuration of the IOSFD, and this issue will be reviewed and discussed later in the “Development of IOSFDs” section.

The response of the scintillator to the radiation is determined by several factors. Taking RTD with an X-/γ-ray radiation source for examples, these factors include the size or the concentration of the scintillation material, the X-ray linear attenuation coefficient, the scintillation rise-decay time, and the overall radiation-to-light conversion efficiency. Focusing on dosimetry for ionizing photon radiation, we next briefly introduce the scintillation mechanism and then look at several physical properties of scintillators that are considered important for IOSFD.

Scintillation mechanism

IOSs normally have broad band gaps where electrons can jump up to higher energy levels via excitation once absorbing ionizing radiation energy. The excited electrons can also drop down to lower energy levels via de-excitation through the emission of (visible) photons. The scintillation process under ionizing X-ray radiation can be divided into three stages: (a) primary photon interaction, ionization of atoms, relaxation, and thermalization of secondary particles; (b) transport of electrons and holes and further relaxation; and (c) luminescence [27, 28]. The attenuation of X-ray photons takes place in stage (a) in the form of four possible primary photon interactions, i.e., photoelectric absorption, Compton scattering, electron-positron pair production, and coherent scattering. Some of the energy absorbed by the scintillator will generate free electron holes; the other is transferred to a lattice vibration (i.e., heat) or secondary X-rays or electrons radiation. In stage (c), the electrons and holes are trapped by the luminescence centers (defects or dopant activators, e.g., Mn2+, Sn2+, Eu3+, and Ti4+), which contributes to radiative recombination in the form of ultraviolet/visible (UV/vis) light emission. For in vivo or under-surface phantom dosimetry, the scintillation material is excited not only by direct X-ray radiation, but also by secondary electrons generated via Compton scattering or photoelectric interaction in the tissue-equivalent material.

Basic physical properties of scintillators

Generally, the IOS for photon radiation detection should have a high light yield, high X-ray linear attenuation coefficient, short scintillation-decay time, low afterglow, good radiation resistance, and linear-response with incident X-ray or electron beams. Due to the long history and consistent research efforts concerning developments in this area, there are numerous types of scintillators available. Many articles and books have systematically reviewed and illustrated the work conducted on scintillator development and the application of scintillators for radiation detection [26, 28,29,30,31,32,33,34]. The synthesis and characterization of scintillators are beyond the scope of this review, and this work briefly introduces the physical parameters describing the scintillator with respect to inorganic scintillation materials used in existing reported IOSFDs [35,36,37].

The parameters describing the scintillation efficiency include the light yield (Nph/Eγ) and the overall scintillation efficiency η. (Nph/Eγ) for photons describes the number of UV/vis photons (Nph) produced in the scintillation conversion per energy Eγ (usually in MeV) of incoming X-/γ-ray [27]. The overall scintillation efficiency η is usually used to evaluate the scintillation ability of X-ray phosphors, and it is the ratio of the energy emitted as scintillation light and the energy absorbed by the scintillation material [38]

$$ \eta ={N}_{\mathrm{ph}}\bullet {E}_{\mathrm{vis}}/{E}_{\gamma } $$
(1)

where Evis is the energy of generated UV/vis photons.

The X-ray linear attenuation coefficient (μ, cm−1) describes how fast the X-ray photon is attenuated in a material. It is determined by the mass-density ρ, the effective atomic number (Zeff), and the ionizing photon energy E. Normally, dense IOSs comprising heavy elements (e.g., Gd, Lu, and I) have relatively higher value of μ than those of OSs, as is shown in Table 1. Term μ is the sum of the linear attenuation coefficients of the primary photon interactions μi (“i” represents “photoelectric,” “Compton,” or “pair”), given by:

$$ {\displaystyle \begin{array}{c}\mu ={\mu}_{\mathrm{photoelectric}}+{\mu}_{\mathrm{Compton}}+{\mu}_{\mathrm{pair}}\\ {}=\left[{\left(\frac{\mu }{\rho}\right)}_{\mathrm{photoelectric}}+{\left(\frac{\mu }{\rho}\right)}_{\mathrm{Compton}}+{\left(\frac{\mu }{\rho}\right)}_{\mathrm{pair}}\right]\bullet \rho =\frac{\mu }{\rho}\bullet \rho \end{array}}\kern0.5em $$
(2)
Table 1 The physical properties of typical IOSs compared with OSs BCF-10 and BCF-12 [28, 30, 34, 39,40,41,42]

where \( {\left(\frac{\mu }{\rho}\right)}_{\mathrm{photoelectric}} \), \( {\left(\frac{\mu }{\rho}\right)}_{\mathrm{Compton}} \), and \( {\left(\frac{\mu }{\rho}\right)}_{\mathrm{pair}} \) are the mass partial interaction coefficients of the three primary photon interactions; μ/ρ is the total mass interaction coefficient. The behavior of photon beam (flux intensity is I) attenuation when transporting a distance l in a uniform material is usually described by the exponential law [43]

$$ I(l)={I}_0\bullet \exp \left(-\mu \bullet l\right) $$
(3)

where I0 is the initial flux intensity.

The total mass attenuation coefficients (μ/ρ) of some IOSs used for reported IOSFD construction or applications are shown in Fig. 2, which are sourced from the XCOM database [44]. For ionizing photons with relatively low energies under 100 keV, photoelectric absorption is the dominant type of photon interaction for IOSs. The energy of the incident photon is totally absorbed during this process. For the water-equivalent material, Compton scattering is the main type of photon interaction when the photon energy is larger than 30 keV [45], and only part of the photon energy is absorbed for light signal generation. As for γ-rays of high energy (more than a few MeVs, as shown in Fig. 2b) used for RT treatment, Compton scattering and electron-positron pair production are predominant, and (μ/ρ) for γ-rays decreases distinctively compared with X-rays in low energy ranges (Fig. 2). More information regarding this topic is published [43, 46, 47]. Due to different photon interaction mechanisms, higher values of (μ/ρ), and ρ, most IOSs exhibit distinctively higher X-ray absorption coefficients than those of OSs in both low (< 100 keV) and high (> 2 MeV) photon energy ranges. We note here that the energy mainly refers to the energy level of incident X-rays, which is determined by the accelerating voltage of the X-ray tube or linac.

Fig. 2
figure 2

The photon mass attenuation coefficients μ/ρ (g ∙ cm−3) of IOSs Gd2O2S, NaI, ZnS, and Al2O3 for photons in the energy ranges 10–100 keV (a), and 2—22 MeV respectively. The insert in (a) is the mass attenuation coefficient of H2O

The mass attenuation coefficients (μ/ρ) of most materials decreases with respect to photon energy in 10–100 keV energy range (Fig. 2a). However, for scintillators comprised of heavy elements such as Gd or I, there is an abrupt rise in the (μ/ρ) − E curve, owning to the characteristic photoelectric absorption at K-edge (e.g., I: 33.17 keV; Gd:50.02 keV). The relationship between (μ/ρ) and E may explain the energy-dependent sensitivity of IOSFD in low-energy X-ray radiation.

The scintillation decay time relates to the scintillation kinetics. The “Scintillation mechanism” section has introduced the scintillation mechanism. The photon absorption and generation of e-h+ pairs are instant; thus, the emission time is determined by the relaxation, thermalization, and transport of e-h+ pairs and the luminescence process. Taking NaI:Tl for example, the time profile of the scintillation is shown in Fig. 3 [48]. Once irradiated, the scintillation pulse rises quickly to the top with a typical rise time within 1 ns. The decay time, however, can be as slow as a few nanoseconds and as high as several milliseconds. The decay can be approximated using an exponential [28]

$$ I(t)\sim \exp \left(-t/\tau \right) $$
(4)
Fig. 3
figure 3

Averaged scintillation pulse recorded for NaI:Tl irradiated with a 137Cs source [48]. Two exponential terms fit to the data are presented with solid line

where τ is the decay time. The decay time of NaI:Tl is shown in Fig. 3 as about 230 ns. Rare-earth elements (e.g., Tb, Pr, and Eu) are often introduced as dopants into IOSs and normally act as luminescence centers. The decay time will also change accordingly with respect to the dopants. Table 1 shows that with the same matrix of Gd2O2S, the scintillator doped with europium (Eu) exhibits the longest decay time (τ ~ 106 ns). Gd2O2S with the dopant praseodymium (Pr) has the shortest decay time (τ ~ 7 μs). This is because the allowed 5d → 4f transitions of Pr3+ ions (the luminescence centers) are fast (tens of nanoseconds), while the lifetime of forbidden and faint f → f transitions (5D0 → 7Fj) of Eu3+ is much slower. To achieve real-time dosimetry, the IOSs chosen for IOSFD should have decay time that is short enough so as to secure accuracy, due to its effect on the sampling frequency.

Normally, it is required that scintillator material used for real-time monitoring has short decay time. However, the long decay time of some scintillators relative to the Cherenkov effect has also inspired solutions for eliminating the Cherenkov interference. For example, the BC-444 scintillator (Saint-Gobain Crystals) was chosen by Archer et al. [49] to fabricate a PSFD because of its slow rise and decay time compared to the Cherenkov effect. They applied an algorithm-based temporal Cherenkov removal technique to eliminate Cherenkov interference. The decay time of NaI:Tl is similar to that of BC-444 (285 ns). Other inorganic scintillators such as Gd2O2S:Tb and YVO4:Eu have much longer decay time; thus, the same temporal Cherenkov removal technique can also be applied when using IOSFD.

IOSs with high light yields, e.g., Gd2O2S:Tb and CsI:Tl, have already shown good performance in X-ray computed tomography. The high X-ray linear attenuation coefficient and overall scintillation efficiency endow IOSs with great potential in low dose detection application. Moreover, numerous scintillators or phosphors with maximum emission peaks in red to infrared wavelength range provide convenience for the stem effect removal. However, there is still significant concern over IOS applied for real-time dosimetry. In the following sub-sections, the factors that limit the IOSFD application in medical radiation dosimetry are presented along with the advantages of utilizing IOSs for SFD dosimetry.

Development of IOSFDs

To the best of our knowledge, Swinth et al. [50] first applied IOSFD for low-background, low-energy biomedical photon counting. The biomedical radiation-sensitive probe comprised a large piece of NaI : Tl crystal coupled to a bundle of glass optical fibers. The probe exhibited a higher sensitivity to low-energy 239Pu photons (17 keV, 60 keV) than the diode dosimeter developed for the same purpose.

Besides ionizing X-/γ- ray detection, research efforts have also been paid to explore the potential of IOSFDs for the dosimetry of biomedical particle radiations. For example, Jang et al. [51] compared the β-ray (electrons) detection ability of different scintillator-coupled optical fiber dosimeters. As shown in Fig. 4a, the sensor tip is fabricated using IOSs and plastic optical fiber bundles. With a metal hydride type of tritium source, three kinds of scintillators (Gd2O2S:Tb, Y3Al5O12:Ce, and CsI:Tl) were used for sensor fabrication to select the best scintillator. Among these three scintillators, the IOSFD with Gd2O2S:Tb exhibited the best scintillation response in terms of generated photons (Fig. 4b).

Fig. 4
figure 4

IOSFD for detection of tritium [51]. a A schematic diagram of sensor tip. b Measured amounts of scintillation photon with three kinds of IOS

In the recent decades, IOSFDs using various kinds of scintillators and configurations were developed, and their potential application for RTD such as external RTD, small-field in-vivo dosimetry, and BT has been investigated. In the following sections, the research on IOSFDs for X-ray RTD will be reviewed; efforts of utilizing IOSFDs for BT dosimetry, PT, and BNCT would also be concluded.

IOSFD for X-ray RTD

A IOSFD based on Gd2O2S:Tb was presented by McCarthy et al. [37, 52]. The schematic illustration of the probe structure is shown in Fig. 5. The cladding layer of the PMMA fiber tip was removed, and the scintillator powders were coated surrounding the fiber core using an epoxy resin-based molding method. Though this dosimeter was less sensitive to X-ray energy below 50 kVp, it exhibited good sensitivity, stability, and repeatability of measurement for various levels of low-energy ionizing X-ray (50–140 kVp). The dosimeter also demonstrated excellent spectral response and repeatability for γ-ray radiation as high as 6 MV and 15 MV. This research carried out an initial characterization measurement of IOSFD in both air and water phantom and is a useful study of IOSFD concerning the linearity, repeatability, stability, and effective coupling between scintillator powders and optical fiber. However, it would still need more research effort into the angular dependence, radiation field-size dependence, and the PDD measurement, for a more comprehensive assessment of this dosimeter for real-time in vivo dosimetry.

Fig. 5
figure 5

The fiber optic radiation dosimeter sensor design reported by McCarthy et al. [52]

The coupling efficiency between the scintillator and the optical fiber is important for the light collection. To improve the coupling efficiency, O’Keeffe et al. [53] proposed a new design by embedding a piece of Gd2O2S:Tb scintillator into the core of an optical fiber (Fig. 6). This design was further confirmed and investigated by Qin et al. [19]. The configuration not only reduced the transmission loss, but also isolated the scintillator from the surrounding environment. Except from the excellent signal-to-noise ratio, linearity, and repeatability, the detector showed a great isotropic response to both radical and axial changes of the incident radiation as well.

Fig. 6
figure 6

Embedded structure of an IOSFD. a The IOSFD (photograph) presented by O’Keeffe et al. [53]. b The IOSFD (schematic representation)

Another effort to increase the light coupling efficiency between the scintillation domain and optical fiber of Gd2O2S:Tb IOSFD was presented by Andres et al. [54]. The basic idea of the sensor configuration optimization is accelerating the light extraction efficiency from scintillation domain via expanding the useful surface in contact with the scintillator. The sensor tip (Fig. 7) is modified by either chemical etching (LM2 in Fig. 7a) or thermomechanical tapering (LM3 in Fig. 7a). Precautions to minimize the air-bubble density were also taken, aiming to increase the optical signal intensity. All fabricated devices have been tested under X-ray irradiation (6 MV) from a linac at a dose rate of 300 monitor units/min with a readout every 100 ms. The results of LM2 (chemical etching) shown in Fig. 7b demonstrated a signal improvement of up to 43 times compared with the previously reported proposals using IOSs and polymer fibers (LM1). Moreover, a scintillator format comparison between scintillator powders and scintillator-epoxy mixture demonstrated that the IOS used in powder format could achieve higher sensitivity. This is due to the higher packing density of the scintillator in the scintillation domain. This is another successful illustration of improving the light coupling efficiency between the scintillator and optical fiber through sensor tip modification.

Fig. 7
figure 7

a Schematic draw of the fabricated transducers LM-1 (plain fiber), LM-2 (etched fiber), and LM-3 (tapered fiber). b Sensor signal recorded for single fiber tip (dotted line), tapered fiber tip (dashed line), and etched fiber tip (solid line) [54]

Research on ruby-based IOSFD was reported by Jordan et al. in 1996 [35]. Here, a ruby (Al2O3:Cr, 1 mm in diameter)-coupled PMMA optical fiber was used to evaluate the integrated dose of both X-ray and electron beams generated with Varian CL600C and 2100A medical linacs. The time-delayed method combined with optical filtering method was used to remove the stem effect (Cherenkov effect and fluorescence) signal from the optical fiber. The results demonstrated that this time-delayed gated signal integration method is effective for eliminating the fluorescent and Cerenkov light origination in the irradiated optical fiber. Irradiated by 4 MV photon beams and 9–12 MeV electron beams, the PDD measurements with this probe agree well with values using an ionization chamber. Therefore, it is possible to apply ruby-based IOSFD for radiation dosimetry. This temperature dependence of Al2O3:Cr was also investigated in this study, and a temperature correction on the order of 0.6% per degree was suggested for clinical application.

The application of Al2O3:Cr ruby-based IOSFD in radiotherapeutic field was further investigated by Teichmann et al. [55]. A ruby (Al2O3:Cr) crystal (1-mm diameter) was adhered to a fused silica light guide (BFH-400, Thorlabs) with optically transparent epoxy (EPO-TEK® 301-2, Epoxy Technology). An optical single-band pass filter (Semrock 689/23 nm BrightLine®) was adopted to remove the unwanted stem signals from the fiber guide. The probe was tested under irradiation of both electron and photon radiation with a radiation dose up to 0.5 kGy. The radioluminescence signal shows a slight linear rise with an accumulated dose up to 2 Gy, which is stable for large doses up to 0.5 kGy. The sensor also demonstrated good sensitivity to dose as low as 2.6 ± 0.2 mGy/h. The stem effect produced in optical fibers covers a wide range of wavelengths [20, 56], and it cannot be completely eliminated. Thus, the stem signal below and above the energy threshold for the generation of Cherenkov radiation could be modeled by a linear function according to their research.

Yttrium phosphors doped with Eu such as Y2O3:Eu, Y2O2S:Eu, and YVO4:Eu emit bright red-light under UV or X-ray excitation; these have been used widely in the industry (e.g., white LEDs). Attempts have also been made to utilize them for dosimetry [57,58,59,60]. One merit of these Eu-doped yttrium phosphors is that they provide the convenience of the stem effect removal with the simplest optical-filtering method. The stem effect (including Cherenkov effect and fluorescence of the plastic optical fiber) dominates in the blue/green spectral region; therefore, combining a red-emission phosphor with a long-pass optical filter can effectively suppress or remove these spurious signals [61, 62].

Molina et al. investigated the feasibility of utilizing Eu-doped yttrium phosphors for fiber optic dosimetry [57]. The radio-luminescence response of three commercial phosphors (Y2O3:Eu, Y2O2S:Eu, and YVO4:Eu) was characterized. All three materials showed no changes in RL response as dose accumulates with no afterglow decay. The same research group further studied the feasibility of using Y2O3:Eu-based IOSFD for real-time dosimetry under 60Co irradiation [58]. The stem effect signal was successfully suppressed via a long-pass optical filter (610 nm). Figure 8 shows that the PDD profile of the IOSFD was affected not only by the irradiation field size but also by the coating thickness of the Y2O3:Eu phosphor. The cause of this PDD deviation was attributed to the higher atomic number of Y2O3:Eu in this research. The larger proportion of phosphor acts as a scintillator with respect to the larger irradiation field size.

Fig. 8
figure 8

The PDD curves from Y2O3:Eu-based FOD compare to that of a standard IC under 60Co irradiation. Surface to source distance is 80 cm and the irradiation field is set to a 10 × 10 cm2 and b 5 × 5 cm2 respectively [58]. Different phosphor coating thickness is applied

A more reasonable explanation, according to the research of Qin et al. [63], is that the inorganic scintillators are more sensitive to photons with energies under 1 MeV. Due to multiple Compton scattering events during the photon transporting in water, the photon energy keeps decreasing with respect to the increasing water depth (i.e., softening of the photon spectrum). The energy eventually reaches below 1 MeV where the photon electronic interaction occurs. As a result, the inorganic scintillator exhibits an over-response compared to IC and PSFD. Nevertheless, this result demonstrated the potential of Y2O3:Eu-based IOSFD for in vivo and real-time dosimetry; however, the coating thickness of the phosphor attached to the fiber is not given, and the angular dependence requires further illustration regarding this kind of dosimeter.

The consistent work on YVO4:Eu-based IOSFD was reported by Martínez et al. [59, 64]. A temporal discrimination method was applied to eliminate the stem effect. The Cherenkov effect is a fast process and decays rapidly without radiation; scintillation is a delayed process. Therefore, the temporal discrimination method can effectively remove the Cherenkov signal by using a scintillation material with relatively longer radio-luminescence lifetime [20]. In this case, a pulse radiation mode was set with the linac, and only the optical signal starting 20 μs after each linac pulse reaching the detector was recorded. The YVO4:Eu-based IOSFD demonstrated temperature-independence, good spatial resolution, and capability for real-time dosimetry. The azimuthal angle experiment test showed that the geometry of the detector determines angular-dependent response to the radiation [59]. The Monte-Carlo simulation method was successfully used to verify the PDD profiles measured with the IOSFDs and the standard IC in the fields with different sizes.

IOSFD for BT dosimetry

To date, commercially available in-vivo dosimeters for BT can only measure the total dose post-treatment, which might introduce errors and uncertainties during the process. Given the small size, durability, and fast and linear response of PSFDs, there have been reports of using PSFD or PS-based radiation detector for in vivo BT dosimetry and source tracking to improve the dose delivery accuracy [65,66,67,68]. Compared to PS, IOSs are more sensitive to low dose radiation and have much higher scintillation efficiency; thus, IOS-based radiation detectors have great potential for in vivo BT dosimetry.

Kertzscher and Beddar tested the potential of the Al2O3:Cr-based IOSFD for dosimetry in 192Ir BT [69]. This ruby-based IOSFD was compared with common PSFDs. The doping concentration of Cr3+ is 0.5%, and the OSs chosen are BCF-12 and BCF-60. Also, in this experiment, three types of IOSFD configurations were tested (Fig. 9): one without reflective paint and optical filter, one with reflective paint but no filter, and one with both a filter and reflective paint. The optical coupling effect of different ruby morphologies was investigated, namely, sphere and half-sphere. The air-kerma strength of the 192Ir source is between 17.6 and 40.0 mGy m2·h−1 throughout the experiments. The contribution of the stem effect and photoluminescence signal from the fiber was quantified. By incorporating a long pass filter between the scintillator and optical fiber, the unwanted photoluminescence from fiber was suppressed from 1 to 5% as the radiation source dwelled 0.5 cm away from the fiber-optic cable. The stem effect was suppressed as low as < 3% using a band-pass filter because the distance from the scintillator to the source is smaller than 7 cm. The IOS exhibited a much stronger scintillation signal than that of OSs—up to 20-fold that of BCF-12-based PSFDs. The result shows that by choosing IOSs with insignificant time-dependent luminescence and afterglow, the inorganic IOSFDs are promising candidates for in vivo high-dose-rate BT dosimetry.

Fig. 9
figure 9

The 3 types of detector probe tips that were used in the experiments in ref [69]

In addition to the Al2O3:Cr ruby, Kertzscher and Beddar also studied four types of IOSs-based fiber dosimeters for real-time BT dosimetry [60, 70]. The scintillation property of commercial Eu-doped phosphors Y2O3:Eu, YVO4:Eu, Y2O2S:Eu, and Gd2O2S:Eu was assessed and compared to plastic scintillators BCF-12 and BCF-60. The IOSFD exhibited scintillation intensities about 16–134 times of that of BCF-12-based PSFD. More importantly, this article showed that the time-dependent response caused by the afterglow of Eu-doped scintillators can be suppressed by the mixture of Y2O3:Eu and YVO4:Eu with an appropriate weight ratio. The same research group later compared ruby (Al2O3:Cr), Y2O3:Eu, YVO4:Eu, ZnSe:O, and CsI:Tl-based IOSFDs for 192Ir BT and also used BCF-12 and BCF-60 as a reference [70]. According to their test, ZnSe:O-based IOSFD exhibited the most favorable characteristics considering its high and stable scintillation intensity, dose linearity, insignificant memory effects, and negligible stem interference. However, all IOSFDs tested are heavily energy dependent because of the non-water equivalent characteristic of IOSs and the distant-dependent energy spectra of 192Ir BT source in water. As a result, further attempts to acquire signal intensity-dependent correction factors are needed.

The possibility of doing time-resolved in vivo dosimetry with the IOSFD for source tracking in BT was also explored [67, 71, 72]. Belley et al. [71] adopted an inorganic nanocrystalline composition (Y2O3:Eu,-Li) as the scintillation part and connected it to a silica-based optical fiber. The clinical performance of real-time dose rate monitoring during high-dose-rate (HDR) BT was evaluated and compared to thermoluminescent dosimeters and the treatment planning system. The result conferred the feasibility of using Y2O3:Eu,-Li-based IOSFD for real-time measurements during HDR BT. Johansen et al. [67] used Al2O3:C-based IOSFD as an in vivo dosimeter for source tracking BT. The IOSFD was placed in a delicate BT needle in the prostate near a 192Ir radiation source. The delivered dose rate and total dose were recorded with a single IOSFD, and the distance between the source and dosimeter was derived from the measured dose rate. The shift of the treatment needle was successfully detected with the IOSFD. This work demonstrated the great application potential of IOSFD to source tracking in BT. The same research group later used the Al2O3:C-based IOSFD to verify the dwell time in HDR BT [72]. By analyzing the dose rate at different sampling rates, dwell times in the identified dwell positions were calculated with high accuracy. Though the detection efficiency needs to be improved, the research proposed an effective method to detect dwell time offsets in BT. Overall, IOSFD-based in vivo dosimetry has promising applications in BT dosimetry.

IOSFD for other types of particle radiation dosimetry

Except for photon-based RT and BT dosimetry, IOSFDs also demonstrate the potential for dosimetry in other types of therapy treatments such as PT and BNCT [73,74,75,76,77,78,79]. PT offers improved dose distribution compared to ionizing photon radiation. The deposited dose reaches the maximum near the end of the transporting range (the Bragg peak). However, applying SFD for PT dosimetry faces challenges whereby SFD exhibits a non-linear response to proton radiation dose at high-energy deposition of the particles within the fiber. These are due to the ionization-quenching phenomenon resulting from non-radiative de-excitation occurring under conditions of high-density energy deposition [80]; efforts have been made to find a scintillator with minimized quenching phenomenon. The IOSFDs reported in this area include a Gd2O2S:Tb-based detector and a lanthanide-doped silica scintillator-based detector [73,74,75]. Penner et al. [73] investigated Gd2O2S:Tb-based IOSFD for photon and proton dosimetry. For proton dose detection, the detector exhibited excellent sensitivity, signal-to-noise ratio, and reproducibility. However, further research effort is required to address the errors caused by the significant quenching.

A Ce-doped silica scintillation fiber dosimeter was tested under 74 MeV proton radiation [74]. The experiment demonstrated that the Ce-doped silica IOSFD has a good linear response to accumulated dose, irradiated dose length, and total dose rate (Fig. 10a–c). At high-density energy deposition of protons, the quenching effect is significant with a peak-to-plateau signal ratio of 2.79 for the proton raw Bragg peak profile; nevertheless, this value is still lower than most standards (≥ 3.7) indicating less severe quenching as shown in Fig. 10d. Hoehr et al. [75] reported an innovative Gd-doped amorphous silica bulk scintillator for dosimetry in PT. This Gd-doped silica IOSFD was tested with 8.2–62.9 MeV protons and 2–6 nA of extracted beam current and then compared with Ce- and Cu-doped silica IOSFDs fabricated with the same sol-gel technique. Three types of IOSFDs all exhibited strong luminescence and linear response to proton radiation. Of these, the Gd-doped silica IOSFD shows superior resolution of the Bragg peak with the lowest Birks’ constant (kB = (0.0162 ± 0.0003) cm/MeV), which is also lower than common PSFD, indicating a significant reduced quenching. Their research demonstrated that this Gd-doped silica IOSFD could be a promising candidate for PT dosimetry, and future steps to minimize the detector size were suggested.

Fig. 10
figure 10

The response of a Ce-doped silica based IOSFD to a 74 MeV proton radiation source [74]. a Integrated counts as a function of set dose for the fiber. b Signal from fiber as a function of its irradiation length. c Count rate from the fiber as a function of dose rate. d Depth dose profiles for the Spread Out Bragg Peak using various dosimeters

During the BNCT process, 6Li-based inorganic scintillators have been adopted to measure the neutron flux. Thermal and epi-thermal neutrons are captured by 6Li(n,t)α reactions after which α particles and tritons are emitted. Other scintillation components are absorbed by these secondary particles and emit light photons. Ito et al. [76] fabricated a 6LiF-ZnS(Ag)-based IOFD to monitor a thermal neutron flux for BCNT. Two types of probes were fabricated: one comprised a thin ZnS(Ag) film and a thin 6LiF film, and another comprised a wavelength-shift (WLS) fiber coated with a ZnS(Ag) film and a thin 6LiF film. These two probes aimed to improve gamma-neutron discrimination ability and neutron detection efficiency, respectively. Watanabe et al. [81] used a similar 6LiF-ZnS(Ag)-WLS combination to increase the neutron sensitivity. The effective length of the detector has a positive effect on the sensitivity. Watanabe et al. [78] reported a small Eu:LiCaAlF6 scintillator-based IOSFD for neutron detection in BNCT. The light yield, the light collecting efficiency, and the transparency of the scintillator were improved compared with the aforementioned 6LiF-ZnS(Ag)-based IOSFD. The experiment confirmed an enhancement on the gamma-neutron discrimination ability. However, the output linearity and dynamic range in real BNCT field requires further investigation.

The “Development of IOSFDs” section reviews IOSFDs with different scintillation materials and sensor configurations for dosimetry in ionizing photon RT, 192Ir BT, PT, and BNCT. Clinical applications still require more research to identify proper scintillation materials, scintillator fiber-coupling configurations, and correction factors as well as protocols for IOSFD for real-time in vivo dosimetry.

Challenges of IOSFD for RTD application

Recent years have witnessed promising progress in IOSFD research, but there still remains some concern over the clinical utilization of IOSFDs for in vivo dosimetry. This section will discuss three major challenges in this research area: PDD deviation from the standard IC measurement, the angular dependence, and the Cherenkov effect.

PDD deviation from the standard IC measurement

The PDD curve is used to measure the percentage of dose deposited in water at different depths, with respect to the point of maximum dose. The curves measured with SFDs are normally compared with standard IC. One attractive merit of PSFD is the water-equivalent property of the plastic organic scintillators and plastic optical fibers; therefore, the PSFD normally has no interference on the dose distribution. Early research on PSFD has demonstrated a good agreement with IC regarding the PDD test [15]. A maximum error of 0.3% was achieved by neglecting the Cherenkov effect when the detector was placed perpendicular to the central axis of a 10 MV X-ray radiation beam (field size 10 × 10 cm2). However, most of IOSFDs exhibited non-ignorable deviation of depth-dose characteristics compared to the IC measurement. To date, this remains the major impediment to the clinical application of IOSFD. For example, the Gd2O2S:Tb-based IOSFD exhibited overresponse to the X-ray (6 MV, 100 MU) when the field size is larger than 10 × 10 cm2 (at 1.5 cm in water equivalent material where the radiation dose reaches maximum (Dmax) for a 6-MV photon beam measured by the IC) [27]. The deviation increased as the field size increased. The magnitude of the deviation depends not only on the field size but also on the type of scintillator, the incident radiation beam profile, and the sub-surface depth [27, 58, 63, 82,83,84,85].

Multiple factors may give rise to the PDD deviation using IOSFD sensors compared with ICs. In the study on YVO4:Eu, Gd2O2S:Tb, and ZnSe(Te)-based IOSFDs, the overresponse of the IOSFD to radiation in PDD experiment was attributed to the higher Zeff of these inorganic scintillators [59, 83, 86]. The high Zeff results in an over-response of IOSFD with respect to water or water-equivalent material due to the secondary radiation of low energy. A recent study reported a new model investigating the overresponse phenomenon in PDD measurement with IOSFD [63]. The experiment verified that the influence of the energy effect and Cherenkov effect only contributed to a small part of PPD deviation from the IC measurement. To explain the overresponse phenomenon, the new model considers the softening of the energy spectrum of the incident X-ray/γ-ray and the secondary electrons, as well as the sensitivity of IOS to photons and secondary electrons with lower kinetic energies. As the photons penetrate into the water phantom, the mean energy of the photon beam and secondary electrons decreases due to photon interactions [87, 88]. The IOS is more sensitive to photons with lower energy in the deeper depth. The model also suggests that the scintillation efficiency of IOS corresponding to lower energetic secondary electron absorption is also higher considering the inner scintillation mechanism. This group used the shielding of the low energy particles and different beam size experiments to prove the theory. However, further Monte-Carlo simulation-based research (including a thorough consideration of the incident beam profile, secondary electrons, the radiation field size, and the probe size of the IOSFD) would be helpful to understand this theory.

This same group proposed a solution to correct the PDD curves of IOSs [85]. This method utilized the fact that different scintillators have unique PDD characteristics and near-perfect dose linearity for a fixed depth. By applying a parallel-paired fiber light guide structure where the two different scintillator materials (Gd2O2S:Tb and NaI:Tb in this case) are separately embedded in two fiber tips, the information of water depth and absorbed dose at the point of measurement can be extracted. This method provides one solution to overcome the PDD curve uncertainty and errors using IOSFD. However, the accuracy of applying this method to a clinical trial with respect to varying field sizes still needs further investigation. Overall, for the wide clinical application of IOSFD, more efforts are needed to study the cause and solution of the PDD deviation of various IOSFDs with respect to the IC measurements.

Angular dependence

Angular dependence was also constantly mentioned in the research concerning IOSFDs [19, 59, 89, 90]. The angular dependence arises from the geometrical asymmetry of the scintillator material in the sensor tip. For better coupling with the optical fiber, the scintillation part is normally shaped in the form of cylinder geometry with a small diameter (smaller than 1 mm) and relatively long length (several millimeters; Figs. 4, 5, 6, 7, and 8). As the angle changes, photons or electrons which enter the cylindrical sensor tip from different locations and directions may have different initial energies and travel further through paths of different lengths in the scintillator material. Hence, it is reasonable to presume that the angular changes may cause a fluctuation of response to the radiation.

Considering the geometrical symmetry, response of IOSFD to radiation with respect to angular changes along axial direction and azimuthal direction of the scintillator and optical fiber (Figs. 11) have different characteristics [59]. IOSFDs with cylindrical sensor tips normally exhibit great axial angular independence and noticeable azimuth angular dependence as shown by Qin et al. [19] and Alharbi et al. [90]. A solution to this issue was suggested from the aspect of geometry [59]. The scintillator powders were dispersed in a quasi-sphere droplet compared with a cylinder-shaped scintillator tip. The probe with this quasi-spherical geometry exhibited better azimuth angle independence with an observed variation less than 2%. Therefore, it is possible to achieve angular independence by optimizing the geometry of the scintillator material pack.

Fig. 11
figure 11

The schematic of the angler dependence measurement showing the azimuthal angular change and the axial angular change [59]. The arrows by the angle 90°, 180°, and 270° give the direction of incident radiation

Cherenkov effect removal techniques

The Cherenkov effect is considered as a major interference when using SFDs for dosimetry. It describes the phenomena that electromagnetic radiation is emitted when the charged particles (e.g., electrons and positrons) passes through a material with the speed exceeding the phase velocity of light in that medium. It occurs in almost every clear material (e.g., silica or plastic optical fiber) at electron energies approximately 180 keV [56]. The Cherenkov light signal covers a broadband spectrum ranging from UV to infrared wavelength. The Cherenkov radiation from the optical fiber and scintillators limits the accuracy of the dosimeter. The intensity of the Cherenkov radiation is of the same order of magnitude as the scintillation signal of the plastic scintillator, thus it must be removed if PSFD is applied [91, 92]. In IOSFD, the intensity of the scintillation signal is significantly stronger than that of the Cherenkov signal. Nevertheless, it is still necessary to remove the Cherenkov effect and only account for the scintillation signal. Here, we briefly summarize the solutions to remove the Cherenkov effect including examples of both inorganic and organic scintillator-based SFDs.

Studies on the generation of Cherenkov radiation have been carried out for both silica and plastic optical fiber [62, 93,94,95]. At least four different methods have been devised to eliminate the Cherenkov effect. These methods either remove Cherenkov physically or take advantage of the characteristic differences of the scintillation and Cherenkov radiation (such as the spectrum and the rise-decay time).

The first method, presented by Beddar et al. [14], removed the Cherenkov background signal by employing a parallel optical fiber located close to the signal fiber that coupled with a piece of scintillator. The light signal detected in the second fiber is an approximation (or a reference) of the Cherenkov background in the signal fiber. The scintillation signal was corrected by subtracting this background signal detected by the reference fiber. The limitation of this method is that any positional differences between the two fibers will introduce errors [95], and it increases the size of the dosimeter.

The second method exploits the spectral difference between the scintillation light and the Cherenkov radiation [61]. Importantly, the intensity of the Cherenkov spectrum is proportional to λ−3 (where λ is the wavelength of the Cherenkov light), and the emission peak is in the blue/green light region. By using scintillation coupled with a longer wavelength emission (e.g., red emission peak) and filtering out the light with a shorter wavelength, the Cherenkov emission was decreased from 6.5 to 2.8% of the scintillator’s signal. Still, the Cherenkov signal cannot be totally eliminated using this method, as Cherenkov light spectrum is continuous and comprises photons with long wavelength. Later, Fontbonne et al. [96] demonstrated that by measuring the absolute dose and the intensities of green and blue emission at two different locations, the Cherenkov background signal can be, in principle, eliminated using linear equations to calculate the contribution of the Cherenkov signal and the scintillation signal. Further research regarding the correction of the Cherenkov effect based on this spectral discrimination method was presented by Frelin et al. [97], Guillot et al. [21], and Ishikawa et al. [98].

The third method, which has been mentioned in the “Basic physical properties of scintillators” section, was presented by Clift et al. [20]. It is a temporal method, and it relies on the fact that the Cherenkov emission is a prompt process, whereas scintillation is a delayed process. The Cherenkov signal could be reduced by reading the signal from the dosimeter between the linac pulses when the Cherenkov radiation was decayed to almost zero. An organic scintillator BC-444G with a long decay constant (264 ns) was used to maximize the amount of scintillation light emitted in between the linac pulses, and a sampling time of 700–705 ns was adopted. This setup successfully eliminated most of the Cherenkov and fluorescent radiation.

The last method, proposed by Lambert et al. [91], corrected the Cherenkov signal “physically” by removing the media where the Cherenkov signal is produced. In this case, the fiber core exposed in radiation was removed. The experiment used an air core light guide to transport the light from the scintillator to the light detector. This theoretically eliminates the generation of Cherenkov light at its source and thus provides a novel solution to the Cherenkov effect.

Conclusion

This work reviewed the work on IOSFDs for medical radiation dosimetry. First, some basic knowledge regarding scintillator material was introduced. The fundamentals of IOSs were reviewed in the “Fundamentals” section in terms of overall scintillation efficiency, decay time, mass attenuation coefficients, and energy dependence. Dense IOSs with high photon attenuation coefficient and scintillation efficiency are favorable for the application of IOSFD for low-dose rate in vivo dosimetry. The “Development of IOSFDs” section detailed the development of IOSFDs by applying different scintillator materials and configurations. Research concerning embedded-structure IOSFDs and sensor tip modification certifies that by optimizing the scintillator coupling method, it is possible to improve the coupling efficiency between the scintillator and the optical fiber. Furthermore, various IOSs with the bright red emission maximum provide convenience to suppress the Cherenkov effect and stem effect in the optical fiber using a spectrum filtering method. IOSFD also offers promising potential for dosimetry in BT, PT, and BNCT, but problems like the quenching phenomenon must be addressed. The response characteristics of different IOSs to radiation dose from various charged particle sources should be investigated in the future. In the “Challenges of IOSFD for RTD application” section, challenges for IOSFD development including PDD deviation from standard IC measurement and angular dependence were discussed. In the “Cherenkov effect removal techniques” section, typical Cherenkov removal techniques were summarized. Clinical applications still require more research effort to identify a proper scintillator material, scintillator-fiber coupling configuration, and correction factor; protocols using IOSFD for in vivo and real-time dosimetry are also still needed. Nevertheless, it is reasonable to expect more use of IOSFDs for in vivo real-time RTD considering their good linearity and repeatability, high sensitivity, and convenience for stem effect removals.

Availability of data and materials

The datasets generated during the photon attenuation coefficient analyses as shown Fig. 2 are available from XCOM: Photon Cross Sections Database on the website https://www.nist.gov/pml/xcom-photon-cross-sections-database

Abbreviations

BT:

Brachytherapy

BNCT:

Boron neutron capture therapy

Z_eff:

Effective atomic number

IOSFD:

Inorganic scintillator-based optical fiber dosimeter

IOS:

Inorganic scintillators

IMRT:

Intensity-modulated radiotherapy

IC:

Ion chamber

linac:

Linear accelerator

max:

Maximum

OS:

Organic scintillator

PDD:

Percent depth dose

PSFD:

Plastic scintillator-based optical fiber dosimeter

PMMA:

Polymethyl methacrylate

PT:

Proton therapy

QA:

Quality assurance

RT:

Radiotherapy

RTD:

Radiotherapy dosimetry

SFD:

Scintillator-based optical fiber dosimeter

SABR:

Stereotactic ablative radiotherapy

UV/vis:

Ultraviolet/visible

WLS:

Wavelength-shift

References

  1. Lillicrap SC, Owen B, Williams JR, Williams PC. Code of Practice for high-energy photon therapy dosimetry based on the NPL absorbed dose calibration service. Phys Med Biol. 1990;35:1355–60.

    Article  Google Scholar 

  2. Thwaites DI, Burns DT, Klevenhagen SC, Nahum AE, Pitchford WG. The IPEMB code of practice for electron dosimetry for radiotherapy beams of initial energy from 2 to 50 MeV based on an air kerma calibration. Phys Med Biol. 1996;41:2557–603.

    Article  CAS  Google Scholar 

  3. Symonds RP, Deehan C, Meredith C, Mills JA. Walter and Miller’s textbook of radiotherapy: radiation physics, therapy and oncology. seventh. Elsevier Health Sciences; 2012.

  4. Mayles WPM, Lake R, McKenzie A, Macaulay E, Morgan H, Jordan T. IPEM report 81: physics aspects of quality control in radiotherapy. York, UK Inst Phys Eng Med. 1999;.

  5. Bogdanich W. Radiation offers new cures, and ways to do harm. New York Times. 2010;23:A1.

    Google Scholar 

  6. Bogdanich W. As technology surges, radiation safeguards lag. New York Times. 2010:1–13.

  7. Garcia R, Nyström H, Fiorino C, Thwaites D. Does a too risk-averse approach to the implementation of new radiotherapy technologies delay their clinical use? Br J Radiol. 2015;88.

  8. Tanderup K, Beddar S, Andersen CE, Kertzscher G, Cygler JE. In vivo dosimetry in brachytherapy. Med Phys. 2013;40:070902.

    Article  PubMed  Google Scholar 

  9. O’Keeffe S, McCarthy D, Woulfe P, Grattan MWD, Hounsell AR, Sporea D, et al. A review of recent advances in optical fibre sensors for in vivo dosimetry during radiotherapy. Br J Radiol. 2015;88:20140702.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Ciocca M, Piazzi V, Lazzari R, Vavassori A, Luini A, Veronesi P, et al. Real-time in vivo dosimetry using micro-MOSFET detectors during intraoperative electron beam radiation therapy in early-stage breast cancer. Radiother Oncol. 2006;78:213–6.

    Article  CAS  PubMed  Google Scholar 

  11. Butson M, Cho G, Odgers D, Poder J. In vivo dosimetry. Radiochromic Film. CRC Press; 2017. p. 301–316.

  12. Legge K, Greer PB, O’Connor DJ, Wilton L, Richardson M, Hunter P, et al. Real-time in vivo rectal wall dosimetry using MOSkin detectors during linac based stereotactic radiotherapy with rectal displacement. Radiat Oncol. Radiation Oncology. 2017;12:1–9.

    Article  Google Scholar 

  13. Byfield J, Chan P. D JEMP. A flexible fiber optics scintillation probe for intracavitary dosimetry 1. Radiology. 1969;92:1120–3.

    Article  CAS  PubMed  Google Scholar 

  14. Beddar AS, Mackie TR, Attix FH. Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical considerations. Phys Med Biol. 1992;37:1883–900.

    Article  CAS  PubMed  Google Scholar 

  15. Beddar AS, Mackie TR, Attix FH. Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: II Properties and measurements. Phys Med Biol. 1992;37:1901–13.

    Article  CAS  PubMed  Google Scholar 

  16. Archambault L, Arsenault J, Gingras L, Beddar AS, Roy R, Beaulieu L. Plastic scintillation dosimetry: optimal selection of scintillating fibers and scintillators. Med Phys. 2005;32:2271–8.

    Article  PubMed  Google Scholar 

  17. Buranurak S, Andersen CE, Beierholm AR, Lindvold LR. Temperature variations as a source of uncertainty in medical fiber-coupled organic plastic scintillator dosimetry. Radiat Meas. 2013;56:307–11.

    Article  CAS  Google Scholar 

  18. Beddar AS, Law S, Suchowerska N, Mackie TR. Plastic scintillation dosimetry: optimization of light collection efficiency. Phys Med Biol. 2003;48:1141–52.

    Article  PubMed  Google Scholar 

  19. Qin Z, Hu Y, Ma Y, Zhao W, Sun W, Zhang D, et al. Embedded structure fiber-optic radiation dosimeter for radiotherapy applications. Opt Express. 2016;24:5172.

    Article  Google Scholar 

  20. Clift MA, Johnston PN, Webb DV. A temporal method of avoiding the Cerenkov radiation generated in organic scintillator dosimeters by pulsed mega-voltage electron and photon beams. Phys Med Biol. 2002;47:1421–33.

    Article  CAS  PubMed  Google Scholar 

  21. Guillot M, Gingras L, Archambault L, Beddar S, Beaulieu L. Spectral method for the correction of the Cerenkov light effect in plastic scintillation detectors: a comparison study of calibration procedures and validation in Cerenkov light-dominated situations. Med Phys. 2011;38:2140–50.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Beddar AS. Plastic scintillation dosimetry and its application to radiotherapy. Radiat Meas. 2006;41:124–33.

    Article  CAS  Google Scholar 

  23. O’Keeffe S, Fitzpatrick C, Lewis E, Al-Shamma’a AI. A review of optical fibre radiation dosimeters. Sens Rev. 2008;28:136–42.

    Article  Google Scholar 

  24. Mijnheer B, Beddar S, Izewska J, Reft C. In vivo dosimetry in external beam radiotherapy. Med Phys. 2013;40:070903.

    Article  PubMed  Google Scholar 

  25. Woulfe P, Sullivan FJ, O’Keeffe S. Optical fibre sensors: their role in in vivo dosimetry for prostate cancer radiotherapy. Cancer Nanotechnol. 2016;7:7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  26. Beaulieu L, Beddar S. Review of plastic and liquid scintillation dosimetry for photon, electron, and proton therapy. Phys Med Biol. 2016;61:R305–43.

    Article  CAS  PubMed  Google Scholar 

  27. Rodnyi PA, Dorenbos P, Van Eijk CWE. Energy loss in inorganic scintillators. Phys status solidi. 1995;187:15–29..

    Article  CAS  Google Scholar 

  28. Nikl M. Scintillation detectors for x-rays. Meas Sci Technol. 2006;17:R37–54.

    Article  CAS  Google Scholar 

  29. Knoll GE, Wiley J. Radiation detection and measurement. 3rd ed. John Wiley & Sons, Ltd.; 2000.

  30. Van Eijk CWE. Inorganic scintillators in medical imaging. Phys Med Biol. 2002;47:85–106.

    Article  Google Scholar 

  31. Moszyński M. Inorganic scintillation detectors in γ-ray spectrometry. Nucl Instruments Methods Phys Res Sect A Accel Spectrometers, Detect Assoc Equip. 2003;505:101–10.

    Article  CAS  Google Scholar 

  32. Ishii M, Kobayashi M. Single crystals for radiation detectors. Prog Cryst Growth Charact1. 1991;23:245–311.

    Article  Google Scholar 

  33. Weber MJ. Inorganic scintillators: today and tomorrow. J Lumin. 2002;100:35–45.

    Article  CAS  Google Scholar 

  34. Greskovich C, Duclos S. Ceramic scintillators. Annu Rev Mater Sci. 1997;27:69–88.

    Article  CAS  Google Scholar 

  35. Jordan KJ. Evaluation of ruby as a fluorescent sensor for optical fiber-based radiation dosimetry. Proc SPIE. 1996;2705:170–8.

    Article  CAS  Google Scholar 

  36. Sporea D, Mihai L, Tiseanu I, Vâţă I, McCarthy D, O’Keeffe S, et al. Multidisciplinary evaluation of X-ray optical fiber sensors. Sensors Actuators A Phys. 2014;213:79–88.

    Article  CAS  Google Scholar 

  37. McCarthy D, O’Keeffe S, Lewis E, Sporea DG, Sporea A, Tiseanu I, et al. Radiation dosimeter using an extrinsic fiber optic sensor. IEEE Sens J. 2014;14:673–85.

    Article  Google Scholar 

  38. Nikl M. Wide band gap scintillation materials: progress in the technology and material understanding. Phys status solidi. 2000;178:595–620.

    Article  CAS  Google Scholar 

  39. Kalivas N, Costaridou L, Kandarakis I, Cavouras D, Nomicos CD, Panayiotakis G. Optical gain signal-to-noise ratio transfer efficiency as an index for ranking of phosphor-photodetector combinations used in X-ray medical imaging. Appl Phys A Mater Sci Process. 2004;78:915–9.

    Article  CAS  Google Scholar 

  40. Zhang WW, Zhang WP, Xie PB, Yin M, Chen HT, Jing L, et al. Optical properties of nanocrystalline Y2O3:Eu depending on its odd structure. J Colloid Interface Sci. 2003;262:588–93.

    Article  CAS  PubMed  Google Scholar 

  41. Tseng YH, Chiou BS, Peng CC, Ozawa L. Spectral properties of Eu3+-activated yttrium oxysulfide red phosphor. Thin Solid Films. 1998;330:173–7.

    Article  CAS  Google Scholar 

  42. Cavouras D, Kandarakis I, Tsoukos S, Kateris A, Nomicos CD, Panayiotakis GS. A theoretical model for calculation of the detective quantum efficiency in granular scintillators. Appl Radiat Isot. 2001;55:831–42.

    Article  CAS  PubMed  Google Scholar 

  43. Ahmed SN. Interaction of radiation with matter. Phys Eng Radiat Detect. 2nd ed. Academic Press; 2015. p. 27–29.

  44. XCOM: Photon cross sections database [Internet]. Available from: https://www.nist.gov/pml/xcom-photon-cross-sections-database.

  45. Shahmohammadi Beni M, Krstic D, Nikezic D, Yu KN. Monte Carlo studies on photon interactions in radiobiological experiments. PLoS One. 2018;13:e0193575.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  46. Baró J, Roteta M, Fernández-Varea JM, Salvat F. Analytical cross sections for Monte Carlo simulation of photon transport. Radiat Phys Chem. 1994;44:531–52.

    Article  Google Scholar 

  47. Hubbell JH. Review of photon interaction cross section data in the medical and biological context. Phys Med Biol. 1999;44:R1–22.

    Article  CAS  PubMed  Google Scholar 

  48. Swiderski L, Moszynski M, Syntfeld-Kazuch A, Szawlowski M, Szczesniak T. Measuring the scintillation decay time for different energy depositions in NaI:Tl, LSO:Ce and CeBr3 scintillators. Nucl Instruments Methods Phys Res Sect A Accel Spectrometers, Detect Assoc Equip. 2014;749:68–73.

  49. Archer J, Madden L, Li E, Wilkinson D, Rosenfeld AB. An algorithmic approach to single-probe Cherenkov removal in pulsed x-ray beams. Med Phys. 2019;46:1833–9.

    Article  PubMed  Google Scholar 

  50. Swinth KL, Ewins JH. Biomedical probe using a fiber-optic coupled scintillator. Med Phys. 1976;3:109.

    Article  CAS  PubMed  Google Scholar 

  51. Jang KW, Cho DH, Yoo WJ, Seo JK, Heo JY, Park JY, et al. Fiber-optic radiation sensor for detection of tritium. Nucl Instruments Methods Phys Res Sect A Accel Spectrometers, Detect Assoc Equip. 2011;652:928–31.

    Article  CAS  Google Scholar 

  52. McCarthy D, O’Keeffe S, Lewis E, Sporea D, Sporea A, Tiseanu I. Optical fibre X-ray radiation dosimeter sensor for low dose applications. Proc IEEE Sensors. 2011:121–4.

  53. O’Keeffe S, Zhao W, Sun W, Zhang D, Qin Z, Chen Z, et al. An optical fibre based sensor for real-time monitoring of clinical linear accelerator (Linac) radiotherapy delivery. IEEE J Sel Top Quantum Electron. 2015;22:35–42.

    Article  CAS  Google Scholar 

  54. De Andres AI, O’keeffe S, Chen L, Esteban O. Highly sensitive extrinsic X-ray polymer optical fiber sensors based on fiber tip modification. IEEE Sens J. 2017;17:5112–7.

    Article  Google Scholar 

  55. Teichmann T, Sommer M, Henniger J. Dose rate measurements with a ruby-based fiber optic radioluminescent probe. Radiat Meas. 2013;56:347–50.

    Article  CAS  Google Scholar 

  56. Beddar AS, Mackie TR, Attix FH. Cerenkov light generated in optical fibres and other light pipes irradiated by electron beams. Phys Med Biol. 1992;37:925–35.

    Article  Google Scholar 

  57. Molina P, Santiago M, Marcazzó J, Spano F, Henniger J, Cravero W, et al. Radioluminescence of red-emitting Eu-doped phosphors for fiberoptic dosimetry. Appl Radiat Isot. 2012;71:12–4.

    Article  CAS  PubMed  Google Scholar 

  58. Molina P, Sommer M, Kattner F, Henniger J. Response characterization of an Y2O3:Eu-based radioluminescence probe under 60Co irradiation. Radiat Meas. 2013;56:338–41.

    Article  CAS  Google Scholar 

  59. Martínez N, Rucci A, Marcazzó J, Molina P, Santiago M, Cravero W. Characterization of YVO4:Eu3+ scintillator as detector for fiber optic dosimetry. Radiat Meas. 2017;106:650–6.

    Article  CAS  Google Scholar 

  60. Kertzscher G, Beddar S. Inorganic scintillation detectors based on Eu-activated phosphors for 192Ir brachytherapy. Phys Med Biol. 2017;62:5046–75.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  61. De Boer SF, Beddar AS, Rawlinson JA. Optical filtering and spectral measurements of radiation-induced light in plastic scintillation dosimetry. Phys Med Biol. 1993;38:945–58.

    Article  Google Scholar 

  62. Lambert J, Yin Y, McKenzie DR, Law S, Suchowerska N. Cerenkov light spectrum in an optical fiber exposed to a photon or electron radiation therapy beam. Appl Opt. 2009;48:3362–7.

    Article  PubMed  Google Scholar 

  63. Qin Z, Xie T, Dai X, Zhang B, Ma Y, Khan IU, et al. New model for explaining the over-response phenomenon in percentage of depth dose curve measured using inorganic scintillating materials for optical fiber radiation sensors. Opt Express. 2019;27:23693.

    Article  CAS  PubMed  Google Scholar 

  64. Martinez N, Teichmann T, Molina P, Sommer M, Santiago M, Henniger J, et al. Scintillation properties of the YVO4:Eu3+ compound in powder form: its application to dosimetry in radiation fields produced by pulsed mega-voltage photon beams. Z Med Phys. 2015;25:368–74.

    Article  PubMed  Google Scholar 

  65. Suchowerska N, Lambert J, Nakano T, Law S, Elsey J, McKenzie DR. A fibre optic dosimeter customised for brachytherapy. Radiat Meas. 2007;42:929–32.

    Article  CAS  Google Scholar 

  66. Linares Rosales HM, Duguay-Drouin P, Archambault L, Beddar S, Beaulieu L. Optimization of a multipoint plastic scintillator dosimeter for high dose rate brachytherapy. Med Phys. 2019;46:2412–21.

    Article  CAS  PubMed  Google Scholar 

  67. Johansen JG, Rylander S, Buus S, Bentzen L, Hokland SB, Søndergaard CS, et al. Time-resolved in vivo dosimetry for source tracking in brachytherapy. Brachytherapy. 2018;17:122–32.

    Article  PubMed  Google Scholar 

  68. Al Towairqi M, Cutajar DL, Braddock T, Li E, Wadi-Ramahi S, Moftah B, et al. Characterisations of a fibre optic dosimetry system for source tracking during HDR brachytherapy. J Phys Conf Ser. 2019;1154.

  69. Kertzscher G, Beddar S. Ruby-based inorganic scintillation detectors for 192Ir brachytherapy. Phys Med Biol. 2016;61:7744–64.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  70. Kertzscher G, Beddar S. Inorganic scintillation detectors for 192Ir brachytherapy. Phys Med Biol. 2019;64:225018.

    Article  CAS  PubMed  Google Scholar 

  71. Belley MD, Craciunescu O, Chang Z, Langloss BW, Stanton IN, Yoshizumi TT, et al. Real-time dose-rate monitoring with gynecologic brachytherapy: results of an initial clinical trial. Brachytherapy. 2018;17:1023–9.

    Article  PubMed  Google Scholar 

  72. Johansen JG, Kertzscher G, Jørgensen EB, Rylander S, Bentzen L, Hokland SB, et al. Dwell time verification in brachytherapy based on time resolved in vivo dosimetry. Phys Medica. 2019;60:156–61.

    Article  CAS  Google Scholar 

  73. Penner C, Hoehr C, O’Keeffe S, Woulfe P, Duzenli C. Characterization of a terbium-activated gadolinium oxysulfide plastic optical fiber sensor in photons and protons. IEEE Sens J. 2018;18:1513–9.

    Article  CAS  Google Scholar 

  74. Savard N, Potkins D, Beaudry J, Jirasek A, Duzenli C, Hoehr C. Characteristics of a Ce-doped silica fiber irradiated by 74 MeV protons. Radiat Meas. 2018;114:19–24.

    Article  CAS  Google Scholar 

  75. Hoehr C, Morana A, Duhamel O, Capoen B, Trinczek M, Paillet P, et al. Novel Gd3+-doped silica-based optical fiber material for dosimetry in proton therapy. Sci Rep. 2019;9:4–11.

  76. Ito Y, Katano G, Harano H, Matsumoto T, Uritani A, Kudo K, et al. Development of a tiny neutron probe with an optical fibre for BNCT. Radiat Prot Dosimetry. 2004;110:619–22.

    Article  CAS  PubMed  Google Scholar 

  77. Bartesaghi G, Conti V, Prest M, Mascagna V, Scazzi S, Cappelletti P, et al. A real time scintillating fiber dosimeter for gamma and neutron monitoring on radiotherapy accelerators. Nucl Instruments Methods Phys Res Sect A Accel Spectrometers, Detect Assoc Equip. 2007;572:228–30.

    Article  CAS  Google Scholar 

  78. Watanabe K, Kawabata Y, Yamazaki A, Uritani A, Iguchi T, Fukuda K, et al. Development of an optical fiber type detector using a Eu:LiCaAlF6 scintillator for neutron monitoring in boron neutron capture therapy. Nucl Instruments Methods Phys Res Sect A Accel Spectrometers, Detect Assoc Equip. 2015;802:1–4.

    Article  CAS  Google Scholar 

  79. Ishikawa A, Ymazaki A, Watanabe K, Yoshihashi S, Uritani A, Tsurita Y, et al. A comparison between simulation and experimental results for depth profile of 6Li reaction rate in a water phantom of BNCT using a small 6Li-based scintillator neutron detector with an optical fiber. Radiat Meas. 2020. p. 106270.

  80. Darafsheh A, Taleei R, Kassaee A, Finlay JC. Proton therapy dosimetry using the scintillation of the silica fibers. Opt Lett. 2017;42:847.

    Article  CAS  PubMed  Google Scholar 

  81. Yagi T, Misawa T, Pyeon CH, Shiroya S. A small high sensitivity neutron detector using a wavelength shifting fiber. Appl Radiat Isot. 2011;69:176–9.

    Article  CAS  PubMed  Google Scholar 

  82. Nascimento LF, Veronese I, Loi G, Mones E, Vanhavere F, Verellen D. Radioluminescence results from an Al2O3:C fiber prototype: 6 MV medical beam. Sensors Actuators, A Phys. 2018;274:1–9.

    Article  CAS  Google Scholar 

  83. Ramírez M, Martínez N, Marcazzó J, Molina P, Feld D, Santiago M. Performance of ZnSe(Te) as fiberoptic dosimetry detector. Appl Radiat Isot. 2016;116:1–7.

    Article  PubMed  CAS  Google Scholar 

  84. Hu Y, Qin Z, Ma Y, Zhao W, Sun W, Zhang D, et al. Characterization of fiber radiation dosimeters with different embedded scintillator materials for radiotherapy applications. Sensors Actuators, A Phys. 2018;269:188–95.

    Article  CAS  Google Scholar 

  85. Qin Z, Hu Y, Ma Y, Lin W, Luo X, Zhao W, et al. Water-equivalent fiber radiation dosimeter with two scintillating materials. Biomed Opt Express. 2016;7:4919.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  86. Woulfe P, O’Keeffe S, McCarthy D, Grattan M, Hounsell A, Cronin J, et al. Characterisation of radioluminesence based optical fibre dosimeter in radiotherapy beam applications. 2013. IEEE SENSORS. 2013:1–4.

  87. Pietrzak R, Konefał A. Determination of energy spectra in water for 6 MV X rays from a medical linac. Acta Phys Pol B. 2016;47:783.

    Article  Google Scholar 

  88. Skrobala A, Adamczyk S, Kruszyna-Mochalska M, Skórska M, Konefał A, Suchorska W, et al. Low dose out-of-field radiotherapy, part 2: calculating the mean photon energy values for the out-of-field photon energy spectrum from scattered radiation using Monte Carlo methods. Cancer/Radiothérapie. 2017;21:352–7.

    Article  CAS  Google Scholar 

  89. Aznar MC, Andersen CE, Bøtter-Jensen L, Bäck SÅ, Mattsson S, Kjær-Kristoffersen F, et al. Real-time optical-fibre luminescence dosimetry for radiotherapy: physical characteristics and applications in photon beams. Phys Med Biol. 2004;49:1655–69.

    Article  PubMed  Google Scholar 

  90. Alharbi M, Gillespie S, Woulfe P, McCavana P, O’Keeffe S, Foley M. Dosimetric characterization of an inorganic optical fiber sensor for external beam radiation therapy. IEEE Sens J. IEEE. 2018;19:2140–7.

    Article  Google Scholar 

  91. Lambert J, Yin Y, McKenzie DR, Law S, Suchowerska N. Cerenkov-free scintillation dosimetry in external beam radiotherapy with an air core light guide. Phys Med Biol. 2008;53:3071–80.

    Article  CAS  PubMed  Google Scholar 

  92. Beddar S. On possible temperature dependence of plastic scintillator response. Med Phys. 2012;39:6522.

    Article  PubMed  Google Scholar 

  93. Intermite A, Putignano M, Welsch CP. Feasibility study of an optical fibre sensor for beam loss detection based on a SPAD array. Proc 9th Eur Work Beam Diagnostics Instrum Part Accel Basel, Switz. 2009. p. 228–30.

  94. Jang KW, Yoo WJ, Seo JK, Heo JY, Moon J, Park JY, et al. Measurements and removal of Cerenkov light generated in scintillating fiber-optic sensor induced by high-energy electron beams using a spectrometer. Opt Rev. 2011;18:176–9.

    Article  CAS  Google Scholar 

  95. Liu PZY, Suchowerska N, Lambert J, Abolfathi P, McKenzie DR. Plastic scintillation dosimetry: comparison of three solutions for the Cerenkov challenge. Phys Med Biol. 2011;56:5805–21.

    Article  CAS  PubMed  Google Scholar 

  96. Fontbonne JM, Iltis G, Ban G, Battala A, Vernhes JC, Tillier J, et al. Scintillating fiber dosimeter for radiation therapy accelerator. IEEE Trans Nucl Sci. 2002;49:2223–7.

  97. Frelin AM, Fontbonne JM, Ban G, Colin J, Labalme M, Batalla A, et al. Spectral discrimination of Čerenkov radiation in scintillating dosimeters. Med Phys. 2005;32:3000–6.

    Article  CAS  PubMed  Google Scholar 

  98. Ishikawa M, Nagase N, Matsuura T, Hiratsuka J, Suzuki R, Miyamoto N, et al. Development of a wavelength-separated type scintillator with optical fiber (SOF) dosimeter to compensate for the Cerenkov radiation effect. J Radiat Res. 2015;56:372–81.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

Not applicable

Funding

This study was funded by China Scholarship Council (grant number 201306230101).

Author information

Authors and Affiliations

Authors

Contributions

Richard M. Perks and Liang Zhao conceived and designed the study. Liang Ding and Qiong Wu wrote the paper. Qun Wang and Yamei Li reviewed and edited the manuscript. All authors read and approved the manuscript.

Corresponding authors

Correspondence to Richard M. Perks or Liang Zhao.

Ethics declarations

Ethics approval and consent to participate

Not applicable

Consent for publication

Not applicable

Competing interests

All the authors declare that they have no competing interests to this work.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ding, L., Wu, Q., Wang, Q. et al. Advances on inorganic scintillator-based optic fiber dosimeters. EJNMMI Phys 7, 60 (2020). https://doi.org/10.1186/s40658-020-00327-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40658-020-00327-6

Keywords