Abstract

The study explores the nitty-gritty of infrared spectroscopy. Firstly, the review gives a concise history of infrared discovery and its location in the electromagnetic spectrum. Secondly, the infrared spectroscopy is reported for its mechanism, principles, sample preparation, and application for absence and presence of functional groups determination in both ligands and coordination compounds. Thirdly, it helps in purity determination of unknown samples. Additional studies regarding this study entail infrared spectroscopy-based synchrotron radiation. It serves as a giant microscope to give detailed information of samples under investigation compared to the conventional infrared instrument. Infrared will continue to be useful to both chemical and pharmaceutical industries, in order to make chemical products and manufactured drugs put on wholesome integrity.

1. Introduction

Infrared is a convenient tool relevant to chemistry and chemistry related disciplines (inorganic and organic chemistry and pharmacy) [1, 2]. Reports of its relevancy could be seen in studies indicating the detection of absence and presence of functional groups in ligands and evidence of coordination in the proposed compounds [3]. Infrared has versatile applications to differentiate between the hydroxyl and amine groups from the spectra [4] and determine the presence of water in case of no thermal analyzer [5], and it can be used for purity determination [5, 6] in case of no elemental analyzer. The study aimed to provide the history, pragmatic uses, maintenance of infrared spectroscopy, and concise infrared spectroscopy based synchrotron radiation.

2. Electromagnetic Spectrum and Spectroscopy

The electromagnetic spectrum is made up of an array of regions useful to chemists and the scientific world [7]. It is also a range of frequencies of electromagnetic radiation and their corresponding wavelengths with photon energies [8] as shown in Figure 1. The range has electromagnetic waves with minimum frequencies of below one g hertz to a maximum of over 1028 hertz, equivalent to wavelengths [9] or from a fraction of an atomic nucleus size to a thousand kilometre [8]. This frequency range is divided into distinct regions and the electronic waves within every frequency region [10]. They are named differently starting from the low frequency (long wavelength) to the spectrum end [11]. The spectrum entails radio waves, microwaves, infrared, visible light, ultraviolet, X-rays, and gamma rays at the high frequency (short wavelength) as shown in Figure 1. The electromagnetic waves in each of the regions possess various characteristics, such as mode of production, mode of interaction with matter, and their pragmatic applications [12].

2.1. Spectroscopy

A characterization technique known as spectroscopy could be used physically to separate electromagnetic waves of various frequencies [13, 14]. Therefore, it generates a spectrum which shows the constituents’ frequencies [15, 16]. Spectroscopy is used to study the interactions of the electromagnetic waves with matter [17, 18].

3. Infrared

3.1. Infrared Spectrum and Spectroscopy (IR Spectroscopy)
3.1.1. Infrared (IR) Spectrum

The infrared region of the electromagnetic spectrum is in three portions, namely, the near-, mid-, and far-IR [3, 19]. They are named based on their relationships with the visible spectrum [20]. The near-IR has a high energy of approximately 14000–4000 cm−1, which can excite harmonic vibrations [3]. The mid-IR has approximately 4000–400 cm−1 while the far-IR has approximately a low energy of 400–10 cm−1, which is next to the radar and microwave regions (Figure 1), and could be used for rotational spectroscopy [3].

3.1.2. Infrared (IR) Spectroscopy

Infrared spectroscopy also referred to as IR spectroscopy deals with the interaction of absorbed radiant energy in the infrared region of the electromagnetic spectrum with matter [21]. Infrared radiation has a longer wavelength and low frequency. The IR radiation is centered on spectroscopy [22]. It is used to identify chemical compounds with a current instrument called Fourier Transform Infrared (FTIR) spectrometer, which implements the principle of IR spectroscopy [19, 23, 24]. The FTIR is also a technique which measures and records infrared spectra [19].

In summary, IR spectroscopy uses IR spectrum to probe matter.

3.2. Mechanism of Action and Principles of Infrared Spectroscopy

The mechanism of action and principles of infrared spectroscopy complement each other [25, 26] as shown in Figure 2.

3.2.1. Infrared Mechanism of Action

Inside the infrared light source is an apparatus called interferometer. It guides the IR sample. The interferometer consists of a moving mirror which changes the IR light distribution when it passes through the sample [27, 28]. The resulting signal called, “interferogram,” is recorded directly, signifying the light output as a function of mirror position [29].

A data-processing technique called “Fourier transform” transforms the raw data into the expected results of the sample’s spectrum [24, 30]. Light output is a function of infrared wavelength. The sample’s spectrum is mostly compared with a reference spectrum [Figure 2] in order to remove fixed-pattern noise [31]. The IR spectrometer having both reference and sample spectra are referred to as “a two-beam absorption spectrometer” [31].

3.3. Principles of IR Spectroscopy

The principle involves a process whereby a beam of IR light from IR source (Globar) is passed through a sample and the infrared spectrum is recorded [22, 24]. The process is directed with the help of a monochromator (monochromatic beam of light), which is placed after IR source and before the IR detector, which interchanges in wavelength over time [32]. The sample absorbs light when there is resonant frequency [33]. Resonant frequency is when the IR frequency is the same as the vibrational frequency of the sample’s bond [3436]. The transmitted light shows the quantity of energy absorbed at every wavelength [37]. The Fourier transform instrument measures the entire wavelength at once and produces an absorbance or transmittance spectrum revealing the IR wavelengths, which the sample absorbs (Figure 2) [24]. Analysis of the spectrum gives details of the sample’s molecular structure. This technique has been used on samples with covalent bonds; high levels of purity few IR active and complex molecular structures [38].

3.4. Theory of Infrared Spectroscopy

The IR spectroscopy is centered on the fact that molecules absorb specific frequencies which are characteristics of their chemical structures [22]. Hence, the resonant frequencies are related not only to the bonds but also to the atoms [39]. In other words, vibrational frequency could be connected with a particular bond type [40].

3.4.1. Vibrational Modes

A vibrational mode, also called vibrational degree of freedom, deals with ways where a molecule undergoes vibrations [41, 42]. A vibration mode could be IR active in a molecule if changes in permanent dipole are fundamental for a vibrational mode in a molecule [43, 44].

Molecules, such as carbon (IV) oxide, have 3N-5 degrees of vibrational with N atoms in the molecules [44, 45]. Linear molecules, like simple diatomic molecules (e.g., CO2; O=C=O), have solely one bond and solely one vibrational bond calculated from 3N-5 degrees of vibrational modes. On the other side, nonlinear molecules, such as water, have 3N-6 degrees of vibrational modes [44, 46]. Here, water has 3 atoms; that is, 3 × 3 − 6 = 3 degrees of freedom. Symmetrical molecules, such as O2, are observed solely in the Raman spectrum, not in the IR spectrum unlike unsymmetrical diatomic molecules, such as CO2 in the IR spectrum [47]. In the case of complex molecules, they have several bonds and they are observed in the IR spectrum [35]. The atoms in methylene (groups, found usually in organic compounds vibrate in six various ways, viz, rocking, scissoring, symmetric and antisymmetric stretching, twisting, and wagging [48].

4. Sample Preparation for IR Characterization

Unlike liquid state nuclear magnetic resonance (NMR), which requires right solvent, the sample for IR characterization does not require a solvent to dissolve it.

4.1. Solid Samples

Solid samples for IR characterization are prepared in four various ways. Firstly, a quantity of the studied sample is ground with fine purified salt, usually potassium bromide (KBr) in an agate mortar with pestle to form a mixture. The KBr is used to record the IR spectra because it has a larger transmission window (100%) in the IR compound than sodium chloride. The agate mortar with pestle is used to prevent contamination [18]. Thin film of the mull produced is smeared onto salt plates and measured by the infrared spectrophotometer [44]. The mixture gotten is then mechanically pressed to form a translucent pellet which allows the beam of light of the spectrophotometer to pass through it [44]. The second method entails a mixture of sample and an oily mulling agent which is ground also in a mortar with a pestle to form a Nujol [44]. The third method is adopted for polymeric materials [44]. This method is called the “cast film” technique [44, 49]. Here, the sample is first dissolved in a suitable, nonhygroscopic solvent [44]. A drop of the solution gotten from the sample dissolution in a suitable nonhygroscopic solvent is deposited on the surface of sodium chloride or sodium bromide cell [44]. Analysis is conducted directly on the film formed on the cell after evaporation of the solution to dryness [44]. It is important to take caution to ensure the film is thin (not too thick) for the light to pass through; otherwise light cannot pass through [44]. This technique is suitable for qualitative analysis [44]. The last method is using microtomy to cut a thin (20–100 µm) film from a solid sample [44]. It is one of the most essential ways to analyze the integrity of a preserved solid, such as a failed plastic product [44].

4.2. Liquid Samples

Here, liquid samples are placed in between two plates of a salt, such as sodium chloride, potassium bromide, or calcium fluoride [44]. The plates are usually transparent to the infrared light so that they do not introduce any lines onto the spectra [44].

4.3. Gaseous Samples

Sample cells with a long path length are needed for gaseous samples in order to make up for the dilution [44]. The path length of the sample cell depends on the concentration of the compound [44]. A sample gas concentration down to several hundred ppm requires a simple glass tube with a length of 5 cm to 10 cm walled with an ending of infrared windows, while a White cell with a mirror guided with infrared light can measure sample gas concentration below ppm [44]. The White cells are available with optical path length ranging from 0.5 m up to 100 metres [44].

In summary, samples prepared for IR characterization can be in solid, liquid, or gaseous states.

4.4. Photoacoustic Spectroscopy

In this spectroscopy, the solid or liquid sample requires minimal sample treatment [44, 50]. The sample is placed into the sample cup with an insertion of a sealed photoacoustic cell for the spectral measurement [44]. The sample may be one solid piece, powder, or basically in any form for the measurement [44]. It is important to note that spectra obtained from different sample preparation methods will look slightly different from each other [44]. Differences in the samples’ physical states bring about slight differences in spectra because of different sample preparation methods [44].

5. Applications of Infrared Spectroscopy

Infrared spectroscopy, being a simple as well as a reliable characterization technique, is applicable to all aspects of organic and inorganic chemistry [51]. It is also useful in research institutes and industries [52]. The purpose of adopting this characterization technique is to determine quality control, dynamic measurement, and monitoring applications such as the long-term unattended measurement of CO2 concentrations in greenhouses, growth chambers by infrared gas analyzers, and forensic analysis [53]. Changes in the character of a certain bond are measured and assessed at a specific frequency over a period of time [11]. Infrared measurements up to 32 times per second could be achieved with modern research instruments, as well as with other techniques simultaneously [5457]. This enables the observations of chemical reactions and processes to be faster and more accurate [54].

5.1. Microelectronics and Semiconductors Infrared

Infrared spectroscopy has been used successfully in the field of microelectronics, as well as semiconductors infrared, such as amorphous silicon, silicon arsenide, and other gallium and silicon compounds [58, 59]. Nowadays, the durability of the instruments has made them useful for field trials [23, 55].

5.2. Isotope Effects

The different isotopes in a particular species may give fine details in infrared spectroscopy, obtainable with species possessing different isotopes [6062]. For instance, the O–O stretching frequency (in reciprocal centimetres) and the wave number of absorbance, , can be calculated using the formula  = 1/2πc√k/µ, where ν is the wavenumber in cm−1, k is the spring constant for the bond, c is the speed of light in ms−1, and μ is the reduced mass of the A-B system as shown in the following:

Presently, the effect of isotopes on vibration and decay dynamics was observed compared to before. Antisymmetric stretch mode decay of interstitial oxygen entails the symmetric stretch mode with strong isotope dependence in some substances, such as germanium and silicon.

5.3. Dimensional Infrared
5.3.1. Two-Dimensional Infrared

The two types of two-dimensional IR are linear and nonlinear.

5.3.2. Linear Two-Dimensional Infrared Spectroscopy

Linear two-dimensional infrared bond spectroscopy analysis entails the use of two-dimensional infrared bond analyses on infrared spectra [63, 64]. This is done in order to increase the spectra of a disturbed sample for simpler spectral analysis and improved resolution [65, 66]. The pulse sequence is used to get a two-dimensional Fourier Transform Infrared spectrum [67, 68].

5.3.3. Nonlinear Two-Dimensional Infrared Spectroscopy

Nonlinear two-dimensional infrared spectroscopy is the infrared version of correlations spectroscopy [69]. Femtosecond infrared laser pulses development came along with nonlinear two-dimensional infrared spectroscopy [70, 71]. The experiment entails the first set of probe pulses useful to the sample, followed by a delay time for system relaxation [7173]. The usual duration for the delay time requires zero to some picoseconds, which could be influenced by tens of femtosecond revolutions. A probe pulse is applied subsequently which results in the emission of a signal from the sample. Graphically, the nonlinear two-dimensional infrared spectrum is a two-dimensional correlation plot of the frequency ɷ1 excited by the first pump pulses and frequency ɷ3 excited by the probe pulse after the delay time. This supports the coupling observation between different vibrational modes because of its extremely high time resolution [74]. It could also be used to observe molecular dynamics on a picoseconds timescale, though still mainly a new technique, but its knowledge is increasing for fundamental research [74, 75]. As said previously, infrared spectroscopy could be used with other characterization techniques, such as two-dimensional nuclear magnetic resonance (NMR) [76, 77]. The nonlinear two-dimensional infrared technique spreads the spectrum in two dimensions and supports the observation of cross peaks, consisting of information of coupling between different modes [78, 79].

6. Difference between Nonlinear Two-Dimensional Infrared Spectroscopy and 2D Nuclear Magnetic Resonance

The nonlinear two-dimensional infrared spectroscopy is different from 2D NMR due to its involvement in the excitation to overtones [35]. The excitation outcome gives an excited state absorption peak located below the diagonal and cross peaks [73]. In 2DNMR, the two techniques, COSY and NOESY, are used often [80]. The cross peaks in the first are relevant to the scalar coupling, whereas, in the later, they are relevant to the spin transfer between diverse nuclei [81, 82]. The 2DNMR techniques analogs are imitated in nonlinear two-dimensional IR spectroscopy, where the nonlinear two-dimensional IR spectroscopy with zero delay time corresponds to COSY [83]. In addition, nonlinear two-dimensional IR spectroscopy with fixed delay time permitting vibrational population transfer which corresponds to NOESY [84]. The COSY alternative of nonlinear two-dimensional IR spectroscopy has been used to determine the secondary structure of proteins [85].

7. Purity Determination

The challenge of values higher than 0.5% obtained from samples analyzed with elemental analyzer is giving a chance to infrared to be considered for purity determination of samples [86, 87]. The application of the infrared technique to analyze samples for purity determination could be observed in three different ways, namely, old way, new way, and lack of reference [24, 8890].

7.1. The Old Way

In the old way, the infrared spectrum of the unknown sample is taken occasionally and compared with the infrared spectrum of the known pure sample (reference) [91]. Any irrelevant peak suggests the presence of impurities [92]. If the peaks matched those of known impurities, they could apply the intensities to evaluate sample purity.

7.2. The New Way

The introduction of computers and Fourier Transform technology probably led to the use of differential spectroscopy in many industries [24]. Here, an unknown sample spectrum is also obtained occasionally and a stored spectrum of the known pure sample is subtracted digitally from it [24]. It leaves spectrum impurities, which could be compared with impurities’ stored database [24]. The Fourier transform recognizes the impurities and generates a report on the sample composition automatically [63].

7.3. Lack of Reference

In case of lack of reference (infrared of the known sample), an essential interpretation of spectrum of unknown sample could be effected from the first principles [90]. The first principles method is centered on the structural features of the molecule entailing the moieties attached to the molecules produce distinct and repeatable absorptions in the spectrum [90].

8. Maintenance of Infrared Instrument

Maintenance is essential for infrared spectrometer in order to obtain durability and availability when needed. Availability when needed is hindered in most cases because of not having a standby power stabilizer to prevent power failure. In addition, the instrument in most cases is uncovered. Covering them will prevent it from dust and, as well, ensure a good result.

9. Concept of Infrared Spectroscopy-Based Synchrotron Radiation

9.1. Infrared Using Synchrotron Radiation

Infrared based synchrotron radiation provides an efficient technique to characterize chemical compounds crosswise surfaces to determine the sample’s chemical composition [93, 94]. Synchrotron radiation (SR) is the intense light produced by a synchrotron [9597]. It covers a very wide electromagnetic spectrum from radio waves to infrared, visible light, ultraviolet light, X-rays, and gamma rays [98]. Its distinct properties are intense brightness, broadband spectrum, polarization, and pulsed nature [99101]. The intense brightness (brilliance) enables the detection of small quantities of various chemical compositions with a better-quality signal-to-noise ratio, whereas the extremely collimated light allows high-resolution special mapping of the chemical distribution [102104].

9.2. Synchrotron Facility

A synchrotron is a basis for very bright (brilliant) light, which operates like a gigantic microscope to allow the matter to be seen at the atomic level [105, 106]. Synchrotron light is a million times brighter than sunlight and a million times more intense than conventional X-rays [107]. According to Blaustein, synchrotron produces X-ray which is a trillion times brighter than the conventional X-rays [108]. Globally, there are over seventy synchrotrons [109, 110]. Several of them are partly or fully devoted to synchrotron radiation generation for research purposes [110].

Synchrotron, which produces the synchrotron radiation (SR), is lay-down accessories referred to as synchrotron facility [108]. The accessories entail three main components, namely, a particle accelerator, series of magnets, and electric field [111]. The particle accelerators (linear and booster accelerators) accelerate charged particles (usually electrons) [112]. The series of magnets (focusing magnets, bending magnets, insertion devices (undulators), and wigglers) provide magnetic fields to rotate the particles in a circular path called storage ring [113]. The electric field accelerates the particles, and all the three main components are synchronized with the travelling particle beam [114, 115]. A beamline is an experimental location, which uses a particular energy domain released by the synchrotron for research [116118].

9.3. Infrared (IR) Microspectroscopy

Fourier Transform Infrared (FTIR) spectroscopy is a distinguished technique which probes the vibrational modes of molecules [119, 120]. In this way, it provides a spectrum which is structure-specific (MFBU). The FTIR is used generally to study bulk homogeneous materials [121]. However, for the past ten years, the developments in infrared sources, optics, and detectors have led to the merging of FTIR spectroscopy with light spectroscopy to allow spectroscopic microscopy in complex environment of chemical compounds. In addition, since FTIR is nonionizing, compositional variations can be investigated as time goes by on time scales from a few milliseconds to hours. Infrared (IR) microspectroscopy, as implied in the name, is a combination of vibrational mid-infrared (IR) spectroscopy and microscopy [23, 122124]. It enables chemical composition of materials to be determined at microscopic level [125, 126]. In general, an IR transmission microscope, with all reflecting optics is attached to a Fourier Transform IR (FTIR) spectrometer, allowing mid-IR absorption spectra to be collected from measurements as small as 5–10 µ min, a sample to be analyzed [127]. This method depends on the use of a high numerical aperture (NA) condenser and objective optics to center the IR beam onto the sample, with ensuring beam collection which passed through the sample [121]. Optical diffraction limit at an analyzing wavelength is used to set the sample dimensions’ limit to be studied [23]. Aperture insertion into the optical path of the IR beam at a central position paced either before or after the sample usually determines the selected region of interest for analysis [128]. This leads to a loss of most intense beam from reaching the detector in a conventional IR source, such as GlobarTM source [129]. On another note, multipixel IR imaging detectors, such as linear array and focal plane array (FPA) detectors, could be used to allow simultaneous collection of FT-IR spectra from various positions in the sample [23, 130, 131]. However, this method usually depends on less than 0.1% of the IR source total intensity, being aimed at every sensing element of the FPA detector [23]. This other method merges the intense brightness of the wideband synchrotron source with an FT-IR spectrometer and IR microscope, which allows the full intensity of the IR source to be focused on a diffraction-limited position on the sample [23, 132]. The outcome is a high signal-to-noise ratio (SNR) when the samples are analyzed with a magnitude of 5–10 µm [133, 134]. A disadvantage associated with FTIR microspectroscopy (FTIRM) is the long wavelength of infrared light [24, 135, 136]. Here, the diffraction-limited spatial resolution is predicted theoretically as 2–10 µm in the mid-infrared region (4000–500 cm−1) [136, 137]. However, the brightness of the conventional thermal infrared source in laboratory FTIR spectrometers limits the spatial resolution [130, 136, 138, 139]. At this very moment, synchrotron radiation plays an essential role because it is hundred to thousand times brighter than a conventional thermal source, therefore, enabling diffraction-limited spatial resolution to be attained [130, 136, 138, 139].

In summary, a synchrotron infrared source offers very much improved data which possess high signal-to-noise ratio and faster acquisition times for brightness-limited experiments, such as those involving high spatial or spectral resolution, or time-resolved measurements.

10. Conclusion and Ongoing Research

The relevance of infrared spectroscopy is worthwhile in order to achieve detailed result of the analyzed sample in research institutions and industries. This indicates that IR is a very versatile characterization technique. Distinct characteristics of intense brightness, high-resolution, and high signal-to-noise ratio (SNR) qualities of infrared based synchrotron radiation made it more excellent than FT-IR.

11. Future Research

At present, synchrotron IR beamlines committed to high spatial resolution analysis of microscopic materials are functioning in developed countries globally. It will be launched in South Africa (developing country) in 2020.

Data Availability

The data in the document and figures used to support the findings in this study are included within the research article.

Conflicts of Interest

The author declares no conflicts of interest.

Acknowledgments

The author appreciates Govan Mbeki Research and Development Centre (GMRDC) for financial assistance.