Real time analysis of trace volatile organic compounds in ambient air: a comparison between membrane inlet single photon ionization mass spectrometry and proton transfer reaction mass spectrometry

Abstract

Real-time monitoring of volatile organic compounds (VOCs) is critical for a better understanding of chemical processes in ambient air or making minute-by-minute decisions in emergency situations. Proton transfer reaction mass spectrometry (PTR-MS) is nowadays the most commonly used technique for real-time monitoring of VOCs while membrane single photon ionization mass spectrometry (MI-SPI-MS) is a promising MS technique for online detection of trace VOCs. Here, to evaluate the potential of MI-SPI-MS as a complementary tool to PTR-MS, a comprehensive comparison has been performed between MI-SPI-MS and PTR-MS. By using two sets of standard gas mixtures TO15 and PAMS, SPI-MS shows advantages in the detection of ≥C5 alkanes, aromatics and halogens; especially for aromatics, the LODs can reach the ppt level. PTR-MS has performed better in the detection of alkenes, ketones and aldehydes. For outdoor measurements, a number of VOCs have been detected while using MI-SPI-MS and PTR-MS in parallel. Consistent temporal variations have been observed for toluene, C8-aromatics and C9-aromatics by the two instruments, with a more sensitive response from the MI-SPI-MS. Thus by measuring both standard gas mixture and complex ambient air samples, we have successfully demonstrated that MI-SPI-MS will be a helpful tool to provide important complementary information on aromatics and alkanes in air, and proper application of MI-SPI-MS will benefit the real-time monitoring of trace VOCs in relative fields.

Graphical abstract: Real time analysis of trace volatile organic compounds in ambient air: a comparison between membrane inlet single photon ionization mass spectrometry and proton transfer reaction mass spectrometry

Supplementary files

Article information

Article type
Paper
Submitted
01 Jun 2020
Accepted
16 Jul 2020
First published
26 Aug 2020

Anal. Methods, 2020,12, 4343-4350

Real time analysis of trace volatile organic compounds in ambient air: a comparison between membrane inlet single photon ionization mass spectrometry and proton transfer reaction mass spectrometry

Z. Yu, C. Liu, H. Niu, M. Wu, W. Gao, Z. Zhou, Z. Huang and X. Li, Anal. Methods, 2020, 12, 4343 DOI: 10.1039/D0AY01102A

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements